Whenmeasureing point topoint resistance(Rtt):
lPlace twoelectrodesno closer than 2"fromthe
edge ofthe surfacebeing measured.
lPlace twoelectrodesno closer than 3"toany
groundablepoint.
lPlace twoelectrodesabout 10"apart fromeach
other forRtt measurementsofaworksurface.
lPlace twoelectrodesabout 3'apart fromeach
other forRtt measurementofafloor.
Preferableelectrode placementsinclude:
lMost commonlyused areaofasurface
lMost wornarea
lCenter ofsurface
lFurthest areafromagrounded point
Ifthe surface tobe measured has sections (i.e. floor
tiles, garment panels), place twoelectrodes on
different sections for Rtt measurements.
Clean the material’ssurface for test lab
measurements, but do not clean thesurface for
materialsthat arealreadyinstalled. Onlyclean and
re-test the installedmaterialiffailureoccurs.
MeasureResistance-to-Ground (Rtg)
Test Procedureinaccordance withANSI/ESDS4.1
section6.4PeriodicWorksurface Testing:
1.Donot cleanthe surface.
2.Removefromthe surfaceonlythose itemsthat
might interferewiththe test. ESD sensitive
devices shallalso beremoved.
3.Connect the blacktest leadtoagrounded point.
4.Connect the red test lead tothe Single-Surface-
Electrode, thenplacethe electrode
convenient point on the surface.
5.Pushthe red pushbutton toperformatest.
Shouldthe 15-secondelectrification period
appear, it maybe bypassedbypushingthe red
pushbutton asecond time.
6.Pushthe right blackpushbutton tosavethe
Measurement ifdesired.
7.Performadditionalmeasurementsbyplacingthe
electrode on the most commonlyusedor worn
area.
SCS - 926 JR Industrial Drive,Sanford,NC 27332
East: (919)718-0000 |West: (909) 627-9634 •Website: StaticControl.com
TB-9092 Page 5of8December 2019©2019 DESCOINDUSTRIES INC
MeasureResistancePoint-to-Point (Rtt) onthe
Surface
1.Donot cleanthe surface.
2.Removefromthe surfaceonlythose itemsthat
might interferewiththe test. ESD sensitive
devices shallalso beremoved.
3.Connect onetest leadtothe Single-Surface-
Electrode, another test leadtothe Dual-Surface-
Electrode’sblackjack, thenplacetwoelectrodes
on the most commonlyused area ofthe surface.
4.The diskside ofthe Dual-Surface-Electrode
shouldbe used inthiskindoftest.
5.The electrodes shouldalsobe 2"awayfromany
edge and 3"awayfromany grounded point. If
the most used area isnot obvious, usetwo
pointsnear the center ofthe surface.
6.Pushthe red pushbutton toperformatest.
Shouldthe 15-secondelectrification period
appear, it maybe bypassedbypushingthe red
pushbutton asecond time.
7. Push the right blackpushbuttonto savethe
measurement ifdesired.
8. Performadditionalmeasurementsbyplacing the
electrode on the most commonlyused orworn
area.
MeasureSurfaceResistanceandResistivity
1.Place the materialtobe tested on an insulative
surface.
2.Place the Dual-Surface-Electrode on topofthe
materialtobe tested.The concentricring side
shouldbe used inthiskindoftest.
3.Connect bothtest leadstothe Dual-Surface-
Electrode’sblackandwhitejacks.
4.Pushthe red pushbutton toperformatest.
Shouldthe 15-secondelectrification period
appear, it maybe bypassedbypushingthe red
pushbutton asecond time.
5.Pushthe right blackpushbutton tosavethe
measurement ifdesired.