DentalEZ StarDental Titan-E User manual

866.DTE.INFO (383.4636)
www.DentalEZ.com
0344
StarDental®
1816 Colonial Village Lane
Lancaster, PA 17601 U.S.A.
Printed in the U.S.A.
©2019 DentalEZ, Inc.
Emergo Europe
Prinsessegracht 20
2514 AP The Hague
The Netherlands
EC REP
StarDental® Instruments
Caution: Federal law restricts this device to
sale by or on the order of a dentist.
065810 E Date of Issue 2019-05-24
Titan®-E Low-Speed
Handpiece Motor
Instruction Manual
Non-Sterile
Medical Device
Contents
Specications
20,000 rpm Motor: StarDental®Part #266131 (Fixed Back End Part #266132)
•NoLubricationRequired
•RecommendedAirPressure:40-45psi(swivelmotor)
•AirFlow:2.1SCFM@40-45psiatMaximumPowerPoint(swivelmotor)
•Rotation:ForwardandReverse
•4-LineConnection(FixedBackEnd)
Installation
1. Selectthemotorappropriatefortubingbasedonpowerrequirements.Examineconnection
ofthemotorandswiveltubingcongurationasthesemustmatchcongurationofdelivery
systemtubing.
DriveAir Water
Exhaust
Air Air
4-Line Tubing 2/3-Line Tubing
Drive
Air
Water
•Specications
•Installation
•Operation
•Cleaning&Sterilization
•Troubleshooting
•ReplacementPartNumbers
•Warranty
Sterilizableupto135°C
135°C
Titan E
20,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
®
StarDental
®
20KMotor
(StarDental Part #266131) Swivel
BEFORE OPERATING THE MOTOR, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS. SAVE
THESE INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL WARNINGS AND CAUTIONS.
Indications For Use
Usedintraorallybytraineddentalprofessionalsfordrillingandpreparationofdentalcavities
forrestoration,suchasllings.Theintendeduseextendstodisking,cavityandcrownpreps,
polishing,postandpindrillingandpinsetting.
Swivel
BearingAssembly
ISOConnector
Tubing
Coupler
Tubing
2. Disconnect swivel from handpiece if necessary. Grasp the swivel in front of the bearing
assemblybetweenthethumbandindexngerasshown.
3. SlidetubingcoupleruntiltheendoftheISOconnectorisexposed;holdinhandasshown
withngerssecuringthetubing.
4. AlignandinsertswiveltubesintotheISOconnectorholes.
5. Whileapplyingslightpressuretoholdttingsintheconnectorholes,slidecouplerforward
andengagethreads.Continuethreadingcouplerclockwiseuntilhandtightagainsttheswivel.
Wrench Tubing
Coupler TurnCouplerWhileHolding
SwivelinPlace
6. Usingthewrenchasshown,tighten theswivel to thetubing coupler.Besuretoturntubing
coupler,notswivel.DONOTOVERTIGHTEN.
Note: Thisjointloosensovertime,resultinginlossofpower.Checkfortightnessperiodically.
7. Connecthandpiecetoswivel.
8. Usinganairpressuregaugefor4-line,StarDentalPart#262591,betweenswivelandtubing,
