Enviro ETS-TSN-S02N08 User manual

Manual
(V1.2)
(Thermoscan)

TABLE OF CONTENT
1]
History: .............................................................................................................................................
2
2]
Product Selection (Model No.): .......................................................................................................
2
3]
Introduction: ....................................................................................................................................
3
4]
Operation Keys: ...............................................................................................................................
4
5]
Display Message: .............................................................................................................................
4
6]
Product Specifications and Features: .............................................................................................
5
7]
Keypad Functions: ...........................................................................................................................
6
7.1 Admin Login: ...........................................................................................................................................
6
7.2
Display Mode: ..........................................................................................................................................
6
7.3
Alarm Param. Set: ..................................................................................................................................
7
7.4
Calibration offset: ...................................................................................................................................
7
7.5
Contrast Control: ....................................................................................................................................
7
7.6. Set Scan Time: ........................................................................................................................................
8
7.7
System Info: .............................................................................................................................................
8
7.8
Back to main: ...........................................................................................................................................
8
8]
ETS-TSN V1.1Connection Details: ................................................................................................
9
8.1
System Connection Details: ....................................................................................................................
9
8.2
Installation Note: ...................................................................................................................................
10
8.3Troubleshooting Chart: .........................................................................................................................
10
1 of 10

1] History:
Sr.
Manual
Date
Author
Approved By
Remarks
No
Version
1.
1.0
05/01/2013
Pramila
1.
Setting of temperature set point and
Humidity set point has been made common.
2.
Fast key response auto increment and
Decrement.
3.
Faster scanning and averaging of sensor
Readings.
4.
Fast display updating.
5.
3rd wire compensation
2.
1.1
08/07//2013
Pramila
6.
System info menu added.
3.
1.2
22/12/2018
Akshada
7.
Model No. Added
2] Product Selection (Model No.):
Sr. No.
Model No.
Description
1.
ETS-TSN-S02N08
(0202SCAN01)
ETS-TSN-Standard/Stability/Nonlinear/8Temp+8RH
2.
ETS-TSN-S02L08
(0202SCAN02)
ETS-TSN-Standard/Stability/linear/8Temp+8RH

2 of 10

3] Introduction:
Thermoscan (ETS- TSN) is microcontroller based electronic device which interfaced with thermolog with
RS485 port. It is an updated hardware plus firmware version of thermoscan, with fast sensor scanning and
three wire compensation facility. Can be configured for 8 channel stability (8 channel temperature+8 channel
humidity) also it has enhanced with easy to use tactile membrane keypad plus sticker.
3 of 10

4] Operation Keys:
SEL
•
To enter into main Menu.
•
Press this key two times to exit from main menu.
• This key used for exit from parameter setting.
INC
•
To Scroll the menu parameter
• Increments numerical data for parameter setting.
• To fast increment press key for 3 sec it will increment automatically by one, further
pressing of key for 8-10 sec will increment count by 10.
DEC
•
Scroll the menu parameter
• Decrements numerical data for parameter setting.
• To fast decrement press key for 3 sec it will decrement automatically by one, further
pressing of key will decrement count by 10.
ENTER
•
To enter into menu or parameter.
• To set selected values
5] Display Message:
CH
Channel No
PV
Process Value
SV
Set Value
Display Header
At the TOP Display header (factory settable)
°C
Temperature unit.
%
Humidity unit in percentage.
4 of 10

6] Product Specifications and Features:
Sr.
Specifications
Description
1
Type
:
Microcontroller based Electronic device with 16 channel configuration.(no
memory storage)
2
Display
:
128 X 64 Alphanumeric Graphical Display
3
Keypad
:
Tactile membrane sticker with keypad
4
Power supply
:
12VDC @ 1A
12 watts
6
Communication Port
:
RS 485 (PC Communication) and
Thermoscan to thermolog communication
8
Size
:
186*107mm
9
Weight
:
~270gm
10
Keypad functions
Admin Login
:
To Access the menu of unit.
Display Mode
:
To set channel format for either 8 channel temperature or 8 channel
stability.
Alarm Parameter. Set
:
To set alarm set point and hysteresis of temperature and humidity.
(Hysteresis can be set upto 10.0 counts.)
Calibration Offset
:
To set calibration/user offset(can set user offset upto 99.9 )
Contrast Control
:
Set contrast of display.
Scan PV Time
:
To set sensor scanning time.(can set 1 to 4 steps)
System Info
:
To verify system info.
Back To Main
:
This function used to exit the Menu with logoff.
5 of 10

