Eastech Flow Controls Vantage 2210 User manual

Vantage Series 2200
Scope
This manual contains information concerning the installation, operation and maintenance
of the Vantage 2210/2220. To ensure proper performance of the unit, the instructions
should be thoroughly understood and followed.
Keep the manual in a readily accessible location for future reference.
Changes and additions to the original edition of this manual will be covered by a
“CHANGE NOTICE” supplied with the manual. The change notice will identify the sections in this
manual where the changes have occurred.
Vantage 2210/2220
Table of Contents
General Specifications............................................................................................................ 1-2
General Description................................................................................................................ 1-3
Installation Procedure.......................................................................................................... 2-1
Enclosure Mounting............................................................................................................... 2-1
Sensor Mounting .................................................................................................................... 2-2
Wiring Connections................................................................................................................ 2-4
Splice Procedure..................................................................................................................... 2-6
Programming........................................................................................................................ 3-1
QuikCal Menu Functions ....................................................................................................... 3-1
Program ............................................................................................................................. 3-2
Level Mode...................................................................................................................... 3-2
Flow Mode....................................................................................................................... 3-3
Totalizer........................................................................................................................... 3-5
4-20 Output...................................................................................................................... 3-6
Setpoints........................................................................................................................... 3-6
Sensor Calibration............................................................................................................ 3-6
Damping........................................................................................................................... 3-7
Lost Echo......................................................................................................................... 3-7
Flow Simulation............................................................................................................... 3-7
Integrator.......................................................................................................................... 3-7
Pump Alternation............................................................................................................. 3-7
Relay Assignment............................................................................................................ 3-8
Status.................................................................................................................................. 3-8
Data Logger....................................................................................................................... 3-9
System Setup...................................................................................................................... 3-10
Calibration......................................................................................................................... 3-11
Sensor Mounting for Primary Elements ............................................................................ 3-12
Parts List............................................................................................................................... 3-17
Section
1
1-1 942001Rev
1.06 04/07

Vantage Series 2200
General Specifications
Span Range
FB5: 0 to 25 feet. Maximum total range including offset distance is 26 feet.
Minimum offset 1 foot.
FB7: 0 to 15 feet. Maximum total range including offset distance is 16 feet.
Minimum offset 1 foot.
FB3: 0 to 50 feet. Maximum total range including offset distance is 52 feet.
Minimum offset 2 feet.
Outputs
4-20 mA DC isolated; 800 ohms max.
Up to Five programmable relays, SPDT .25 amp @ 120 VAC, .5 amp @ 24 VDC.
RS-232 Serial Port, 9600 – 36500 Baud, Modbus™ Protocol
RS-485 Serial Port optically isolated, Modbus™ Protocol (2220 only)
Display 4 line, 20 characters per line backlit LCD display.
Programming Front panel mounted 16 button keypad.
Power 90/240 VAC, 50/60 Hz, or 12 VDC @ 150 mA continuous.
Accuracy FB5: ±0.02" or ±0.05% of target distance
FB7: ±0.02" or ±0.05% of target distance
FB3: ±0.1" or ±0.1% of target distance
Sensors
FB5:
Temperature Range: -20° to 160° F (-30° to 70° C) 50 kHz
Operating Frequency: 60 KHz
Beam Angle: 6° included at -3dB Boundary
Housing: PVC body, PVC cap
Cable: 2 twisted pair, foil shielded, standard lengths of 32 feet
(10 meters) or 65 feet (20 meters). May splice up to
1000 ft maximum of Belden 8728 or equal
FB7:
Temperature Range: -20° to 160° F (-30° to 70° C) 50 kHz
Operating Frequency: 51 KHz
Beam Angle: 8° included at -3dB Boundary
Housing: Tefzel™ body, Teflon™ cap
Cable: 2 twisted pair, foil shielded, standard lengths of 32 feet
(10 meters) or 65 feet (20 meters). May splice up to
1000 ft maximum of Belden 8728 or equal
FB3:
Temperature Range: -40° to 200° F (-40° to 90° C) 30 kHz
Operating Frequency: 30 KHz
Beam Angle: 14° included at –3dB Boundary
Housing: Glass filled polyester / Glass reinforced epoxy face
Cable: 100 feet of 2 twisted pair, foil shielded. May splice up
to 300 ft maximum of Belden 8728 or equal.
The maximum range will be reduced for cable lengths over 300 feet. The
range is reduced 1 foot for every 100 feet of cable over 300 feet.
Electronic
Enclosure IP66/NEMA 4X standard, temperature range: -4° to 158° F (-20° to 70° C)
Optional with heater, temperatures down to -40° F (-40°C)
Optional Modem 14400 BBS data speed
1-2

