dacell DN-130L User manual

Ver 4.0
MODEL : DN-130L
DIGITAL INDICATOR
USER MANUAL
DACELL CO.,LTD.
Address : 681-1 Cheoksan-Ri, Nami-Myeon, Cheongweon-Gun,
Chung-buk
TEL :82-43-260-2242
FAX : 82-43-260-2245
http://www.dacell.com

1
Contents
1. Sepcial Features 2
1-1. High-speed conversion method 2
1-2. Calibration method 2
1-3. HOLD & PEAK HOLD 2
1-4. Comparison Output 2
1-5. Data Back-up 2
1-6. Watch dog 2
1-7. Option 3
1-8. Power 3
2. Attention 4
2-1. Attention for installation 4
2-2. Attention for use 4
3. Specification 5
4. Front Panel 6
5. Rear Panel 7
6. Wiring Diagram 9
7. Components & Function 10
7-1. Flow Chart 10
7-2. How to use Hold mode 11
7-3. How to use comparison output function 12
8. Setting mode 13
8-1. Types of Setting mode & Set-up 13
8-2. Function mode (Function Set-up) 14
8-3. Digital calibration (Calibration by sensor output value) 21
8-4. Actual load calibration (Real-weight Calibration) 22
8-5. SPAN constant calibration (Calibration by Shunt CAL value) 23
8-6. Data Back-up & Restore 24
8-7. Lock Set-up 25
9. Product Inspection 26
10. OPTION (OP-02, OP-03)
Option 02 (RS232C) 27
Option 03 (RS485) 29

2
1. Special Features
We are grateful to you for picking up our product. If you experience problems while
using it, you may refer to the user manual or contact the Technical Support Department
of the company.
This product is an indicator which displays Torque and RPM in digital mode and it
has the following features.
1-1. High-speed Conversion Method
24bit High-speed A/D converter that can detect the sensor’s input signal at a rate of
1000times per second is used. For analog output, 16bit D/A converter with 1000times per
second is used so that there is almost no devuatuib betweeb display value and output value.
1-2. Calibration Method
2 types of calibration systems are adopted. Calibration by actual load (standard weight) and
calibration by rated output of sensor (mV/V).
1-3. Hold and Peak Hold
You can choose Peak hold or Sample hold as per your need.
1-4. Comparison Output
You can do set-up the maxium and lower limit value by the keys on the front panel and output
contact point signal of the rear panel.
1-5. Data Back-up
All the set-up values will be memorized on the flash memory so the inputted data can be
saved and no need to do the re-setting even in case of the interruption of the electric power or
power disconnection.
1-6. Watch dog
This function is for automatic reset in case the system is stopped due to the external factors
such as noise.

3
1-7. Option
You can use RS232C and RS485 output as the optional specification.
1-8. Power
AC 85 ~ 265V 50/60Hz Free voltage

4
2. Attention
For the efficient and safe use, please carefully read and be fully aware of the following
details before using this product.
It is strictly forbidden to use this product for any other purpose of use or to attempt to
make any alteration on this product.
2-1. Attention for installation
Please keep it out of wet places.
Do not set it up near vibration & impulse, high temperature and humidity.
Keep it out of the direct rays of the sun. Set it up where there is less dust, and Keep
it out of direct air including salt and ion.
Do not use when there is inflammable gas or heavy machinery, and smog.
Ground earth-terminal ( ).
Make wire separately from power system wiring and noise wiring.
Make sure the use of 4 line sealed cable as a sensor cable. Too long cable leads to
measurement error due to wiring resistance (around 10meters).
2-2. Attention for use
During calibration, Do not input free-load state and real-weight load until it becomes
stable. Pressing Enter Key in unstable condition leads to calibration error.
Do not press any Key in use at one's discretion. Please refer to 7. Setting-up mode for
the function and method of Key.

