Diapro T?bbi Ürünler Across System User manual


Incubator
Version 1.0

Foreword
Thank you for purchasing our Products: Card Incubator , This
Manual for users contains function and operation of the Instrument.
In order to use the instrument properly, please read this manual
carefully before using the Instrument.
Opening Check
Please check the Instrument and Appendix with the packing list
when you first open the instrument packing case. If you find there
is something wrong with the Instrument and the Appendix, do
contact the vendor or the producer.

Contents
1. Introduction………………………………………………......1
1.1 Delivery package…………………………….…………….1
1.2 Structure Description………………………………….......2
1.3 Installing the device………………………………….…....3
2. Technical data……………………………………………….4
3. Safety precautions……………..……………………….......5
4. Operation guide..…………………………………………….6
4.1 Control elements……….…………………………………..6
4.2 Setting the timing time……... ..…………………………...6
4.3 Setting the temperature…………………………..……....6
4.4 Switching to the Fahrenheit display……………………..7
4.5 Start and Stop……………………………………………..7
4.6 Temperature calibration…………………………………..8
5.Maintenance and cleaning…………………………………10
6. Troubleshooting……………………………......................10
Annex 1 Wiring Diagram for Incubator ………………….11
Packing list………………………………………………….12

1 Introduction
This Incubator designed for workstations in training and research
laboratories in the fields of bioscience, medicine and chemistry.
Before starting up this incubator for the first time, please read the
rest of this operations manual.
1.1 Delivery package
Card Incubator 1pcs
Power line 1pcs
Operations manual 1pcs
1

1.2 Structure Description
2
Right Display
Panel
Power Switch
Power connector
lid
Left Block
Right Block
Left Display

1.3 Installing the device
1.3.1 Place the card incubator onto a level , horizontal surface.
1.3.2 Insert the column connector of the power line to power connector of
the device ,and insert another connector of the adapter to mains power
supply.
1.3.3 Power on the main switch. The incubator is ready to operate when
the display becomes visible.
1.3.4 Put the test card into the block, then close the lid.
3
AC220V

2 Technical data
Model K37-24
Power supply AC220V 22V 50Hz
Fuse 250V 3A 5X20
Power supply AC220V 22V 50Hz
Maximum Power 200 W
Temperature range 3550
Timing range 1sec 99min59sec
Accuracy of the temperature 0.5
Display accuracy 0.1
Heating time
(from 20 to 37 )5min
Ambient temperature 530
DimensionsWDH280mm380mm115mm
Weight 6.0 kg
4

3 Safety precautions
This product is a normal and an indoor Instrument.
5
Read the Manual carefully before operation, The expert of wiring
equipment can operate this Instrument.
The operator should not open or repair the Instrument by himself,
which will result in losing the qualification of repair guarantee or
occur accident. If there is some wrong with the Instrument, the
company will repair it.
The Instrument should be put in the place of low temperature, little
dust, no water and no sun or strong lamp. What’s more, the place
should be good aeration, no corrosively gas or strong disturbing
magnetic field, far away from central heating, camp stove and
other hot resource. Don't put the Instrument in wet and dusty place.
Mains switch is on the rear of the device, push “I” to pow
er on the device, and push “O” to power off the device.
Power off when you finish your work. Pull off the connector plug when
there’s long time no use of the Instrument and cover it with a cloth or
plastic paper to prevent from dust.

4. Operation guide
4.1 Control elements
4.2 Setting the timing time
press “” or “” key at the left ,you can set the timing time from 1sec to
99min59sec for the left block. And you can press “” or “”key at the
right ,set the timing time for the right block too.
6
4.3 Setting the temperature
The default setting value of the block is 37.0
,if you need other
temperature. For example, you need the temperature of the left block is 40
,
you can press left “START/STOP” key and left “” key simultaneously.
When press “” or “” key, the units digit will
flicker, then you can modify the timing time, About 6
seconds, the flicker will disappear, exit setting
interface.
Then the units digit will flicker,
you can press “” or “” key to
modify the temperature, About 6
seconds, the flicker will
disappear, exit setting interface.
Temperature
of left block
Time of left
block
Start/stop
the left timer
Setting time
of left block
Temperature
of right block
Time of right
block
Start/stop the
right timer
Setting time
of right block

7
4.4 Switching to the Fahrenheit display
Press “” and “” key simultaneouslythe display switches from
Centigrade to Fahrenheit. Also press “” and “” key simultaneously again,
the display switches from Fahrenheit to Centigrade.
4.5.3 Press “START/STOP” key for 1 seconds continuously to stop
the time .
4.5 Start and Stop
The instrument start to heat automatically according to the previous set
temperature as soon as power on. The default temperature is 37 .
the left “START/STOP” key control the left block and the right “START/STOP ”
key control the right block.
4.5.1 When the temperature of left block reach 37.0 , you can press
left “START/STOP” key to start the timing time of the left block. Also,
When the temperature of right block reach 37.0 , you can press the
right “START/STOP” key to start the timing time of the right block
4.5.2 When the timing time of left block reach 0, the buzzer alarm “du.
du. du…….” and the display appear “OVER’. The controller hold the
temperature too. Press any key to return, and press the “START/ST O
P” key to time again.

