ST6InstructionManual
Page06LIVEVOLUMEDISPLAY
modeldependent
LiveVolumecanonlybeactivatedwhenAreaSettingshavebeenstored. SeeUserMenupage14.
ToaccesstheLiveVolumeDisplaypressthe▼keywheninVelocityMode.
ToreturntoVelocityModepressthe▲key.
PT:1.000PitotTubeFactor.
Upto5differentPitotTubeFactorscanbestored.
ToinputPitotTubeFactorsseeUserMenupage15.
ToselectaPitotTubeFactorseeOptionMenupage20.
A:1AreaSetting.
Upto5differentAreaSettingscanbestored.
ToinputAreaSettingsseeUserMenupage14.
ToselectAreaSettingsseeOptionMenupage19.
S1Speed.
Thereare5differentspeedsettings.
Usethe‘speed’keytotogglebetweenspeedsettings.
L1Locationnumber.
StoredatainuptoninedifferentLocationsseeOptionMenupage19.
Theinstrumentwillgiveanaveragevalueinthemodeofoperationfordatathatis
storedinthesameunitsofmeasurementandlocation.
WhendataisdownloadedtoaPCitwillbelistedundertheLocationinwhichitisstored.
AutoZero...WhenevertheAutoZerosequenceisinitiatedmanuallyoratpreset‘AutoZero...’will
bedisplayeduntilthecycleiscomplete.
0.0L/s
UnitofMeasurement. TheLiveVolumeistheVelocitymultipliedbytheArea.
Form/stheresultsareinL/s;forft/mintheresultsareincfm.
InLiveVolumeModetheunitsofmeasurementcannotbechanged.
23.3°CTemperatureProbereading.
Thisisnotdisplayedwhenthetemperatureprobeisnotattached.
HWhenthe‘holdtemp’keyispressed‘H’willbedisplayedafterthetemperaturereading,
signallingthatthetemperatureisbeingheld.
Thetemperaturewillbeheldeveniftheprobeisdetached,untilthe‘holdtemp’keyis
pressedasecondtime.
Pressingthe‘holdtemp’keyasecondtimewillreleasethetemperatureandiftheprobe
isattachedreadingswillcontinuetobetakenasnormal.Iftheprobeisnotattached,the
instrumentwillusethedefaultreadings.
Store
5
Numberofreadingsstored.
Upto4000readingscanbestored.
ForinformationonstoringdatamanuallyseeKeypadControlspage02.
ForinformationondataloggingseeOptionMenupages21.
AvVol:47.5AverageVolume ofthestoredreadings.
TheAverageVolumeisafunctionofVelocitymultipliedbyArea,seepage11.
PT:1.000 A:1 S1
L1AutoZero...
0.0L/s
23.3°CH Store
AvVol: 47.5 5