Dr. Neuhaus Sagemcom ZDUE-450-PLUS-VII User manual

ZDUE-450-PLUS-VII Installation Instructions
1
Safety Notes
Incorrect installation, commissioning or operation
can lead to serious damage to property and
personal injury and to death.
Please read these instructions carefully before installing,
commissioning and using the ZDUE-450-PLUS-VII. Use the
appliance only as described in the data sheet and the
manual.
The ZDUE-450-PLUS-VII meets the requirements of safety
standard EN 62368-1 (protection class 2).
Qualified personnel:
Installation and commissioning of the ZDUE-450-PLUS-VII
and the accessories may only be carried out by electricians.
They must be trained and qualified to detect and avoid
hazards and to install, commission, ground and mark
appliances, systems and circuits in accordance with the
standards and recognized rules of technology.
Connection to the house installation:
An easily accessible, all-pole circuit breaker in the power
supply circuit is required for the house installation.
Alternatively, a single-pole circuit breaker can be used in the
outer conductor as long as a distinct neutral conductor has
been integrated into the supply line. Observe country-
specific regulations. In Germany, the circuit breaker must at
least meet the requirements of the DIN VDE series 0100
standards (see CENELEC HD384; IEC 60364).
For the house installation, there must be an installation fuse
that complies with the DIN VDE series 0100 standards (see
CENELEC HD384; IEC 60364) and is properly adapted to
the cable cross-section of the power supply line. The
additional short-circuit protection must have a selectivity of I
> 1500A.
The appliance is not suitable for connection to an electrical
supply network with IT configuration (IT system).
Cable Routing:
The space between antenna/data lines and lines carrying
dangerous voltages must be at least 10 mm.
Location:
Never use the ZDUE-450-PLUS-VII in areas where the
operation of radio equipment is prohibited. The appliance
contains a radio transmitter, which can possibly impair the
function of medical electrical appliances such as hearing
aids or cardiac pacemakers.
The ZDUE-450-PLUS-VII is suitable for use in overvoltage
category III. If the appliance is likely to be exposed to higher
transient over voltages than those in overvoltage category III
when connected, it will be necessary to take further safety
precautions for the installation.
Antenna Installation
Indoor and outdoor antennas of the appliance may be
installed and operated only with a minimum distance of 20
cm from humans.
The antennas must be installed and operated so that they
do not interfere with other antennas or devices.
When using a directional antenna, observe the legal limit for
electromagnetic fields (0 Hz to 300 GHz) in public areas.
For details, see Council Recommendations 1999/519/EC of
12 July 1999.
It is mandatory, that the installation of an outdoor antenna
must be carried out by a qualified electrician. The standards
EN 60728-11 and VDE 0855-300 as well as any other
national installation guidelines for lightning protection must
be followed.
Intended Use
The ZDUE-450-PLUS-VII is an appliance designed for the
remote inquiry and the remote monitoring of electricity, heat,
gas and water meters. It is design to be used in mobile radio
networks.
Opening the device
In order to insert the SIM card or to access the DIP switch,
you first have to open the device:
1. First disconnect all the poles of the device from the
electricity supply if it is plugged in there.
2. Now unscrew the screw in the terminal cover and
remove the cover.
3. Now remove the device lid.
1
Mobile Radio Net Status
/ Error LED
2
Status LEDs
3
Service connector
(optional)
4
DIP switch
5
SIM card holder
6
SMA antenna connector
(SMA or FAKRA-D)
7
Service push button
8
Terminal block (RS485)
9
RS232 interface
10
Terminal block
11
Current Loop LED
Altering the device settings
Using the DIP switch (4) you toggle between
the both operating modes, meter mode and
AT command mode.
In meter mode (OFF), the device can be
adjusted locally via one of the meter
interfaces or remotely via communication
commands according to EN 62056-21. In this
case, the device will then operate using the
parameters that have been set.
Setting the PIN of the SIM card
In order to operate the ZDUE-450-PLUS-VII, you will require
a plug-in SIM card (1,8V only) from a mobile radio network
operator.
A ZDUE-450-PLUS-VII having factory default settings
expects a SIM card with a PIN 0000. The PIN setting of the
ZDUE-450-PLUS-VII can be changed in the device settings.
Inserting the SIM card
Insert the SIM card as follows:
1. Open the SIM card holder by gently pressing the lid
towards the edge of the device, flip it up and insert the
SIM card into the slot in the holder. The gold-plated
contacts on the SIM card have to lie on the gold-plated
contacts of the holder when the holder is closed.
2. Close the lid of the SIM card holder and lock it by
pressing it gently back towards the center of the device.
You will feel it click into place.
