Eickemeyer LIFEVET O2 User manual

LIFEVET 02
SAUERSTOFFMESSGERÄT/OXYGEN ANALYZER
BEDIENUNGSANLEITUNG/USER MANUAL
TELEPHONE 0049 7461 96580-0
www.eickemeyer.com
GERMANY
EICKEMEYER KG
Eltastraße 8
78532 Tuttlingen
T +49 7461 96 580 0
F +49 7461 96 580 90
E info@eickemeyer.de
www.eickemeyer.de
SWITZERLAND
EICKEMEYER AG
Sandgrube 29
9050 Appenzell
T +41 71 788 23 13
F +41 71 788 23 14
E info@eickemeyer.ch
www.eickemeyer.ch
UNITED KINGDOM
EICKEMEYER Ltd.
3 Windmill Business Village
Brooklands Close
Sunbury-on-Thames
Surrey, TW16 7DY
T +44 20 8891 2007
F +44 20 8891 2686
E info@eickemeyer.co.uk
www.eickemeyer.co.uk
DENMARK
EICKEMEYER ApS
Lysbjergvej 6, Hammelev
6500 Vojens
T +45 7020 5019
F +45 7353 5019
E info@eickemeyer.dk
www.eickemeyer.dk
NETHERLANDS
EICKEMEYER B.V.
Bedrijventerrein
Pavijen-West Bellweg 44
4104 BJ Culemborg
T +31 345 58 9400
F +31 345 58 9401
E info@eickemeyer.nl
www.eickemeyer.nl
ITALY
EICKEMEYER S.R.L.
Via G. Verdi, 8
65015 Montesilvano (PE)
T +39 0859 35 4078
F +39 0859 35 9471
E info@eickemeyer.it
www.eickemeyer.it
CANADA
EICKEMEYER Inc.
250 Briarhill Dr.
Stratford, Ont. Canada
N5A 7S2
T +1 519 273 5558
F +1 519 271 7114
E info@eickemeyer.com
www.eickemeyerveterinary.com
USER CONTROLS
There are only 2 buttons on the LifeVet 02unit.The left hand button is labelled 'CALIBRATE' and
the right hand button is labelled ‚ON / OFF‘. For routine use only the ON/OFF button need be used
to turn the unit on and off.
Button Use
ON / OFF
Press this button once to turn on the unit. When the unit is turned on, 6 horizontal lines will appear
on the screen for 3 seconds and then be replaced by the measured oxygen percentage value.
CALIBRATE
Press and hold this button for 5 seconds to start the calibration procedure.
CALIBRATION PROCEDURE
The calibration procedure can be run in two parts – a simple 21 % recalibration or a 21–100 %
span
calibration. To run a span calibration you must have a guaranteed source of 100% oxygen,
such as from
an oxygen bottle or hospital oxygen line.
Performing a 21 % recalibration
1) Press and hold the CALIBRATE button until the screen fl ashes CS
2) Release the button
3)
Press the CALIBRATE button for 1 second – the screen changes to a fl ashing 21
4) Ensure that the unit is in clean room air for at least 60 seconds and then press the CALIBRATE
button again for 1 second
5) The screen changes to a fl ashing 100
6) Turn the unit off using the ON/ OFF button
Performing a 21–100 % span calibration
1) Press and hold the CALIBRATE button until the screen fl ashes CS
2) Release the button
3)
Press the CALIBRATE button for 1 second – the screen changes to a fl ashing 21
4) Ensure that the unit is in clean room air for at least 60 seconds and
then press the CALIBRATE button again for 1 second
5) The screen changes to a fl ashing 100
6) Flow 100% oxygen through the LifeVet 02connector for 60 seconds
7) After 60 seconds press the CALIBRATE button for 1 second
8) The screen changes to a fl ashing CC, indicating Calibration Complete
9) Press the CALIBRATE button one more time to leave the calibration routine
MAINTENANCE
Changing the battery
When the battery level is low, the screen will fl ash slowly about once every 2 seconds. If the fl ashing
is continuous then the battery is below the minimum level and should be replaced. Undo the back
cover of the device by undoing the securing screws in the corner of the backplate. Remove the battery
from the battery holder and replace with a long life, low drain 1.5v Alkaline battery. Replace the
backplate and tighten all the corner securing screws.
