Edison DAM075 User manual

Auto Range Digital Multimeter
- 1000V DC /750V AC
OPERATING MANUAL
FFuunnccttiioonnss
•NNooccoonnttaaccttvvoollttaaggeeddeetteeccttoorr..
•WWiirriinnggffaauullttss..
•TTeemmppeerraattuurreetteesstteerrffrroomm--2200ttoo11000000°°CC
ORDER CODE: EDI-516-3100K
MODEL DAM075
5163100K_Instructions.qxd 17/09/2009 13:54 Page 1

2
CONTENT
1. GENERAL INSTRUCTIONS ......................................................................................2
2. DESCRIPTION ..........................................................................................................4
3. TECHNICAL SPECIFICATIONS ..................................................................................5
4. OPERATING INSTRUCTION ......................................................................................9
5. MAINTENANCE ......................................................................................................13
6. PACK CONTENT ....................................................................................................14
1.
GENERAL INSTRUCTIONS
This multimeter has been designed according to IEC1010 standards for
electronic measuring instruments with an over-voltage category (CAT II 1000V,
CAT III 600V) and pollution level 2.
To get the best service from this instrument, read carefully this user manual and
respect the detailed safety precautions.
1.1. Precautionary safety measures
Preliminary
• When using this multimeter, the user must observe all normal safety rules
concerning:
- Protection against the dangers of electric current.
- Protection of the multimeter against misuse.
• For your own safety, only use the test probes supplied with the instrument.
Before use, check the probes are in good condition.
During use
• If the meter is used near noise generating equipment, be aware that the
display may become unstable or indicate large errors.
• DO NOT use the meter or test leads if they look damaged.
• Use the meter only as specified in this manual; otherwise, the protection
provided by the meter may be impaired.
• USE EXTREME CAUTION when working around bare conductors or bus bars.
• DO NOT operate the meter around explosive gas, vapour, or dust.
• BEFORE EACH MEASUREMENT make sure the main function dial is at the
correct position.
• When the range of the value to be measured is unknown, check that the range
initially set on the multimeter is at the correct test position.
• To avoid damage to the instrument, DO NOT exceed the maximum limits of the
input values.
5163100K_Instructions.qxd 17/09/2009 13:54 Page 2

3
Auto Range Digital Multimeter - 1000V DC/750V AC
• CAUTION when working with voltages above 60V DC or 30V AC RMS. Such
voltages pose a shock hazard.
• When using the probes, keep your fingers behind the finger guards.
•
Before changing functions, disconnect the test leads from the circuit under
test.
• Disconnect circuits power & discharge all high-voltage capacitors before
testing resistance, continuity or diode.
• Before measuring current, turn power off to the circuit before connecting the
meter to the circuit.
• Change the battery when the symbol appears to avoid incorrect data.
1.2. Symbols
Symbols used in this manual and on the instrument:
CAUTION: Refer to the instruction manual. Incorrect use may result
in damage to the device or its components.
Earth.
This instrument has double insulation.
Fuse: F 500mA/250V.
1.3 Instructions
• Before opening up the multimeter, always turn OFF the device, disconnect
from all sources of electric current and make sure you are not charged with
static electricity, which may destroy internal components.
• Any adjustment, maintenance or repair work carried out on the meter should
only be carried out by a qualified person.
• If any faults or abnormalities are observed, take the instrument out of service
and ensure that it is NOT USED until it has been checked and/or re-calibrated
by a qualified person.
• If the meter is not going to be used for a long time, take out the battery and do
not store the meter in high temperature or high humidity environment.
• For continuous protection against fire, replace fuse only with the specified
voltage and current rating: F500mA/250V.
• NEVER use the meter unless the back cover and battery cover are in place and
fastened fully.
5163100K_Instructions.qxd 17/09/2009 13:54 Page 3