4.6 Temperature calibration
The temperature of the instrument has been calibrated before it is sold out.
But if there is deviation between the actual temperature and the displayed
temperature due to some reasons, you can do as follows to correct the error.
The temperature veracity will be within 0.5after the temperature adjustment.
The circumstances temperature should be lower than 35
Adjustment methods as follows:
4.6.1 Put a high precision sensor into the 2.0mm hole in the middle of the left
block , and put another high precision sensor in right block (Make sure the
precision of the sensor should be within 0.1). It is very important that the
sensor must touch the hole of the block fully, maybe need add silicone heat sink
compound in the hole. And must close the lid too. Seeing from Fig a.
Fig a
lid
Block
8
Note! The temperature can be corrected only after the instrument
reaches the set temperature for 20 minutes to ensure the precise of the
temperature.
High precision sensor

4.6.2 Press left “START/STOP ” key and right
“START/STOP” simultaneously, the display come in the
calibration interface. The practical temperature shows 37.0,
if the block do not reach 37.0 C, it will rise to 37.0 at once.
When the practical temperature reach 37.0, the units digit
will flicker ceaselessly. The left display and the right display
come in calibration interface at the same time.
4.6.3 After 20 minutes, if the actual temperature of high
precision sensor in left block is 36.2, then press left
“”or “”key to amend the left display value to 36.2.
4.6.5 Then press left “START/STOP” key to
confirm, it exit the calibration interface. You also
can press right “START/STOP” key to confirm.
After Temperature calibration, the temperature
displayed is the same with the practical temperature
of block.
Note! During Temperature calibration, press left “START/STOP” and right
“START/STOP” key simultaneously to cancel the calibration. The system keeps the
former calibration.
So don’t press left “START/STOP” and right “START/STOP” key
simultaneously unless need calibrate the temperature!
9
4.6.4 If the actual temperature of high precision sensor
in right block is 36.4, then press right “”or “”key to
amend the right display value to 36.4.
Left display
Right display

5 Maintenance and cleaning
The well in the block should be cleaned by the cloth stained with
alcohol to assure good heat translation between the block and
the test tube and no pollution. If there are smutches on the
Instrument, clean them with cloth.
Power off when cleaning the Instrument.
When cleaning the well, don’t drop the cleaning liquid in the well.
Corrosive cleaning liquid is strongly prohibited.
6 Troubleshooting
Error Cause Solution
No display
No main power
connection.
Power failure.
Plug in mains cable on
both sides.
Check the mains fuse
“oPEn” in the
display with the
alarm of “du...”
Broken sensor or
loose contact of
the module
Contact service.
“Shor” in the
display with the
alarm of “du...”
The sensor is
short Contact service.
“hhhh” in the
display with the
alarm of “du...”
The temperature
of the block over
than 60
Contact service.
“ ” in the
display with the
alarm of “du...”
temperature
fluctuation is more
than 1
Contact service.
No heating of the
block Heater failure Contact service.
Press invalid Keyboard failure Contact service.
10

Annex 1 Wiring Diagram for Incubator
11
AC220V
KEY an d D is p l ay
PT1
Power
HEA T1
HEA T2 PT2
MAIN
BOARD
Key an d Di sp l ay
L
PE
N
HEAT PT100
Block1
HEA T PT100
Block2

Packing List
No. Item Type QTY
1Incubator K37-24 1
2Power line 250V 6A 1
3Operation manual 1
12
Other manuals for Across System
1
Table of contents
Other Diapro T?bbi Ürünler Card Reader manuals
Popular Card Reader manuals by other brands

UTC Fire and Security
UTC Fire and Security Interlogix ACL800DIN-RDRMF Installation sheet

HID
HID iCLASS R10 Installation

Texas Instruments
Texas Instruments TRF7960 manual

Panasonic
Panasonic FZ-VSC551 Series operating instructions

Alien
Alien ALR-9680 Hardware setup guide

Nexus
Nexus Cidron Standard Reader installation guide