Replacing the 22 mm Adaptor seal
There is an O-ring seal in the 22 mm Adaptor that prevents the leakage of gas from the anaesthetic
circuit. If this seal becomes damaged or is lost it must be replaced. Use a small screwdriver to ease
out the old seal and then press the new seal into place. New seals can be obtained from EICKEMEYER®
under Replacement Part number E21319501.
TROUBLESHOOTING
Problem: There are 6 horizontal lines on the screen all the time.
Solution: The oxygen cell is depleted. Contact EICKEMEYER®to arrange a unit service.
Problem: The display is constantly fl ashing although the numbers seem correct.
Solution: The internal AA battery is low. Replace the battery.
Problem: There is nothing on the display and no response to the ON / OFF button.
Solution: Fit a new Long-life AA battery. If this does not resolve the problem, contact EICKEMEYER®
to arrange a unit service.
Art. No. 213195

ALLGEMEINES
Das Sauerstoffmessgerät LifeVet 02ist ein batteriebetriebenes Gerät zur Überwachung des prozentualen
Sauerstoffgehalts bei der Anästhesie oder in freier Luft. Das LifeVet 02verwendet eine langlebige galva-
nische Zelle und angeschlossene Elektronik, um den Sauerstoffgehalt zu messen und anzuzeigen.
Das Gerät benötigt eine einzelne 1,5 Volt Alkalibatterie und wird damit bei einer Nutzung von 8 Stunden
täglich etwa ein Jahr lang laufen.
MERKMALE
• Einfach einzusetzen
• Der Grundkörper ist am Adapter um 360 Grad in die gewünschte Richtung drehbar
• Einfach zu kalibrieren und anzuschließen
• Verwendbar im Kreisteil oder in freier Luft
• Adapter zum leichten Anschluss in 22 mm / 15 mm Kreisteil
• Langlebiger Sensor
• Langlebige Batterie
BEDIENELEMENTE
Am LifeVet 02gibt es nur 2 Bedienelemente. Die linke Taste heißt CALIBRATE und die rechte Taste heißt
ON / OFF. Im Routineeinsatz ist nur die Taste ON / OFF zum Ein- und Ausschalten des Geräts nötig.
Tastenfunktion
ON / OFF
Zum Einschalten des Geräts drücken Sie einmal auf diese Taste. Nach dem Einschalten erscheinen 3 Sekun-
den lang 6 waagerechte Linien, bevor der gemessene Sauerstoffanteil in Prozent angezeigt wird.
CALIBRATE
Halten Sie diese Taste 5 Sekunden lang gedrückt, um den Kalibriervorgang zu starten.
KALIBRIERVORGANG
Der Kalibriervorgang kann auf zwei Arten erfolgen – eine einfache Neukalibrierung bei 21 % (Raumluft)
oder eine Kalibrierung im Bereich 21–100%. Für eine Kalibrierung im Bereich 21–100 % müssen Sie
eine zuverlässige Quelle mit reinem 100 % Sauerstoff haben, etwa eine Sauerstofffl asche.
