Adaptor A4/2
Instructions
Adaptor with 4 mm male and 2 mm female banana
connector.
The adaptor should be used to connect equipment with 2 mm
sheathed plugs to 4 mm safety output jacks. The 2 mm connectors
are found on certain older equipment and on newer equipment rated
for voltages more than 1000 V. The 4 mm output jacks are found on
newer high voltage power supplies rated for less than 1000 V.
Important safety precautions
Observe the following safety precautions to prevent exposure to high
voltages:
•Check carefully the instruction manual for the used power supply
and the connected equipment.
•This adaptor is designed to use with Cytiva high voltage power
supplies with 4 mm sheathed output jacks, for example EPS
2A200, EPS300 and EPS600.
•This adaptor is approved for 500 V relative protection ground.
•The power supply output must not exceed ±500 V relative to
protection ground. This means a maximum setting of 1000 V in
case the power supply's outputs are symmetrical against
protection ground, and a maximum setting of 500 V in case the
power supply has outputs connected to protection ground.
•The power supply setting must never exceed the maximum
rating of the connected equipment which should be less than
1000 V.
•Switch the power supply off before connecting or disconnecting
the adaptor.
•Use this adaptor only with the sheathed plugs illustrated above.
Do not use with an exposed plug or a plug with a retractable
shroud.
Important user information
WARNING
Meaning: Consult the instruction manual to avoid
personal injury or damage to the product or other
equipment.
WARNING! The Warning sign is used to call attention to the
necessity to follow an instruction in detail to avoid personal injury. Be
sure not to proceed until the instructions are clearly understood and
all stated conditions are met.
WARNING
Risk of exposure to high voltage unless safety
precautions are followed.
Power supply
4 mm safety
output jack
A4/2 adaptor
Instrument lead
2 mm sheathed plug
Specifications
Max. voltage: 500 V
Product certifications: EN61010-1, EN61010-2-031,
UL3101-1, CSA C22.2 1010.1, CE
This declaration of conformity is only valid when the adaptor is used
in laboratory locations, used as delivered from Cytiva, and connected
to other CE labeled instruments or products recommended or
approved by Cytiva.
cytiva.com 56307307 AF