ENJI PRIME Happy Care INSTANT FLEXIBLE DIGITAL... User manual

INSTANT FLEXIBLE DIGITAL
THERMOMETER
USER’S
MANUAL
*Read instructions carefully before using the Digital
Thermometer.
WARNING
Choking Hazard:
Thermometer cap may be fatal when swallowed. Do not
allow children to use this device without parental
supervision.
Do not use thermometer in ear. Designed for use in oral,
rectal and armpit (axilla) readings only.
Do not place thermometer near extreme heat, it may
explode.
Use of the Probe Cover may result in a 0.2oF (0.1oC)
discrepancy from actual temperature.
Remove battery from the device when not in operation
for a long time.
The use of temperature readings for self-diagnosis is
dangerous. Consult your doctor for the interpretations of
the results. Self-diagnosis may lead to the worsening of
the existing disease conditions.
Do not attempt to get measurements when the
thermometer is wet as inaccurate readings may result. Do
not bite the thermometer. Doing so may lead to breakage
and/or injury.
Do not attempt to disassemble or repair the
thermometer. Doing so may result in inaccurate
readings.
After each use, disinfect the thermometer especially in
case the device is used by more than one person.
Do not force the thermometer into the rectum. Stop
insertion and abort the measurement when pain is
present. Failure to do so may lead to injury.
Do not use thermometer orally after being used rectally.
We advise you to purchase our Probe Covers for better
hygiene when used by the whole family.
For children who are two years old or younger, please do
not use the devices orally if the unit has been stored at
temperatures over 41℉~104℉(5℃~40℃), leave it in
41℉~104℉(5℃~40℃) ambient temperature for about
15 minutes before using it.
This digital thermometer provides a quick and highly accurate reading of an individual's body
temperature. The digital thermometer is intended to measure the human body’s temperature in
regular mode orally, rectally or under the arm, and the device is reusable for clinical or home use
on people of all ages. To better understand its functions and to provide years of dependable results,
please read all instructions first.
This appliance conforms to the following standards:
ASTM E1112 Standard Specification for Electronic Thermometer for Intermittent Determination of
Patient Temperature, ISO 80601-2-56 Medical electrical equipment —Part 2-56:Particular
requirements for basic safety and essential performance of clinical thermometers for body
temperature measurement, IEC 60601-1-11 Medical electrical equipment —Part 1-11: General
requirements for basic safety and essential performance –Collateral Standard: Requirements for
medical electrical equipment and medical electrical systems used in the home healthcare
environment and complies with the requirements of IEC 60601-1-2(EMC), AAMI/ANSI ES60601-
1(Safety) standards. And the manufacturer is ISO 13485 certified.
DIRECTIONS
1. Press the On/Off Button next to LCD display. A tone will sound as the screen shows,
followed by last recorded temperature. After showing the self-test temperature, the
thermometer is now in the testing mode.
2. Position the thermometer in desired location (mouth, rectum, or armpit).
a. Oral Use: Place thermometer under tongue as indicated by “” position shown in
Figure 2. Close your mouth and breathe evenly through the nose
to prevent the measurement from being influenced by
inhaled/exhaled air. Normal temperature between 96.3oF and
99.1oF (35.7oC and 37.3oC).
b. Rectal Use: Lubricate probe tip with petroleum jelly for easy
insertion. Gently insert sensor approximately 1 cm (less than ½”)
into rectum. Normal temperature between 97.2oF and 99.9oF
(36.2oC and 37.7oC).
c. Armpit Use: Wipe armpit dry. Place probe in armpit and keep arm pressed firmly at
side. From a medical viewpoint, this method will always provide inaccurate readings, and
should not be used if precise measurements are required. Normal temperature between
95.4oF and 98.1oF (35.2oC and 36.7oC).
3. The degree sign flashes throughout the testing process. When flashing stops, a soft alarm
will beep. The reading may take from 10 to 30 seconds depending on where the computer
is used. The measured reading will appear on the LCD simultaneously. The minimum
measurement time until the signal tone (beep) must be maintained without exception.
The measurement continues even after the buzzing notification. So that in order to
achieve better body temperature measurement result, recommend to keep the probe in
mouth and rectum about 2 minutes, or in armpit about 5 minutes regardless of the beep
sound and at least 30 seconds measurement interval should be maintained in the first few
uses.
4. To prolong battery life, press the On/Off Button to turn unit off after testing is complete.
If no action is taken, the unit will automatically shut off after around 10 minutes.

TROUBLESHOOTING
ERROR
MESSAGE
PROBLEM
SOLUTION
Lo
Temperature taken is
lower than 90.0°F(32.0°C)
Turn off, wait one minute and take a
new temperature via close contact and
sufficient rest.
