DNTLworks Equipment ProCart II Mobile Treatment Unit User manual

1
ProCart IITM
Mobile Treatment Console
OPERATION MANUAL
© Copyright 2011 DNTLworks Equipment Corporation
09/20/2011
All Rights Reserved
DNTLworks Equipment Corporation
7300 South Tucson Way
Centennial, Colorado 80112 USA
Toll Free: (800) 847-0694
Main: (303) 693-1410
Fax: (303) 693-6189

2
Contents
Introduction ..................................................................................................................................................3
Customer Service ..........................................................................................................................................4
DNTLworks Limited Warranty.......................................................................................................................4
Returns......................................................................................................................................................5
Description of Unit........................................................................................................................................ 5
Air System ................................................................................................................................................. 5
Water System............................................................................................................................................5
Vacuum System.........................................................................................................................................5
Electrical System .......................................................................................................................................5
Specifications ................................................................................................................................................5
Unit Features.................................................................................................................................................6
Top View ................................................................................................................................................... 6
Front View.................................................................................................................................................7
Back View.................................................................................................................................................. 8
Operation......................................................................................................................................................9
Setup .........................................................................................................................................................9
Handpiece Operation..............................................................................................................................10
Three-way Syringe ..................................................................................................................................11
Vacuum System.......................................................................................................................................11
Water Bottle............................................................................................................................................13
Accessories..............................................................................................................................................13
Dental Office Infection Control Guidelines (CDC).......................................................................................14
Barrier Techniques..................................................................................................................................14
Cleaning and Disinfection of Dental Unit and Environmental Surfaces .................................................15
Use and Care of Handpieces and Other Intra-oral Dental Devices.........................................................15
Other Important Issues...........................................................................................................................16
Shut Down Instructions............................................................................................................................... 16
Maintenance...............................................................................................................................................19
Electrical Diagram ...................................................................................................................................20
Air Diagram .............................................................................................................................................21
Water Diagram........................................................................................................................................22

3
Vacuum Diagram.....................................................................................................................................23
Back, Inside View ....................................................................................................................................24
Pumps, Inside View.................................................................................................................................25
Air and Vacuum Tanks, Inside View........................................................................................................26
Shelf, Inside View....................................................................................................................................27
Control Panel, Inside View......................................................................................................................28
Troubleshooting Flow Charts......................................................................................................................29
Handpiece Water ....................................................................................................................................29
Syringe Water .........................................................................................................................................30
Air Regulator Adjustment .......................................................................................................................30
Waste Leak..............................................................................................................................................31
Poor Vacuum...........................................................................................................................................31
Compressor Will Not Shut Off or Will Recycle Often..............................................................................32
Compressor Will Not Start ......................................................................................................................33
Compressor Will Not Recycle or Stalls....................................................................................................33
Introduction
Thank you for purchasing the ProCart IITM Mobile Treatment Unit from DNTLworks Equipment
Corporation. The information contained in the manual should answer any questions regarding service
and operation of your ProCart IITM unit.
All authorized personnel who operate, maintain, or service your ProCart IITM unit should carefully review
this manual before attempting to operate, perform maintenance on or service the unit. Your ProCart IITM
unit should be operated and maintained by trained personnel only. Should questions or problems arise,
contact our Customer Service Department.
Although your ProCart IITM unit has been designed and tested for maximum safety and optimum
performance, it is sold with the express understanding that DNTLworks, its subsidiaries, agents and
representatives will not accept any responsibility for the following, including, but not limited to:
1) Operator’s lack of knowledge, negligence or carelessness in the operation of this equipment.
2) Equipment not properly maintained or serviced.
3) Injury to personnel or patients from improper use.
4) Modification or tampering of any kind.