obtainareading.Adjusttheairpressureatdeliveryhead until gaugereads40-45 psiwhile
operatinghandpieceinfullforwardpositionandwithfootcontrolcompletelydepressed.
Note:Ifdeliveryunithasoil/mistlubrication,detachanddrain;handpieceneedsnolubrication.
Installation - Fixed Back End Motor
Titan E
20,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
®
StarDental
®
Direction&Speed
ControlRing
RearEndConnector
20KFixedBackEndMotor
(StarDental®Part #266132)
Note:Ifdeliveryunithasoil/mistlubrication,detachanddrain;motorisLUBEFREE.
NOTICE Topreventdamagetobearings,beforeinstallingalwayscheckdriveairpressureby
usingairpressuregauge,StarDentalPart#262591(4/6-Line)betweenmotorandtubingand
obtainareading.Adjustairpressureatdeliveryheaduntilgaugereads40-45psiwhileoperating
motorandwithfootcontrolcompletelydepressed.
1. Examinetubing.Verifytubinguses4-Lineconnection.
2. Toconnectmotortotubing,aligndriveairandexhausttubeswithmatingISOconnectorholes
incoupler.
3. Whileapplyingslightpressuretoholdtubinginconnectorholes,slidetubingcouplerforward
andengagethreads.Continuethreadingtubingcouplerclockwiseuntil
handtightagainstthemotor.DONOTOVERTIGHTEN.
Note:Thisjointloosensovertime,resultinginairleakageandlossofpower.Checkfor
tightnessperiodically.
Operation
The120°rotationspeedadjustmentringsetsdirectionandspeed.Torunmotorforwardatfull
speed,rotateringtowardstheletter‘F’whileoperatingthefootcontrol.Rotatetotheletter‘R’to
yieldfullspeedinreverse.Varyspeedbyrotatingringineitherdirectiontowardscenter.Note
thesmalldepressionforthecurrentsetting.Aftersetting,themotorrunsatcurrentspeedand
directionuntilreset.Pausingthemotorisnotneededtoalterdirectionorspeedusingthering.
Titan E
20,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
®
StarDental
®
SmallDepression
120°RotationSpeed
AdjustmentRing
ATTACHMENTS
TheTitan-E20KMotorisdesignedtoworkwithallstandardE-Typeattachments.
Titan E
20,000 rpm- Lube Free
Made in the U.S.A.
JXXXXX
®
StarDental
®
StraightNoseCone
Motor-to-Angle
Contra-AngleAttachment
20KTitan-EMotor
ANGLE INSTALLATION
StarDental Titan angles can be attached to E-Type Motor-to-Angle
Contra-AngleAttachments
.
ToinstallaTitanAngle:
1. Unthreadknurledring.
2. Pullknurledringenoughtoshowsplines,remove
knucklefromangle*. Splines
Knuckle
Knurled Ring
Titan 3 LubeFree Ball Bearing
Auto Latch Angle Head
*Note:Savetheknuckleinasafeplace.Youwill
needtheknuckletouseanglewithTitan
attachments.
StraightNoseCone:
Attachtomotor.Installburaccordingtomanufacturer'sinstructions.
NOTICE AnyseriousincidentoccurringfromtheuseofthisdeviceshouldbereportedtoStarDental
immediately.IfincidentoccurredinaEUMemberStatethecompetentauthoritymust
benotied.