5
3. 사양
Available sensor
Strain Gauge sensor (Bridge 120, 350, 700)
Upper limit Display : -19999 ~ +99999
A/D Converter: 24bit, 500 times/sec
D/A Converter : 16bit
Temperature Characteristic (Amp Characteristic)
Zero : 0.5μV/℃
Span : 50ppm/℃
Front Panel
Measured value display: 7 segment 5 Digit, Character Height 14mm
Status display : Red LED 6 pcs
Key Switch : 4 pcs
Output
Comparison output: Upper limit(HI), Lower limit(LO)
Contact Capacity - AC 125V 0.5A/ DC 24V 1A
Analog Output : DC 0~10V (4~20mA available by internal operation)
• Temperature Range: -10℃∼60℃, less than 80% RH (no dewing.)
External dimension : 97 ×97 ×141 mm (W ×H ×D)
Panel cutting size : 92.5 ×92.5 mm
Temperature Range : About 650g
Power Supply : AC 85~265V 50/60Hz free voltage
Option : RS232C (OP-02), RS485 (OP-03)

6
4. Front Panel
①HI, LO Indication LED
: Hi or LO LED will be lighted when measured value exceeds the setting value.
②HOLD indication LED
: This LED will be lighted when the measured value is on Hold.
③Measured value Indication
: It indicates the measured value and each setting value..
④
Measruing Mode : Once this Key is pressed for more than 1 second, the current
measuring value will be Zero (0) and the Analog output will be 0V (4mA) as well.
SET-UP Mode : Once this is pressed on the Function Set-up Mode, you will
return to the measuring mode.
⑤
Measuring Mode : Once this Key is pressed, the lower limit setting value will be
displayed and this value also can be changed.
SET-UP Mode : The location of row for the number flickering can be moved.
⑥
Measuring Mode : Once this Key is pressed, the upper limit setting value will be
displayed and the value can be changed.
SET-UP Mode : The flickering number will be increased by 1 and 1.

7
⑦
Measuring Mode : Once this Key is pressed, HOLD will be selected and once this
key is pressed again, HOLD will be cancelled.
SET-UP Mode : Save each set-up value.

8
5. Rear Panel
Please check the location of terminal and its use. While the projected button is being
pressed, please insert the cable into the lower hole completely. As soon as you release
the button, the connection will be completed. At the point, please slightly pull the calbe
and check whether the cable is come off or not.
(The most suitable calbe is Φ0.5~1. Linking cable must be brazing or used with 1
terminal)
①AC_L, N : Main Power Supply Wiring Terminal
②: Ground Terminal (as an independent ground connection.)
③E1+ : CH1 Sensor Supply Voltage + Connection Terminal(FED)
④E1- : CH1 Sensor Supply Voltage - Connection Terminal(White)
⑤S1+ : CH1 Sensor Ouput Signal + Connection Terminal(Black)
⑥S1- : CH1 Sensor Output Signal –Connection Terminal(Green)
⑦GND : SHIELD connection terminal of CH1 sensor(Tare)
⑧E2+ : CH2 Sensor Supply Voltage + Connection Terminal(Red)
⑨E2- : CH2 Sensor Supply Voltage - Connection Terminal(White)
⑩S2+ : CH2 Sensor Ouput Signal + Connection Terminal(Black)
⑪S2- : CH2 Sensor Output Signal –Connection Terminal(Green)
⑫GND : SHIELD connection terminal of CH2 sensor(Tare)

9
⑬OUT1, 2 : Analog (DC 0 ~10V/DC 4 ~ 20mA) Output Terminal
⑭GND : External Input Common Terminal
⑮IN1 : External HOLD Input Terminal (CH1)
⑯IN2 : External ZERO Input Terminal (CH1)
○
17 IN3 : External HOLD Input Terminal (CH2)
○
18 IN4 : External ZERO Input Terminal (CH2)
○
19 COM : RELAY Output Common Terminal
○
20 RY1 : CH1 Lower Limit Output Terminal
○
21 RY2 : CH1 Upper Limit Output Terminal
○
22 RY3 : CH2 Lower Limit Output Terminal
○
23 RY4 : CH2 Upper Limit Output Terminal
◀Attention▶
1. When wiring, please pull out the power plug. (Main power supply)
2. Please use the thick calbe for ground terminal ( ) so that you can avoid any trouble
due to the impulse voltage or surge. If possible, please keep it as an independent
ground.
(This must be used in the area with heavy noise. If the ground is made with other
device, it will be affected by noise.)
3. Please check the function of the terminal before you do the wiring so that you can
prevent any malfunction in advance.
4. Manufacture will be relieved of its responsibility for any damge or injury due to the
disassembly and alteration made without any agreement. Also, no A/S will be made.

