Easy at Home EBP-017 User manual

User Manual
Wrist Blood Pressure Monitor EBP-017
Version:1.0
Please do read the user manual carefully and thoroughly so as to ensure
the safe usage of this product.
Please keep this manual in order to reference in future.
Thank you very much for selecting Wrist Blood Pressure Monitor
EBP-017.
Wrist Type
EASY AT HOME MEDICAL,LLC
Any questions,please call toll-free :
1-855-822-6999 M-F 9 a.m.-5 p.m. CST
E-MAIL:[email protected]

1
CATALOGUE INTRODUCTION
INTRODUCTION ............................................................... 1
Safety Information
LCD Display Signal
Monitor Components
BEFORE YOU START .......................................................... 5
Installing and Replacing the Batteries
Setting Date, Time and Measurement Unit
MEASUREMENT ............................................................... 8
Tie the Cuff
Start Measurement
DATA MANAGEMENT 10
Recall the Records
Delete the Records
INFORMATION FOR USER ..................................................... 13
Tips for Measurement
Maintenance
ABOUT BLOOD PRESSURE.................................................... 15
What are systolic pressure and diastolic pressure?
What is the standard blood pressure classification?
Why does my blood pressure fluctuate throughout the day?
Why do I get a different blood pressure at home compared to the hospital?
Is the result the same if measuring on the right wrist?
TROUBLESHOOTING ......................................................... 17
SPECIFICATIONS ............................................................. 18
CONTACT INFORMATION...................................................... 19
Safety Information
Thank you for selecting wrist type blood pressure Monitor (EBP-017).
The monitor features blood pressure measurement, pulse rate measurement and
the result storage. The design provides you with two years of reliable service.
Readings taken by the EBP-017 are equivalent to those obtained by a trained
observer using the cuff and stethoscope auscultation method.
This manual contains important safety and care information, and provides step-by-
step instructions for using the product.
Read the manual thoroughly before using the product.
FEATURES:
Systolic Blood Pressure
Diastolic Blood Pressure
Pulse Rate
Memory: Up to 60 pieces of records
The signs below might be in the user manual, labeling or other components. They are the
requirement of standard and using.
Symbol for “THE OPERATION
GUIDE MUST BE READ”
Symbol for “MANUFACTURER”
Symbol for “SERIAL NUMBER”
Symbol for “TYPE B APPLIED
PARTS”
Symbol for “DIRECT CURRENT”
Symbol for “ENVIRONMENT
PROTECTION - Electrical waste
products should not be disposed of
with household waste. Please recycle
where facilities exist. Check with your
local authority or retailer for recycling
advice”
Table of Contents
..........................................................
COMPLIED STANDARDS LIST ......................................19
FCC STATEMENT 20
General Description
General Description
.............................................................
Symbol for “MANUFACTURE
DATE”
EMC GUIDANCE 20.............................................................