Ausführung der 21 % Neukalibrierung
1) Halten Sie die Taste CALIBRATE gedrückt, bis die Anzeige CS blinkt
2) Lassen Sie die Taste los
3)
Drücken Sie 1 Sekunde lang die Taste
CALIBRATE
– die Anzeige wechselt zu einer blinkenden 21
4) Stellen Sie sicher, dass sich das Gerät mindestens 60 Sekunden lang in sauberer Raumluft
befi ndet und drücken dann wieder 1 Sekunde lang die Taste CALIBRATE
5) Die Anzeige wechselt zu einer blinkenden 100
6) Schalten Sie das Gerät mit der ON/ OFF-Taste aus
Ausführung der Kalibrierung für den Bereich 21–100 %
1) Halten Sie die Taste CALIBRATE gedrückt, bis die Anzeige CS blinkt
2) Lassen Sie die Taste los
3)
Drücken Sie 1 Sekunde lang die Taste
CALIBRATE
– die Anzeige wechselt zu einer blinkenden 21
4) Stellen Sie sicher, dass sich das Gerät mindestens 60 Sekunden lang in sauberer Raumluft
befi ndet und drücken dann wieder 1 Sekunde lang die Taste CALIBRATE
5) Die Anzeige wechselt zu einer blinkenden 100
6)
Lassen Sie 100% reinen Sauerstoff 60 Sekunden lang durch den Anschluss des LifeVet 02strömen
7) Nach 60 Sekunden drücken Sie 1 Sekunde lang die Taste CALIBRATE
8) Die Anzeige wechselt zum blinkenden CC für Kalibrierung komplett
9) Drücken Sie nochmal die Taste CALIBRATE, um den Kalibriermodus zu verlassen
WARTUNG
Batterie wechseln
1) Bei geringem Batteriestand blinkt die Anzeige langsam etwa alle 2 Sekunden. Wenn die Anzeige
ständig blinkt, ist die Batterie zu schwach und muss gewechselt werden.
2) Entfernen Sie die Rückwand des Geräts, indem Sie die Sicherungsschrauben an den Ecken
der Rückwand lösen. Entnehmen Sie die Batterie aus dem Batteriefach und ersetzen Sie diese
mit einer langlebigen, auslaufsicheren 1,5V-Alkalibatterie. Montieren Sie die Rückwand wieder
und ziehen alle Schrauben an den Ecken fest.
Dichtung des 22 mm Adapters wechseln
1) Ein Dichtungsring am 22 mm Adapter verhindert einen Gasaustritt aus dem Anästhesiekreislauf.
Falls diese Dichtung beschädigt ist oder fehlt, muss sie ersetzt werden. Heben Sie die alte
Dichtung mit einem kleinen Schraubendreher heraus und drücken dann die neue Dichtung hinein.
2) Neue Dichtungen erhalten Sie von
EICKEMEYER®
unter der Ersatzteilnummer E21319501.
FEHLERBEHEBUNG
Problem: Die Anzeige zeigt stets 6 waagerechte Linien an.
Lösung: Die Sauerstoffzelle ist erschöpft. Wenden Sie sich für eine Wartung des Geräts
an EICKEMEYER®.
Problem: Die Anzeige blinkt ständig, obwohl die Werte richtig erscheinen.
Lösung: Die interne AA-Batterie ist schwach. Tauschen Sie die Batterie aus.
Problem: Die Anzeige bleibt leer und das Gerät reagiert nicht auf die ON / OFF-Taste.
Lösung: Setzen Sie eine
neue langlebige AA-Batterie ein. Falls das Problem nicht behoben ist,
wenden Sie sich zwecks Wartung
des Geräts an
EICKEMEYER®
.
GEBRAUCHSANWEISUNG
LIFEVET 02SAUERSTOFFMESSGERÄT
LifeVet 02Grundkörper
Hauptanzeige
CALIBRATE-Taste
ON/OFF-Taste
LifeVet 02Rückwand
22 mm/15 mm Adapter
GENERAL
The LifeVet 02Oxygen Analyser is a battery-powered device suitable for monitoring the percentage
of oxygen in an anaesthetic or in free-air. The LifeVet 02uses a longlife galvanic cell and associated
micro-power electronics to measure and display the oxygen level. The unit is powered by a single
1.5 volt alkaline battery and is designed to run for approximately one year, based on a usage rate of
8 hours per day.