Vantage Series 2200
General Description
The following description applies to both the Vantage 2210 and 2220.
The Vantage series 2200 is an ultrasonic level/flow meter. Its' design allows it to be easily programmed
as a level meter or an open channel flow meter. It can operate two ultrasonic sensors and can be
programmed for two level applications, two flow applications or one level and one flow application.
The Vantage series 2200 is supplied with a backlit LCD display with 4 lines and 20 characters per line.
In the normal mode the display has two pages with up to 4 lines that can be assigned to each page. The
pages are switched by pressing the UP/NEXT key. The information for each line can be assigned and
arranged at the user’s discretion. The backlight of the display can be programmed on or off or timed
off.
The programming of the unit is accomplished with the 16 button keypad by means of a drill down type
menu structure. The meter stores the steps taken when it was previously programmed. So when a
programming parameter needs to be changed, the user can quickly get to the screen to make the change.
The display screens can be viewed in three languages: English, Spanish and German.
The Vantage series 2200 can be programmed to operate on four different types of sensors. The standard
sensors used with the meter are the FB7, FB3 and FB5. The FB7 sensor is normally used for flow or
level measurements up to 15 feet. The FB4 sensor is used for flow or level measurements up to 25 feet.
The FB3 sensor is used for level measurements of up to 50 feet. The fourth sensor that can be used with
the meter is the FB1 sensor that was used with the previous Models 2100 and 2500.
When used as an open channel flow meter, the Vantage series 2200 has most of the commonly used
flumes and weirs stored in memory. For special open channel primary devices, the user can input a
Head vs Flow table, or an equation with a K and power factor.
The Vantage series 2200 is also capable of being programmed for pump alternation control when used
as a level meter. Up to three setpoints and four relays can be used for this function.
The meter also contains a data logger. It will display daily summaries for totals
over the last eight days. The logged data can also be displayed in graphic form on the
display. With the Vantage DDS software the logged data can be downloaded and converted to a csv
(comma separated variable) file to be imported into a spread sheet program such as Excel™.
The 2200 has self diagnosis and any faults, or tripped setpoints, will be displayed if the alarms are
assigned to one of the display lines. The following are the alarms that could be displayed:
LP#1 – Indicates that the 4-20 mA output loop is open.
Int – Will flash on when contact integrator activates.
Ovr – Indicates that the flow, or level, is above the maximum flow or level.
SP# – Indicates that a setpoint has been tripped.
Sig – Indicates that the meter is not receiving a signal from the sensor.
1-3

Vantage Series 2200
Installation
Enclosure Mounting
The enclosure is rated IP 66 (NEMA 4X). A sunshade is recommended for outdoor installation. There
are two stainless steel mounting brackets factory assembled to the enclosure. The mounting feet have slots
for ¼" bolts (4 places). The electronics should be mounted with the display at eye level or lower. There are
three ½ inch holes in the bottom of the enclosure for conduit fittings. These holes have rubber plugs
installed at the factory. The holes used for wiring must be properly prepared and sealed to maintain rating. If
you do not use all three holes for conduit, leave the rubber plugs in the holes to protect the enclosure ratings.
Opening the Enclosure:
There are two hinged door clasps on the front cover of the enclosure. To open, put thumb on one of the
hinges, pull toward the outside of the enclosure. Once the hinge pops to the outside it will lower allowing
the clasp at the bottom of the hinge to release. Swing the cover towards the front to open. The opposite side
will act as a hinge to swing the door freely. To close, clasp the bottom side of the hinge and push the top of
the hinge toward the enclosure until it locks.
Hinge Lock and Optional Door Lock
There are two plastic gray plugs supplied with the Vantage series 2200. These plugs may be used to
permanently disable one side of the hinged handles. If an optional door lock was supplied with the unit then
one side of the hinge handle should be plugged and the other side will have the key lock used. Either side
hinge handle may be disabled. Insert the gray plug into the keyhole. Warning: This will permanently
disable the hinge handle. The other side can be used for the key provided for the optional lock.
Note: The key will need to be left in the hinge handle if the door is to remain unlocked. The only way
the key can be removed is if the hinge handle is locked.
Note: When supplied with the optional modem the enclosure height is 12.875” instead of 9.319”.
Section
2
2.756"
8.25"
9.319"
8.917"
9.496"
2-1

Vantage series 2200
4.00"
2.75"
1.88"
3.00"
5.00"
.38"
.38"
FB7
0.43"
2.08"
2.71"
3.50"
2.50"
ORing
1"NPT
2"NPT
FB5
2-2
Sensor Mounting Bracket Dimensions:
The Vantage series 2200 is supplied with a stainless
steel mounting bracket. The mounting bracket should
be leveled in both plains. The 2200 sensor will be
mounted to the 1 inch hole in the mounting bracket.
Remove the top 1 inch nut from the sensor, slide the
cable through the slot in the bracket, and slide the 1
inch nipple on the sensor up through the 1 inch hole
in the bracket. Replace the 1 inch nut on the nipple
and tighten, or screw conduit fitting to nipple and
tighten to secure sensor. Adjust other nut if
necessary. Do not over tighten the nut.
FB7 Sensor Dimensions:
The FB7 sensor is used with flow or level applications
where the maximum head rise (maximum level) is
15.00 ft or less. (See specifications Page 1-2.) There is
a 1 inch NPT threaded nipple on top of the sensor for
mounting on optional mounting bracket and a 2 inch
NPT thread on the bottom barrel of the sensor for tank
mounting. An o'ring is provided on the sensor if
mounting with 2" NPT thread. This o'ring must be
used or sensor operation will be affected.
FB5 Sensor Dimensions:
The FB5 sensor is used with flow or level applications
where the maximum head rise (maximum level) is
25.00 ft or less. (See specifications Page 1-2.) There is
a 1 inch NPT threaded nipple on top of the sensor for
mounting on the mounting bracket.