32
INTRODUCTION INTRODUCTION
LCD Display Signal
Systolic
blood pressure High pressure result
Diastolic
blood pressure Low pressure result
BPM Pulse/minute; heartbeats/minute
Memory
mmHg Measurement unit the blood pressure
(1mmHg=0.133kPa)
kPa Measurement unit of the blood pressure
(1kPa=7.5mmHg)
Year/Month/Day,Hour/Minute
Current time
Movement error
symbol Shocking will result in inaccurate
CUFF air is exhausting of deflatingDeflating
Low battery
Batteries are low and need to be replaced
Grade The grade of the blood pressure
SYMBOL DESCRIPTION EXPLANATION
Arrhythmia Irregular heartbeat
Heartbeat Heartbeat detection during the
measurement
The displayed measurement values
is from the memory.
This device is intended for adult use only.
This device is intended for non-invasive measuring and monitoring of arterial blood pressure. It is
not intended for use on extremities other than the wrist or for functions other than obtaining a blood
pressure measurement.
Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood
pressure.Do not begin or end medical treatment without asking a physician for treatment advice.
If you are taking medication,consult your physician to determine the most appropriate time to
measure your blood pressure. Never change a prescribed medication without consulting your
physician.
If the cuff pressure exceeds 40 kPa (300 mmHg),the unit will automatically deflate. Should the cuff
not deflate when pressures exceeds 40 kPa (300 mmHg), detach the cuff from the wrist and press
the START/STOP button to stop inflation.
To avoid measurement errors, carefully read this manual before using the product.
The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable
anesthetic mixture with air of with oxygen or nitrous oxide.
The operator shall not touch output of batteries and the patient simultaneously.
The user must check that the equipment functions safely and see that it is in proper working
condition before being used.
CAUTION
This device is contraindicated for any female subject who may be suspected of, or is pregnant.
Besides providing inaccurate readings, the effects of this device on the fetus are unknown.
Please use ACCESSORIES and detachable partes specified/ authorised by MANUFACTURE.
Otherwise, it may cause damage to the unit or danger to the user/patients.
Manufacturer will make available on request circuit diagrams, component parts list etc.
This unit is not suitable for continuous monitoring during medical emergencies or operations.
Otherwise, the patient’s wrist and fingers will become anaesthetic, swollen and even purple due to a
lack of blood.
Please use the device under the environment which was provided in the user manual. Otherwise,
the performance and lifetime of the device will be impacted and reduced.
During use, the patient will be in contact with the cuff.The materials of the cuff have been tested and
found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any
potential alergic reaction or contact injury.
The device doesn’t need to be calibrated within the two years of reliable service.
Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to the
local guidelines.
When the device was used to measure patients who have common arrhythmias such as atrial or
ventricular premature beats or atrial fibrillation, the best result may occur with deviation. Please consult
your physician about the result.
The device has been evaluated clinically using manual cuff/stethoscope auscultation as the reference.
Blood pressure measurements determined with this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultatory method, within the limits prescribed by the
American National Standard, Manual, electronic, or automated sphygmomanometers.”
Do not use the monitor under the conditions of strong electromagnetic field (e.g. medical RF
equipment) that radiates interference signal or electrical fast transient / burst signal.

CUFF
(Type B applied part)
BATTERY
COMPARTMENT
Monitor Components
List
1.Wrist Blood Pressure Monitor
(EBP-017)
3.User manual2. 2×AAA batteries
Component List
1.PCBA
2.Air Pipe
3.Pump
4.Valve
5.Cuff
54
BEFORE YOU START
Installing and Replacing the Batteries
CAUTION
1.Open the battery
door.
2.Insert the batteries
according to the
polarity indications
.
(Always select the
authorized / specified
2×AAA batteries.)
3.Close the battery
door.
displays on the LCD.
The LCD display dims.
When powering on the monitor, the LCD doesn’t light up.
Replace the batteries under following circumstances:
Remove batteries if the device is not likely to be used for some time.
Worn batteries are harmful to the environment. so please do not dispose with other daily trash.
Remove the old batteries from the device following your local recycling guidelines.
INTRODUCTION
Do not dispose of batteries in fire. Batteries may explode or leak.
LCD DISPLAY
SYSTOLIC
DIASTOLIC
MEM BUTTON
SET BUTTON
START/STOP BUTTON
PULSE RATE
GRADE
TIME
Memory
Norm
OptimOptim
Time
Date

Setting Date, Time and Measurement Unit
Please proceed to time setting before your initial use so as to
ensure each piece of record are labled with a time stamp. (Year
Range: 2010-2050)
1.When the monitor is OFF, press
the SET button to enter into the
clock mode, then press and hold
“SET” button for about 3 seconds
to enter year setting.
2.Press “MEM” button to change
the numeral. Each press will
increase the numeral by one in a
cycling manner.
3.Press “SET” button to confirm the
[YEAR]. Then the monitor diverts
to [MONTH] and [DAY] setting.
76
BEFORE YOU START
4.Repeat steps 2 and 3 to set
the [MONTH] and [DAY]. Then
it will enter into time format
setting.
5.Select the 24 hours
time format or the 12
hours time format.
Repeat steps 2 and 3
to set the [HOUR]
and [MINUTE].
6.Repeat steps 2 and
3 to set the
measurement unit.
7.After confirming the meausrement unit, the LCD will display all the
settings you have done one by one and the monitor will shut off.
BEFORE YOU START