4
Customer Service
In the event you require assistance with your unit, please call 1-800-847-0694 or 303-693-1410 and
speak with one of our customer service representatives. Our service hours are from 8:00 a.m. to 5:00
p.m., Mountain Standard Time, Monday through Friday.
In most instances, service problems may be solved over the telephone. If service is required, you may
ship the unit to our manufacturing facility for repair. Warranty service will be performed in accordance
with the DNTLworks’ Limited Warranty. Non-warranty service will be provided at reasonable parts and
labor costs.
DNTLworks Limited Warranty
DNTLworks warrants to the purchaser that these products are free of defects in materials and/or
workmanship for three (3) full years from date of delivery, on a “parts only” basis. In addition,
DNTLworks extends a ninety (90) day labor warranty from the date of delivery for all products we
manufacture. Shipping charges incurred to the factory under warranty purposes will be the
responsibility of the owner.
During the warranty period, all parts which, upon inspection and examination by DNTLworks, are proven
to be defective, will be replaced free of charge. All decisions concerning whether a part will be repaired
or replaced and the manner, method, and extent of such repair or replacement will be at the sole
discretion of DNTLworks. The responsibility of DNTLworks does not include repair and replacement cost
resulting from misuse, abuse, improper maintenance, or normal wear and tear.
DNTLworks will pay for labor costs for warranty service for a period of 90 days from the date of
purchase. DNTLworks sole obligation under said warranty is to repair, or, at its option, replace the
defective part. The buyer will have no options.
Warranties for products not manufactured by DNTLworks, but sold in combination with DNTLworks
products, will be honored by DNTLworks for the entire duration of the original manufacturer’s warranty
period.
The warranty will be voided by alterations, tampering with, improper installation or maintenance,
accident or modification of the equipment, with the exception of work performed by DNTLworks or one
of its authorized service agents. This warranty expressly excludes all damage to the products resulting
from careless or neglectful transportation. DNTLworks will in no event be responsible for any work done
without first obtaining DNTLworks’ written consent.
This warranty is made expressly in lieu of all other warranties, expressed or implied, including any
implied warranties of merchantability or fitness for a particular purpose. No employee, agent, franchise,
dealer or other person is authorized to give any warranties of any nature on behalf of DNTLworks.
Except as provided herein, DNTLworks will have no liability or responsibility to the customer or any
other person or entity with respect to any liability, loss or damage caused or alleged to be caused
directly or indirectly by equipment sold, leased, or furnished by DNTLworks, including, but not limited

5
to, any interruption of services, loss of business or anticipatory profits or consequential damage arising
out of or connected with the sale, lease, use, or anticipated use of equipment. Notwithstanding the
above limitations and warranties, DNTLworks liability hereunder for damages incurred by customer or
other will not exceed the amount paid by customer for the particular equipment involved.
Returns
Purchased goods may not be returned without the express written consent of DNTLworks and a Return
Goods Authorization Number (RGA#). All items must be returned within 14 days of initial delivery and
are subject to a 15% restocking charge. Special order items cannot be returned for credit consideration.
Freight charges on approved return items shall be borne by the customer.
Description of Unit
ProCart II™ is an easy-to-clean, self-contained operatory system that sets up in just a few minutes. It
requires no plumbing and is mobile for ease of movement between operatories or for storage.
Air System
The internal air compressor will activate when main power switch is turned on, filling the air tank and
pressurizing the water reservoir. The compressor will deactivate when air pressure reaches 100psi and
will recycle at 70psi.
Water System
Water container should be filled 3/4 full. Water to handpieces is controlled by the foot control toggle
and the valves on the front of the unit. The water for the air/water syringe is controlled by depressing
the water button on the syringe. Syringe water pressure is controlled by the water regulator at a
maximum of 40psi.
Vacuum System
The internal vacuum system is activated by removing HVE or saliva ejector from holder. You may use
either the saliva ejector or HVE as needed. There is a vacuum solids trap and a waste purge pump. In the
vacuum container is an automatic shut down system to prevent spillage. Cannot used Saliva Ejector and
HVE at the same time.
Electrical System
There are two electrical outlets on your system, one internal outlet and one external, to accommodate
accessories. You may use a maximum of 2 amps on the internal and external outlets. The unit is
protected by a 15 amp circuit breaker.
Specifications
Dimension
Length
20” (51cm)

6
Width
21” (53cm)
Adjustable Height
30” (76cm) to 34” (86cm)
Total Weight
100lbs (45.5 Kg)
Electrical
Voltage
115 or 220Vac
Frequency
60 or 50 Hz
Wattage
1200 to 1440 W
Compressor
Horsepower
1/3 Hp
Working Pressure
70 to 100psi
Flow Rate
2.5scfm @ 40psi
Air Tank
1 Gallon (4 Liter)
Vacuum Pump
Horsepower
1/3 Hp
Pressure
12 inHg
Flow Rate
4.6 scfm
Container
1 Gallon (4 Liter),7.5 Pints Useable.
Unit Sound
Full Load
48dB @ 3’
Unit Features
Top View
Vacuum Relief Valve
DO NOT ADJUST
Vacuum Solids Trap
Vacuum Purge Switch
Water Container
Water Regulator
Handpiece
Pressure
Adjustments
Handpiece Pressure Gauge
Set Up Guide
& Manual
Repair Kit
Handpiece Water
Adjustment Knobs