FEATURES
• Simple to use
• Main body can be rotated through 360 degrees on adaptor for preferred viewing position
• Simple calibration and span procedures
• Use in-circuit or in free-air
• Adaptor allows easy connection in 22 mm/ 15 mm circuits
• Long sensor life
• Long battery life
USER MANUAL
LIFEVET 02 OXYGEN ANALYSER
LifeVet 02Main Body
Main screen
CALIBRATE Button
ON/OFF Button
LifeVet 02Back Panel
22 mm/15 mm Adaptor

LIFEVET 02
SAUERSTOFFMESSGERÄT/OXYGEN ANALYZER
BEDIENUNGSANLEITUNG/USER MANUAL
TELEPHONE 0049 7461 96580-0
www.eickemeyer.com
GERMANY
EICKEMEYER KG
Eltastraße 8
78532 Tuttlingen
T +49 7461 96 580 0
F +49 7461 96 580 90
E info@eickemeyer.de
www.eickemeyer.de
SWITZERLAND
EICKEMEYER AG
Sandgrube 29
9050 Appenzell
T +41 71 788 23 13
F +41 71 788 23 14
E info@eickemeyer.ch
www.eickemeyer.ch
UNITED KINGDOM
EICKEMEYER Ltd.
3 Windmill Business Village
Brooklands Close
Sunbury-on-Thames
Surrey, TW16 7DY
T +44 20 8891 2007
F +44 20 8891 2686
E info@eickemeyer.co.uk
www.eickemeyer.co.uk
DENMARK
EICKEMEYER ApS
Lysbjergvej 6, Hammelev
6500 Vojens
T +45 7020 5019
F +45 7353 5019
E info@eickemeyer.dk
www.eickemeyer.dk
NETHERLANDS
EICKEMEYER B.V.
Bedrijventerrein
Pavijen-West Bellweg 44
4104 BJ Culemborg
T +31 345 58 9400
F +31 345 58 9401
E info@eickemeyer.nl
www.eickemeyer.nl
ITALY
EICKEMEYER S.R.L.
Via G. Verdi, 8
65015 Montesilvano (PE)
T +39 0859 35 4078
F +39 0859 35 9471
E info@eickemeyer.it
www.eickemeyer.it
CANADA
EICKEMEYER Inc.
250 Briarhill Dr.
Stratford, Ont. Canada
N5A 7S2
T +1 519 273 5558
F +1 519 271 7114
E info@eickemeyer.com
www.eickemeyerveterinary.com
USER CONTROLS
There are only 2 buttons on the LifeVet 02unit.The left hand button is labelled 'CALIBRATE' and
the right hand button is labelled ‚ON/OFF‘. For routine use only the ON/OFF button need be used
to turn the unit on and off.
Button Use
ON / OFF
Press this button once to turn on the unit. When the unit is turned on, 6 horizontal lines will appear
on the screen for 3 seconds and then be replaced by the measured oxygen percentage value.
CALIBRATE
Press and hold this button for 5 seconds to start the calibration procedure.
CALIBRATION PROCEDURE
The calibration procedure can be run in two parts – a simple 21 % recalibration or a 21–100 %
span
calibration. To run a span calibration you must have a guaranteed source of 100% oxygen,
such as from
an oxygen bottle or hospital oxygen line.
Performing a 21 % recalibration
1) Press and hold the CALIBRATE button until the screen fl ashes CS
2) Release the button
3)
Press the CALIBRATE button for 1 second – the screen changes to a fl ashing 21
4) Ensure that the unit is in clean room air for at least 60 seconds and then press the CALIBRATE
button again for 1 second
5) The screen changes to a fl ashing 100
6) Turn the unit off using the ON/ OFF button
Performing a 21–100 % span calibration
1) Press and hold the CALIBRATE button until the screen fl ashes CS
2) Release the button
3)
Press the CALIBRATE button for 1 second – the screen changes to a fl ashing 21
4) Ensure that the unit is in clean room air for at least 60 seconds and
then press the CALIBRATE button again for 1 second
5) The screen changes to a fl ashing 100
6) Flow 100% oxygen through the LifeVet 02connector for 60 seconds
7) After 60 seconds press the CALIBRATE button for 1 second
8) The screen changes to a fl ashing CC, indicating Calibration Complete
9) Press the CALIBRATE button one more time to leave the calibration routine
MAINTENANCE
Changing the battery
When the battery level is low, the screen will fl ash slowly about once every 2 seconds. If the fl ashing
is continuous then the battery is below the minimum level and should be replaced. Undo the back
cover of the device by undoing the securing screws in the corner of the backplate. Remove the battery
from the battery holder and replace with a long life, low drain 1.5v Alkaline battery. Replace the
backplate and tighten all the corner securing screws.