10
6. Wiring Diagram
<Wiring Diagram>
<Example - External Input PLC Connection>

11
7. Components & Function
7-1. Flow Chart
Sensor input
A/D conversion
1000 times/s
Fixed filter
Average : 32 times
Output count : 32 times/s
Display filter
Display update rate : 1~32times/s
Average : 1 ~ 32times/s
Display peak hold
Display D/A conversion
1 ~32 times/s
Comparator
OP-02(RS232C)
OP-03(RS485)
Display
Analog output
Relay output
Serial output

12
7-2. How to use Hold mode
Hold mode is largely divided into Peak Hold and Sample Hold. Please select Analog,
Digital or Display Hold as per your purpose of use (Hi or Low Speed)
To input Hold, you can use Hold key on the front panel or external input. For the
operation method, please refer to the drawing below.
1) Peak Hold : To Hold the maxium value among the measured values.
Peak hold mode : It is a high-speed Peak hold. It is to hold and display the
maximum value of (+) direction.
Absolute peak hold mode : It is to hold and display the the maximum value of the
absolute value (+/-).
2) Sample hold : It is to hold and display the value at the time of Hold signal input
among the measured values.
On
Off
On
Off
On
off
①−:sensor input value
②−: Display value &
analog output value
Front panel HOLD LED
External HOLD input termanal
Front panel HOLD Key
Measured value
time
①
②
Measured value
time
①
②
<Peak HOLD> <Sample HOLD>

13
7-3. How to use comparison ouput function
For comparison output function, there are 3 different modes such as Decision, High
limit, Low limit. It displays through the relay of rear panel comparing each setup value.
On High limit and Low limit mode, Hysteresis can be used.
To set up upper limit (High) and lower limit(Low), please use the key on the front panel.
1) Decision mode :
Measured Value ≤Lower limit setup value ⇒CH1 : RY1 ON, CH2 : RY3 ON
Measured Value ≥Upper limit setup value ⇒CH1 : RY2 ON, CH2 : RY4 ON
2) High limit mode :
Measured Value ≥RY1 setup value ⇒CH1 : RY1 ON, CH2 : RY3 ON
Measured Value ≥RY2 setup value ⇒CH1 : RY2 ON, CH2 : RY4 ON
Measured Value < RY1 setup value
–
Hysteresis value⇒CH1:RY1 OFF, CH2:RY3 OFF
Measured Value < RY2 setup value - Hysteresis value⇒CH1:RY2 OFF, CH2:RY4 OFF
3) Low limit mode :
Measured Value ≤RY1 setup value ⇒CH1 : RY1 ON, CH2 : RY3 ON
Measured Value ≤RY2 setup value ⇒CH1 : RY2 ON, CH2 : RY4 ON
Measured Value > RY1 setup value + Hysteresis value⇒CH1:RY1 OFF, CH2:RY3 OFF
Measured Value > RY2 setup value + Hysteresis value⇒CH1:RY2 OFF, CH2:RY4 OFF

14
8. Setting Modes
8-1. Types of Setting mode & Set-up
For setting mode, there are 4 different types of mode such as Function mode, Digital
calibration mode, Actual load calibration mode and SPAN constant calibration mode.
Mesuring mode
While key is being
Press, press key
Function mode
END
①
Digital calibration
mode
END
②
END
③
Actual load
calibration mode
①Function mode
Access to each function setup mode. Please refer 8-2 function mode.
②Digital calibration mode
It is to calibrate into the sensor’s output value. No need to prepare for the actual
load (stand weight). Please refer 8-3 how to calibrate.
③Actual load calibration mode
It is to calibrate by adding the actual load (standard weight or the load you know).
Please refer 8-4 how to calibrate.
④SPAN constant calibration mode
It is to calibrate with the S.CAL value written down for load calibration. Please
refer 8-5 how to calibrate.
END
④
SPAN constant
calibration mode

15
8-2. Function mode
1) How to set function.
Measuring mode
While key is being, Press, press
key
END

16
1) How to set RELAY data
key : Lower limit setup key
key : Upper limit setup key
Measuring mode
Press key or press key
END
Input value

