ECH2O Check
Quick Start Guide

Select your probe type:
Press and hold the left (I) button,
then press the right (II) button
repeatedly to cycle through types:
E5E5
E5E5
E5is for the EC-5 probe
10H10H
10H10H
10H is for the 10HS probe
E10E10
E10E10
E10is for the EC-10 probe
E20E20
E20E20
E20is for the EC-20 probe
USrUSr
USrUSr
USris for user-set calibration
 Press the right (II) button to
scroll through display modes:
*PP
PP
Pctct
ctct
ct: percent water content (cm/m)
*iPfiPf
iPfiPf
iPf: inches per foot (water content)
AA
AA
Adcdc
dcdc
dc:Analog/digital converter value
ECH2O Check
Quick Start Guide

Select your probe type:
Press and hold the left (I) button,
then press the right (II) button
repeatedly to cycle through types:
E5E5
E5E5
E5is for the EC-5 probe
10H10H
10H10H
10H is for the 10HS probe
E10E10
E10E10
E10 is for the EC-10 probe
E20E20
E20E20
E20 is for the EC-20 probe
USrUSr
USrUSr
USris for user-set calibration
 Press the right (II) button to
scroll through display modes:
*PP
PP
Pctct
ctct
ct: percent water content (cm/m)
*iPfiPf
iPfiPf
iPf: inches per foot (water content)
AA
AA
Adcdc
dcdc
dc:Analog/digital converter value
ECH2O Check
Quick Start Guide

Select your probe type:
Press and hold the left (I) button,
then press the right (II) button
repeatedly to cycle through types:
E5E5
E5E5
E5is for the EC-5 probe
10H10H
10H10H
10H is for the 10HS probe
E10E10
E10E10
E10is for the EC-10 probe
E20E20
E20E20
E20is for the EC-20 probe
USrUSr
USrUSr
USris for user-set calibration
 Press the right (II) button to
scroll through display modes:
*PP
PP
Pctct
ctct
ct: percent water content (cm/m)
*iPfiPf
iPfiPf
iPf: inches per foot (water content)
AA
AA
Adcdc
dcdc
dc:Analog/digital converter value
ECH2O Check
Quick Start Guide

Select your probe type:
Press and hold the left (I) button,
then press the right (II) button
repeatedly to cycle through types:
E5E5
E5E5
E5is for the EC-5 probe
10H10H
10H10H
10H is for the 10HS probe
E10E10
E10E10
E10is for the EC-10 probe
E20E20
E20E20
E20is for the EC-20 probe
USrUSr
USrUSr
USris for user-set calibration
 Press the right (II) button to
scroll through display modes:
*PP
PP
Pctct
ctct
ct: percent water content (cm/m)
*iPfiPf
iPfiPf
iPf: inches per foot (water content)
AA
AA
Adcdc
dcdc
dc:Analog/digital converter value
ECH2O Check
Quick Start Guide

Select your probe type:
Press and hold the left (I) button,
then press the right (II) button
repeatedly to cycle through types:
E5E5
E5E5
E5is for the EC-5 probe
10H10H
10H10H
10H is for the 10HS probe
E10E10
E10E10
E10 is for the EC-10 probe
E20E20
E20E20
E20 is for the EC-20 probe
USrUSr
USrUSr
USris for user-set calibration
 Press the right (II) button to
scroll through display modes:
*PP
PP
Pctct
ctct
ct: percent water content (cm/m)
*iPfiPf
iPfiPf
iPf: inches per foot (water content)
AA
AA
Adcdc
dcdc
dc:Analog/digital converter value
ECH2O Check
Quick Start Guide

Select your probe type:
Press and hold the left (I) button,
then press the right (II) button
repeatedly to cycle through types:
E5E5
E5E5
E5is for the EC-5 probe
10H10H
10H10H
10H is for the 10HS probe
E10E10
E10E10
E10is for the EC-10 probe
E20E20
E20E20
E20is for the EC-20 probe
USrUSr
USrUSr
USris for user-set calibration
 Press the right (II) button to
scroll through display modes:
*PP
PP
Pctct
ctct
ct: percent water content (cm/m)
*iPfiPf
iPfiPf
iPf: inches per foot (water content)
AA
AA
Adcdc
dcdc
dc:Analog/digital converter value
ECH2O Check
Quick Start Guide