Vantage series 2200
FB3
3.25"
5.50
"
4.00"
VmT
Zero
Span
Offset
2-3
FB3 Sensor Dimensions:
The FB3 sensor is used for applications where the
maximum level is 50 feet. (See Specifications Page
1-2.) There is a 1 inch NPT threaded nipple on top
of the sensor for mounting on bracket or customer
provided flange.
Sensor Mounting Terms:
Whether the sensor is to be used for flow in
conjunction with a primary element such as flumes
or weirs, or used for level measurement only, there
are two terms that must be understood to mount the
sensor properly in either application.
1. The VMt is the vertical mounting distance
between zero level (zero flow level of flumes or
weirs or zero level of tank) and the bottom of the
sensor. This is the offset plus the span.
Consideration must be given when mounting the
sensor in respect to sidewalls. The distance from a
wall to the FB7 sensor is 0.875 inch per foot of VMt.
For the FB3 it is 1.5 inches per foot of VMt.
2. The HMt is utilized in flow applications. It is the
upstream horizontal distance that the sensor needs to
be placed from a reference point of the primary
element. The HMt dimension is displayed on the
2200 when programming the unit. Refer to Page 3-
12 for mounting detail for various primary elements.
Mt
Span
Offset Minimum
FB5&7 – 12"
FB3 ----
–
24"
V
ertical
Mounting
Dimension
1" NPT

Vantage Series 2200
Sensor Wiring
All of the terminal connections for the output signals are depicted in the drawing. The wiring
connections are also on the inside of the enclosure terminal access cover.
TERMINALS A: BOTTOM ROW TERMINALS B: TOP ROW
1=Hi, Red, Single sensor wire, transmit 1= Red, Hi, Sensor #2 dual, transmit
2= Lo, Black, Single sensor wire, receive and Shield wire 2= Black, Lo, Sensor #2 dual, receive and Shield wire
3= White, Hi, Single sensor wire, temperature 3= White, Hi, Sensor #2 dual, Temperature
4= Green, Lo Single sensor wire, ground 4= Green, Lo, Sensor #2 dual, Ground
5= #1 Compensator 5= Red, Hi, Sensor #1 dual, transmit
6= #2 Compensator 6= Black, Lo, Sensor #1 dual, receive and Shield wire
7= GND Compensator 7= White, Hi, Sensor #1 dual, temperature
8= NO, Relay 2 8= Green, Sensor #1 dual, Ground
9= C, Relay 2 9= NO, Relay 5
10= NC, Relay 2 10= C, Relay 5
11= NO, Relay 1 11= NC, Relay 5
12= C, Relay 1 12= NO, Relay 4
13= NC, Relay 1 13= C, Relay 4
14= N/A 14= NC, Relay 4
15= N/A 15= NO, Relay 3
16= +(Positive), Powered 4-20mA DC Output #1 16= C, Relay 3
17= -(Negative), Powered 4-20mA DC Output #1 17= NC, Relay 3
18= Hi, DC Battery Input 18= +(positive) 12VDC Output
19= Lo, DC Battery Input 19= Gnd (negative) 12VDC Output
20= TX, Rs232, Computer’s RX 20= +(Pos.), Powered 4-20mA DC Output #2, 2220 only
21= RX, Rs232, Computer’s TX 21= -(Neg.), Powered 4-20mA DC Output #2, 2220 only
22= RTS, Rs232 22= N/A
23= CTS; Rs232 23= N/A
24= N/A 24= N/A
Note: All relays are optional and can be added in the field. PN 544718-0001.
VAC Input= TB1,
HI (Hot),
LO (Neutral),
GND (Ground)
Relay 1Relay 2 Relay 3
Relay 4
Relay 5
Fuse,
4-20ma -
Fuse,
4-20ma +
4-20ma
Test Points
2-4
AC Power
Fuse .25 Amp
Heater
Fuse
Optional
Heater
AC Power
Connection
Green LED
Indicates ac
power is applied

Vantage Series 2200
Inserted Components:
There may be a component added to the terminal strips on the Compensator
inputs. The component size and type is determined by various
Cable lengths and the style of sensor chosen. For cable lengths 5 to 150ft
A 6800pf capacitor will be installed on terminals 5 & 7 for a single sensor.
For Single Sensor: The component will be attached to Row A,
Terminals 5 & 7.
For Dual Sensors: The components will be attached to Row A,
Terminals 5 & 7 and 6 & 7.
Adding or Cutting Cable Lengths:
If adding or cutting cable lengths in the field the component style or
Value may change. Be sure to due a frequency tune on any new sensor
Or electronics that are replaced. (Page 11)
Other Wiring:
FB sensor with Grey cable: FB sensors with Orange Cable:
Belden 9365 Belden 3124A
Terminal Wire Terminal Wire
A1 Red A1 Red
A2 Black A2 Black
A3 White A3 White
A4 Shield A4 Green & Shield
FB1 sensor with Yellow Triax cable: Belden 9222
Terminal Wire
A1 Center conductor
A2 Middle shield
A3 Outer shield
A4 Open
If the unit is supplied with the modem option, the telephone connection is made on the Relay #5 NO
(TIP) and NC (RING) terminals as shown below.
2-5
Top Row B:
For Dual Sensors Only
Bottom Row A:
For Single sensor only.
2-5
Telephone connections for modem
TIP RING

Vantage Series 2200
Vantage series 2200 Splice Procedure
When additional cable length is required, cable can be spliced up to a total length of 1000 feet. The
cable provided with the 2200 sensor has 2 twisted pairs with shields around each pair and a shield
around both pairs. Eastech Flow Controls Inc. splice kit part number is 544700-0001 which will include
butt splice connectors and coax seal strips. Prepare the sensor end wire and the wire to be spliced per
the following instructions. Must use Belden type 8728 or equal, 2 twisted pairs, 22 awg (7x30)
shielded wire.
1. Slice the outer cover on the wire and spread open to expose the foil on the two wire pairs. Be
careful not to slice into the foil or the inner wires.There will be two pairs of wires, each being
covered by a colored foil. Split
the two pairs of wire.
2. Use electrical tape and tape
the end of the foil on each
bundle to keep the bundles
separate.
3. Place the end of the wires
into the butt splice connectors.
Use pliers and crimp the round
part of the butt splice
connectors. There will be seven
(7) splices which includes the
four colored wires and three
shields.
4. Wrap each spliced pair with
its shield wire with electrical
tape.
5. Use the two strips of coax
seal to wrap the entire splice
after verifying operation of the
unit.
2-6
2-6