2.
DESCRIPTION
2.1 Multimeter Features
ON/OFF - This button is used to turn the
meter ‘ON’ or ‘OFF’.
Rotary Dial - This dial is used to select
functions and desired ranges.
FUNC. - This button is operative in
and ‘AA’ (amps) ranges.
11..In range a beep will sound when
changing from Diode (default) to Audible
continuity test, another beep will sound
when pressed again to switch back to
diode.
22..In ‘AA’ (current range) a beep will sound
when changing from DC (default) to AC,
another beep will sound when pressed
again to switch back to DC reading.
RANGE - This button is operative in VVand
ΩΩranges.
This key selects Automatic (default) or manual
mode.
11..To switch from auto to manual mode:
aa..Select your desired test range (VVor ΩΩ).
bb..Press the RRAANNGGEEbutton for 1 sec, a beep will sound.
cc..Manual mode is now set.
22..To switch from manual back to auto mode:
aa..Press the RRAANNGGEEbutton for 2 sec, 2 beeps will sound.
bb..Auto mode is now set.
HOLD - This button is operative in all dial ranges.
11..To fix the display on the current value press the HOLD key once.
22..A second press returns the meter to normal mode.
REL. - This button is operative in VV,,AA,,ΩΩand CCAAPPranges.
11..Select your desired test range.
22..Press on the RREELL..button, 1 beep will sound to indicate selection.
4
5163100K_Instructions.qxd 17/09/2009 13:54 Page 4

5
Auto Range Digital Multimeter - 1000V DC/750V AC
button (back light) - This button is operative in all ranges except mode.
Used the key to illuminate the LCD screen.
11..Select your desired test range.
22..Press the button for 2 sec, a beep will sound and the backlight will be
turned ‘ON’.
33..Press it again, 1 beep will sound and the backlight will be turned ‘OFF’.
2.2 LCD Display
• W43 x H29mm (1 ¾” x 1”) LCD screen.
• Display: 3 3/4digits LCD display has a max.
reading of 3999.
2.3 Terminals (footer of multimeter)
10A - Terminal receiving RED lead for 10A measurement.
µAmA - Terminal receiving RED lead for µAmA measurement.
COM - Terminal receiving the BLACK lead as a common reference.
V, ΩΩ, , ,°C, Hz - Terminal receiving the RED lead for voltage, resistance,
capacitance, diode, continuity measurements, temperature and frequency.
3.
TECHNICAL SPECIFICATIONS
3.1 General Specifications
• Environment conditions:
- CAT III 600V and CAT II 1000V.
- Pollution degree: 2.
- Altitude < 2000m.
- Operating temperature: 0~40°C, (<80% RH, non-condensing)
- Storage temperature: -10~60°C, (<70% RH, battery removed)
- Temperature Co-efficient: 0.1 x (specified accuracy) / °C (<18°C or >28°C).
• MAX. Voltage between terminals and earth ground: 750V AC rms or 1000V DC.
• Fuse Protection: uAmA: F 500mA/250V ∅5X20
• Sample Rate: 3 times/sec for digital data.
• Display: 3 3/4digits LCD display with max. reading 3999.
• Automatic indication of functions and symbols.
• Range selection: automatic and manual.
• Low battery indication: The symbol is displayed when the battery is under
the proper operation range.
• Over range indication: LCD will display ‘OOLL’.
5163100K_Instructions.qxd 17/09/2009 13:55 Page 5