Tie the Cuff
1
.Remove all accessories (watch, bracelet, etc) from your left wrist. If your
physician has diagnosed you with poor circulation in your left wrist, use your
right wrist.
2.
Roll or push up your sleeve to expose the skin.
3.
Apply the cuff to your left wrist with your palm facing up.
4
. Position the edge of the cuff about 1-1.5cm
(
0.4 -0.6
) from wrist joints
.
5
. Fasten the wrist cuff around your wrist, leaving no extra room between the cuff
and your skin. If the cuff is too loose, the measurement will be inaccurate.
Rest for 5 minutes before measuring.
Wait at least 3 minutes between measurements. This allows your blood
circulation to recover.
For a meaningful comparison, try to measure under similar conditions. For
example, take daily measurement at approximately the same time, on the same
wrist, or as directed by a physician.
98
MEASUREMENT
Start Measurement
1.After correctly positioning the cuff,
press START/STOP button to turn
on the monitor, and it will complete
the measurement process.
Adjust to zero.
Inflating and measuring.
Display and save the
measuring result.
2.Press START/STOP button to turn
off the monitor. Otherwise, the
monitor will shut off within 1
minute.
MEASUREMENT
LCD Display.
Norm
Optim
Time
Date
Keep the BP monitor the same level of your heart

Recall the Records
CAUTION
The most recent record (1) is shown first. Each new measurement
is assigned to the first (1) record. All other records are pushed back
one digit (e.g., 2 becomes 3, and so on), and the last record (60) is
dropped from the list.
The exact time and date
of the measurement
will display alternatively.
2. Press “MEM” button again, it will
display the latest measurement
record. Press “MEM” button to rotate
the history records. Press SET button
to look back the record. You can
check 60 groups of record totally.
When it displays the record, it will
shutdown in 1minute if has no
operation.
1.When the blood pressure monitor is off,
press “MEM” button to access the
memory, it will display the latest record
first when the records are less than
three groups. When there are three or
more than three groups ,it will display
the average value of the latest three
records first.
1110
DATA MANAGEMENT
3.When there has no record, the LCD
displays “---”.
DATA MANAGEMENT
UP
DOWN
The corresponding
time is 10:38.
The corresponding
date is January 1st.

Tips for Measurement
Delete the Records
When you did not obtain the accurate measurement, you can clear all the
measuring results by following steps below.
1.Press and hold both the “MEM” button
for 3 seconds when the monitor is in the
memory recall mode, the flash display
“dEL ALL” will show.
2.When the LCD display blinking “dEL
ALL”, press “SET” button to confirm
clearing.
The LCD will display “dEL dONE”,
indicating that memory clearing is
complete.
3.If you wish to give up clearing, press the
START/STOP button to turn off the
monitor.
Measurements may be inaccurate if taken in the following
circumstances.
Within 1 hour
after dinner or drinking
Within 20 minutes
after taking a bath
In a very cold environment
Immediate measurement
after tea, coffee, smoking
When talking or moving your fingers
When you want to discharge urine
1312
INFORMATION FOR USER
DATA MANAGEMENT

