DESCO EUROPE - 2A DUNHAMS LANE, LETCHWORTH, HERTFORDSHIRE, SG6 1BE, UK
TB-3040.E Page 3 of 5 © 2018 DESCO INDUSTRIES INC
Employee Owned
Figure 3. Reading the Digital Static Field Meter while in
the ±20kV range
Thousands
Hundreds
Tens
HOLDING THE LAST READING
With the meter positioned 25mm from the object being
measured, press the HOLD button. This will freeze the
reading from the object on the display and the analogue
output signal. This feature allows the operator to move
the meter where it may be more easily read or saved for
later reference.
Note:Theredranginglightswillbeowhilethemeteris
in HOLD mode. It is advised to do this between
measurements to prolong battery life.
ANALOGUE OUTPUT
The analogue output jack labelled “OUT” on the face of
the meter accepts a standard 2.5mm monaural phone
plug and is provided so the output of the Digital Static
Field Meter may be connected to an oscilloscope,
strip chart recorder, external meter or other device.
The voltage at this output is 1/1000th (±2kV range)
or 1/10,000 (±20kV range) of the measured voltage.
Contact Customer Service for more information.
BATTERY REPLACEMENT
The Digital Static Field Meter operates from a standard
9 VDC alkaline battery. Battery life is in excess of 50
hours under normal use. When the battery voltage drops
below 6.5V, “BAT” will appear on the display. To change
the battery, slide the battery cover down at the back of
the Meter and remove the battery from the battery clip.
Replace the battery with a fresh one and reinstall the
battery cover. The battery should be removed from the
Meter if its is to be stored for an extended period of time.
Maintenance
The Digital Static Field Meter is factory calibrated and no
maintenance is required. If for any reason you believe
the Meter is not working correctly, please contact Desco
Europe Customer Service. CAUTION - There are no
user serviceable parts. Any unauthorised service will
void the warranty and result in additional repair charges.
Note: This Meter is a precision instrument and should
not be subjected to dropping as damage would not be
covered by the limited warranty.
Figure 4. Reading the Digital Static Field Meter while in
the ±2kV range
Hundreds
Tens
Ones
Note: This aspect ratio should be at least 3 for best
accuracy, i.e. the object should be at least a 3 inch
(75mm) square when measuring at a 1 inch (25mm)
distance. Accurate measurements may be made at other
measurement distances by scaling the meter range and
observing the proper aspect ratio. For example, at a
measurement distance of 3 inch (75mm), multiply the
meter reading by 3 to give a range of 0 to 60 kilovolts.
For accuracy, the object being measured at this distance
should be at least a 9 inch (230mm) square.