Vantage Series 2200
Section
3
3-1
Flw1 00 GPM
1T 00x10 GAL
Lvl1 00 In
Alm Sig 4-20
QuikCal Menu Functions
The screen to the left represents the normal screen. Up to eigh
t
lines may be assigned to the normal screen. Pressing the UP/Next
key will switch to the second four lines and back. To program,
recalibrate or change any function in the Vantage series 2200,
p
ress the “MENU” key. This will display the main menu
selections for all of the functions of the Vantage series 2200
QuikCal firmware. Below is a quick reference for the main menu
and a brief description of each to allow the user to navigate to the
MENU
>01) Review Meter Selection of this will display the parameters that the meter is programmed. (e.g. Max
level, Offset, VMt, Totalizer, Lo
gg
er, etc>
)
>02) Program 01) Level/Vol To program for use as a level meter.
02) Flow To program for use as a flow meter.
03) Totalizer To select totalizer engineering units and multiplier.
04) 4-20 Out To adjust or assign to 4-20ma output.
05) Setpoints To assign setpoints. (e.g. Hi or Lo alarms)
06) Sensor Cal To calibrate distance calibration from target to face of sensor.
07) Damping To adjust damping time.
08) Lost Echo To adjust Lost echo time and Fail to zero or span.
09) Simulation To simulates flow or level outputs.
10) Integrator To set contract integrator time for relay.
11) Pump Alternation Selection of setpoint for pump alternations.
12) Relays Relay assignment for all relays.
>03) Status 01) Sensor To review signal strength, temperature and gain.
02) Level To review distance level.
03) Alarms/Relays To review alarms tripped and 4-20 loop.
04) Logger To review time, store at times, amount stored and amount left
for logging.
05) History To review logged channel history.
06) Daily Sum To review daily total, minimum and maximum flows.
>04) Data logger 01) Set Time/date To set the time and date for the Vantage series 2200
02) Storage Rate To set logger storage intervals.
03) Secondary To set secondary logging interval based on a set point.
04) Log channels To set channels to log and values to log.
05) Clear data To clear all stored logger data.
>05) System Setup 01) Language To set unit to display language to be used.
02) Display To set display contrast and backlighting.
03) Communications To set communication parameters and enable modem.
04) Display lines To assign up to eight lines to be displayed on the main screen.
05) Sensors Used To select the type and quantity of height sensor to be used.
06) Rly Pulse Wdt To set contact closure time of relays.
07) Totals Reset To reset the totalizer.
08) New Password To change password.
09) Summary Reset To clear daily summary.
10) Meter reset To reset to factory defaults.
11) New Firmware To upload new firmware into meter.
>06) Calibration 01) Flow Simulation To check flow simulation of H vs Q.
02) 4-20 Adjustment To adjust 4-20ma output signal.
03) Sensor Cal. To adjust distance calibration from target to face of sensor.
Same as sensor cal. under program menu.
Use the UP or
DOWN key to scroll
through the
selections. Press the
numbers to make a
selection.

Vantage Series 2200
>2) Program
Programming for Level/Volume Applications
From the main screen press the MENU key, then the number 02 keys.
Enter Security ID (00000000 from the factory), press the ENTER key
and then the number 01 key. The screen to the left will be visible on
the display.
Press the number on the keypad that corresponds to the engineering
units desired. Use the UP or DOWN button to move the list up or
down.
The next screen shown to the left is for selecting the volume units if the
meter is to be setup to display volume. If None is selected, the next
screen will be the entry of the maximum level to be measured and the
sensor offset. See page 2-3 for picture defining Max Level and Offset.
If a volume unit is selected, the screen to the left will appear. This is to
select the number of decimal to the right to display the volume units.
The next screen gives the choices for the type of tank being monitored.
The linear is for vertical standing circular tanks of rectangular tanks.
The Data Points selection allows the user to input up to 32 level verses
volume special curve. When this is selected a data entry screen will
appear. The Horiz Circ selection is for horizontal circular tanks.
The next screen is to enter the maximum tank level and the volume for
the maximum level. The cursor will be under the first digit of the level
value. Use the number keys to enter the desired value. If you need to
enter a larger number than the one displayed, use the DOWN/Left
arrow key to move the cursor to the left. After the last digit is entered
or the Enter Key is pressed the next screen will appear.
The sensor # and the previously programmed level units will be
displayed. If two sensors are used, the screen will prompt for which
sensor to calibrate. Move the cursor to the desired location in Max
level by using the DOWN/LEFT arrow key. Enter the maximum range
that the unit is to be programmed. Use the UP/NEXT key to drop the
cursor to program the offset region. Enter the sensor offset value.
The next screen will allow the user to assign the 4-20mA output. The unit can be programmed to
have 4.00mADC at zero level and 20.00mADC at span level or 20.00mADC at zero level and
4.00mADC at span level. Press 1 on the keypad to toggle the desired assignment. Press the
ENTER key.
If two sensors are being used, reenter the Program menu and repeat the previous programming
procedure if second sensor is for level or go the next page if for flow.
3-2
Level Units
01)Inches
02)Feet
03)Meters
Sensor #1
Units- Inches
Max Level 50.00
Offset 12.00
Volume Units
01)None
02)GAL
03)MET3
Program/Cal.
01)Level/Vol
02)Flow
03)Totalizer
Display Format
01)#.
02)#.#
03)#.##
Choose Tank Type
01)Linear
02)Data Points
03)Horiz Circ
Enter Tank Maximum
Level and Volume
Lvl= 50.00 In
Vol= 7500 Gal