Select your probe type:
Press and hold the left (I) button,
then press the right (II) button
repeatedly to cycle through types:
E5E5
E5E5
E5is for the EC-5 probe
10H10H
10H10H
10H is for the 10HS probe
E10E10
E10E10
E10is for the EC-10 probe
E20E20
E20E20
E20is for the EC-20 probe
USrUSr
USrUSr
USris for user-set calibration
 Press the right (II) button to
scroll through display modes:
*PP
PP
Pctct
ctct
ct: percent water content (cm/m)
*iPfiPf
iPfiPf
iPf: inches per foot (water content)
AA
AA
Adcdc
dcdc
dc:Analog/digital converter value
ECH2O Check
Quick Start Guide

Select your probe type:
Press and hold the left (I) button,
then press the right (II) button
repeatedly to cycle through types:
E5E5
E5E5
E5is for the EC-5 probe
10H10H
10H10H
10H is for the 10HS probe
E10E10
E10E10
E10 is for the EC-10 probe
E20E20
E20E20
E20 is for the EC-20 probe
USrUSr
USrUSr
USris for user-set calibration
 Press the right (II) button to
scroll through display modes:
*PP
PP
Pctct
ctct
ct: percent water content (cm/m)
*iPfiPf
iPfiPf
iPf: inches per foot (water content)
AA
AA
Adcdc
dcdc
dc:Analog/digital converter value
ECH2O Check
Quick Start Guide

Select your probe type:
Press and hold the left (I) button,
then press the right (II) button
repeatedly to cycle through types:
E5E5
E5E5
E5is for the EC-5 probe
10H10H
10H10H
10H is for the 10HS probe
E10E10
E10E10
E10is for the EC-10 probe
E20E20
E20E20
E20is for the EC-20 probe
USrUSr
USrUSr
USris for user-set calibration
 Press the right (II) button to
scroll through display modes:
*PP
PP
Pctct
ctct
ct: percent water content (cm/m)
*iPfiPf
iPfiPf
iPf: inches per foot (water content)
AA
AA
Adcdc
dcdc
dc:Analog/digital converter value
To make a measurement:
Press the left (I) button
*Note: If probe is measured in air
while in “IPF” or “PCT” mode, a
negative value will be displayed.
To adjust calibration, refer to the
operations guide.
To replace the battery, unscrew the
back of case. Battery size: CR2
The ECH2O Check will shut off
automatically when not in use for
more than 5 minutes.
To make a measurement:
Press the left (I) button
*Note: If probe is measured in air
while in “IPF” or “PCT” mode, a
negative value will be displayed.
To adjust calibration, refer to the
operations guide.
To replace the battery, unscrew the
back of case. Battery size: CR2
The ECH2O Check will shut off
automatically when not in use for
more than 5 minutes.
To make a measurement:
Press the left (I) button
*Note: If probe is measured in air
while in “IPF” or “PCT” mode, a
negative value will be displayed.
To adjust calibration, refer to the
operations guide.
To replace the battery, unscrew the
back of case. Battery size: CR2
The ECH2O Check will shut off
automatically when not in use for
more than 5 minutes.
© 2009 Decagon Devices, Inc. 10688-06
© 2009 Decagon Devices, Inc. 10688-06 © 2009 Decagon Devices, Inc. 10688-06
To make a measurement:
Press the left (I) button
*Note: If probe is measured in air
while in “IPF” or “PCT” mode, a
negative value will be displayed.
To adjust calibration, refer to the
operations guide.
To replace the battery, unscrew the
back of case. Battery size: CR2
The ECH2O Check will shut off
automatically when not in use for
more than 5 minutes.
To make a measurement:
Press the left (I) button
*Note: If probe is measured in air
while in “IPF” or “PCT” mode, a
negative value will be displayed.
To adjust calibration, refer to the
operations guide.
To replace the battery, unscrew the
back of case. Battery size: CR2
The ECH2O Check will shut off
automatically when not in use for
more than 5 minutes.
To make a measurement:
Press the left (I) button
*Note: If probe is measured in air
while in “IPF” or “PCT” mode, a
negative value will be displayed.
To adjust calibration, refer to the
operations guide.
To replace the battery, unscrew the
back of case. Battery size: CR2
The ECH2O Check will shut off
automatically when not in use for
more than 5 minutes.
© 2009 Decagon Devices, Inc. 10688-05
© 2009 Decagon Devices, Inc. 10688-06 © 2009 Decagon Devices, Inc. 10688-05
To make a measurement:
Press the left (I) button
*Note: If probe is measured in air
while in “IPF” or “PCT” mode, a
negative value will be displayed.
To adjust calibration, refer to the
operations guide.
To replace the battery, unscrew the
back of case. Battery size: CR2
The ECH2O Check will shut off
automatically when not in use for
more than 5 minutes.
To make a measurement:
Press the left (I) button
*Note: If probe is measured in air
while in “IPF” or “PCT” mode, a
negative value will be displayed.
To adjust calibration, refer to the
operations guide.
To replace the battery, unscrew the
back of case. Battery size: CR2
The ECH2O Check will shut off
automatically when not in use for
more than 5 minutes.
To make a measurement:
Press the left (I) button
*Note: If probe is measured in air
while in “IPF” or “PCT” mode, a
negative value will be displayed.
To adjust calibration, refer to the
operations guide.
To replace the battery, unscrew the
back of case. Battery size: CR2
The ECH2O Check will shut off
automatically when not in use for
more than 5 minutes.
© 2009 Decagon Devices, Inc. 10688-06
© 2009 Decagon Devices, Inc. 10688-06 © 2009 Decagon Devices, Inc. 10688-06

