Eagle Nebula User manual

USER MANUAL
Weighing Indicators
Version - 1.00 Manual Rev No - 002
Nebula & Splash


Table of Content
1. Introduction 1
1.1 Nebula Indicator Features 1
1.2 Specifications 1
1.3 Connections 1
1.3.1 RS232 1
1.3.2 Power supply 2
1.3.3 Loadcell Connections 2
1.4 Physical Dimensions 3
1.5 Display and Key Descriptions 4
1.5.1 Display LED 4
1.5.2 Keypad LED 5
1.5.3 Display LCD 6
1.5.4 Keypad LCD 7
1.5.5 Keys and Key Press Description 8
1.5.6 Two Keys Pressing Together Functions 8
2. User Setting (uSEr) 9
2.1 Display Speed (SPd) 9
2.2 Display Brightness (brt) 9
2.3 Display Backlight (dbL) 10
2.4 Display Backlight Colour (bLCL) 10
2.5 Zero Tracking (Ztrc) 11
2.6 Zero Lock (ZLOC) 11
2.7 Power On Zero (PZro) 12

2.8 Peak Hold (PHLd) 12
2.9 Auto Sleep Time (ASLP) 13
2.10 Auto Tare (AtrE) 13
2.11 Liter Factor (LF) 14
2.12 Stability (Stb) 14
2.13 Sensitivity (SEn) 14
2.14 Accumulation (ACC) 15
2.15 Check weighing and Counting (CHC) 15
3. System Setting (SySt) 20
3.1 Date Format (dtF) 20
3.2 Set Date (dtE) 21
3.3 Time Format (tIF) 21
3.4 Set Time (tIE) 21
3.5 Serial Number (SrnO) 22
3.5.1 Serial Number Prefix (SrPF) 22
3.5.2 Serial Number Interval (SrLn) 23
3.5.3 Serial Number Reset (rSt) 23
3.6 Report Format (rPt) 23
3.7 Factory Reset (FrSt) 24
3.8 Unit conversion (UC) 24
3.9 Password (PW) 25
4. Communication Setting (CS) 26
4.1 RS232 (rS232) 26
4.1.1 Baud Rate 26

4.1.2 Data Bits (db) 26
4.1.3 Stop Bit (Sb) 27
4.1.4 Parity (Pr) 27
4.1.5 Data Out (dOUt) 27
4.1.6 Transmission mode (tr) 28
4.1.6.1 OFF 28
4.1.6.2 Continuous transmission (Con) 28
4.1.6.3 Print (Prt) 28
4.1.6.4 Once After Stable (oSt) 28
4.1.6.5 Only Once After Stable (oSto) 28
4.1.6.6 Command (Cd) 29
4.1.7 Starting Character (StCH) 29
4.1.8 String Direction (di) 30
4.1.9 Decimal Point Transmit (dp) 31
4.1.10 Transmit Sign (SG) 31
4.1.11 Leading Zero (LZ) 32
4.1.12 Unit (Ut) 33
4.1.13 Line Feed (LF) 33
4.2 Bluetooth (bLE) 34
4.2.1 Function (FUnC) 34
4.2.2 Bluetooth Name (NA)- 35
4.2.3 Pin (Pn)- 35
4.2.4 Data Out (dOUt) 35
4.2.5 Transmission mode (tr) 35

4.2.6 Starting Character (StCH) 36
4.2.7 String Direction (dI) 36
4.2.8 Decimal Point Transmit (dp) 36
4.2.9 Transmit Sign (SG) 36
4.2.10 Leading Zero (LZ) 36
4.2.11 Unit (Ut) 36
4.2.12 Line Feed (LF) 36
4.3 TTL (ttL) 36
5. Factory Settings 37
5.1 Scale Setting (SCSt) 37
5.1.1 - 5.1.4 Intentionally kept blank
5.1.5 Range (rng) 37
5.1.5.1 SR-Single Range 37
5.1.5.2 DR-Dual Range 38
5.1.5.3 TR-Triple Range 38
5.1.7 Capacity 38
5.1.8 Division (di) 38
5.1.9 Decimal Point (DP) 39
5.1.10 Unit (UNT) 39
5.2 Factory Calibration (FCAL) 39
5.3 Standard Calibration (SCAL) 40
6. MODES 40
6.1 Simple Weighing 40
6.2 Simple Counting 41

