Dabi Atlante EAGLE.PS User manual


Owner’s Manual
2
MANUAL PRESENTATION
EQUIPMENT MANUAL (OPERATING INSTRUCTIONS)
TechnicalName:Radiographic images scanner
BusinessName:EAGLE.PS
Model:EAGLE.PS
Brand:DABI ATLANTE
Supplier/Manufacturer:
Alliage S/A Indústrias Médico Odontológica
Rod. Abrão Assed, Km 53+450m - Cx. Postal 782 CEP 14097- 500
Ribeirão Preto - S.P. - Brasil
Telephone +55 (16) 2102-5000 - Fax +55 (16) 2102-5001
CNPJ: 55.979.736.0001/45
TechnicalManager:Ricardo J. Ravaneli
CREA-SP: 5060714523
ANVISA Register Nº: 10101139027
ATTENTION
For more Safety:
Read and understand all instructions contained in these operating instructions before installing or operating
this equipment.
Note: These operating instructions shall be read by all operators of this equipment.

Radiographic image scanner EAGLE.PS
3
INDEX
02MANUALPRESENTATION
04EQUIPMENTIDENTIFICATION
04 Equipment Indication
04 Equipment Purpose
05 Principles and grounds applied to the operation of the product
05 Equipment Description
06MODULES,ACCESSORIES,OPTIONALSANDCONSUMERMATERIALS
09TECHNICALSPECIFICATIONS
10 Packaging symbology
10 Product Symbology
12EQUIPMENTINSTALLATION
17EQUIPMENTOPERATION
71 CAUTIONS, RESTRICTIONS AND WARNINGS
71 Recommendations for equipment conservation
71 Transport, storage and operation conditions
71 Sensitivity to predictable environmental conditions under normal use
71 Cautions and Attention “during equipment installation”
72 Cautions and Attention “during equipment use”
72 Cautions and Attention “after” equipment use
72 Cautions and Attention during "cleaning and disinfection" of the equipment
73 Cautions in case of changes to equipment operation
73 Cautions to be adopted against predictable or unusual risks related to the deactivation and
abandonment of equipment
73CONSERVATIONANDPREVENTIVEANDCORRECTIVEMAINTENANCE
73 Additional procedures for reuse
73 General cleaning
74 Disinfection
74 Preventive maintenance
74 Corrective maintenance
75UNFORESEEN‐TROUBLESHOOTING
91EQUIPMENTWARRANTY
91FINALCONSIDERATIONS
92APPENDIXA–WIA(WINDOWSIMAGEAQUISITION)
96APPENDIXB–WINDOWSFIREWALLCONFIGURATION–SERVERMODE

Owner’s Manual
4
EQUIPMENT IDENTIFICATION
Dearcustomer
Congratulations for your excellent choice. While purchasing "DABI ATLANTE" quality equipment, you can
be sure you have Retrieved technology products compatible with the world’s best product in its class. This
manual gives you an overview of your equipment. It describes important details that can guide you in its
correct use, as well as in the solution of small problems that may occur.
We advise you to read it completely and keep it for future reference.
PRODUCT IDENTIFICATION
TechnicalName:Radiographic images scanner
BusinessName:EAGLE.PS
Model:EAGLE.PS
Brand:DABI ATLANTE
EquipmentIndication
This equipment is for dental use only and should be used and handled by a qualified person (properly
regulated professional, according to the local legislation of the country) observing the instructions contained
in this manual.
It is the obligation of the user to use the equipment only under perfect operating conditions and to protect
himself, patients and third parties from possible hazards.
EquipmentPurpose
The Radiographic images scanner Eagle.PS is intended for intraoral oral radiographic examination by
optical reading and processing of image data of phosphorus plaques for the diagnosis of diseases of the
teeth, mandible and oral structures.