7
Front View
Handpiece Water
Adjustment Knobs
Handpiece Tubings
Air/Water Syringe
Saliva Ejector
HVE
Foot Control
Water Toggle

8
Back View
Air/Water
Accessory QD’s
Air Tank
Purge
Waste Purge
Power Cord &
Waste Purge Line
Power Switch
Circuit Breaker
Accessory Outlet
Outlet
Power Receptacle
Cooling Fans

9
Operation
Setup
[Type a quote from the document or
the summary of an interesting point.
You can position the text box
anywhere in the document. Use the
Text Box Tools tab to change the
formatting of the pull quote text box.]
Place foot control near
the operator.
Install handpieces.
Install saliva ejector
tip. Install HVE tip.
Plug power cord into
power receptacle and
into wall outlet.
Make sure the ProCart II
is the only unit on the
circuit.
Connect waste
purge tubing.
Turn on power
switch.
Fill water
container 3/4
from the top.
Connect water
container to
unit.
Remove handpiece and
depress foot control.
Adjust pressure to
manufacturer’s setting.
Repeat for other
handpieces.
Turn on water toggle.

10
Handpiece Operation
Always clean handpieces between patients.
See Dental Office Infection Controls Guidelines CDC. Below.
Remove handpiece from
holder. Depress foot
control and adjust
handpiece water.
Repeat for other
handpieces.
Test air and water
on syringe.
Remove HVE or saliva
ejector from holder.
Remove handpiece
from holder.
Depress foot control.
When using a handpiece, make sure air pressure does not
exceed handpiece manufacturer’s recommended pressure.
When setting the maximum handpiece pressure, always
make sure foot control is fully depressed, and adjust
handpiece air adjustment valve.
Turn on foot control
toggle valve. If the
handpiece does not
require water, turn
toggle off.
For handpiece coolant
water, adjust water valve
while fully depressing foot
control.
Turn toggle on holder
to the ON position.

11
Three-way Syringe
Vacuum System
Water only.
Air only.
Air and water (mist).
You can use disposable syringe tip kits. (A-dec type disposable tip kit)
Always clean syringe between patients.
See Dental Office Infection Controls Guidelines CDC, below.
When using either saliva ejector or HVE, make sure the other valve
is closed. Cannot used Saliva Ejector and HVE at the same time.
Always clean vacuum system between patients.
See Dental Office Infection Controls Guidelines CDC, below.
Remove HVE or saliva
ejector and open
valve.
Move toggle to ON
position.
Before using saliva ejector always check screen.
Replace solids trap
as needed.

12
Purge Vacuum System
Cleaning Between Patients and End of Day
1) Remove any waste in vacuum container by purging and disposing of contents in a CDC-approved
manner. See Dental Office Infection Control Guidelines CDC, below.
2) Mix one quart of evacuation cleaner, turn on evacuation system and suction cleaner into waste
bottle using the HVE valve, and the saliva ejector.
3) Dispose of cleaner waste in CDC-approved manner.
4) Wipe down outside of the bottle and tubings with a surface cleaner.
5) Clean tubings and handpieces in CDC-approved manner.
6) Clean unit surfaces CDC-approved manner.
Never adjust vacuum relief valve.
When using saliva ejector, vacuum relief valve will have an air leak
sound.
Make sure purge tubing is installed.
Always clean vacuum system between patients.
See Dental Office Infection Controls Guidelines CDC, below.
To clean vacuum system, use a non-foaming vacuum system
cleaning solution.
Hold end of purge
line over a container.
Turn on waste purge.
Never leave purge pump
running when not purging
system.

13
Water Bottle
Accessories
Disconnect red air line.
Depress red pressure relief valve.
Disconnect blue water line.
Make sure o-ring is not
missing and is seated
in the groove.
Lubricate o-ring as
needed.
Fill about 3/4 full.
We recommend
distilled water
only.
Reconnect
water and air
lines.
You can adjust the water pressure to the air/water syringe.
The inside outlet is used for accessories built into the unit, such as
fiber optic systems or water heaters, etc.
Never use more than 2 amps between the inside and outside
outlets.
Questions: Call Customer Service.