3. CarefullymeshthesplinesontheanglewiththesplinesoftheMotor-to-AngleContra-Angle
Attachment
(seeillustrationbelow).
4. Tightenknurledringontothreadsuntilsnug.
5. Installburaccordingtotheangle’sinstructionmanual.
7LWDQ (
USP/XEH)UHH
0DGHLQWKH86$
-;;;;;
6WDU'HQWDO
Splines
Angle
Motor-to-Angle
Contra-AngleAttachment
Titan-EMotor
TransmissionGear
CAUTIONWhenangledoesnotseatonattachment,gearsarenotmeshedcorrectly.Remove
angle;spinangletransmissiongearandrepeatStep3.
Titan E
®
StarDental
®
Internal
Surfaces
SwivelBearing/
SpringAssembly
Clean
Cleaning & Sterilization
CAUTION
INSTRUCTIONS
Maintenance &Lubrication: TheTitan®-EMotorisLubeFreeand,therefore,requiresnolubrication.Motorshouldberoutinely
maintained,checkedforcorrosion,wornparts,etc.andshouldbereplacedonanasneededbasis.
WARNING
Donotexceed135°C
Donotuse:
Limits on Reprocessing
Thehandpiecemotormustbecleanedandsterilizedbeforerstuseandbetweenpatientuse.
Cleaning-Manual: TheTitan®-EMotorisLubeFreeand,therefore,requiresnolubrication.
Attimesdebrisfromthedeliverysystem’sdriveairtubingcollectsinthemotor.Thisbuild-up
maycausesluggishmotoroperation.Toease,place3dropsofDentaLube II,StarDental®Part
#262539intothedriveairtubeofmotor,andalternatelyoperatethemotorinforwardand
reverse.
Cleaning the Swivel, the Mating Motor Part
Cleanallinternalsurfacesatrearofhandpiecemotorwithacottonswabsoakedin
isopropylalcoholuntilclean.
Cleanallexternalsurfacesusingasmalltoothbrushwithisopropylalcohol.
Tooptimizeoperationoftheswivelconnector,cleanallinternalsurfacesofthemotorwith
acottonswabsoakedinisopropylalcoholuntilclean.Cleanareaaroundbearing/spring
assemblyatrearofswivelusingisopropylalcoholandatoothbrush.Drybothusingsyringeair.
Cleaning-Automated: Washer-disinfectorsarenotrecommended.
Wrapped Instrument
Sterilization:
GravityTypeAutoclave:Pre-VacuumAutoclave:
•Minimumof15minutesat132°C•Minimumof4minutesat132°C
•Minimumof30minutesat121°C
Donotexceed135°C
•dryheatsterilizer
•immersiontechniques
•coldsterilizers •ultrasoniccleaners
•sterilizerbagscontainingdisinfectants •handsoaps
•detergents
•chemicaldisinfectors
•surfacesprays •surgicalmilk
•surfacewipes
Additional Information: Do not exceed the sterilizer manufacturer’s guidelines for load capacity. Use only paper or
paper/plasticbags.
Storage: Storeinbagaftersterilizationuntilreadytouse.
Repeatedprocessinghasminimaleffectontheseattachments.Endoflifeisnormally
determinedbywearanddamageduetouse.
Drythemotorfor30minutesminimumusingthesterilizationdrycycleandallowthemotorto
cooltoroomtemperatureinbagpriortouse.Ifmoistureispresent,increaseyoursterilizerdry
cycleappropriately.
Drying:
Containment andtransportation: Itisrecommendedthathandpiecesarereprocessedassoonasisreasonablypracticalfollowinguse.
Preparation for cleaning: Wearpuncture-resistantglovesandprotectiveeyewear.Aftereachpatient,ushairbyrunning
handpiecemotorfor20seconds.Removemotorfromswivelifapplicableortubing.
Usesteamsterilizationexclusivelyforcompletesterilityasoutlinedbelow.
Disinfection:
SeeInstallation,OperatingRecommendations,and/orTroubleshootingsectionsofthismanual.
Inspection & function testing:
Single:Astandardsterilizationpackagingmaterialmustbeused.Ensurethatthepackislarge
enoughtocontaintheitemwithoutstressingtheseals.
Packaging:
Manufacturer Contact: Seerstorlastpanelsofthismanual.
StarDental®
Instrument Solutions
StarDental®
Instrument Solutions
7LWDQ (
6WDU'HQWDO
7LWDQ (
6WDU'HQWDO
Drive
Air
RedTip
Applicator
ItistheresponsibilityoftheusertouseonlyFDAcompliantsterilizingequipmentandaccessories
thatarecapableofmeetingthemanufacturersrecommendationsbelow.
CAUTION
Troubleshooting
Problem: LACK OF POWER
Cause: Airpressure
Correction: Verifyproperhandpieceairpressure.RefertoInstallation.
Cause: Crimpedtubing
Correction: Straightentubinginsideandoutsidedeliveryhead.
Cause: Swivelnotthreadedfullyontotubing
Correction: RefertoInstallation.
Cause: Brittlemotororswivelsiliconerubberseal
Correction: Replacesiliconerubbersealfor2/3-lineswivelwithStarDentalPart#203103
and4-linewithStarDentalPart#202897.
Cause: SwivelO-ringsworn
Correction: ReplaceswivelO-rings,StarDentalPart#259884.
Cause: Debrisbuild-upinhandpiece
Correction: Cleanhandpiece.RefertoCleaning.
Problem: SWIVEL WILL NOT DISENGAGE FROM HANDPIECE
Cause: Build-upofdebris
Correction: Place3dropsofDentaLube II,StarDentalPart#262539,intotheswivel/tubing
connector.Rotateswiveltocirculatelube.Disengagehandpiece.Contactdealer
ifunabletodisengageswivel.Topreventreoccurrence,refertoCleaning.
Note: Ifaproblempersists,contactdealerforfurtherassistanceorcontactStarDental technical
serviceat1-866-DTE-INFO.
AlwaysinsistonStarDentalpartsforbestperformance.
