DRAFT 19 May 2005
2 - 4
Air Mode
In Air Mode, the air temperature can be maintained from 20°C (68°F) to 37°C (99°F) as indicated by the
Set Temp °C display setting. In Temperature Override Mode, the temperature can be maintained from
37°C (99°F) to 38.5°C (101.3°F). A probe located below deck monitors the air temperature, and
compares it with the Set Temp °C display setting. The information from this probe is supplied to the
heater control circuitry, which proportions the heater output to maintain the air temperature at the Set
Temp °C display setting. Actual air temperature is displayed by the Air Temperature °C display. A
second sensor within the air temperature probe serves as a backup to limit the incubator temperature to
between 39°C (102°F) and 40°C (104°F). At this temperature, an alarm activates, and the heater shuts off.
If desired, an auxiliary air temperature probe can control the air temperature. This probe is suspended
above the mattress through the weighing scale hole, and plugged into a special receptacle on the side of
the incubator. When plugged in, the primary air temperature control probe is disconnected, but the
backup sensor within the primary temperature probe remains connected. Thus, the auxiliary probe
controls the air temperature.
In Air Mode, the infant's temperature is a function of the air temperature and the infant's ability to
establish and maintain its own temperature. A small infant, or one with underdeveloped homeostatic
control, may not be able to maintain a stable temperature at the desired level.
Baby Mode (Model C450 QT®Isolette® Infant Incubator Only)
In Baby Mode, the infant’s skin temperature can be maintained from 34°C (93°F) to 37°C (99°F) as
indicated by the Skin Temp °C display. In Temperature Override Mode, the temperature can be
maintained from 37°C (99°F) to 37.9°C (101.3°F). A temperature sensing probe is attached directly to the
infant's skin. The information from the probe is supplied to the heater control circuitry, which proportions
the heater output to maintain the baby's temperature at the Set Temp °C setting. The actual baby
temperature appears on the Baby Temperature °C display. The Set Temp °C setting does not control air
temperature while in Baby Mode, but air temperature is still displayed. The air temperature is still limited
to 38.5°C (101.3°F). If Air Mode is selected while the skin probe remains connected, the Baby
Temperature °C display continues to display actual skin temperature, but it does not control.
If the probe is disconnected from its receptacle during Baby Mode, the Baby Temperature °C display
goes blank, and a Probe alarm activates. The high temperature alarm sensor within the air temperature
probe remains in the circuit to limit the air temperature to less than 40°C (104°F).
Alarms
Alarms are provided for power failure, system failure, inadequate air flow, probe failure, high
temperature, and variation from the set temperature. Each time the unit is turned on, the unit
automatically goes through an alarm check sequence to verify proper alarm function. After the automatic
alarm check, the low Set Temp alarm is disabled for 60 min or until the temperature reaches the Set
Temp °C setting, whichever occurs first. Each of these alarms is described below.
Air Flow
A sensor located below deck in the normal air path of the fan controls this alarm. If air flow stops due to
a fan failure, the temperature of the self-heated sensor rises, causing the Air Flow alarm indicator on the
front panel to light and produce a pulsating tone. A short-circuited, air flow sensor failure also actuates
the alarm within 10 s of the failure. This alarm is not self-resetting and cannot be silenced or canceled by
the Silence/Reset key until the alarm condition is corrected.