Vantage Series 2200
>2) Program Continued
Programming for Flow Applications
From the main screen press the MENU key, then the number 02 keys.
Enter Security ID (00000000 from the factory), then press the ENTER
key and then the 02 keys. The Level Units screen will be visible on the
display. If two sensors are being used, the next screen requires the
selection of the sensor for which the programming applies.
Press the numbers on the keypad that corresponds to the engineering
units desired. Use the UP or DOWN button to move the list up or down.
Select the flow engineering unit desired by pressing the number in front of
the selection. Units available are:
01) GPM, gallons/minute 06) CFD, cubic foot/day 11) MS3, cubic meters/second
02) GPD, gallons/day 07) LPS, liters/second 12) M3H, cubic meter/hour
03) MGD, million gallons/day 08) LPM, liters/minute 13) M3D, cubic meter/day
04) CFS, cubic foot/second 09) LPD, liters/day 14) IGM, imperial gallons/minute
05) CFM, cubic foot/minute 10) MLD, million liters/day 15) BPH; barrels/hour
The FLOW DISPLAY FORMAT screen asks how many digits you want to show to the right of the
decimal point. Press the number that corresponds to your selected value:
01) #. 02) #.# 03) #.## 04) #.###
Example: GPM, #., will show a direct flow reading (e.g. 100 GPM)
The next three screens will be the selection for the type and size of primary element:
3-3
Program/Cal.
01)Level/Vol
02)Flow
03)Totalizer
>01) Flumes 01) Parshall 1) 2 inch 4) 9 inch 7) 24 inch 10) 60 inch
2) 3 inch 5) 12 inch 8) 36 inch 11) 72 inch
3) 6 inch 6) 18 inch 9) 48 inch 12) 84 inch
13)96inch
02) Manhole 1) 4 inch 3) 8 inch 5) 12 inch
2) 6 inch 4) 10 inch
0
3) Palmer Bowlus 1) 6 inch 4) 12 inch 7) 21 inch
2) 8 inch 5) 15 inch 8) 24 inch
3) 10 inch 6) 18 inch
Plasti-Fab HQ curves are used for the above, if other manufacturer use SPECIAL.
0
4) Trapezoidal 1) Small V60 3) X-Large V60
2) Large V60 4) 3.0 Ft V60
Plasti-Fab HQ curves are used for the above.
0
5) H Flume 1) H 4.5 Ft 4) HS .6 Ft
2) HL 4.5 Ft 5) HS .8 Ft
3) HS .4 Ft 6) HS 1.0 Ft
Plasti-Fab HQ curves are used for the above.
0
6) Lagco 1) 6 inch 4) 12 inch 7) 21 inch
2) 8 inch 5) 15 inch 8) 24 inch
3) 10 inch 6) 18 inch
Level Units
01)Inches
02)Feet
03)Meters
Level Units
01)Inches
02)Feet
03)Meters

Vantage Series 2200
>2) Program Continued
3-4
>02)Weirs 01) V-Notch 1) 11.25 degree 4) 45 degree
2) 22.5 degree 5) 60 degree
3) 30 degree 6) 90 degree
0
2) Contracted 1) 12 inch 5) 36 inch 9) 96 inch
2) 18 inch 6) 48 inch 10) 120 inch
3) 24 inch 7) 60 inch 11) Other . . .
4) 30 inch 8) 72 inch
0
3) Suppressed 1) 12 inch 4) 36 inch 7) Other . . .
2) 18 inch 5) 48 inch
3) 24 inch 6) 60 inch
0
4) Cipolletti 1) 12 inch 4) 36 inch 7) Other . . .
2) 18 inch 5) 48 inch
3) 24 inch 6) 60 inch
NOTE: If you have a special length weir plate you will need to select Other. Enter the width of the weir.
>03)Nozzles 01) Open Flow 1) 6 inch 4) 12 inch 7) 18 inch
2) 8 inch 5) 14 inch 8) 20 inch
3) 10 inch 6) 16 inch 9) 24 inch
0
2) Kennison 1) 8 inch 2) 10 inch 3) 12 inch
>04) SpecialNote: If the equation or data input is to be used, you must program the flow
engineering units and the level engineering units into the unit first before using this function. (e.g. Q
= CFS and H = head in feet, program the 2200 for CFS and FT. Once the SPECIAL program is
completed you may change the flow units and level units to the desired units.
1) Q=KH^PWR Enter the K value by using the number and decimal keys. Use the LEFT
arrow key to position the cursor for the number of digits to be entered. Once the last digit is
entered in the K selection the cursor will drop to the Power input. Enter the Power function.
Press the ENTER key.
2) Data Input Enter Level and Flow in selected engineering units by using the number
and decimal keys. Use the LEFT arrow key to position the cursor for the number of digits to
be entered. Once the last digit is entered in the Level section the cursor will drop to the Flow
input. There are a maximum of 32 points available for H/Q input. It is recommended that
you do not use less than 10 points. Use the last point input as zeros, this will automatically
advance to the next screen.