7] Keypad Functions:
7.1 Admin Login:
ENVIRO TECHNOLOGY
>> Admin Login
Admin User: 00010
Password: 00001
Steps:-
To login press SEL (select) key.
Select admin login menu then press enter.
Put admin user ID (10) and press enter key.
Now put password (1) and press enter.
Now it will show message “Admin Login OK”. For correct user ID and password.
For wrong admin login it shows “Admin Login Fail” message.
To log out go to Admin Login then press enter to logout.
If any key is not pressed within 60 sec then you will exit from program mode.
7.2 Display Mode:
ENVIRO TECHNOLOGY
>> Display Mode
Stability Type
Steps:
This parameter is used to set display formats:
Select proper display format using increment –decrement keys.
Press enter key to save it.
To exit from this menu press SEL key.
Following are the display formats:
(a) Temperature Type: to diplay1 to 8 temperature channels only.
(b) Stability Type :to display1 to 8 temperature + humidity channels.
6 of 10

7.3 Alarm Param. Set:
ENVIRO TECHNOLOGY
>>AlarmParParam.Set. Set
Temperature Alarm
Steps:
Can be use to set Temperature alarm set point and hysteresis of channel 1 to channel 8.also can set
humidity alarm set point and hysteresis of channel 9 to channel 16.
Go to this parameter using SEL key and set Temperature set point using increment decrement key.
Press ENTER to save those values.
Then put hysteresis and press enter key.
Do same procedure for humidity settings.
Temperature max range:- -200°C to 850°C.
Humidity range:-1) for EE_1V=0 to 1 V
2) For Humidity= 0 to 3.3V.
7.4 Calibration offset:
ENVIRO TECHNOLOGY
>>Calibration Offset
Channel No: 01
Cal. Offset : 00.0
Steps:
Can set calibration offset channelwise.
Press SEL and go to calibration offset menu.
Set offset and press enter.
Can set calibration offset or user offset upto 99.9
7.5 Contrast Control:
ENVIRO TECHNOLOGY
>>Contrast Control
15
Steps:
To set contrast of display from 0 to 25.
Go to Contrast Control menu set contrast using increment/decrement keys.
Press Enter and save it.
7 of 10

7.6. Set Scan Time:
ENVIRO TECHNOLOGY
>>Set Scan Time
Scan time: 001
Steps:
Can set scanning time from 1 to 4.
If we set scan time to 4 it will scan 1 channel four times and will show average of it on display.
Go to this menu and press enters .set scan time then presses enter.
7.7 System Info:
This menu is used to verify the system’s hardware and firmware information ,
it is read only.
Go to this menu using incr/decre key.
Press enter.
7.8 Back to main:
Use to come back to main screen.
We can also come to main screen by pressing SEL key two times from main menu and by
pressing SEL key three times if you are in sub menu.
8 of 10

8] ETS-TSN V1.1Connection Details:
8.1 System Connection Details:
TEMPERATURE INPUT SIGNAL
THERMOLOG
HUMIDITY INPUT SIGNAL
12VOLT
SUPPLY THERMOSCAN
RS485
Converter
PC
9 of 10

8.2 Installation Note:
While installing this system one must take care of following points.
All cables connecting to “thermoscan “must lay separately, they should not mix up with high
voltage & high current cables (like Motor; compressor; contactor; Heater etc.
There should be a proper insulation for sensor wires to avoid noise interrupt.
Take care while laying cables for Thermoscan interface with thermolog; as these cables are very
sensitive to noise. Do not connect RS-485 cables in Star network; connect in loop network.
o Maximum distance for thermolog to thermoscan communication cable is 600 mtr. This cable
must not be routed along with power cable.
o Cable specification for RS-485 => 14/36; 22AWG; 2 core shielded twisted cable.
RTD-PT100 sensor Cable length: 100metre MAX.
0 to 1V humidity sensor cable length: 3 meter max (keep as short as possible.)
0 to 3.3V humidity sensor cable length: 3 meter max (keep as short as possible.)
4 to 20mA sensor cable length: 3 meter max (keep as short as possible.)
8.3Troubleshooting Chart:
Following are few trouble shooting point which will guide for rectifying problem while dealing with
Scanner Unit.
No.
Problem
Probable solutions
1.
Display Doesn’t show reading
1.
Check CTx and CRx of thermoscan and thermolog.
on thermoscan or on
2.
Check for Broken communication line between thermolog and
thermolog.
Thermoscan.
3.
Set thermoscan and thermolog display mode properly.
4.
Check all sensor pins are connected properly.
2.
Junk data on thermolog
1. Check communication between thermoscan and thermolog.
2.
Check wire.
10 of 10