WIKA DPT-10 User manual

Operating Instructions
Dierential pressure transmitter DPT-10
Dierential pressure transmitter DPT-10
Metallic measurement diaphragm
4 … 20 mA/HART

2
Contents
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
Contents
1 About this document
1.1 Function ........................................................................................................................... 4
1.2 Target group ..................................................................................................................... 4
1.3 Symbolism used ............................................................................................................... 4
2 For your safety
2.1 Authorised personnel ....................................................................................................... 5
2.2 Appropriate use ................................................................................................................ 5
2.3 Warning about incorrect use ............................................................................................. 5
2.4 General safety instructions ............................................................................................... 5
2.5 Safety label on the instrument .......................................................................................... 5
2.6 CE conformity ................................................................................................................... 6
2.7 Fulllment of NAMUR recommendations ......................................................................... 6
2.8 Safety instructions for oxygen applications ....................................................................... 6
3 Product description
3.1 Conguration .................................................................................................................... 7
3.2 Principle of operation........................................................................................................ 8
3.3 Adjustment ..................................................................................................................... 11
3.4 Packaging, transport and storage ................................................................................... 11
4 Mounting
4.1 General instructions to use the instrument ..................................................................... 13
4.2 Instructions for oxygen applications ............................................................................... 14
4.3 Mounting and connection instructions ............................................................................ 14
4.4 Measurement setup ow ................................................................................................ 18
4.5 Measurement setup level ............................................................................................... 21
4.6 Measurement setup density and interface ...................................................................... 25
4.7 Measurement setup dierential pressure ....................................................................... 27
4.8 Mounting external housing ............................................................................................. 29
4.9 Installation control .......................................................................................................... 30
5 Connecting to power supply
5.1 Preparing the connection ............................................................................................... 31
5.2 Connection procedure .................................................................................................... 32
5.3 Single chamber housing ................................................................................................. 33
5.4 Double chamber housing ............................................................................................... 34
5.5 Double chamber housing Ex d ....................................................................................... 37
5.6 Version IP 66/IP 68, 1 bar ............................................................................................... 38
5.7 Switch-on phase............................................................................................................. 39
6 Adjustment with the display and adjustment module
6.1 Short description ............................................................................................................ 40
6.2 Insert display and adjustment module ............................................................................ 40
6.3 Adjustment system ......................................................................................................... 41
6.4 Set parameters ............................................................................................................... 42
6.5 Menu schematic ............................................................................................................. 52
6.12 Saving the parameter adjustment data ........................................................................... 55
7 Setup with the adjustment program AMS™
7.1 Parameter adjustment with AMS™................................................................................. 56

3
Contents
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
8 Setup
8.1 Select the mode ............................................................................................................. 57
8.2 Flow measurement ......................................................................................................... 57
8.3 Level measurement ........................................................................................................ 59
8.4 Density and interface measurement ............................................................................... 63
8.5 Dierential pressure measurement ................................................................................ 63
9 Maintenance and fault rectication
9.1 Maintenance .................................................................................................................. 66
9.2 Rectify faults ................................................................................................................... 66
9.3 Instrument repair ............................................................................................................ 67
10 Dismounting
10.1 Dismounting steps.......................................................................................................... 68
10.2 Disposal ......................................................................................................................... 68
11 Supplement
11.1 Technical data ................................................................................................................ 69
11.2 Dimensions .................................................................................................................... 79
Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and op-
eration in Ex areas. These safety instructions are part of the operating
instructions manual and come with the Ex-approved instruments.
Editing status: 2013-07-11

4
1 About this document
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
1 About this document
1.1 Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important instruc-
tions for maintenance and fault rectication. Please read this informa-
tion before putting the instrument into operation and keep this manual
accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained specialist
personnel. The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
• List
The dot set in front indicates a list with no implied sequence.
→ Action
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat-
teries and accumulators.

5
2 For your safety
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
2 For your safety
2.1 Authorised personnel
Mount and set up the pressure transmitter only if you know the appli-
cable national regulations and have the appropriate qualication. You
must be aquainted with the regulations and instructions for hazard-
ous areas, measurement and control technology as well as electrical
circuits because the pressure transmitter is "electrical equipment"
according to EN 50178. Depending on the application conditions, it
is necessary that you have appropriate knowledge, e.g. concerning
corrosive products or high pressure.
2.2 Appropriate use
DPT-10 is a dierential pressure transmitter for measurement of ow,
level, dierential pressure, density and interface.
You can nd detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions
manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbi-
trary conversions or modications are explicitly forbidden.
2.3 Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to
application-specic hazards, e.g. vessel overll or damage to system
components through incorrect mounting or adjustment.
2.4 General safety instructions
This is a high-tech instrument requiring the strict observance of stand-
ard regulations and guidelines. The user must take note of the safety
instructions in this operating instructions manual, the country-specic
installation standards as well as all prevailing safety regulations and
accident prevention rules.
The instrument must only be operated in a technically awless and
reliable condition. The operator is responsible for trouble-free opera-
tion of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula-
tions.
2.5 Safety label on the instrument
The safety approval markings and safety tips on the device must be
observed.