6
• Polarity indication: ‘__’ displayed automatically.
• Auto power off: If there is no key or dial operation for 15 minutes, the meter
will power itself ‘off’ to save battery consumption.
• 1 minute before power off, the beeper will sound 5 times. It will then beep
again before turning itself off.
• Power source: 4.5V battery (DC1.5 x 3 AAA batteries).
• Accessories: Operating manual. Test probes and a ‘K’ type thermocouple.
3.2 Measurement Specifications
• Accuracy: ± (% of reading + number of digits) at 18°C to 28°C (64°F to 82°F)
with relative humidity to 80%.
No Contact AC Voltage Detector
DC Voltage
• Input impedance: 10MΩ.
• Maximum input voltage: 1000V DC or 750V AC rms.
AC Voltage
• Input impedance: 10MΩ.
• Maximum input voltage: 1000V DC or 750V AC rms.
• Frequency Range: 40Hz-400Hz.
• Response: Average, calibrated in rms of sine wave.
SENSITIVITY FREQUENCY DISTANCE
>50V 50Hz < 150mm
RANGE RESOLUTION ACCURACY
400mV 0.1mV
4V 1mV ±(0.8% of rdg +3 digits)
40V 10mV
400V 100mV
1000V 1V ±(1.0% of rdg +3 digits)
RANGE RESOLUTION ACCURACY
4V 1mV
40V 10mV ±(1.0% of rdg +3 digits)
400V 100mV
750V 1V ±(1.2% of rdg +3 digits)
5163100K_Instructions.qxd 17/09/2009 13:55 Page 6

7
Auto Range Digital Multimeter - 1000V DC/750V AC
DC Current
• Overload protection: F 500mA/250V fuse for uAmA range.
• Maximum input current:
- 400mA DC or 400mA AC rms for µA and mA ranges.
- 10A DC or 10A AC rms for 10A ranges.
AC Current
• Overload protection: F 500mA/250V fuse for uAmA range.
• Maximum input current:
- 400mA DC or 400mA AC rms for µA and mA ranges.
- 10A DC or 10A AC rms for 10A ranges.
• Frequency Range: 40Hz-400Hz.
• Response: Average, calibrated in rms of sine wave.
Resistance
• Open circuit voltage: approx. 0.25V
• Overload protection: 250V DC or 250V AC rms.
RANGE RESOLUTION ACCURACY
400Ω0.1Ω
4kΩ1Ω
40kΩ10Ω±(1.2% of rdg +3 digits)
400kΩ100Ω
4MΩ1kΩ
40MΩ10kΩ±(2.0% of rdg +5 digits)
RANGE RESOLUTION ACCURACY
4000µA 1µA ±(1.2% of rdg +3 digits)
400mA 0.1mA
10A 10mA ±(2.0% of rdg +8 digits)
RANGE RESOLUTION ACCURACY
4000µA 1mµA ±(1.5% of rdg +5 digits)
400mA 0.1mA
10A 10mA ±(3.0% of rdg +8 digits)
5163100K_Instructions.qxd 17/09/2009 13:55 Page 7

8
Diode and Audible Continuity Test
• Overload protection: 250V DC or 250V AC rms.
Temperature
• Overload protection: 250V DC or 250V AC rms.
Capacitance
• Overload protection: 250V DC or 250V AC rms.
RANGE RESOLUTION ACCURACY
Built-in buzzer sounds Open circuit voltage:
if resistance is less approximate 0.5V
than approx. 40Ω
Display read approx. Forward DC Current:
forward voltage of diode Approx. 1mA
Reversed DC Voltage:
approx. 1.5V
RANGE RESOLUTION ACCURACY
-20°C to 0°C 1°C ±(5% of rdg +4 digits)
0°C to 400°C 1°C ±(1% of rdg +3 digits)
400°C to 1000°C 1°C ±(2% of rdg +3 digits)
RANGE RESOLUTION ACCURACY
4nF 1pF ±(5.0% of rdg +5 digits)
40nF 10pF
400nF 100pF
4µF 1nF ±(3.0% of rdg +5 digits)
40µF 10nF
200µF 100nF
5163100K_Instructions.qxd 17/09/2009 13:55 Page 8