14
Dental Office Infection Control Guidelines (CDC)
DENTAL OFFICE INFECTION CONTROL GUIDELINES FOR THE PROTECTION OF
PATIENTS AND DENTAL CARE PERSONNEL AS RECOMMENDED BY THE UNITED
STATES CENTERS FOR DISEASE CONTROL (CDC)
NOTE: A medical history should be taken at the initial patient visit and on recall, updated
with notations made on the chart.
Barrier Techniques
1) Dentists, hygienists and assistants should wear new gloves for each patient treated. Hands
should always be washed with liquid soap before and after treatment, contact with patients or after
touching inanimate objects likely contaminated by blood or saliva, and before leaving the operatory.
Remove torn, cut or punctured gloves immediately, wash hands, and re-glove before completion of
dental procedures.
2) During all treatment procedures, dentists, hygienists and assistants should wear face masks
and protective eyewear, or in lieu of both of these, a chin-length plastic face shield.
The outside outlet is used for external accessories use.
Never use more than 2 amps between the inside and outside outlets.
The air 1/4”QD is used for accessories that require air. The
pressure will be between 70 to 100psi. The accessory should not
be used at the same time you are using one of the handpieces.
You may connect to an external air or nitrogen system. Make sure
the pressure does not exceed 100psi.
The water 1/4” QD is used for accessories that require water. The
pressure will be 65psi.
You may connect to an external water system. Make sure the
pressure does not exceed 65psi. You will need to disconnect the
water bottle from the unit.

15
3) Reusable and/or disposable gowns, laboratory coats or uniforms should be worn when
street clothing may be soiled with blood or other body fluids. Gowns should be changed at least
daily or when visibly soiled with blood.
Cleaning and Disinfection of Dental Unit and Environmental Surfaces
1) After treatment of each patient and at the completion of daily work activities, countertops
and dental unit surfaces that may have become contaminated with patient material should be cleaned
with disposable toweling, using an appropriate cleaning agent and water as necessary. Surfaces then
should be disinfected with a suitable chemical germicide.
2) A chemical germicide registered with the EPA as a “hospital disinfectant” and labeled for
“tuberculocidal” (i.e., mycobactericidal) activity is recommended for disinfecting surfaces that have
been soiled with patient material. These intermediate level disinfectants include phenolics,
iodophors, and chlorine-containing compounds. Because mycobacteria are among the most
resistant groups of microorganisms, germicides effective against mycobacteria should be effective
against many other bacterial and viral pathogens. A fresh solution of sodium hypochlorite
(household bleach) prepared daily is an inexpensive and effective intermediate-level germicide.
Concentrations ranging from 500 to 800 ppm of chlorine (a 1:100 dilution of bleach and tap water
or 1/4 cup of bleach to 1 gallon water) are effective on environmental surfaces that have been
cleaned of visible contamination. Caution should be exercised, since chlorine solutions are corrosive
to metals, especially aluminum.
3) Low-level disinfectants - EPA registered “hospital disinfectants” that are not labeled for
“tuberculocidal” activity (e.g., quatemary ammonium compounds) - are appropriate for general
housekeeping purposes such as cleaning floors, walls and other housekeeping surfaces. Intermediate
and low level disinfectants are not recommended for reprocessing critical or semi-critical dental
instruments.
4) Before high-level disinfection or sterilization, and while wearing heavy duty rubber
(household) gloves, ultrasonically clean (preferably) or scrub instruments in order to remove debris.
Use and Care of Handpieces and Other Intra-oral Dental Devices
1) Routine between-patient use of a heating process capable of sterilization (i.e., steam under
pressure (autoclaving), dry heat, or heat/chemical vapor) is recommended for all highspeed dental
handpieces, lowspeed handpiece components used intra-orally, and reusable prophylaxis angles.
Manufacturers’ instructions for cleaning, lubrication, and sterilization procedures should be followed
closely to ensure both the effectiveness of the sterilization process and the longevity of these
instruments. According to manufacturers, virtually all highspeed and lowspeed handpieces in
production today are heat tolerant and most heat-sensitive models manufactured earlier can be
retrofitted with heat-stable components.
2) Internal surfaces of highspeed handpieces, lowspeed handpiece components, and
prophylaxis angles may become contaminated with patient material during use. This retained patient
material then may be expelled intra-orally during subsequent uses. Restricted physical access -
particularly to internal surfaces of these instruments - limits cleaning and disinfection or sterilization
with liquid chemical germicides. Surface disinfection by wiping or soaking in liquid chemical
germicides is not an acceptable method for reprocessing highspeed handpieces, lowspeed handpiece
components used intra-orally, or reusable prophylaxis angles.