6
2 For your safety
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
2.6 CE conformity
The device fullls the legal requirements of the applicable EC guide-
lines. By axing the CE marking, we conrm successful testing of the
product.
2.7 Fulllment of NAMUR recommendations
The device fullls the requirements of the applicable NAMUR recom-
mendations.
2.8 Safety instructions for oxygen applications
For instruments in oxygen applications the special instructions in
chapters "Storage and transport", "Mounting" as well as "Technical
data" under "Process conditions"must be noted. Furthermore the valid
national regulations, implementation instructions and memorandums
of the professional assocations must be noted.

7
3 Product description
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
3 Product description
3.1 Conguration
The scope of delivery encompasses:
• DPT-10 dierential pressure transmitter
• Depending on the version, ventilation valves and/or screw plugs
(details see chapter "Dimensions")
• Optional accessory
• Documentation
– this operating instructions manual
– Test certicate for pressure transmitters
– Operating instructions manual "Display and adjustment mod-
ule" (optional)
– Ex-specic "Safety instructions" (with Ex versions)
– if necessary, further certicates
The following illustration shows the components of DPT-10:
+ –
1
2
3
Fig. 1: DPT-10 in basic version
1 Housing cover, optionally with integrated display and adjustment module
2 Housing with electronics
3 Process component with measuring cell
The components are available in dierent versions.
The nameplate contains the most important data for identication and
use of the instrument:
Scope of delivery
Constituent parts

8
3 Product description
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
1
9
10
2
3
4
5
6
7
8
Fig. 2: Layout of the type label (example)
1 Instrument type
2 Field for approvals
3 Measuring range
4 Signal output/Supply voltage
5 Process pressure - Process temperature
6 Seal material
7 Product code
8 Order number
9 Serial number of the instrument
10 ID numbers, instrument documentation
3.2 Principle of operation
DPT-10 is a dierential pressure transmitter for measurement of ow,
level, dierential pressure, density and interface. Measured products
are gases, vapours and liquids.
Flow measurement
QQ
p1p1
p2p2
Q~ ∆p Q~ ∆p
++ ––
1 2
Fig. 3: Flow measurement with DPT-10 and DP ow element, Q = ow, Δp =
dierential pressure, Δp = p1- p2
1 Orice
2 Pitot tube
Application area

9
3 Product description
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
Level measurement
123
h= ∆p
ρg
+
+
–
–
h
+ –
Fig. 4: Level measurement with DPT-10. Δp = dierential pressure, ρ = density
of the medium, g = acceleration of gravity
1 Basic version with eective pressure lines
2 Version with ange chemical seal
3 Version with capillaries and cell chemical seals
Dierential pressure measurement
+
2
1
Fig. 5: Dierential pressure measurement with DPT-10
1 Filter
2 DPT-10
Density measurement
=∆p
hg
+
–
h
1
Fig. 6: Density measurement with DPT-10, h = dened mounting distance, Δp =
dierential pressure, ρ = density of the medium, g = acceleration of gravity
1 DPT-10

10
3 Product description
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
Interface measurement
1
+
–
h
2
3
Fig. 7: Interface measurement with DPT-10
1 DPT-10
2 Liquid with higher density
3 Liquid with lower density
A metallic measuring cell is used as sensor element. The process
pressures are transmitted via the separating diaphragms and lling
oils to a resistance measuring bridge (semi-conductor technology).
The dierence between the acting pressures generates a change in
the bridge voltage. This change is measured, further processed and
converted into a corresponding output signal.
The markings "+" and "-" on the process component in chapter
"Mounting and connection instructions" must therefore be noted when
connecting to the process. The pressure acting on "+" goes positive,
the pressure acting on "-" goes negative into the calculation of the
pressure dierence.
The conguration of the measuring cells diers depending on the
measuring range:
1
2
3
4
5
1
p2
p
Fig. 8: Metallic measuring cell 10 mbar and 30 mbar - p1and p2process pres-
sures
1 Measuring element
2 Silicone diaphragm
3 Separating diaphragm
4 Filling oil
5 Integrated overvoltage arrester
Functional principle