3. Now replace the device lid and the terminal cover.

ZDUE-450-PLUS-VII Installation Instructions
2
Connecting the device
First connect the meter to the device and then connect the
power supply.
The device is equipped with 3 different meter interfaces,
which are wired in parallel.
RS232 interface (RJ45 jack):
Pin
Signal
Description
2
DSR
Output; Always active
4
GND
Signal ground
5
TXD
Output; Data to meter
6
RXD
Input; Data from meter
Signals and levels according to V.24 / V.28.
All other pins are reserved.
RS485 interface (terminal block):
Signal
Description
RT-
Negative RS-485 interface signal
RT+
Positive RS-485 interface signal
Dual-wire RS485 interface to connect up to 32 transceivers. The
bus connection is terminated to Z=120 Ω (nominal) (RT+ to RT-)
and the cable length is limited to 1000m.
CL1 interface (terminal block) / optional:
Signal
Description
RTX-
Negative CL1 interface signal
RTX+
Positive CL1 interface signal
20mA power interface (current loop) to connect meters with the
power supply interface according to DIN EN 62056-21. Approx. 4
meters can be connected to this interface.
The CL1 interface is optional. Devices without CL1 interface do
not contain RTX- / RTX+ terminals. Please refer to the label of
the terminal block.
Important:
If no meter is connected to the CL1 interface, the RTX- and
RTX+ contacts need to be short-circuited. Otherwise, the
other interfaces will not operate.
Auxiliary power source (HS; terminal block):
Signal
Description
HS-
Negative pole of the auxiliary power supply
HS+
Positive pole of the auxiliary power supply
Auxiliary power supply of 9V / max. 100mA
The auxiliary power supply is not available for all the ZDUE-450-
PLUS-VII models.
Power supply (terminal block):
Signal
Description
L
AC: Unom = 100VAC ... 230VAC
DC: Unom = 60VDC ... 230VDC
N
Inom = 120mA / 40mA
Connecting the antenna
Observe the safety instructions for antenna mounting.
Follow the instructions that came with your antenna. The
antenna must have the following characteristics: passive,
gain <2 dBi, VSWR <2,0:1, impedance 50 Ω, adapted for
the used frequency bands. Use antennas from the
accessory range of the ZDUE.
The antenna is connected to the type SMA or type FAKRA-
D antenna jack (ANT). The antenna should be installed to
ensure adequate signal quality. Ensure that there are no
large metal objects (e.g. reinforced concrete) near the
antenna, as these will adversely affect the signal quality.
Light-emitting diodes (LEDs)
Status LED (2) / Green & Orange with light pipe
Permanently green
Supply voltage available
Permanently orange
Mobile radio module on
Depending on the firmware version, both LEDs may be off in
case of a missing or defect SIM card.
Current Loop LED (11)
Permanently green
Meter connected to the CL1
interface or CL 1 interface short-
circuited.
Mobile Radio Status-/Error LED (1) / Orange
3s on / 3 x flashing
Network search
3s on / 2 x flashing
Error status
3s on / 1 x flashing
SIM/PIN error
Permanently off
Signal strength unknown
3s off / 1-4 x flashing
Signal strength (1=low; 4=high)
Permanently on
Mobile radio connection active
This LED operates in meter mode only.
Service push button (7)
Push the button 1x
for more than 5
seconds.
The parameters being set by EN 62056-
21 commands will be deleted and the
factory settings are reloaded.
Push the button
when switching
the power on.
The actual firmware version will be output
at the local interface (19200bps; 8N1).
The service push button operates in meter mode only.
Simplified EU Declaration of Conformity
Hereby, Sagemcom Dr. Neuhaus GmbH declares that the radio
equipment ZDUE-450-PLUS-VII is in compliance with Directive
2014/53/EU. The full text of the EU declaration of conformity is
available at the following internet address:
www.sagemcom.com/neuhaus
Frequency bands: LTE-Cat-M1
450/700/800/900/1800/2100 MHz
Transmit power: max. 23dBm = 0,2 W (@ 450 MHz)
max. 21dBm = 0,12W (Others)
Copyright Statement
The information published in this document is under copyright. Any
translations, reprints, reproduction and storage in data-processing systems
require the express prior consent of the manufacturer.
Subject to technical modification.
All trademarks and product names are trademarks, registered trademarks or
product names of the respective owners.
All information is based on the information provided by the manufacturer.
No guarantee or liability will be assumed for any error or omissions.
The contents of this handbook and the technical specifications may change
without prior notice. The descriptions of the specifications in these
instructions do not in any way represent a contract.
© Sagemcom Dr. Neuhaus GmbH
Doc.-No. 8239AD002 / Version 1.2 / November 2023