16
3) Highspeed handpieces should be run to discharge water and air for a minimum or 20-30
seconds after use on each patient. Handpieces, in addition, should be heat sterilized between uses
on patients. This procedure is intended to aid in physically flushing out patient material that may
have entered the turbine and air or water lines. Use of an enclosed container or high-velocity
evacuation should be considered to minimize the spread of spray, splatter, and aerosols generated
during discharge procedures. Additionally, there is evidence that overnight or weekend microbial
accumulation in water lines can be reduced substantially by removing the handpiece and allowing
water lines to run and to discharge water for several minutes at the beginning of each clinic day.
Sterile saline or sterile water should be used as a coolant/irrigation when surgical procedures
involving the cutting of bone are performed.
Other Important Issues
1) A “no-touch” technique (e.g., hemostats or needle holders), should be utilized when using
“sharps” (needles, scalpels, blades, etc.).
2) In the operatory, sterilized and decontaminated instruments, charts, and other objects should
be protected from patient contact.
3) “Sharps” should be disposed of in puncture proof containers; hazardous and/or infectious
waste materials, which include “sharps,” should be disposed of in a manner consistent with
prevailing local laws.
4) All dental personnel should be encouraged to receive immunization protection whenever
possible, e.g., hepatitis B immunization.
5) All impressions, models and devises should be disinfected before submission and upon
receipt from the dental laboratory.
Shut Down Instructions
Depressurize
and remove
water
container.
Wipe dry.
Replace water
container.
Turn on foot control
toggle valve.

17
Open all water valves.
Place handpieces
over container and
depress foot control
until all water is
purged.
Depress syringe
water button until
all water is purged.
Remove all
handpieces and
vacuum
attachments.
Turn on vacuum.
Mix non-foaming
vacuum cleaning
solution and run
solution through
HVE valve.
Run cleaning
solution through
saliva ejector
valve.
Turn off vacuum.
Make sure
purge tubing
is installed.
Hold end of purge
line over a
container.

18
Turn on waste
purge.
Wipe water
container
dry.
Place air tank
(red) purge line
over a container.
Depress tank
purge valve
until all air
and moisture
is removed
from air tank.
Before closing lid, disconnect water container.
Turn off power switch.
Place paper towel under air regulator bowl.
Depress pin valve at the bottom of bowl to drain moisture.
Unit must be pressurized.

19
Maintenance
The following visual checks should be performed before operating the unit:
1) Look for mechanical damage that could affect safe operation, including, but not limited to, the
following:
a. Cracks in power cord.
b. Splits or kinks in air or water lines.
c. Check for cracks in water container.
d. Cracks, kinks or splits in handpiece tubing, syringe tubing and vacuum tubings.
2) Look for loose or missing items, including, but not limited to, the following:
a. Loose or missing screws, nuts and/or bolts.
b. Loose handles.
Should mechanical or other damage be noted that would affect safety or operation, the unit should not
be used until repair or replacement of defective items is completed. You may call Customer Service for
help.
!!CAUTION!!
Handpieces:
Follow manufacturers’ instructions for maintenance.
Syringe:
Follow manufacturer’s instructions for maintenance.
Air Pressure Source:
Do not exceed 100 psi.

20
Electrical Diagram
Table of contents
Other DNTLworks Equipment Dental Equipment manuals
Popular Dental Equipment manuals by other brands

3M ESPE
3M ESPE Elipar DeepCure-S operating instructions

Acteon
Acteon SATELEC Air Max user manual

Dentsply Sirona
Dentsply Sirona CEREC Primescan AC operating instructions

DEXIS
DEXIS FOCUS user manual

Durr Dental
Durr Dental VistaScan Mini View 2.0 XPS07.2D Series Installation and operating instructions

Carestream
Carestream CS 1200 User and installation guide