11
3 Product description
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
1
p2
p
12
3
4
Fig. 9: Metallic measuring cell from 100 mbar - p1and p2process pressures
1 Measuring element
2 Overload diaphragm/Middle diaphragm
3 Filling oil
4 Separating diaphragm
4 … 20 mA/HART two-wire electronics for voltage supply and meas-
ured value transmission on the same cable.
The supply voltage range can dier depending on the instrument
version.
The data for power supply are specied in chapter "Technical data".
The optional background lighting of the display and adjustment
module is powered by the sensor. A certain level of operating voltage
is required for this. You can nd the exact voltage specications in
chapter "Technical data".
3.3 Adjustment
The instrument can be adjusted with the following adjustment media:
• With the display and adjustment module
• the suitable WIKA DTM in conjunction with an adjustment software
according to the FDT/DTM standard, e.g. PACTware and PC
• with manufacturer-specic adjustment programs AMS™ or PDM
• With a HART handheld
The entered parameters are generally saved in DPT-10, optionally
also in the display and adjustment module or in PACTware.
3.4 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging of standard instruments consists of environment-
friendly, recyclable cardboard. For special versions, PE foam or PE
foil is also used. Dispose of the packaging material via specialised
recycling companies.
Caution:
Instruments for oxygen applications are sealed in PE foil and provided
with a label "Oxygen! Use no Oil". Remove this foil just before mount-
ing the instrument! See instruction under "Mounting".
Voltage supply
Packaging

12
3 Product description
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con-
cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
• Not in the open
• Dry and dust free
• Not exposed to corrosive media
• Protected against solar radiation
• Avoiding mechanical shock and vibration
• Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
• Relative humidity 20 … 85 %
Transport
Transport inspection
Storage
Storage and transport
temperature

13
4 Mounting
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
4 Mounting
4.1 General instructions to use the instrument
Make sure that all parts of the instrument coming in direct contact
with the process, especially the sensor element, process seal and
process tting, are suitable for the existing process conditions, such
as process pressure, process temperature as well as the chemical
properties of the medium.
You can nd the specications in chapter "Technical data" and on the
nameplate.
Use the recommended cables (see chapter "Connecting to power
supply") and tighten the cable gland.
You can give the instrument additional protection against moisture
penetration by leading the connection cable downward in front of the
cable entry. Rain and condensation water can thus drain o. This ap-
plies mainly to outdoor mounting as well as installation in areas where
high humidity is expected (e.g. through cleaning processes) or on
cooled or heated vessels.
The ventilation for the electronics housing is realised via a lter ele-
ment in the vicinity of the cable glands.
2
1
1
Fig. 10: Position of the lter element with single and double chamber housing
1 Filter element for ventilation of the electronics housing
2 Blind plug
Information:
Make sure that the lter element is always free of buildup during
operation. A high-pressure cleaner may not be used for cleaning.
DP ow elements are calculated for certain pipeline and operating
data. Therefore, check the pipeline data before installation at the
measuring point and compare the measurement loop number.
Detailed instructions for mounting the DP ow element are stated in
DIN EN ISO 5167 as well as in the instrument documentation from the
respective manufacturer.
You will nd general recommendations for the installation of eective
pressure lines in the corresponding national or international stand-
ards. When installing eective pressure lines outdoors, consider
applying suitable anti-freeze protection, e.g. tube heating. Install
eective pressure lines with a monotonic downward slope of at least
10 %.
Suitability for the process
conditions
Moisture
Ventilation
Eective pressure trans-
mitter
Eective pressure lines

14
4 Mounting
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
In case of strong vibrations at the application position, the instrument
version with external electronics should be used.
Higher process temperatures often mean also higher ambient tem-
peratures for electronics and connection cable. Make sure that the
upper temperature limits stated in chapter "Technical data" for the
environment of the electronics housing and connection cable are not
exceeded.
4.2 Instructions for oxygen applications
Oxygen and other gases can be explosive when brought into contact
with oils, grease and plastics, so the following measures must also be
taken:
• All components of the plant, such as e.g. measuring instru-
ments must be cleaned according to the requirements of BAM
(DIN 19247)
• Depending on the seal material, certain temperatures and pres-
sures must not be exceeded in oxygen applications, see chapter
"Technical data"
Danger:
Instruments for oxygen applications must be unpacked just before
mounting. After removing the protective cover of the process tting,
the label "O2" will be visible on the process tting. Penetration of oil,
grease and dirt should be avoided. Danger of explosion!
4.3 Mounting and connection instructions
When connecting the DPT-10 to the measurement loop, take note
of the plus/minus side of the process component. The plus side is
marked with a "+", the minus side with a "-" on the process compo-
nent next to the oval anges.
+ –
2
1
Fig. 11: Marking for plus/minus side on the process component
1 Plus side
2 Minus side
The following illustration shows the elements for a tube mounting and
an example for a mounting arrangement with valve block.
Vibrations
Temperature limits
Oxygen applications
Connection plus/minus
side
Mounting arrangement