9
Auto Range Digital Multimeter - 1000V DC/750V AC
Frequency
• Overload protection: 250V DC or 250V AC rms.
• Input voltage range: 0.6~3V AC rms (Input voltage must be enlarged with
increasing frequency under measurement).
4.
OPERATING INSTRUCTION
No contact AC voltage detector
TThheerreeiissnnoo‘‘AAuuttooPPoowweerrOOffff’’ffuunnccttiioonnoonntthhiisssseettttiinngg..
TThhee((bbaacckklliigghhtt))iissnnoottooppeerraattiivveeiinntthhiissrraannggee..
11..Set dial to the symbol.
22..When meter is switched on a Green LED light will appear lit.
33..Put the ‘EF-DETECT AREA’ (top end) close to a lead or power socket, if AC
electrical voltage is present the Red LED indicator will flash and you will hear
a repetitive warning beep, figure will also be displayed on the LCD screen.
Voltage Measurement
TTooaavvooiiddeelleeccttrriiccaallsshhoocckkaanndd//oorrddaammaaggeettootthheemmuullttiimmeetteerr,,
DDOONNOOTTaatttteemmppttttoottaakkeeaannyyvvoollttaaggeemmeeaassuurreemmeenntttthhaattmmiigghhtt
eexxcceeeedd11000000VVDDCCoorr775500VVAACCrrmmss..DDoonnoottaappppllyymmoorreetthhaann
11000000VVDDCCoorr775500VVAACCrrmmssbbeettwweeeenntthheeccoommmmoonntteerrmmiinnaallaannddtthheeeeaarrtthhggrroouunndd..
11..Set rotary switch to the V (ACV or DCV) range.
22..Press “RANGE” key to select manual mode.
33..
Connect the Black and Red test leads to the COM and V terminals respectively.
44..Connect the test leads to the circuit being measured.
55..Read the displayed value. The polarity of red test lead connection will be indicated
when taking a DC measurement.
RANGE RESOLUTION ACCURACY
9.999Hz 0.001Hz
99.99Hz 0.01Hz
999.9Hz 0.1Hz ±(0.1% of rdg +1 digit)
9.999kHz 1Hz
99.99kHz 10Hz
199.9kHz 100Hz
>200kHz 100Hz Unspecified @ >200kHz
5163100K_Instructions.qxd 17/09/2009 13:55 Page 9

10
Resistance Measurement
TTooaavvooiiddeelleeccttrriiccaallsshhoocckkaanndd//oorrddaammaaggeettootthheemmuullttiimmeetteerr,,
ddiissccoonnnneeccttcciirrccuuiittppoowweerraannddddiisscchhaarrggeeaallllhhiigghh--vvoollttaaggee
ccaappaacciittoorrssbbeeffoorreetteessttiinngg..
11..Set the rotary switch to ΩΩrange.
22..Connect the Black test lead to the CCOOMMand the Red test lead to the VVΩΩ
terminals respectively.
33..Connect the test leads to the circuit being measured and read the displayed value.
44..In order to ensure a more accurate measurement of low resistance, short
the test leads before testing. This is necessary to subtract for the resistance of
the test leads.
NNOOTTEE::For measuring resistance above 1MΩΩ, the meter may take a few seconds
to stabilise reading. This is normal for high resistance measuring.
When the input is not connected, i.e. at open circuit, the figure ‘OOLL’ will be
displayed for the overrange condition.
Diode Measurement
TTooaavvooiiddeelleeccttrriiccaallsshhoocckkaanndd//oorrddaammaaggeettootthheemmuullttiimmeetteerr,,
ddiissccoonnnneeccttcciirrccuuiittppoowweerraannddddiisscchhaarrggeeaallllhhiigghh--vvoollttaaggee
ccaappaacciittoorrssbbeeffoorreetteessttiinnggddiiooddee..
11..Set the rotary switch to range.
22..Press ‘FFUUNNCC..’ button to select measuring mode.
33..Connect the Black test leads to the CCOOMMand the Red test leads to the ΩΩ
terminals respectively.
44..Connect the Red test lead to the anode and the Black test lead to the cathode
of the diode under testing.
55..The meter will show the approx. forward voltage of the diode. If the lead
connection is reversed, only figure ‘OOLL’ will be displayed.
Continuity Measurement
11..Set the rotary switch to range.
22..Press ‘FFUUNNCC..’ button to select measuring mode.
33..Connect the Black test leads to the CCOOMMand the Red test leads to the ΩΩ
terminals respectively.
5163100K_Instructions.qxd 17/09/2009 13:55 Page 10