Vantage Series 2200
>2) Program Continued
3-5
Flow Primary Element
Max Flow ***.**
VMt= **.**
HMt= **.**
Once the primary element type and size is selected the screen at
the left will appear. This screen displays the Maximum flow of
the Primary Element, the suggested VMt (vertical mounting
distance of the bottom of the sensor to zero flow, and the HMt
(horizontal mounting distance of the sensor for the primary
element chosen). The VMt dimension is a recommended
mounting distance. If you choose to relocate the sensor head
change the VMt distance in the next screen. These values will
not change. They are for reference only. Press the ENTER key.
This screen allows the user to change the maximum flow rate
and the vertical mounting distance (VMt). To change the
maximum flow rate use the LEFT arrow key to go to the most
significant digit. Press the number wanted on the keypad, this
will send the cursor to the next number. Once all numbers have
been entered the cursor will drop to the VMt line displayed. If
the user chooses to change the VMt of the sensor, enter the
number by using the keypad. If a VMt value entered is less
than the minimum offset plus the head rise of the selected
maximum flow rate, the minimum the VMT value will not
change. Press the ENTER key and then the MENU key. If the
new parameters are to be stored then press the ENTER key. If
you do not wish to save the new parameters, press any other
key. You are now back at the programming selection list. Press
the MENU key to return to the normal display screen.
Totalizer Setup
Press the 03 keys when in the Program selection list to program the
totalizer. The next screen will be the engineering unit selection. The
available options for the engineering units are:
01)GAL, Gallons 05) BARR, Barrels
02) MET3, Cubic Meters 06) CUFT, Cubic Feet
03) LTRS, Liters 07) ACFT, Acre feet
04) IGAL, Imperial Gallons
Press the numbers on the keypad that corresponds to the
engineering units desired.
The next screen selection is the totalizer multiplier. There are eight
selections for totalizer multiplier. Use the UP or DOWN key to
display all multipliers available. Press the number key that
corresponds to the multiplier required.
Enter Application
Max Flow and Vmt
Max Flow **.**
VMt= **.**
Press ENTER to store
Any changes.
Press any other key
To not store changes.
03) Totalizer

Vantage Series 2200
>02) Program Continued
4-20 Output Assignment and Adjustment
Selection 4 in the programming menu is the 4-20mA output and assignment
adjustment. Press the 04 key to adjust or assign the 4-20mA DC output.
1) Adjustment: To adjust or calibrate the 4-20mA DC output press the 01
key. For the Vantage 2220 another screen will prompt to select which
sensor 4-20 output to adjust.
To adjust Zero: Press the 5 key, the cursor arrow will appear before the 5) 4
mA line. Press the 3 key for coarse adjustment or the 4 key for fine adjust-
ment. Now press the 1 key to adjust the mA upwards or the 2 key to adjust
downwards.
To adjust Span: Press the 6 key, the cursor arrow will appear before the 6) 20
mA line. Press the 3 key for coarse adjustment or the 4 key for fine adjust-
ment. Now press the 1 key to adjust the mA upwards or the 2 key to adjust
downwards.
To assign the 4-20mA loop the level or flow press the 02 keys at the 4-20 Out
selection. To select the 4-20 signal to track level press the 01 key. To select
the 4-20 signal to track flow press the 2 key. Press the ENTER key. If two
sensors are being used, added assignments for Flow 2, Level 2, Flow 1 + 2
and Flow 1 – 2 will be available.
Programming Setpoints
This selection will allow the user to assign up to three setpoints for High or
Low alarm conditions. Press the 05 keys to enter the setpoint selections. Press
the 01 keys for Setpoint #1. Press the 02 keys for Setpoint #2. Press the 03
keys for Setpoint #3. The next screen allows the user to assign the setpoint
selected to level or flow. Press the 01 keys for Level and the 02 keys for
Flow. The level selection will be in the engineering units selected for level.
The flow selection will be in engineering units selected for flow. The next
screen will allow the user to input ON and OFF points for the setpoint
selected. For Low alarm the ON value will be less than the OFF value. For
High alarm the ON value will be greater than the OFF value. To program;
using the DOWN/LEFT arrow key move the cursor to the left most digit.
Enter the number desired by using the keypad. The cursor will advance to the
right after the selection is entered. Press the ENTER key. The Setpoints must
be assigned to a Relay. (14 keys under Program.).
Sensor Calibration
To adjust the sensor calibration, press the 06 keys. The dimension physically
measured from the bottom the sensor to any target or liquid level is the
distance that will be displayed in the next screen. If the dimensions displayed
vary from the distance measured, use the 1 or 3 key to adjust the displayed
length to the measured length.
3-6
04) 4-20 Out
>1) Up 2) Down
. . . . . . . . . . . . . . . . .
>3) Coarse 4) Fine
>5) 4 mA 6) 20 mA
05) Setpoints
>06) Sensor Cal.