Replacing the 22 mm Adaptor seal
There is an O-ring seal in the 22 mm Adaptor that prevents the leakage of gas from the anaesthetic
circuit. If this seal becomes damaged or is lost it must be replaced. Use a small screwdriver to ease
out the old seal and then press the new seal into place. New seals can be obtained from EICKEMEYER®
under Replacement Part number E21319501.
TROUBLESHOOTING
Problem: There are 6 horizontal lines on the screen all the time.
Solution: The oxygen cell is depleted. Contact EICKEMEYER®to arrange a unit service.
Problem: The display is constantly fl ashing although the numbers seem correct.
Solution: The internal AA battery is low. Replace the battery.
Problem: There is nothing on the display and no response to the ON / OFF button.
Solution: Fit a new Long-life AA battery. If this does not resolve the problem, contact EICKEMEYER®
to arrange a unit service.
Art. No. 213195

LIFEVET 02
SAUERSTOFFMESSGERÄT/OXYGEN ANALYZER
BEDIENUNGSANLEITUNG/USER MANUAL
TELEPHONE 0049 7461 96580-0
www.eickemeyer.com
GERMANY
EICKEMEYER KG
Eltastraße 8
78532 Tuttlingen
T +49 7461 96 580 0
F +49 7461 96 580 90
E info@eickemeyer.de
www.eickemeyer.de
SWITZERLAND
EICKEMEYER AG
Sandgrube 29
9050 Appenzell
T +41 71 788 23 13
F +41 71 788 23 14
E info@eickemeyer.ch
www.eickemeyer.ch
UNITED KINGDOM
EICKEMEYER Ltd.
3 Windmill Business Village
Brooklands Close
Sunbury-on-Thames
Surrey, TW16 7DY
T +44 20 8891 2007
F +44 20 8891 2686
E info@eickemeyer.co.uk
www.eickemeyer.co.uk
DENMARK
EICKEMEYER ApS
Lysbjergvej 6, Hammelev
6500 Vojens
T +45 7020 5019
F +45 7353 5019
E info@eickemeyer.dk
www.eickemeyer.dk
NETHERLANDS
EICKEMEYER B.V.
Bedrijventerrein
Pavijen-West Bellweg 44
4104 BJ Culemborg
T +31 345 58 9400
F +31 345 58 9401
E info@eickemeyer.nl
www.eickemeyer.nl
ITALY
EICKEMEYER S.R.L.
Via G. Verdi, 8
65015 Montesilvano (PE)
T +39 0859 35 4078
F +39 0859 35 9471
E info@eickemeyer.it
www.eickemeyer.it
CANADA
EICKEMEYER Inc.
250 Briarhill Dr.
Stratford, Ont. Canada
N5A 7S2
T +1 519 273 5558
F +1 519 271 7114
E info@eickemeyer.com
www.eickemeyerveterinary.com
USER CONTROLS
There are only 2 buttons on the LifeVet 02unit.The left hand button is labelled 'CALIBRATE' and
the right hand button is labelled ‚ON/OFF‘. For routine use only the ON/OFF button need be used
to turn the unit on and off.
Button Use
ON / OFF
Press this button once to turn on the unit. When the unit is turned on, 6 horizontal lines will appear
on the screen for 3 seconds and then be replaced by the measured oxygen percentage value.