15
4 Mounting
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
234
5
6
7
8
9
10
1
Fig. 12: Mounting arrangement with tube mounting
1 Strap for tube mounting
2 Mounting bracket
3 Ventilation valve
4 Fixing screws
5 DPT-10
6 PTFE seal
7 Valve block
8 Oval ange adapter
9 Fixing screws
10 Eective pressure line
Valve blocks enable the simple installation and setup of the dierential
pressure transmitter. They separate the pressure transmitter from the
process side and enable also a check of the measurement loop. They
are available as 3-fold and 5-fold version. The integrated equalization
valve enables a pressure compensation between plus and minus side
during the setup. Thanks to the valve block it is possible to dismount
the DPT-10 without interrupting the process. This means higher plant
availability and simpler setup or maintenance.
The 3-fold valve block with anging on both sides enables a mechani-
cally stable connection between the DPT-10 and e.g. the tapping
points or the ange plate of a pitot tube. With the ve-fold valve block,
two additional valves allow blowing out the process lines or checking
the DPT-10 in installed condition.
The following illustration shows the connection of the 3-fold valve
block.
Valve blocks
Connect 3-fold valve
block

16
4 Mounting
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
A
A:
1 23 45
2 2
1
3 45
Fig. 13: Connection of a 3-fold valve block
1 Process tting
2 Process tting
3 Inlet valve
4 Inlet valve
5 Breather valve
The following illustration shows the connection of the 3-fold valve
block, anging on both sides.
3-fold valve block, ang-
ing on both sides

17
4 Mounting
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
A:
A
1 23 45
2 2
13 45
Fig. 14: Connection of a 3-fold valve block, anging on both sides
1 Process tting
2 Process tting
3 Inlet valve
4 Inlet valve
5 Breather valve
The following illustration shows the connection of the 5-fold valve
block.
5-fold valve block

18
4 Mounting
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
A:
A
1 2
43
5 7 89 6
2 2
3 1
9
4 6
7 8
5
Fig. 15: Connection of a 5-fold valve block
1 Process tting
2 Process tting
3 Check/Ventilate
4 Check/Ventilate
5 Valve for checking/ventilating
6 Valve for checking/ventilating
7 Inlet valve
8 Inlet valve
9 Breather valve
4.4 Measurement setup ow
→ Mount DPT-10 above the measurement loop so that condensate
can drain o in the process cable.
In gases

19
4 Mounting
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
+–
1
2
4
3
Fig. 16: Measurement setup with ow measurement of gases, connection via
3-fold valve block
1 DPT-10
2 3-fold valve block
3 Blocking valves
4 Orice or impact pressure probe
+
1
2
3
Fig. 17: Measurement setup with ow measurement of gases, connection via
3-fold valve block, anging on both sides
1 DPT-10
2 3-fold valve block, anging on both sides
3 Orice or impact pressure probe
1. Mount DPT-10 below the measurement loop
2. Mount condensate vessels at the same height with the discharge
socket and at the same distance to DPT-10
3. Fill the eective pressure lines to the height of the condensate
vessels before setup
In vapours

20
4 Mounting
WIKA Operating Instructions - Dierential pressure transmitter DPT-10
+ –
1
233
55
6
4
Fig. 18: Measurement setup, ow measurement in vapours
1 Condensate vessels
2 Orice or impact pressure probe
3 Blocking valves
4 DPT-10
5 Drain or blow-o valves
6 3-fold valve block
When using a 5-fold valve block, the drain or blow-o valves are
already integrated.
1. Mount DPT-10 below the measurement loop so that the eective
pressure lines are always lled with liquid and gas bubbles can
bubble up to the process line
2. For measurements in products with solid content such as e.g.
dirty liquids, the installation of separators and drain valves is
recommended to enable collection and removal of debris and
sediment.
3. Fill the eective pressure lines to the height of the condensate
vessels before setup
In liquids
Other manuals for DPT-10
1
Table of contents
Other WIKA Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

VWR
VWR OX 4100 H operating manual

Siemens
Siemens SITRANS FUH1010 Product instruction manual

Tornatech
Tornatech GT-FluxFP operating instructions

ICT
ICT SFM1 quick start guide

HunterLab
HunterLab UltraScan VIS Supplemental manual

S+S Regeltechnik
S+S Regeltechnik RHEASREG KLSW Operating Instructions, Mounting & Installation