Vantage Series 2200
>02) Program Continued
Output Damping Adjustment
To adjust the 4-20mA output damping press the 07 keys. This will allow the
user to adjust the damping time. The damping times available are:
01) None 05) 60 Seconds
02) 5 Seconds 06) 2 Minutes
03) 15 Seconds 07) 4 Minutes
04) 30 Seconds 08) 8 Minutes
Lost Echo Setting
To adjust the Lost Echo time: (This is how long the meter will hold the last
value after losing the signal until failing to the Lost Echo 4-20 mA DC
assignment).
To set the Lost Echo time, press the 08 key. The lost echo times available are:
01) 5 Seconds 05) 2 Minutes
02) 15 Seconds 06) 4 Minutes
03) 30 Seconds 07) 8 Minutes
04) 60 Seconds 08) 16 Minutes
After pressing the desired number, or ENTER key, the next screen to appear
is the Lost Echo 4-20mA assignment. In this screen the user will select the
default for the 4-20mA DC output during a lost echo. The selections are:
01) Fail to Zero
02) Fail to Span
03) Hold last value
Press the number desired, this will return to the main program screen.
Simulation
The simulation screen will allow the user to enter a level to simulate
level/volume or flow. Enter the level in the engineering units displayed. The
Flow or Level/Volume line will display the flow or level/volume for that
level. If the flow displayed is different than expected, check the programming
of the flume, weir or special H/Q programming. Pressing the UP key will
allow the user to test the totalizer function. Press the MENU key to return to
the main program screen.
Integrator Setup
The next option in the program menu is the Integrator screen. To select this,
press the 10 keys. This screen will allow the user to assign the contact closure
time for a contact integrator. The cursor will appear on the most significant
digit. Use the number keys to enter the totalized flow value you want to have
for a contact output. Press the ENTER key to return to the main program
screen.
Pump Alternation Setup
The next option in the program menu is the Pump alternation screen. To
select this press the 11 keys. The first screen shows the selection of the three
setpoints and the four relays. The NN indicated that nothing has been selected
for that position. Press the 5 key to enter into the selection of setpoints and
relays.
3-7
07) Damping
08) Lost Echo
09) Simulation.
10) Integrator
11) Pump Alt.

Vantage Series 2200
>02) Program Continued
The first screen is for Setpoint Position #1. Select the number corresponding
to the setpoint desired. After the selection, the next setpoint position will be
shown. After the setpoint position #3 is selected, the relay position screen will
be shown. Select the number corresponding to the relay for that position.
After the selection, the next relay position will be shown. After the relay
position #4 is selected, press the MENU key, then the ENTER key to save
changes.
For example: a sewer line will feed into a wet well at a lift station. The station
employs three pumps. The wet well is 20 feet deep. As the wet well fills, the
operator wants to turn on the pump (Pump 1) when the level reaches 12 feet
and off at 2 feet. If the level in the well continues to rise with only one pump
running, the operator will probably require the second pump (Pump 2) to
come on at a level of 16 feet and off at 8 feet. If the level in the well continues
to rise with both pumps running, the operator will initiate a third pump (Pump
3) to come on at a level of 18 feet and off at 12 feet.
There are three setpoints: on at 12 feet, off at 2 feet (Setpoint Position 1); on
at 16 feet, off at 8 feet (Setpoint Position 2); on at 18 feet, off at 12 feet
(Setpoint Position 3).
The relay positions are then selected depending on which ones are to control
the pumps.
Relay Assignment
The next option in the program menu is the Relays screen. To select this
press 2nd Function then 3 (F23) keys. This option will allow the user to assign
each of the five relays to the following selections:
01) None 05)Lost Signal 09)Contact Integrator
02) Setpoint #1 06)4-20 Loop 10)Tot1
03) Setpoint #2 07)Over range 1 11)Tot2
04) Setpoint #3 08)Over range 2
Press selection desired. Press the ENTER key to save any changes. Should
the Relays screen not show as a selection, go to the Main selection screen and
select 05) System Setup, then 06) Options, then 01) Relays Added. Select the
number of relays you want active.
>03) Status
The status selection allows the user to view the status on the following
options:
01) Sensor: View signal strength, temperature and the signal gain.
02) Level: Indicates the distance between the sensor and the target and the
level.
03) Alarms/Relays: View the Alarms Set and the Relays Energized.
04) Logger: View the logger Time and Time to Store, Amount of logging
Stored and the Amount of free space to Store.
3-8
12) Relays
>03) Status
11) Pump Alt.

Vantage Series 2200
>03) Status Continued
05) History: View logged data in graphic form for each of the eight channels
available to log. Select the channel to be viewed by pressing the number
on the keypad. Press the UP or DOWN key to scroll through the data.
06) Daily Sum: View the Average, Minimum and Maximum flows and the
time of the event for the last eight days of flow.
Press the ENTER key to return to the main program menu.
>04) Data Logger
The next selection in the program menu is the data logger selection. There are
five selections in the data logger menu.
1) Set Time/Date. Press the UP key to move the arrow to the date or time
that is to be changed. Press the number value on the key pad to change.
Note the time is entered and viewed as military time.
2) Storage Rate. This will allow the user to select the storage rate for the
logging. Selections are:
01) 1 minute 03) 10 minute 05) 30 minute
02) 5 minute 04) 15 minute 06) 60 minute
3) Secondary. This will allow the user to select a secondary log rate to store
logging at a different interval than the main interval. This may be used to
store at faster intervals during storms or flow events. The selections
available are:
01) Not active 02) Setpoint #1 03) Setpoint #2 04) Setpoint #3
If setpoints are selected then the next screen will be storage rate times
available.
4) Log Channels. There are up to 8 channels available for logging. The
selections for each channel are:
01) Not Used 04) Flow 1 07) Total 2 10) Sensor2 Temp
02) Level 1 05) Flow 2 08) Setpoints 11) Lvl1-Lvl2
03) Level 2 06) Total 1 09) Sensor 1 Temp
5) Clear Data. Press the 5 key to clear all stored data.
2200 Data Logger Download Program
Refer to the Data Download software manual for instructions in retrieving the data
from the meter.
3-9
>04) Data Logger