CALIBRATE
Press and hold this button for 5 seconds to start the calibration procedure.
CALIBRATION PROCEDURE
The calibration procedure can be run in two parts – a simple 21 % recalibration or a 21–100 %
span
calibration. To run a span calibration you must have a guaranteed source of 100% oxygen,
such as from
an oxygen bottle or hospital oxygen line.
Performing a 21 % recalibration
1) Press and hold the CALIBRATE button until the screen fl ashes CS
2) Release the button
3)
Press the CALIBRATE button for 1 second – the screen changes to a fl ashing 21
4) Ensure that the unit is in clean room air for at least 60 seconds and then press the CALIBRATE
button again for 1 second
5) The screen changes to a fl ashing 100
6) Turn the unit off using the ON/ OFF button
Performing a 21–100 % span calibration
1) Press and hold the CALIBRATE button until the screen fl ashes CS
2) Release the button
3)
Press the CALIBRATE button for 1 second – the screen changes to a fl ashing 21
4) Ensure that the unit is in clean room air for at least 60 seconds and
then press the CALIBRATE button again for 1 second
5) The screen changes to a fl ashing 100
6) Flow 100% oxygen through the LifeVet 02connector for 60 seconds
7) After 60 seconds press the CALIBRATE button for 1 second
8) The screen changes to a fl ashing CC, indicating Calibration Complete
9) Press the CALIBRATE button one more time to leave the calibration routine
MAINTENANCE
Changing the battery
When the battery level is low, the screen will fl ash slowly about once every 2 seconds. If the fl ashing
is continuous then the battery is below the minimum level and should be replaced. Undo the back
cover of the device by undoing the securing screws in the corner of the backplate. Remove the battery
from the battery holder and replace with a long life, low drain 1.5v Alkaline battery. Replace the
backplate and tighten all the corner securing screws.
Replacing the 22 mm Adaptor seal
There is an O-ring seal in the 22 mm Adaptor that prevents the leakage of gas from the anaesthetic
circuit. If this seal becomes damaged or is lost it must be replaced. Use a small screwdriver to ease
out the old seal and then press the new seal into place. New seals can be obtained from EICKEMEYER®
under Replacement Part number E21319501.
TROUBLESHOOTING
Problem: There are 6 horizontal lines on the screen all the time.
Solution: The oxygen cell is depleted. Contact EICKEMEYER®to arrange a unit service.
Problem: The display is constantly fl ashing although the numbers seem correct.
Solution: The internal AA battery is low. Replace the battery.
Problem: There is nothing on the display and no response to the ON / OFF button.
Solution: Fit a new Long-life AA battery. If this does not resolve the problem, contact EICKEMEYER®
to arrange a unit service.
Art. No. 213195

LIFEVET 02
SAUERSTOFFMESSGERÄT/OXYGEN ANALYZER
BEDIENUNGSANLEITUNG/USER MANUAL
TELEPHONE 0049 7461 96580-0
www.eickemeyer.com
GERMANY
EICKEMEYER KG
Eltastraße 8
78532 Tuttlingen
T +49 7461 96 580 0
F +49 7461 96 580 90
E info@eickemeyer.de
www.eickemeyer.de
SWITZERLAND
EICKEMEYER AG
Sandgrube 29
9050 Appenzell
T +41 71 788 23 13
F +41 71 788 23 14
E info@eickemeyer.ch
www.eickemeyer.ch
UNITED KINGDOM
EICKEMEYER Ltd.
3 Windmill Business Village
Brooklands Close
Sunbury-on-Thames
Surrey, TW16 7DY
T +44 20 8891 2007
F +44 20 8891 2686
E info@eickemeyer.co.uk
www.eickemeyer.co.uk
DENMARK
EICKEMEYER ApS
Lysbjergvej 6, Hammelev
6500 Vojens
T +45 7020 5019
F +45 7353 5019
E info@eickemeyer.dk
www.eickemeyer.dk
NETHERLANDS
EICKEMEYER B.V.