Maintenance What are systolic pressure and diastolic pressure?
press
artery vein
blood discharging
Systolic
relax
blood entering
Diastolic
When ventricles contract and pump blood out
of the heart, the blood pressure reaches its
maximum value in the cycle, which is called
systolic pressure. When the ventricles relax,
the blood pressure reaches its minimum
value in the cycle, which is called diastolic
pressure.
What is the standard blood pressure classification?
INFORMATION FOR USER
To obtain the best performance, please follow the instructions below.
Put in a dry place and avoid the sunshine
Avoid intense shaking
and collisions
Use wet cloths to remove dirt
Avoid touching water,
clean it with a dry cloth in case.
Avoid dusty and unstable
temperature environment
Avoid washing the cuff
Irregular Heartbeat Detector
CAUTION
The appearance of the IHB icon indicates that a pulse irregularity consistent with
an irregular heartbeat was detected during measurement. Usually this is NOT a
cause for concern. However, if the symbol appears often, we recommend you
seek medical advice. Please note that the device does not replace a cardiac
examination, but serves to detect pulse irregularities at an early stage.
The chart on the right is the
standard blood pressure classifi-
cation published by American
Heart Association (AHA).
AHA Home Guideline for
Upper Limit of Normal BP
Please consult a physician if your measuring result falls outside the range.
Please note that only a physician can tell whether your blood pressure value
has reached a dangerous point.
CAUTION
SYS 135 mm Hg
DIA 85 mm Hg
Blood Pressure Category
Normal
Prehypertension
High Blood Pressure
(Hypertension) Stage 1
High Blood Pressure
(Hypertension) Stage 2
Hypertensive Crisis
(Emergency care needed)
Systolic
mmHg (upper#)
Diastolic
mmHg (lower#)
less than 120
120-139
140-159
160 or higher
Higher than 180
and
or
or
or
or
less than 80
80-89
90-99
100 or higher
Higher than 110
This chart reflects blood pressure categories defined by American Heart Association.
14 15
INFORMATION FOR USER ABOUT BLOOD PRESSURE
• If you have any problems with this device, such as setting up, maintaining or using,
please contact the SERVICE PERSONNEL of . Don’t open or repair the
device by yourself.
• Please report to if any unexpected operation or events occur.
CAUTION
Norm
Optim
Norm
Optim
Time
Date
Norm
Optim
Time
Date
Norm
Optim
Norm
Optim
Norm
Optim
An irregular heartbeat is detected when a heartbeat rhythm varies while the unit is
measuring the systolic and diastolic blood pressure.During each measurement,
this equipment records the heartbeat intervals and works out the standard deviation.
If the calculated value is larger than or equal to 15,the irregular heartbeat symbol
appears on the symbol when the measurement results are displayed.
Time
Date
Time
Date
Time
Date
Time
Date

1716
TROUBLESHOOTINGABOUT BLOOD PRESSURE
1. Individual blood pressure varies
multiple times everyday, it is also affect-
ed by the way you tie your cuff and mea-
surement position, so please take the
measurement under the same conditions.
2.If the person takes medicine, the
pressure will vary more.
3.Wait at least 3 minutes for
another measurement.
Why does my blood pressure
fluctuate throughout the
day?
Is the result the same
i f measu r ing o n t h e
right wrist?
It is ok for both wrists, but there
will be some different results for
different people. We suggest you
measure the same wrist every time.
Why do I get a different
blood pressure at home
compared to the hospital?
The blood pressure is different even
throughout the day due to weather
emotion, exercise etc, Also, there is the
“white coat” effect, which means blood
pressure usually increases in clinical
settings.
What you need to pay
attention to when you measure
your blood pressure at home:
If the cuff is tied properly.
If the cuff is too tight or too loose.
If the cuff is tied on the wrist.
If you feel anxious.
Taking 2-3 deep breaths before
beginning will be better for measuring.
Advice:Relax yourself for 4-5
minutes until you calm down.
This section includes a list of error messages and frequently
asked questions for problems you may encounter with your wrist
blood pressure monitor. If the products not operating as you
think it should, check here before arranging for servicing.
PROBLEM SYMPTOM CHECK THIS REMEDY
No power
Low
batteries
Error
massage
Display is dim or
will not light up.
Batteries are exhausted.
Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Show on
the display Batteries are low.
E 1 shows The cuff is not secure.
Refasten the cuff and then
measure again.
E 2 shows The cuff is very tight Refasten the cuff and then
measure again.
E 3 shows The pressure of the
cuff is excess.
Relax for a moment
and then measure again.
E 10 or
E 11 shows
The monitor detected
motion while measuring.
Movement can affect the
measurement.Relax for a
moment and then
measure again.
E 20 shows
E 21 shows Relax for a moment and
then measure again.
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
The measurement
process does not detect
the pulse signal.
Loosen the clothing on
the wrist and then
measure again.
Measure incorrectly.