Radiographic image scanner EAGLE.PS
5
EQUIPMENT IDENTIFICATION
Principlesandgroundsappliedtotheoperationoftheproduct
The Radiographic images scanner Eagle.PS operates as a scanner, converting the image data stored on
the phosphorous plate into a digital radiographic image.
The retrieval process is performed by positioning the phosphorous plate in the inner part of the mouth,
behind the structure to be to examined. The structure should be exposed to a dose of x-rays using an
external source.
Once exposed, the phosphorous plate should be inserted into the input compartment of the equipment to
read the image data.
Dental imaging software performs the interface between the Eagle.PS system and the user, allowing to
view, enhance, store, send, print, among other functions.
Equipmentdescription
The Radiographic images scanner Eagle PS, provides high resolution images that meet all Diagnosis
needs Compact design, easy to handle, ideal for dental offices (General practice).
Body injected in polycarbonate with anti-UV protection, which provides greater safety, resistance and
durability to the set.
Eagle.PS is compatible with any X-ray machine.
Reusable and inexpensive replacement phosphorous plates, available in 4 formats (adult and child).

Owner’s Manual
6
MODULES, ACCESSORIES, OPTIONALS AND CONSUMER MATERIALS
The contents of this page is for information, thus, the
equipment may be different from the one illustrated.
Therefore, when purchasing the product, check the technical
compatibility between equipment, coupling and accessories
01 - Eagle PS Body
02 - Inlet enclosure
03 - On/Off key
04 - Green LED indicator On / Off
05 - Blue LED (status indication)
06 - “Turn Off phosphorous plate” key
07 - White or Orange LED (status indication)
08 - Power supply input
09 - USB port
Status Indication LED’s - See page . 07

MODULES, ACCESSORIES, OPTIONALS AND CONSUMER MATERIALS
Status LED color
Blue White Orange Green Orange/White
Access Established
communication
with the Sof-
tware.
Ready to
start erasure.
Equipment
switched on.
Ready to re-
ceive a new
phosphorous
plate and-
make a new
reading.
Blinking Performing data
exchange with
Software.
Erasing the
Phosphorous
Plate.
An error
occurred.
Blinking alter-
nating
Initializing the
equipment.
Operating LED’s- Status Indication
Attention
If the equipment shows any abnormality, check whether the problem is related to any
of the items listed in the Unforeseen item (situation, cause and solution). If the problem
cannot be solved, switch off the equipment and request a visit of an DABI ATLANTE service
technician at the nearest reseller or call DABI ATLANTE Customer Service + 55 (16) 3512-1288.
Accessories the product
02
04 05
01
03
a b c d
7

8
MODULES, ACCESSORIES, OPTIONALS AND CONSUMER MATERIALS
Attention
- Use of any part, accessory or material not specified
or foreseen in these operating Instructions is entire
responsibility of the user.
01. Image retrieval Software
Eagle.PS system interface, responsible for image management.
02. Phosphorous plates*
The Phosphorous plate serves as a means of exposure to X-rays for image storage. Eagle.PS works
with four plate dimensions: 0, 1, 2 and 3 (on request).
a. For periapical examination (for pediatric use)
b. For bitewing examination (for pediatric use)
c. For periapical examination (for adult use)
d. For bitewing examination (for adult use)
03. Protective covers*
For each plate size a single use disposable hygienic protection is used to prevent cross contami-
nation. The transparent side allows you to see the point, while the opaque side protects the plate.
04. USB cable
For communication between PC / Eagle.PS.
05. Power cable
24V connection - provides power for equipment operation.
06. “Asepsi” cleaner*
Cleaner for sanitizing the phosphor plates.
07. Phosphor plates support*
Organizer support for phosphor plates.
07
06
* Optional item

Radiographic image scanner EAGLE.PS
9
TECHNICAL SPECIFICATIONS
Power supply
SupplyVoltage
127/220V~ (100-240V~)
Frequency
50/60 Hz
Powerconsumption
<40W
Eagle.PS
SupplyVoltage
24 (V, CC)
Electricshockprotection
Class I Equipment
Operationmode
Continuous
Protectionagainstharmfulpenetrationofwaterorparticulatematter
IPX0
Netweight
5,3 Kg
Dimensions
A = 285 mm x L = 175 mm x P = 240 mm
General
FramingaccordingtoAnvisa
Class I
Laserprotectionclass
3B
Attention
Use caution when using this equipment in conjunction with other moving equipment to
avoid collisions.
Warning
Manufacturing materials used for construction of this equipment are Biocompatible.