11
Auto Range Digital Multimeter - 1000V DC/750V AC
44..Connect the test leads to the resistance on the circuit being measured. When
the test lead on the circuit is below 40Ω, a continuous beeping will sound.
NNOOTTEE::Continuity test is available to check an open/short circuit.
Capacitance Measurement
TTooaavvooiiddeelleeccttrriiccaallsshhoocckkaanndd//oorrddaammaaggeettootthheemmuullttiimmeetteerr,,
ddiissccoonnnneeccttcciirrccuuiittppoowweerraannddddiisscchhaarrggeeaallllhhiigghh--vvoollttaaggeeccaappaacciittoorrss
bbeeffoorreetteessttiinngg..UUsseetthheeDDCCvvoollttaaggeeffuunnccttiioonnttooccoonnffiirrmmtthhaatttthhee
ccaappaacciittoorriissddiisscchhaarrggeedd..
11..Set the rotary switch to range.
22..Connect the Black test lead to the CCOOMMand the Red test lead to the
terminals respectively.
33..Connect the test leads to the capacitor and read the displayed value.
NNOOTTEE::The meter may take a few seconds (200mF range, 30 seconds) to stabilise
reading. This is normal for high capacitance measuring.
To improve the accuracy of measurements less than 4nF, subtract the residual
capacitance of the meter and leads.
Frequency Measurement
11..Set the rotary switch to HHzzrange.
22..
Connect the Black and Red test leads to the CCOOMMand HHzzterminals respectively.
33..Connect the test leads across the source or load under measurement, and
read the displayed value.
NNOOTTEE::Readings are possible at input voltages above 3V rms, but the accuracy is
not guaranteed.
In noisy environment, it is preferable to use shield cable for measuring small signal.
Temperature Measurement
TTooaavvooiiddeelleeccttrriiccaallsshhoocckk,,ddoonnoottccoonnnneecctttthheetthheerrmmooccoouupplleesswwiitthh
aalliivveeeelleeccttrriiccaallcciirrccuuiitt..
11..Set the rotary switch to °°CCrange. The LCD display will show the current
environment temperature.
22..Connect the Red K-type thermocouple lead into the °°CCterminal and the Black
K-type thermocouple into the CCOOMMterminal.
33..Contact the test subject with the thermocouple probe.
44..Read temperature value on the LCD display.
5163100K_Instructions.qxd 17/09/2009 13:55 Page 11

12
Current Measurement
TTooaavvooiiddddaammaaggeettootthheemmeetteerr,,cchheecckktthheemmeetteerrssffuusseebbeeffoorree
pprroocceeeeddiinngg..UUsseetthheepprrooppeerrtteerrmmiinnaallss,,ffuunnccttiioonn,,aannddrraannggeeffoorryyoouurr
mmeeaassuurreemmeenntt..
11..Set the rotary switch to the 44000000mmAA,,440000mmAAor 1100AArange.
22..Press ‘FFUUNNCC..’ button to select DCA or ACA mode.
33..Connect the Black test lead to the CCOOMMterminal and the Red test leads to the
mmAAterminal for a maximum of 400mA. For a maximum of 10A, move the Red
test lead to the 1100AAterminal.
44..Connect test leads in series with the load in which the current is to be measured.
55..Read the displayed value. The polarity of red test lead, connection will be indicated
when making a DC measurement.
66..When the figure ‘OL’ is displayed a beep will sound continually, this indicates
an overrange situation and the higher range has to be selected.
Misplug Alarm
TTooaavvooiiddddaammaaggeettootthheemmeetteerr,,tthhiissmmuullttiimmeetteerrhhaassaa‘‘MMiisspplluuggAAllaarrmm’’
ffuunnccttiioonn..WWhheenntthheeaallaarrmmssoouunnddss,,pplleeaasseecchheecckktthheerraannggeeaannddtthhee
tteerrmmiinnaalloofftthheelleeaadd.
The following operations will sound the alarm.
11..If you...
Set the rotary dial to any range EXCEPT the symbol, then connect the
test lead to the µµAAmmAAor 1100AAterminal the meter will sound the alarm.
22..If you...
Set the rotary dial to the µµAAor mmAArange, connect the test lead to the 1100AA
terminal the meter will sound the alarm.
33..If you...
Set the rotary switch to the 1100AArange, connect the test lead to the µµAAmmAA
terminal the meter will sound the alarm.
5163100K_Instructions.qxd 17/09/2009 13:55 Page 12

