Upkeep of the Machine
5 Upkeep of the Machine
The device should be cleaned once a week. In order to allow for
a simple cleaning of the device, the shrouding cover over the
polishing area can be removed. To that end, the captivated
screw, located in the upper left area of the cover, must be
loosened. The shrouding cover can now be tipped back or, by
pushing it to the left, it can be removed from the hinges. Before
removal of the hood, the plug of the lamp must be
disconnected. The device and the interior of the shrouding
cover can now easily be cleaned with a soft cloth. When it is
heavily soiled, a mild cleaning agent can be used. To
reassemble the device, the shrouding cover is placed upon the
device and tightened down with the captivated screw. Once the
shrouding cover is remounted, the lamp must be re-connected
with the aid of the plug.
If the output of the extractor decreases markedly, the filter mat
must be changed. The filter is to be found in the filter drawer on
the front right-hand side of the machine. To change the filter
mat, the two screws are each turned a quarter rotation to the
left and the filter drawer is pulled forward out of the machine.
The old filter should be removed and the new filter inserted with
the printed side down. By emptying the gauze filter that is
located under the polishing wheel, gross detritus such as brush
hairs, etc., can be removed.
Filter met information: the filter mat in use is a fine filter and is
extremely well suited to screen out dust arising from the
polishing process. The surface is adhesive with a degree of
separation of 97% in accordance with DIN 24 185.
This filter mat is not washable. If filters capable of being washed
out are desired, they are available on request. These filter mats
belong to the coarse filter category with a filtration efficiency of
94% according to DIN 24 185. Disadvantage: fine dust particles
are not filtered out.
When the polishing wheel is being changed, the equipment
should be switched off and the plug removed.
The polishing engine does not require any attendance.
For your own safety, we recommend that you check the mains
cable and the mains connection jack at the unit for damage at
regular intervals.
There are no fuses in the machine.
In case of technical faults please contact your supplier or the
manufacturer of the unit.
Repair and maintenance works which require the unit to be
connected and opened must be carried out by authorized and
specialized personnel only.
Mains cable and
mains connection
Opening by
authorized
specialized
personnel only
6BA_Unispeed_EN_Vers.01.2014 © Elma Schmidbauer GmbH