
Model2241-2i&-2i/2 TechnicalManual Section2
Ludlum Measurements, Inc. Page 2-3 January 2018
calibration. Alternatively, expose the detector to a source of known value
and verify that the detector detects greater than or equal to the efficiency
listed in the detector manuals’ specification section. Failure to meet these
criteria may indicate a detector malfunction.
Switch the AUD ON/OFF switch to the ON position and confirm that the
external unimorph speaker produces an audible click for each event detected
(audio divide-by 1 parameter). The AUD ON/OFF switch will silence the
clicks if in the OFFposition; however, an audible alarm condition will still be
heard.
Increase the source activity or lower the alert and alarm points to initiate an
ALERT and ALARM condition. (Refer to Section 8, “Entering or Changing
Switchboard Parameters.”) Depress the RESET switch to acknowledge the
audible alarm. Decrease the radiation activity below the ALERT and ALARM
threshold and depress the RESET switch to clear the alarm conditions. If an
alarm condition is not present, depressing the RESET switch the first time will
reset the alert condition and zero the ratemeter.
Position a check source to produce a ratemeter reading of 100 to 2000
counts per minute or 10-100 µR/hr. While observing the ratemeter
fluctuations, select between the fast and slow response time (F/S) positions
to observe variations in the display. The Sposition should respond
approximately five times slower than the Fposition (for fixed response
mode) and three times slower when in variable response mode. The slow
response position is normally used when the meter is displaying low
numbers, which require a more stable display. The fast response position is
used at high count levels.
Move the OFF/SCALER/RATEMETER switch to the SCALER position. Depress
the COUNT switch located in the end of the carrying handle in order to
initiate a count cycle. The word COUNTING should be flashing on the LCD
during the count cycle and should disappear at the end of the predetermined
count time. If a scaler ALARM condition occurs, the RESET switch can be
depressed to acknowledge the alarm; however, the COUNT switch must be
depressed to clear the visual ALARM and to restart the count cycle.
Depress and release the LIGHT switch. The backlight located behind the
LCD should illuminate (for pre-programmed ON time). Select the desired
F/S, AUD ON/OFF, INT/EXT and RATEMETER or SCALER parameters and
proceed to use the instrument.