
PRELIMINARYADJUSTMENTS OF THE SPECTROMETER :
First focus the cross wires by moving the eye piece in and out. This fixes the position of the
eye piece, relative to the cross wires. Next pointing the telescope through an open
window at some distant object lying at infinity such as an electric pole, adjust the distance
between the eye piece and the objective with the help of helical arrangement, until there is
no parallax between distinct image of distant object (at infinity) and the cross wire. When
this is done the telescope must not be readjusted again during the experiment. It may
happen that a second observer whose sight differs from that of the first is unable to focus
the cross wires easily. He may readjust the eye piece provided that he does not alter any
other part of the telescope.
The telescope is now turned towards the collimator with its slit illuminated with a
monochromatic source of light. The collimator lens is then, adjusted with the help of
helical arrangement until a distinct image of the slit falls on the cross wires and there is no
parallax between the two. The instrument is now ready so that parallel rays pass from the
collimator to the telescope.
Schuster' Method : In the absence of a distant object this method may be used. After
focusing the eye piece on cross wires the prism is so placed on the prism table that it has
maximum illumination from the collimator. The prism table is then rotated slowly, following
the refracted image with the telescope until the position of minimum deviation is found.
Now the table is so rotated, so that the refracting edge is slightly turned towards the
telescope causing the rotation of the image. The telescope is adjusted to give the distinct
image. Again the prism table is turned to the other side of the minimum deviation i.e. the
refracting edge towards the collimator and the collimator is adjusted to give a fine and
distinct image of the slit. This procedure is repeated a number of times (usually three to
four alternate focusing are sufficient) till the image does not blur on rotation.
Levelling of the Table: Place the prism on the table of the spectrometer, so that its
refracting edge is at the center and parallel to the slit of the collimator and one of its clear
faces say AC is perpendicular to the line joining two of the levelling screws P & Q. Adjust
the prism table so that parallel beam of light illuminates both the faces surrounding the
edge, as shown in the Fig. 1.
Receive the light in the telescope in the position T. Adjust the levelling screws P & Q to
1
bring the reflected image in the middle of the field of view of the telescope. Next turn the
telescope to face AB in the position Tand adjust the screw Rto bring the image again in
2
the center of the field of view.
Expt. 1 : To determine the angle of the prism by any of the following two methods:
a) By turning the Telescope : Having adjusted the image symmetrical in the field of
view of the telescope, on either face, allow the telescope to receive rays along AT from
1
the face AC and along AT from the face AB each time the slit falling upon the across
2
wires. The Difference between the two readings (angle T AT ) is equal to twice the angle
1 2
of prism i.e, 2Awhere Ais angle of the prism. Half this difference gives the anglesA.
-2-
© EISCO SCIENTIFIC ●www.eiscolabs.com