Replacement Part Numbers
Titan 3 Titan T
Description Item Number Item Number
AirPressureGaugefor4/6-Line.....................................................262591............................................................262591
SiliconeRubberSeal,4-Line(Pkg.of3).......................................202897...........................................................202897
DentaLube II............................................................................................262539...........................................................262539
ACCESSORIES
Swivel,Motor2/3-Line.......................................................................260152............................................................260152
Swivel,Motor4-Line............................................................................260144............................................................260144
LubeFreeBallBearingAutoLatchAngle......................................264146........................................................... 263950
LubeFreeBallBearingManualLatchAngle................................262738........................................................... 263948
LubeFreeBallBearingAutochuckFrictionGripAngle..................264151........................................................... 263954
LubeFreeBallBearingManualChuckFrictionGripAngle...........262740
BallBearingAutoLatchAngle..........................................................264182
BallBearingManualLatch.................................................................259005...........................................................263947
BallBearingAutochuckFrictionGripAngle...............................264149
TitanEndodonticAngle,AutoLatch..............................................262570............................................................ 262570
TitanEndodonticAngle,ManualLatch........................................262571............................................................ 262571
For new products, manuals and technical information, call
1-866-DTE-INFO or visit our web site at www.dentalez.com.
Important Notice About
Returning Product To The DentalEZ Inc.
Anyone wishing to return any DentalEZ Inc. product to a DentalEZ facility MUST obtain the proper
RETURN AUTHORIZATION NUMBER from the Customer Service Department. The RA Number must
appear clearly marked on both the outside and inside of the returned product carton. When calling for
a return authorization number you will be asked to provide the following information:
To Obtain A
Return Authorization
Number Call
1-866-DTE-INFO
Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation
with this procedure will help us expedite processing of any returns.
• Product Name, Model Number or Part Number
• Serial Number of Product
• Dealer Return Purchase Order Number
• Reason For Return
• Copy of Invoice Showing Purchase
Limited Warranty
O-RingsandSealsarespecicallyexcludedfromthiswarranty.
DentalEZ®warrantstheTitan
®-EMotorfor1yearfromthedateofpurchase*.
Pleasenotethefollowingadditionaltermsofourwarrantypolicy:
• Warrantiescovermanufacturingdefectsonlyanddonotcoverdefectsresultingfrom
abuse,improperhandling,cleaning,careormaintenance,normalwearandtearor
non-observanceofoperating,maintenanceorinstallationinstructions.Failure to use
authorized parts or an authorized repair facility voids this warranty.
• Liability is limited to repair or replacement of the defective product at our sole
discretion. All other liabilities, in particular liability for damages, including,
without limitation, consequential or incidental damages are excluded.
• THISWARRANTYISINLIEUOFALLOTHERWARRANTIES,EXPRESSEDORIMPLIED,
INCLUDINGANYIMPLIEDWARRANTIESOFMERCHANTABILITYORFITNESSFORA
PARTICULARPURPOSE.NOEMPLOYEE,REPRESENTATIVEORDEALERISAUTHORIZED
TOCHANGETHISWARRANTYINANYWAYORTOGRANTANYOTHERWARRANTY.
*Providedconditionsinwarrantyaremet.
WARRANTY REGISTRATION
PLEASEVISITTHEWEBSITEBELOWTOREGISTERYOURSTARDENTALPRODUCT.THISWILL
ENSUREPROPERREGISTRATIONANDPROMPTANDACCURATESERVICESHOULDANYOFYOUR
PURCHASESNEEDEVALUATIONORREPAIRDURINGTHEWARRANTYPERIOD.
StarDental®Products:www.dentalez.com/warranty
This manual suits for next models
2
Popular Engine manuals by other brands

Leroy-Somer
Leroy-Somer Nidec FLSD Series RECOMMENDATIONS FOR INSTALLATION AND MAINTENANCE

MTD
MTD SX800 owner's manual

Willys
Willys MB Technical manual

Novarossi
Novarossi MEPHISTO.21 ON-ROAD manual

Mercury Racing
Mercury Racing HP1200 SCi Dry Sump Operation, maintenance & warranty manual

PCM
PCM DI Catanium CES Owner's operation and maintenance manual