6.3 Check Weighing & Check Counting 41
6.4 Accumulation 43

EAGLE Service
We ensure we will provide essential services for
dependable performance of the Nebula indicator.
Congratulations on choosing the quality and precision of
EAGLE. Proper use of your new equipment according to
this manual and regular calibration and maintenance by
our factory trained service team ensures dependable
and accurate operation, protecting your investment.
Contact us about a service agreement tailored to your needs and
budget.
There are several important ways to ensure you
maximize theperformance of your investment:
1. Contact EAGLE for service: The value of a
measurement is proportional to its accuracy –
an out of specification scale candiminish quality,
reduce profits and increase liability. Timely service
from EAGLE will ensure accuracy and optimize
uptime and equipment life.
a. Installation, Configuration, Integration and
Training: Our service representatives are
factory trained, weighing equipment experts.
We make certain that your weighing
equipment is ready for production in a cost
effective and timely fashion and those
personnel are trained for success.
b. Initial Calibration Documentation: The
installation environment and application
requirements are unique for every industrial
Terminal so performance must be tested and
certified. Our calibration services and
certificates document accuracy to ensure
production quality and provide a quality system
record of performance.
c. Periodic Calibration Maintenance: A calibration
service agreement provides on-going
confidence in your weighing process and

documentation of compliance with
requirements. We offer a variety of service
plans that are scheduled to meet your needs
and designed to fit your budget.
© EAGLE 2023
No part of this manual may be reproduced or
transmitted in any form or by any means, electronic
or mechanical, including photocopying and recording,
for any purpose without the express written
permission of EAGLE.
This documentation is furnished with Restricted Rights.
Copyright 2023 EAGLE. This documentation contains
proprietary information of EAGLE. It may not be
copied in whole or in part without the express written
consent of EAGLE.
EAGLE reserves the right to make refinements or
changes to theproduct or manual without notice.
COPYRIGHT
EAGLE ® is a registered trademark of E.G.
KANTAWALLA PVT. LTD. all other brand or product
names are trademarks or registered trademarks of
their respective companies.
EAGLE RESERVES THE RIGHT TO MAKE
REFINEMENTS OR CHANGES WITHOUT NOTICE.
Statement regarding harmful substances.
We do not make direct use of harmful materials such
as asbestos, radioactive substances or arsenic
compounds. However, we purchase components from
third party suppliers, which may contain some of
these substances in very small quantities.

WARNINGS AND CAUTIONS
CAUTION! Read this manual before operating or
servicing this terminal. Follow these instructions
carefully. Do not allow untrained personnel to
operate, clean, inspect, maintain, service, or tamper
with this terminal. Always disconnect this
equipment from the power source before cleaning
or performing maintenance. Call eagle for parts,
information, and service.
WARNING! Only permit qualified personnel to
service this terminal. Exercise care when making
checks, tests and adjustments that must be made
with power on. Failing to observe these precautions
can result in bodily harm.
Warning! For continued protection against shock
hazard connect to properly grounded outlet only. Do
not remove the ground prong.
Warning! Disconnect all power to this unit before
removing any connection, opening the enclosure or
servicing.
Warning! Before connecting/disconnecting any
internal electronic components or interconnecting
wiring between electronic terminal always remove
power and wait at least thirty (30) seconds before
any connections or disconnections are made. Failure
to observe these precautions could result in damage
to or destruction of the terminal or bodily harm.

Disposal of Electrical and Electronic Equipment
In conformance with the European Directive
2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE) this device may not be
disposed of in domestic waste. This also applies to
countries outside the EU, per their specific
requirements.
Please dispose of this product in accordance with
local regulations at the collecting point specified
for electrical and electronic equipment.
If you have any questions, please contact the
responsible authority or the distributor from which
you purchased this device. Should this device be
passed on to other parties (for private or
professional use), the content of this regulation
must also be related.
Thank you for your contribution to environmental
protection.

1 | P a g e
1. Introduction
1.1 Nebula Indicator Features
1) Light and Compact indicator.
2) Available with Stainless Steel & ABS Housings.
3) Stainless Steel Indicator with water proof protection.
4) Available in LED and LCD variants.
5) LCD Variant with colour changing display.
6) Lithium Ion battery operated
7) RS232 & TTL Port.
8) C type USB charging power adapter in ABS Indicator.
9) Weighing, Check weighing, Piece Counting, Check Counting,
Accumulation, Peak Hold modes.
10) Provision to connect with portable thermal printer.
11) Optional Bluetooth and RTC Card.
1.2 Specifications
Display Colour (LCD Version)- White, Yellow, Green, Red, Blue
Display Colour (LED Version)- Green
Battery -7.4, 2600 mAH
Adapter- 5V 2.4A USB C type in ABS indicator.
1.3 Connections
1.3.1 RS232(SS Indicator)
RS 232 serial connection is done with three wire as indicated below
RS232 serial interface connections
RS232 Pin Connection for ABS Indicator is as per below,
1. VCC
2. Rx
3. TX
4. GND

2 | P a g e
1.3.2 Power supply
The quality of the instrument’s ground will determine the
accuracy and the safety of your measuring system. A proper ground
connection is needed to minimize extraneous electrical noise affects on
the measurement. A poor ground can result in an unsafe and unstable
operation. It is important that the instrument should not share power
lines with noise-generating parts such as heavy load switching relays,
motor control equipment, inductive loads, etc. If the condition of the
power line in the plant is poor, prepare a special power line and grounding.
5 V 2.4 A power supply adapter is used for the power supply. Power
supply pin (C type USB) is located at the rear side of the indicator in the
ABS indicator and and in SS variant the power supply cable will go
directly inside the indicator.
5V 2.4 A c type connector of ABS Nebula indicators.
1.3.3 Load Cell Connections
Suggestion: After load cell connected to the instrument, the welding
Near to the load cell is not recommended. Disconnect all
Connector’s from indicator before welding on the mechanical hardware.
Load cell connection detail is shown in Figure below.
In 4-wire installations the sense and excitation pins with the same
Polarity should be short circuited at the connector side.