Other Decagon Accessories manuals

Decagon 10HS User manual

Decagon

Decagon 10HS User manual

Decagon EC-5 User manual

Decagon

Decagon EC-5 User manual

Decagon 10HS User manual

Decagon

Decagon 10HS User manual

Popular Accessories manuals by other brands

Dynafit ST RADICAL TURN quick guide

Dynafit

Dynafit ST RADICAL TURN quick guide

Quark-Elec AS06 user manual

Quark-Elec

Quark-Elec AS06 user manual

National Instruments WSN-3230 USER GUIDE AND SPECIFICATIONS

National Instruments

National Instruments WSN-3230 USER GUIDE AND SPECIFICATIONS

NeuLog NUL-204 Guide

NeuLog

NeuLog NUL-204 Guide

Rancilio iSteam specification

Rancilio

Rancilio iSteam specification

Emerson Rosemount 3300HT instruction manual

Emerson

Emerson Rosemount 3300HT instruction manual

Outdoor Revolution Journey 260 instructions

Outdoor Revolution

Outdoor Revolution Journey 260 instructions

MESTIC MAC-40 Instructions for use

MESTIC

MESTIC MAC-40 Instructions for use

Velux MMA FS manual

Velux

Velux MMA FS manual

ipf electronic OTQ80576 manual

ipf electronic

ipf electronic OTQ80576 manual

BEMKO TOKAR PLD-260-120 installation instructions

BEMKO

BEMKO TOKAR PLD-260-120 installation instructions

Vega VEGAPULS 63 operating instructions

Vega

Vega VEGAPULS 63 operating instructions

Advantech Advantech Modem ADAM-4550 user manual

Advantech

Advantech Advantech Modem ADAM-4550 user manual

GP U421 instruction manual

GP

GP U421 instruction manual

PRESONUS Quantum 4848 owner's manual

PRESONUS

PRESONUS Quantum 4848 owner's manual

QGenie QG-103N manual

QGenie

QGenie QG-103N manual

Panasonic HL-C211F-MK instruction manual

Panasonic

Panasonic HL-C211F-MK instruction manual

Harris Farms Nurture Right 360 instructions

Harris Farms

Harris Farms Nurture Right 360 instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.