Bedrijventerrein
Pavijen-West Bellweg 44
4104 BJ Culemborg
T +31 345 58 9400
F +31 345 58 9401
E info@eickemeyer.nl
www.eickemeyer.nl
ITALY
EICKEMEYER S.R.L.
Via G. Verdi, 8
65015 Montesilvano (PE)
T +39 0859 35 4078
F +39 0859 35 9471
E info@eickemeyer.it
www.eickemeyer.it
CANADA
EICKEMEYER Inc.
250 Briarhill Dr.
Stratford, Ont. Canada
N5A 7S2
T +1 519 273 5558
F +1 519 271 7114
E info@eickemeyer.com
www.eickemeyerveterinary.com
USER CONTROLS
There are only 2 buttons on the LifeVet 02unit.The left hand button is labelled 'CALIBRATE' and
the right hand button is labelled ‚ON/OFF‘. For routine use only the ON/OFF button need be used
to turn the unit on and off.
Button Use
ON / OFF
Press this button once to turn on the unit. When the unit is turned on, 6 horizontal lines will appear
on the screen for 3 seconds and then be replaced by the measured oxygen percentage value.
CALIBRATE
Press and hold this button for 5 seconds to start the calibration procedure.
CALIBRATION PROCEDURE
The calibration procedure can be run in two parts – a simple 21 % recalibration or a 21–100 %
span
calibration. To run a span calibration you must have a guaranteed source of 100% oxygen,
such as from
an oxygen bottle or hospital oxygen line.
Performing a 21 % recalibration
1) Press and hold the CALIBRATE button until the screen fl ashes CS
2) Release the button
3)
Press the CALIBRATE button for 1 second – the screen changes to a fl ashing 21
4) Ensure that the unit is in clean room air for at least 60 seconds and then press the CALIBRATE
button again for 1 second
5) The screen changes to a fl ashing 100
6) Turn the unit off using the ON/ OFF button
Performing a 21–100 % span calibration
1) Press and hold the CALIBRATE button until the screen fl ashes CS
2) Release the button
3)
Press the CALIBRATE button for 1 second – the screen changes to a fl ashing 21
4) Ensure that the unit is in clean room air for at least 60 seconds and
then press the CALIBRATE button again for 1 second
5) The screen changes to a fl ashing 100
6) Flow 100% oxygen through the LifeVet 02connector for 60 seconds
7) After 60 seconds press the CALIBRATE button for 1 second
8) The screen changes to a fl ashing CC, indicating Calibration Complete
9) Press the CALIBRATE button one more time to leave the calibration routine
MAINTENANCE
Changing the battery
When the battery level is low, the screen will fl ash slowly about once every 2 seconds. If the fl ashing
is continuous then the battery is below the minimum level and should be replaced. Undo the back
cover of the device by undoing the securing screws in the corner of the backplate. Remove the battery
from the battery holder and replace with a long life, low drain 1.5v Alkaline battery. Replace the
backplate and tighten all the corner securing screws.
Replacing the 22 mm Adaptor seal
There is an O-ring seal in the 22 mm Adaptor that prevents the leakage of gas from the anaesthetic
circuit. If this seal becomes damaged or is lost it must be replaced. Use a small screwdriver to ease
out the old seal and then press the new seal into place. New seals can be obtained from EICKEMEYER®
under Replacement Part number E21319501.
TROUBLESHOOTING
Problem: There are 6 horizontal lines on the screen all the time.
Solution: The oxygen cell is depleted. Contact EICKEMEYER®to arrange a unit service.
Problem: The display is constantly fl ashing although the numbers seem correct.
Solution: The internal AA battery is low. Replace the battery.
Problem: There is nothing on the display and no response to the ON / OFF button.
Solution: Fit a new Long-life AA battery. If this does not resolve the problem, contact EICKEMEYER®
to arrange a unit service.
Art. No. 213195
Other manuals for LIFEVET O2
1