3 | P a g e
Load cell Pin Connection as per below,
1. Ex+
2. Ex –
3. Sig +
4. Sig –
5. Shield
Load cell Connector SS & ABS Variant 4 wire LC connection
1.4 Physical Dimensions
Please refer below dimension of Nebula SS Indicator and ABS
Indicator
(Nebula SS Indicator Dimensions)

4 | P a g e
(Nebula ABS Indicator Dimensions)
*All dimensions in mm
1.5 Display and Key Descriptions
1.5.1 Display LED
The display of Nebula indicator comes with two variant displays - LED
and LCD.
1. The following are the indications in the LED display.
Gross - Indicates the displayed weight is gross weight.
Net - Indicates the displayed weight is net weight.
Tare - Indicates the displayed weight is Tare weight.
- Indicates the scale counts are in zero condition.
- Indicates the weight value is unstable.

5 | P a g e
Units - g, kg, lb, L, t units are available on the right of the display.
- Indicates the bluetooth function is enabled.
- Indicates the check weighing or counting function is enabled.
- Indicates the accumulation is enabled.
- Indicates the counting function is enabled.
P.Hold- Indicates the peak hold function is enabled.
- It is used in check weighing or check counting function. If this
Yellow LED glows then the weight or count is under range.
- It is used in check weighing or check counting function. If this
Green LED glows then the weight or count is in OK range.
-It is used in check weighing or check counting function. If this Red
LED glows then the weight or count is in over range.
1.5.2 Keypad LED
The keys and the key functions of Nebula are:

6 | P a g e
Nebula LED indicator Keypad
1.5.3 Display LCD
2. The following are the indications in the LCD display.
GW - Indicates the displayed weight is gross weight.
NW - Indicates the displayed weight is net weight.
- Indicates the scale counts are in zero condition.
- Indicates the weight value is unstable.
Units - g, kg, lb, L, t units are available on the right of the display.
- Indicates the bluetooth function is enabled.
- Indicates the check weighing function is enabled.
- Indicates the accumulation is enabled.

7 | P a g e
- Indicates the counting function is enabled.
P.Hold- Indicates the peak hold function is enabled.
- It is used in check weighing or check counting function. If this
Yellow LED glows then the weight or count is under range.
- It is used in check weighing or check counting function. If this
Green LED glows then the weight or count is in OK range.
-It is used in check weighing or check counting function. If this Red
LED glows then the weight or count is in over range.
1.5.4 Keypad LCD
The keys and the key functions of Nebula are:
Nebula LCD Indicator Keypad

8 | P a g e
1.5.5 Keys and Key Press Description
1.5.6 Two Keys Pressing Together Functions:
1) Func/Esc + M+/MR - Show the total accumulated weight.
(This will work when the Accumulation mode in ON)
2) Func/Esc + Tare - Show the total accumulated times
(This will work when the Accumulation mode in ON)
3) Func/Esc + Zero - Clear accumulation memory (including weight and
accumulated times) and also to end the accumulation print (if print format
4 is enabled)
Note: -To come out, press ESC key
4) N/G + Tare - This is for unit Conversion (Till you press and hold, it
will show the converted value. At a time, any one conversion is possible
and it works on First come first serve basis)
5) Func/Esc + N/G - Bluetooth name
Note: -To come out, press ESC key
Keys
Func/Esc
Zero
Tare
N/G
M+/MR
Short press
from home
screen
To enter into
counting
mode
To make
the
display
zero
To tare the
weight kept
on the
platform
To toggle
between Net
and Gross if
tare is used
Print and
accumulate
Long Press
from home
screen
To enter into
parameter
settings
To show
the
firmware
version
number
To show the
machine
serial number
(In the 14
segment display it
will show snr)
To show the
month and
year of sale
(In the 14
segment display it
will show sdt)
ON/OFF
Short press
when inside
the settings
To go to
previous
step
Increment
Decrement
Shift (Right
to Left)
Enter
Long press
during ON
(Count
Down)
To enter into
factory
settings
-
-
To show the
month and
year of
production (In
the 14 segment
display it will show
pdt)
-

9 | P a g e
6) Zero + Tare - Recalling Location number for check weighing or check
counting
(After pressing these two keys together, display will show the last
selected Location and increment key/decrement key is used to scroll from
W1 to W9, C1 to C9 and also to NIL. When NIL is selected, no check
weighing or check counting function is activated)
7) N/G + Tare – For engineering purpose.
This manual suits for next models
1