17
# Function mode list #
Name
Function
Setting
출고시 기준설정값
F-01
Decimal point
0, 1, 2, 3
1
F-02
Division
1, 2, 5, 10, 20, 50
1
F-03
Display filter
0, 4, 8, 16, 32
8
F-04
Hold mode
Sample hold, Peak hold, Absolute peak hold
Peak hold
F-05
Comparison mode
Decision, High limit, Low limit
Decision
F-06
Hysteresis
0 ~ 99
0
F-08
DAC capacity
-19999 ~ +99999
10000
F-09
ID Number
0 ~ 32
0
F-10
Baud rate & PRINT
2400, 4800, 9600, 19200, 38400, 57600
9600
F-11
Auto zero tracking
0 ~ 99
0
F-12
Auto zero tracking time
0.0 ~ 5.0 sec
0.0
F-13
Force unit
Kg(kg/㎠), N, ℓb, bar, MPa
Kg
F-14
Key disabling
Zero key, Lo key, Hi key, Hold key
0000
F-01. Decimal point (Decimal point set up)
(Standard setup value: 1)
Display data
Setting
Remark
0
00000
: No decimal point
CH1, CH2
Individual setup
1
0000.0
: One decimal place
2
000.00
: Two decimal places
3
00.000
: Three decimal places

18
F-02. Division (Minimum display unit set up)
(Standard setup value: 1)
Display data
Setting
Remark
1
Displayed In 1
(0, 1, 2, 3, 4 …….)
CH1, CH2
Individual setup
2
Displayed in 2
(0, 2, 4, 6, 8 …….)
5
Displayed in 5
(0, 5, 10, 15 …….)
10
Displayed in 10
(0, 10, 20, 30 …….)
20
Displayed in 20
(0, 20, 40, 60 …….)
50
Displayed in 50
(0, 50, 100, 150 …….)
F-03. Display filter (Display speed set up)
(Standard setup value: 8)
Display data
Setting
Remark
0
No filter
Only CH1
4
Average time 1/8 sec
8
Average time 1/4 sec
16
Average time 1/2 sec
32
Average time 1 sec
F-04. Hold mode
(Standard setup value: 1)
Display
data
Setting
Remark
0
Sample Hold
: To hold the display value at the time of Hold signal
input.
CH1
setup
1
Peak Hold
: To hold the maximum value of display values during
Hold signal input.
2
Absolute Peak Hold
: To hold the maximum absolute value of display values
during Hold signal input.
F-05. Comparision mode (Comparision output mode set up)
(Standard setup value: 0)
Display data
Setting
Remark
0
Decision mode
: Relay RY1, RY2, RY3, RY4 output
CH1, CH2
Individual
setup
1
High limit mode
: Relay RY1, RY2, RY3, RY4 output
2
Low limit mode
: Relay RY1, RY2, RY3, RY4 output

19
F-06. Hysteresis
(Standard setup value: 00)
Display data
Setting
Remark
00
00
: Hysteresis –not used
CH1,CH2
Individual
setup
~
99
01 ~99
: Hysteresis –used(Decision mode is not applied)
F-08. DAC capacity (Analog Outupt value set up)
(Standard setup value: 10000)
Display
data
Setting
비고
-19999
Analog output capacity set-up
<Setting value & output>
Setting
value
Voltage output(±10V)
Current output(4~20mA)
Display value
output
Display value
+10000
-10000
-10V
-10000
-
0
0V
0
4mA
+10000
+10V
+10000
20mA
-10000
-10000
+10V
-10000
20mA
0
0V
0
4mA
+10000
-10V
+10000
-
1
CH1, CH2
Individual
setup
~
+99999
F-09. ID Number (Communication Device Number set up)
(Standard setup value: 00)
Display data
Setting
Remark
00
00
: Device number is not set-up(Stream mode:always transmit data )
CH1
setup
~
32
01~32
: Device number is set-up(Command mode:Transmit data by command)
CH2 ID Number : CH1 ID Number + 1
F-10. Baud rate & Print (Communication Speed set up)
Display data
Setting
Stream mode
Command mode
비고
2.40
2400 bps
O
O
CH1
setup
4.80
4800 bps
O
O
9.60
9600 bps
O
O
19.20
19200 bps
O
X
38.40
38400 bps
O
X
57.60
57600 bps
O
X
Table of contents