Hi
Temperature taken is
higher than 109.9°F(42.9°C)
Turn off, wait one minute and take a
new temperature via close contact and
sufficient rest.
Err
The system is not functioning
properly.
Unload the battery, wait for 1 minute
and repower it. If the message
reappears, contact the retailer for
service.
Dead battery: Battery icon is
flashing, can’t measure.
Replace the battery.
How To Calibrate Your Thermometer:
The thermometer is initially calibrated at the time of manufacturing. If the thermometer is
used according to the manual, periodic readjustment is not required. However, we
recommend conducting a quick calibration to confirm that your unit didn't get damaged during
shipment. A quick way to calibrate your unit is by doing the water test. Dip the tip of the
thermometer into a cup of hot water. Wait for your thermometer to show 'Hi' or a high
temperature reading number on its LCD display (for example it shows 38 °C or 100 °F). Then
let it sit for a minute or two then dip it into a cup of cold water. Your display should show 'Lo'
(meaning low temperature not low battery) or a low temperature reading number this time
(for example it shows 33 °C or 91°F). Leave it at room temperature for 30 minutes and try using
it again, and if it does not show ‘Hi’ or ‘Lo’ it will mean you have successfully checked your
thermometer and good to go! If you find that your reading is very low and/or not changing on
the LCD display at all, please get in touch with us through email with your Amazon Order ID.
SYMBOL EXPLANATION
Direct Current
Batch Code
Type BF Applied Part
Manufacturer
Consult
Accompanying
Documents
Storage and Transportation
Temperature Limit: -4 oF ~ 131oF (-
20 oC ~ 55oC)
CLEANING & DISINFECTION
Wipe the thermometer with a soft clean cloth. For stubborn stains, wipe the thermometer
with a cloth that has been dampened in water or in neutral detergent solution and then wring
thoroughly. Finish by wiping with a soft dry cloth. For disinfection, 75% Ethanol or Isopropyl
alcohol can be used. Observe the following to prevent damage to the thermometer:
Do not use benzene, thinner, gasoline or other strong solvents to clean the
thermometer.
Do not attempt to disinfect the sensing section (tip) of the thermometer by
immersing in alcohol or in hot water (water over 122oF/50oC).
Do not use ultrasonic washing to clean the thermometer.
oF/oC SWITCHABLE
Temperature readings are available in the Fahrenheit or Celsius scale (oF/oC; located in the
upper right corner of LCD). With the unit off, press and hold On/Off Button for approximately
2 seconds to change the current setting.
SPECIFICATIONS:
Type:
Digital Thermometer (Not Predictive)
Measure range:
90.0oF-109.9oF (32.0oC-42.9 oC) (oC
/oF chosen by manufacturer)
Accuracy:
0.2oF ( 0.1oC) during 95.9oF ~
107.6oF (35.5oC ~ 42.0oC) at 64.4oF ~
82.4oF (18oC ~ 28oC) ambient
operating range 0.4oF ( 0.2oC) for
other measuring and ambient
operating range
Operating mode:
Direct Mode
Display:
Liquid crystal display, 3 ½ digits
Memory:
For storing the last measured value
Battery:
One 1.5 V DC. Button battery (size
LR41 or SR41, UCC 392)
Battery life:
Approx. 200 hours of continuous
operation of 1 year with 3
measurements per day
Dimension:
13.9 cm x 2.2 cm x 1.2 cm (L x W x H)
Weight:
Approx. 12 grams including battery
Expected service
life:
Three years
Ambient operating
range:
Temperature: 41oF ~ 104oF (5oC ~
40oC)
Relative humidity: 15% ~ 95% RH
Atmospheric Pressure: 700hPa ~
1060hPa
Storage and
transportation
condition:
Temperature: -4oF ~ 131oF (-20oC ~
55oC)
Relative humidity: 15% ~ 95 % RH
Atmospheric Pressure: 700hPa ~
1060hPa
Ingress Protection
Rating:
IP 27
Classification:
Type BF
BATTERY REPLACEMENT
1. Replace battery when appears in the lower
right corner of LCD display.
2. Pull battery cover off as shown in Figure 3.
3. Gently pull out plastic circuit board with
battery chamber approximately 1 cm (slightly
less than ½”.) (See Figure 4)
4. Use pointed object such as a pen to remove
old battery. Discard battery lawfully. Replace
with new 1.5 V DC button type LR41 or SR41,
UCC392, or equivalent. Be sure battery is
installed with
“+” polarity facing up. (See Figure 5)
5. Slide battery chamber back into place and
attach cover.
All rights reserved.