Owner’s Manual
10
TECHNICAL SPECIFICATIONS
Packaging symbology
Maximum Stacking, determines the
maximum amount of boxes which
can be stacked during
transportation and storage
Determines that the packaging must
be stored or transported with
moisture protection ( Do not expose
to rain, or water spills )
Determines that the packaging
must be stored or transported with
the arrow side up.
Determines that the packaging must
be stored or transported with light
protection.
Determines that the packaging
must be stored and transported
carefully (it should not suffer falls or
receive impacts).
Determines the temperature limits
among which the package must be
stored or transported.
Product symbology
General Warning
If instructions are not followed
properly, use may cause damage
to the product, the user or the
patient.
Note:
Indicates useful information for
product operation.
ON/OFF button
Operational LED for status
Indication.

Radiographic image scanner EAGLE.PS
11
TECHNICAL SPECIFICATIONS
Product Symbology
Use only one time. Do not reuse
It is mandatory to use protective
gloves.
Dangerous electrical voltage.
Laser alert.
USB connection.
Erase phosphorous plate key

Owner’s Manual
12
EQUIPMENT INSTALLATION
Special Actions or Particular Conditions for Installation
Environmental conditions of the room:
See optimal conditions in "Transport, storage and operation conditions", page.71
System requirements
The following components and accessories are required for proper operation of the equipment:
Eagle.PS (Equipment responsible for scanning the image contained in the x-ray);
Peripherals used to interact with the user:
Keyboard:
Accessory by which the characters (alphanumeric, mainly) are inserted for the interaction between the user
and the computer. It can have varying patterns depending on the language used and desired features.
Mouse:
Data input device of a PC for software with graphical interface.
Monitor:
It is a computer output device whose function is to transmit information to the user through the image.
Monitors designed for view of x-rays should have a higher intensity of light and wide range of contrast.
Strong lights, sunlight and reflexes reduce the diagnostic ability of the x-ray examination
Operational system:
The system will work on the following operating systems:
- Windows 7 (32 and 64 bits) or higher
- Mac OS X v10.8 to v10.9.
General requirements:
To ensure system security, access control to features is based on the user's profile on the system.
The user registration will be done on the system itself. See page .32
- Include User.
However, there is a possibility that the system will work with Single login, i.e., without the need for a
registered user.
In such case all features will be available.
For Software configurations in server mode, the following requirements are needed:
- Minimum application resolution 1280x768;
Minimum hardware requirements:
- Processor: 3GHz Dual Core
- RAM memory 8Gb
For Software Settings in client mode, the following requirements are required:
- Minimum resolution 1280x768;
Minimum hardware requirements:
- Processor: 2GHz Dual Core
- RAM memory 4Gb
- Minimum disk space: 10GB
Above requirements are needed to ensure maximum quality of the software image. If they are not followed,
the quality of the image can be compromised. Whenever the monitor resolution used is smaller than the
displayed image size, image quality may be impaired.

Radiographic image scanner EAGLE.PS
13
EQUIPMENT INSTALLATION
System Installation
Note
Before installing the system, make sure that the computer meets all of the above requirements.
To install the system, follow these steps:
1. Insert the Software DVD-ROM into the PC DVD-ROM drive.
2.The Select Language Configuration dialog box appears. Select the installation language and click OK.
3. Check if this is the version and software to be installed on the presentation screen:
4. This system is protected by an exclusive EAGLE license. Read the agreement and indicate whether or
not you agree to continue:
Note
The installation will only continue if the
agreement is accepted.

Owner’s Manual
14
EQUIPMENT INSTALLATION
5. Select destination location:
Note
On this installer screen, a default location that
can be changed is suggested. However, for a
standardization, t is suggested to remain in
the indicated location.
The disk space required for installation is also
indicated. If you do not have this free space,
cancel the installation and exclude some files
to restart the procedure.
6. Select type of installation
Available installation types:
Client: Where only the application is installed. Suggestion
for network clients. In this case, you will need to Include
the IP address of the database server
For the application to work correctly.
Server: Where client and database server are installed.
Suggestion for Application Server or for single user.
7. Choose additional installation tasks
Additional types of icons:
Create an icon on the desktop: Creates an icon on the
user's desktop for easy access to the application.
Single Sign-On: This option disables the use of Login and
Password for the system.