Power Supply
Battery powered mode:
2×AAA batteries (3V)
Display Mode
Measurement Mode
Oscillographic testing mode
Measurement Range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the wrist
About 13.5cm-21.5cm( )
Weight
External dimensions
Attachment
2×AAA batteries,one storage,user manual
Mode of operation
Continuous operation
Degree of protection
Software version
Device classification
V01
Internally Powered ME Equipment
WARNING: No modification of this equipment is allowed.
Contact Information
Complied Standards List
For more information about our products, please visit ,
or call Customer 1-866-822-6999 M-F 9 a.m.-5 p.m.CST, usual
problems and customer download, will serve you anytime.
Pressure: 0kPa-40kPa(0mmHg-300mmHg)
pulse value:(40-199)beat/minute
Approx.112g(3.95oz)(Excluding batteries)
Type B applied part
IP Classification
IP×0
1918
CONTACT INFORMATIONSPECIFICATIONS
Risk management
Labeling
User manual
General Requirements
for Safety
Electromagnetic
compatibility
Non-invasive
Sphygmomanometers
General
Requirements
ISO 14971:2007
EN 980:2008
EN 1041:2008
IEC 60601-1:2005
IEC 60601-1-2:2007
AASI/AAMI SP10:2002/A1:2003/A2:2006/(R)2008
Digital LCD V.A 31.5mm×44mm(1.25’’×1.73’’)
Pressure:
5℃-40℃(41℉-104℉)within±0.4kPa(3mmHg)
pulse value:±5%
Temperature:5℃to 40℃(41℉to 104℉)
Relative humidity: ≤85%RH
Atmospheric pressure: 86kPa to 106kPa
Temperature:-20℃-60℃
Relative Humidity: 10%RH-93%RH
Atmospheric Pressure: 50kPa-106 kPa
(-4℉- 140℉)
Approx. 68mm×75mm×31mm(2.68”×2.95”×1.22”)
(Excluding the cuff)
5
-
/
1
”
3
8/
1
”
2

20
Table 1 Guidance and MANUFACTURER’s declaration – ELECTROMAGNETIC
EMISSIONS- for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’sdeclaration – electromagnetic emissions
RF emissions
CISPR 11 Group 1
Class B
Not applicable
Not applicable
Compliance
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environment.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions IEC
61000-3-3
RF emissions
CISPR 11
Emissions test Electromagnetic environment - guidance
The device uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
Guidance and manufacturer’s declaration – electromagnetic immunity
IMMUNITY test
±6 kV contact
±8 kV air ±6 kV contact
±8 kV air
±2 kV for
power supply lines Not applicable
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 s
3A/m
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Table 2 Guidance and MANUFACTURER’s declaration – electromagnetic
IMMUNITY – for all ME EQUIPMENT and ME SYSTEMS
IEC 60601 test level Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
3A/m
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC
61000-4-5
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an
environment
EMC Guidance
21
±1 kV for
input/output lines
Not applicable
Not applicable
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the device requires
continued operation during
power mains interruptions,
it is recommended that the
device be powered from an
uninterruptible power supply
or a battery.
EMC GUIDANCE
FCC Statement
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference
that may cause undesired operation.

22
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
Table 6 Recommended separation distances between portable and mobile
RF communications equipment and the ME EQUIPMENT or ME SYSTEM –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances
between portable and mobile RF communications equipment and the device.
The device is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the device
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmittters) and the
device as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter(m)
For transmitters rated at a maximum output power not listed above, the recommended
separation distance din metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where Pis the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
0.01
0.1
1
10
100
Not applicable 0.117
0.369
1.167
3.690
11.67
0.233
0.738
2.333
7.378
23.33
23
P
V
=d ]
5,3
[
1
P=d 1.167 2.333 P=d
Not applicable
Not applicable
Not applicable
Not applicable
Table 4 Guidance and MANUFACTURER’s declaration – electromagnetic IMMUNITY –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration – electromagnetic immunity
IMMUNITY test
Not applicable
3 V/m
Compliance
level
IEC 60601
TEST LEVEL
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 Vrms
150 kHz to
80 MHz
Electromagnetic environment -
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and dis the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
a
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
P
V
=d ]
5,3
[
1
P=d
80 MHz t o 800 MHz
P=d 2.333
800 MHz t o 2, 5 GH z
1.167
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular /
cordless) telephones and land mobile radios, amateur radio, AM and FM radio broad-
cast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which
the device is used exceeds the applicable RF compliance level above, the device
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the device.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
[ ]V/m.