Vantage Series 2200
>05 System Setup
The system setup option will allow the user to set up the Vantage series 2200
for the following options.
01) Language: This will allow the user to select the language displayed in the
Vantage series 2200. The options are 01) English, 02) German, 03)
Spanish.
02) Display: Choosing this feature allows the user to select the contrast of the
display from 01) Highest to 08) Lowest. This feature also allows to
display the back light to turn it off, or to program for a timed “off” of the
display if the key pad is not touched in a selected time interval.
03) Communications: This option will allow the user to set the baud rate,
flow control and slave I.Ds of the RS-232 and RS-485 communications.
01) Baud Rate - Select the baud rate desired to communicate with meter.
02) Flow Control - Hardware .should be selected unless a device
requires no flow control.
03) Slave ID - Select the desired Slave Identification number.
04) Modem Init - If a modem is being used select 02) Enable, otherwise
select 01) Disabled.
04) Display Lines: This option will allow the user to select the eight display
lines to be viewed on the main screen during operation. The options for
the display lines are:
01) Level 1 06) Total 2 11) Signal 1 15) Distance 1
02) Level 2 07) Tot1&2 Dif 12) Signal 2 16) Distance 2
03) Flow 1 08) Tot1&2 Sum 13) Lvl1-Lvl2 17) Date/Time
04) Flow 2 09) Relays 14) Flw1+Flw2 18) Blank line
05) Total 1 10) Alarms
05) Sensor Used: This option will select the unit being programmed for one
or two sensors and the type of sensors being used. Following is the
options for sensors.
01) FB1/FB4 - 60KHZ, 60KHZ, PVC, range 1-30 feet, w/1 ft. offset.
02) FB2 - 51KHZ, white teflon, sensor range 1-15 feet, w/1 ft. offset.
03) FB3 - 30KHZ, black plastic sensor, range 2-50 feet w/2 ft. offset.
This option will also display the temperature of the sensor and give the
option of calibrating the temperature and optimizing the electronics to the
frequency of the sensor.
06) Rly Pulse Wdt: The Relay Pulse Width sets the contact time for the
relays. The selections are 50, 100, 150, 200, 250, or 300 milliseconds.
07) Totals Reset: This option will reset the totalizer to zero. Press 5 to begin.
08) New Password: This option will allow the user to change the password to
enter into the QuikCal programming.
09) Summary Reset: This clears the Daily Summary memory.
10) Meter Reset: This option will reset all parameters to the factory defaults.
11) New firmware: This option will allow the user to upload any new
firmware to the latest revision. This requires connection to the RS232
Data Port with a computer or Palm PDA. Do not enter into this screen
unless you are prepared to upload new firmware.
3-10
>05) System Setup

Vantage Series 2200
>6 Calibration
The next option in the programming menu is Calibration. The options
available in the Calibration menu are:
01) Flow Simulation: The flow simulation screen will allow the user to
check the flow curve programmed into the unit. Enter the flow level in
the engineering units displayed. The Flow line will display the flow at
the entered interval. If the flow displayed is different than expected
check the programming of the flume, weir or special H/Q programming.
Pressing the UP key will allow the user to test the totalizer function.
02) 4-20 Adjustment: To adjust or calibrate the 4-20mA DC output, Press
the 1 key. The screen shown on the left will appear.
To adjust Zero: Press the 5 key, the cursor arrow will appear before the
5) 4 mA line. Press the 3 key for coarse adjustment or the 4 key for fine
adjustment. Now press the 1 key to adjust the mA upwards or the 2 key
to adjust downwards.
To adjust Span: Press the 6 key, the cursor arrow will appear before the
6) 20 mA line. Press the 3 key for coarse adjustment or the 4 key for
fine adjustment. Now press the 1 key to adjust the mA upwards or the 2
key to adjust downwards.
03) Sensor Cal: This option will allow the user to calibrate the system by
measuring the distance between the face of the sensor and the target (or
water) and adjusting the displayed distance value up or down with the 1
or 3 key to calibrate the unit to the correct distance of the target.
There is a Near and Far distance adjustment when calibrating the meter.
The Near distance adjustment should be made with a target being at least
the offset value from the face of the sensor. This distance should be no
more than 36 inches. The meter automatically determines the distance is
less than 36 inches. Use the 1 or 3 keys to adjust the displayed distance
value to the actual target distance. Should the distance be greater than 36
inches, the meter can be forced into the Near adjustment mode by
pressing the 4 or 6 keys. “Near” is displayed when the 4 or 6 key is
pressed.
The Far distance adjustment should be made with a target being at the
Vertical Mounting (Vmt) value or at least 37 inches from the face of the
sensor. Use the 1 or 3 keys to adjust the displayed distance value to the
actual target distance. The meter automatically determines the distance is
than 36 inches. Should the distance be less than 36 inches, the meter can
be forced into the Far adjustment mode by using the 7 or 9 keys instead
of the 1 and 3 keys. Far is displayed when the 7 or 9 key is pressed.
The Near and Far adjustment may need to be made several times until
distance readings are correct for both adjustments without making
adjustments to either.
3-11
>06)
Calibration
>1) Up 2) Down
. . . . . . . . . . . . . . . .
>3) Coarse 4) Fine
>5) 4 mA 6) 20 mA
This manual suits for next models
1
Table of contents
Popular Measuring Instrument manuals by other brands

Nibe
Nibe EMK 300 Installer manual

Shakmat
Shakmat Clock O’Pawn manual

Lutron Electronics
Lutron Electronics ARW 4220 user manual

MONARCH INSTRUMENT
MONARCH INSTRUMENT Nova-Pro 300 instruction manual

Surecom
Surecom SF-103 user manual

Endress+Hauser
Endress+Hauser Proline Cubemass 300 Description of Device Parameters