dlb AIJGO-60 User manual

www.dlb.hu +36 27 502 555sales@destsa.hu
INSTRUCTION
MANUAL
AIJGO-60
WALKING TESTER

TABLE OF CONTENTS
Overview ...................................................3
The product and its accessories ..............................3
Technical datas ..............................................3
About construction and use ..................................4
Parts of the product ............................................4
Button functions ...............................................4
Operation instructions and informations......................5
Connecting accessories .........................................5
Switch on ......................................................5
Walking test ................................................6
Display of measurement results ..............................6
Storing measurement results ................................7
Management of measurement results on computer ...........7
Maintenance ................................................9
Replacement of batteries .......................................9
Calibration .....................................................9

3
OVERVIEW
With an AIJGO-60 Walking tester you can measure the electrostatic charges which
are generated on a person who walks on the oor. The device displays the measured
values on an LCD display in numerical or graphical mode.
CE declaration
We declare that the AIJGO-60 product complies with the requirements of IEC 61340-
5-1, ANSI/ESD S20.20 and Directive 2001/95/EC (General product safety).
THE PRODUCT AND ITS ACCESSORIES
ӹAIJGO-60 Walking tester
ӹHand electrode
ӹHand electrode cable
ӹGrounding cable
ӹ2 pcs crocodile clips
ӹMini USB data cable
ӹPendrive with softwares
ӹ4 pcs AA batteries
TECHNICAL DATAS
Sizes 93 x 185 x 35 mm (W x L x H)
Weight ca. 350 g
Measuring range ±2 kV
Measuring accuracy <±2 kV max. 10%; ±2 kV - ±5 kV Indication
Temperature measuring range 0 °C – 45 °C
Temperature measuring accuracy ±0,3%, max. ±0,5%
Temperature measuring frequency ~ 2 s
Humidity measuring range 10 – 85% RH
Humidity measuring accuracy ±2%, max. ±3,5%
Humidity measuring frequency ~ 2 s
Sample rate 250 µs
Data output Via mini USB-B connection
Display Graphic 128 x 64 pixel
Power supply With 4 pieces of 1,5 V batteries or via USB
cable
Operating time ca. 12-14 hour

1
2
3
4
5
LCD display
On/O button
Hand electrode
jack
Mode button
6
Mini USB socket
Grounding cable jack
ABOUT CONSTRUCTION AND USE
Parts of the product
Button functions
4
On/O button (2) You can turn on/o the device with long
keypress (3 sec.).
With short keypress you can delete the MIN/
MAX voltage values during operation.
Mode button (3) You can switch between the graphical/
numerical display of the charge measurement
and the display of temperature and relative
humidity values with short keypress. With
long keypress (3 sec.) you can store the MIN/
MAX values, if there is empty space in the
measurement list.

5
OPERATION INSTRUCTIONS AND INFORMATION
Connecting accessories
1. Connect the grounding cable’s protected yellow banana plug part to the „5”
point of the device and the other end of the cable to the grounding.
2. Connect the hand electrode cable’s protected green banana plug part to the „4”
point of the device and the other end of the cable to the hand electrode.
Switch on
Turn on the device with the On/O button. After power on it shows a text similar to
the following, with information about the device:
AIJGO-60
WALKING TESTER
S/N: 1234
Ver: v1.02
You can see it for 10 seconds, afterwards the device changes to the numerical display
of the charge measurement values. If you do not want to wait this long, push the
„MODE” button to reach the mentioned numerical display immediately.

6
WALKING TEST
Walk in ESD footwear on ESD oor and keep the hand electrode in hand:
When measuring temperature and humidity, make sure that you do not cover the
sensor on the bottom of the device with your hands or other objects, and do not
place the device on a cooled or heated surface, as this may aect the measurement
results.
DISPLAY OF MEASUREMENT RESULTS
You can see an example for the numerical display of the charge measurement here:
+ 0894V
- 0072V
0000V
The large numbers show the measured peak values, the small numbers at the bottom
show the current values. You can switch between the graphical/numerical display
of the charge measurement values and the display of temperature and relative
humidity values with short keypress. In graphical display of the charge measurement
you can see a voltage graph with a maximum oset of ±500 V. Over and under the
graph you can see the measured peak values. The device records the peak values
even when the display mode is changed.

7
MIN/MAX STORED
LIST POS: x
STORING MEASUREMENT RESULTS
You can store the measured results by long pressing the „MODE” button. After this a
notication can be seen on the display for 3 seconds. If there is still an empty space in
the measurement list, a text similar to the following appears on the display:
The value „x” indicates the position in the measurement list where the measured
peak values are stored. This list can hold up to 20 measurements. When it is full, it is
not possible to store any more measurement results and the following message is
displayed after a long press of the „MODE” button:
LIST FULL
When the list is full, it is only possible to store new measurement results after deleting
the list, which can only be done by a computer.
MANAGEMENT OF MEASUREMENT RESULTS ON
COMPUTER
An AIJGO-60 can be connected to a computer with a USB-A – mini USB-B cable.
For the management of measurement results on a computer, you need to install the
driver program supplied on pendrive with the device, which assigns a virtual serial
port to the tester. To do this, run the „CDM v2.08.30 WHQL Certied.exe” application
as administrator (Right mouse button -> Run as administrator). If the installation is
done, run the supplied terminal program (termite-3.1.exe). The program is running
immediately after the installation is done, later you can use the shortcut in the Start
menu (look for “Termite”).

8
On the startup of the terminal program, a window similar to the following image will
appear:
Click on the „Settings” button and set up the terminal program based on the
following:
In the “Port” eld, enter the virtual serial port number assigned to the device. When
you are done, click „OK”.
Two commands can be used in the program to manage the stored data of an
AIJGO-60 product: „LIST” and „DELETE”. The commands can be sent by pressing
“enter” after typing them. The following is a picture of the terminal program after
a user has retrieved the stored list with the “LIST” command, deleted it with the
“DELETE” command, and then retrieved the actual stored list with a “LIST” command
again.

9
The listed data can be selected in the terminal program window and copied to
somewhere else if necessary.
Maintenance
Replacement of batteries
By unscrewing two screws on the back of the device, you can remove the cover plate
and then replace the batteries. Be sure to place them in the correct direction to
avoid, also to avoid possible circuit damage.
Calibration
Calibration of the device is recommended once in every 2 years.
Theimagesinthisdocumentareforinformationpurposesonly.Thedesign,theaccessories,
the technical specications and various details of the product can be changed. We reserve
the right to sell such changed product without notice. In case of improper use other than
as described in this documentation or in case of modication of the product which D és
Tsa. Bt. has not given its written consent, our company may reject any warranty claims.

D és Tsa. Bt.
2600 Vác, Dr. Csányi László Krt. 83.
Hungary
Tel. / Fax: +36 27 502 555
+36 27 200 835
E-mail: sales@destsa.hu
Web: www.destsa.hu
www.dlb.hu
Table of contents
Other dlb Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

ENDEAVOUR
ENDEAVOUR ET2044 instruction manual

TechnipFMC
TechnipFMC Smith Meter AccuLoad IV Series Installation and maintenance manual

Ranger
Ranger 2150 Reference manual

Endress+Hauser
Endress+Hauser Gammapilot M FMG60 operating instructions

Hach
Hach Lange CL17 Instrument manual

ELPRO
ELPRO ECOLOG-PRO Base Operation manual