Radiographic image scanner EAGLE.PS
15
EQUIPMENT INSTALLATION
8. Confirm the information selected in the Installation Summary:
9. Complete the installation of the System
To start System Installation the equipment must be
disconnected from the computer. And it should only be
connected to the end of the installation.
The application should only be opened after confirmation
of the installation of the FTDI Driver by the operating
system.
Note
- Firewall port 5431 must be unlocked in order for the Database to be installed correctly, see
section Appendix B – Windows Firewall Configuration – Server Mode.
- If there is an error in the database installation, the process will be aborted and must be
restarted.
- If an error occurs in the installation of the FTDI Driver, the user will be warned. To continue
with the installation, it must be checked whether the reader is not connected to the computer.
- If the FTDI Driver is uninstalled incorrectly, it will not be automatically to reinstall. Therefore,
always use the installation and un-installation of the system with the executables provided by
EAGLE
- If there is already an installation of the Postgres database the user must close all services to
start the installation of the EAGLE.PS system

Owner’s Manual
16
EQUIPMENT INSTALLATION
Electrical connection
Before starting, check that the 24 V DC voltage and the Supply Voltage 100 V - 240 V are consistent
with the information on the identification plate.
The connection ports are located on the back of the equipment.
Connections
The installation procedure is simple. Follow the steps below to properly install the product.
Eagle.PS is equipped with 2 connecting ports located on the rear of the equipment.
- 24V (b) Connection: Provides power for operation of the equipment.
- USB connection (a): Connects to the computer.
- Cables enclosure (c)..
1. Remove the cover from the enclosure cables.
2. Connect the power supply cord (01) to the power supply
Immediately plug the power supply into an electrical outlet.
3. Connect the USB cable (02) to the enclosed input, the free end, connect to the USB port on the
computer.
.

Radiographic image scanner EAGLE.PS
17
EQUIPMENT OPERATION
Phosphorous plates
General guidelines:
Warnings
Eagle.PS must only be used with EAGLE approved plates. Using other supplier's plates may
influence the image quality.
Phosphorous plates are toxic: for this reason they should be wrapped in protective covers
before being inserted into the patient's mouth. In case the patient swallows a plaque or part of
it, a doctor should be consulted immediately.
If the patient damages the protective cover by biting it, the patient's mouth should be flushed
immediately with water.
Active side: The colored side (usually white or blue) is
the active side of phosphorous plate. This side is also
marked by a positioning symbol.
Inactive side: The Inactive side is black. Some
Manufacturers Mark this side as plate size (0, 1, 2, 3 or 4)
and the text "Back" or "Reverse".
Working with Phosphorous plates
The plates are sensitive to physical damage and, if necessary, must be replaced periodically.
Pay attention for:
Plates are flexible like an x-ray film.
- Do not bend or press hard.
- Do not scratch or get them dirty.
- Do not place in autoclaves
Storage:
- Store them in a dark place
- Do not store in a hot or humid place.
- Store them with protective covers using appropriate sizes
- If an image is stored for more than 24 hours, it must be erased first before the next acquisition.
- Protect from exposure to the sun and ultraviolet rays.
Protective cover
For each size of the imaging plate a disposable, single use, hygienic
protection is used to prevent cross contamination. The transparent side
allows you to see the point, While the opaque side protects the Active side of
the inserted plate to avoid image degradation.
The protective cap is a disposable product, thus, it should not be
used again.

Owner’s Manual
18
EQUIPMENT OPERATION
Preparing to shoot the X-ray
Insert a disinfected plate into the respective Protective cover with
The Inactive side facing the adhesive strip, so that the number of
the plate size is seen through the transparent side of the hygienic
protection. Remove the tape and close the opening.
Warning
The Protective cover should be disinfected with a
disinfectant towel before placed in the patient's
mouth. Alternatively it is possible to use a spray
disinfectant on a soft, lint-free cloth.
The protective cover is a disposable product, it should
not be used once more.
Select the appropriate positioner for the region of Interest and the
size of the phosphorous plate.
Perform the required X-ray procedure in accordance with the clinical
procedure.
Attention: Check that the Active side of the plate is facing the x-ray
tube (as seen on the image)
Warning
Infection risk
Clean and sanitize the hygienic protection after each patient.
Wear gloves when removing the protective cap from the patient's mouth.
Starting image acquisition
The process of image retrieval is controlled by software Installed on
your computer.
It is necessary to open the software, click on "Retrieve Image",
Select patient / select dentist / click Retrieve.