13
Auto Range Digital Multimeter - 1000V DC/750V AC
5.
MAINTENANCE
General Maintenance and Cleaning
TTooaavvooiiddeelleeccttrriiccaallsshhoocckkoorrddaammaaggeettootthheemmuullttiimmeetteerr,,ddoonnoottggeett
wwaatteerriinnssiiddeetthheeccaassee..RReemmoovveetthheetteessttlleeaaddssaannddaannyyiinnppuuttssiiggnnaallss
bbeeffoorreeooppeenniinnggtthheeccaassee..
Cleaning the multimeter:
• Wipe the case with a soft damp (not wet) cloth and mild detergent.
• DO NOT use abrasives or solvents.
• Dirt or moisture in the terminals can affect readings.
• Cleaning the terminals:
• Turn the meter off and remove all test leads.
• Shake out any dirt that may have gathered in the terminals.
• Soak a clean swab or cotton bud with a cleaning and oiling agent (such as WD-40).
• Work the swab/bud around in each terminal until you are happy there are clean.
The oiling agent insulates the terminals from moisture-related contamination.
BBeeffoorreerreeppllaacciinnggtthheebbaatttteerryyaannddffuusseess,,ddiissccoonnnneecctttteessttlleeaaddssffrroomm
aannyycciirrccuuiittuunnddeerrtteesstt,,ttuurrnntthheemmuullttiimmeetteerr‘‘OOFFFF’’aannddrreemmoovveetteesstt
lleeaaddssffrroommtthheeiinnppuutttteerrmmiinnaallss..TToopprreevveennttddaammaaggeettootthheemmeetteerroorr
ppeerrssoonnaalliinnjjuurryy,,aallwwaayyssrreeppllaacceetthheebbaatttteerryyoorrffuusseewwiitthhssppeecciiffiieeddrraattiinngg..
Battery Replacement
When the battery voltage drops below a usable operation range the symbol will
appear on the LCD display, this indicates that the battery will need replacing. To do
this follow these steps.
• Set the meter to ‘OOFFFF’ state.
• Disconnect test leads from any input terminals.
• Use the appropriate screwdriver to unscrew the secured back cover. This will
expose the batteries.
• Replace the battery with 3 new 1.5V batteries (AAA).
• Replace the battery cover and screw securely back into place.
• The multimeter should new be fully functional.
5163100K_Instructions.qxd 17/09/2009 13:55 Page 13

14
Fuse Replacement
• Set the meter to 'OFF' state.
• Disconnect test leads from any inputs terminals.
• Use the appropriate screwdriver to unscrew the secured back cover and then
the battery cover.
• Replace the fuse only with a 500mA/250V 5 20.
• Replace the 2 back covers and screw securely back into place.
• The multimeter should now be fully functional.
6.
PACK CONTENTS
• Digital multimeter
• Test probes
• Temperature sensor probe
• DC1.5V x3 AAA batteries (installed)
• Instruction manual
5163100K_Instructions.qxd 17/09/2009 13:55 Page 14

15
Auto Range Digital Multimeter - 1000V DC/750V AC
5163100K_Instructions.qxd 17/09/2009 13:55 Page 15

Edison Products are continuously being developed and improved.
All details and illustrations are for guidance and may be subject to change.