Radiographic image scanner EAGLE.PS
19
EQUIPMENT OPERATION
Open the protective cover of the phosphorous plate on the adhesive tab;
Insert the phosphorous plate into the front opening to begin image retrieval. Remember to position
downwards the Inactive side of the phosphorous plate.
The equipment gets the mat started by pulling the phosphorous plate inside and the protective cover must
be removed first for the correct release of the phosphorous plate.
The progress of image retrieval should be monitored through software watching the monitoring screen.
After the image retrieval is complete, the phosphorous plate is automatically switched off and expelled
through the front opening.
Points of observation:
No other phosphorous plate should be inserted until the reading process is completed.
This action may damage improperly inserted phosphorous plates in the equipment.
The image saved through the software has been erased from phosphorous plate, not being able to import
again.
If there is no need for a new image acquisition, the equipment must be switched off.
Retrieve Image
The Retrieve Image screen allows the operator to import an image by associating it with the patient current.
This image has a requester Dentist associated with it.
Figure 43 - Image retrieval Screen

Owner’s Manual
20
EQUIPMENT OPERATION
Patient image retrieval is available through the following path:
"Retrieve Image" icon in the Patient search result table.
Icon in the View Image screen
Image Menu - Retrieve
Home Screen– Other Options – Retrieve Image
In this screen the image retrieval is performed. Enables the operator to Retrieve an image of the
phosphorous plate reader equipment.
Patient *: Represents the name of the patient related to the image to be Retrieved.
Applicant *: Represents the name of the requesting Dentist of the x Rays.
CRO: Represents the CRO number of the requesting Dentist. It is automatically populated when the
"Applicant" is selected.
UF: Represents the federative unit issuing the CRO of the requesting Dentist Options are: ‘AC’, ‘AL’, ‘AM’,
‘AP’, ‘BA’, ‘CE’, ‘DF’, ‘ES’, ‘GO’, ‘MA’, ‘MG’, ‘MS’, ‘MT’, ‘PA’, ‘PB’, ‘PE’, ‘PI’, ‘PR’, ‘RJ’, ‘RN’, ‘RO’, ‘RR’, ‘RS’,
‘SC’, ‘SE’, ‘SP’, ‘TO’ e (empty) (default). It is automatically filled out when the "Applicant" Is filled in.
Current User: Represents the option that, when selected, indicates that the requester is the user currently
logged in.
Quality: Represents the quality of the image to be Retrieved. Possible Options are: 'High' and 'Fast'.
Image Type *: Represents the type of image to be obtained. Possible Options are: "Periapical" or
"Interproximal".
Region *: Represents the region to be X-rayed. The Options are in this field depending on the choice made
in the "Image Type " field.
The system provides the following actions:
Counter-clockwise rotation: Represents the rotation functionality in the counter-clockwise direction, on
the Retrieved image.
Clockwise rotation: Represents the rotation functionality in the clockwise direction, on the Retrieved
image.
Auto-adjust: Represents the option that, if selected, the system automatically adjusts the image brightness
and contrast. This is enabled only if the image has already been Retrieved.
View Image: Represents the option that, if selected, the system redirects to “View Image” (See page .45).
Information:
An image Retrieved and exposed to light may contain an image that does not match the original.
To Retrieve a new image it is necessary to wait finishing the retrieval of the first image.
Remarks:
The fields marked with (*) are mandatory;
'Auto Adjust', 'Rotate', 'Image Type', 'Region', 'Save' and 'Cancel’ functions will be only available after
image acquisition. It is possible to select the dentist that requested the X-ray examination through the
Search feature.
When you click this button, the system will display a screen containing a list of previously saved applicants,
with the 'Name', 'CRO' information and the fields (filters), if you wish to search for a specific applicant.
After acquiring an image, the values contained in the 'Applicant *', 'CRO' and 'UF' fields will remain filled
until the retrieval of the image is canceled, that is, until the operator exits the screen However, the operator
can always select another applicant before acquiring an image.
Table of contents
Other Dabi Atlante Scanner manuals
Popular Scanner manuals by other brands

Extech Instruments
Extech Instruments IRT500 user guide

Böwe Bell + Howell
Böwe Bell + Howell Truper 3200 Operator's manual

Kodak
Kodak Ngenuity 9000 Series user guide

Brother
Brother DSmobile 620 quick start guide

TreadReader
TreadReader HANDHELD product manual

DESKO
DESKO PENTA Scanner Technical specifications