ERIS PG ERIS-414 User manual

АПНС.421510.414-00 РЭ
1

Content
2
Content
Introduction ........................................................................................................................ 3
1 Safety precautions............................................................................................................ 5
2 application and functions................................................................................................. 7
3 Scope of supply ............................................................................................................... 9
4 Storage........................................................................................................................... 10
5 First operation (switching on) of the gas detector.......................................................... 11
6 Operation and maintenance ........................................................................................... 12
7 Gas detector description................................................................................................. 13
8 Marking ......................................................................................................................... 16
9 Menu.............................................................................................................................. 17
10 Gas detector design...................................................................................................... 21
11 Technical specifications............................................................................................... 22
12 Alarm........................................................................................................................... 25
13 BUMP TEST ............................................................................................................... 31
14 Battery charge.............................................................................................................. 33
15 Calibration................................................................................................................... 36
16 Verification.................................................................................................................. 43
17 Data exchange with PC................................................................................................ 44
18 Repair .......................................................................................................................... 45
19 Warranty statement and claims information ................................................................ 48
20 Disposal....................................................................................................................... 50
Appendix A. Metrological characteristics......................................................................... 51
Appendix B. Failures........................................................................................................ 54

Introduction
3
INTRODUCTION
The present operation manual is the guideline document for handling of the
portable gas detector PG ERIS-414 (hereinafter referred to as the gas detector,
device). The certificate of the approved type is always supplied along with the
device. The device serial number, detected gases, alarm thresholds, operation
conditions, scope of supply, verification mark and other device characteristics are
indicated in the certificate.
The gas detector is admitted for use in the Russian Federation and has the pattern
approval certificate issued by Federal Agency on Technical Regulating and
Metrology RU.C.31.373.A No. 63539 and is listed in State Register of Measuring
Instruments of the Russian Federation under No. 65108-16.
The gas detector meets requirements of Customs Union Technical Regulation
TR TS 012/2011 "The safety of equipment for explosive environments", conformity
certificate No. TS RU S-RU.MY62.V.04130. Validity period: till 07.07.2021
inclusively.
The gas detector meets requirements of Customs Union Technical Regulation
TR TS 020/2011 "Electromagnetic compatibility of technical equipment",
registration number of conformity declaration TS No. RU D-RU.МА10.V.00243.
Validity period: till 07.12.2018 inclusively.

Introduction
4
Explanation of the warning signs of the present manual:
ATTENTION
Reference to the potential hazardous situation that in case of non-
observance of the corresponding preventive measures could lead
to personnel injury, device damage or environment incidents.
Warning on device mishandling.
INFORMATION
Additional information on device handling.
ATTENTION TEMPERATURE CONDITIONS
Special attention shall be paid to the temperature during the work
performance.
DANGER
Reference to the immediately dangerous situation that in case of
non-observance of the corresponding preventive measures could
lead to serious consequences.

1 Safety precautions
5
1SAFETY PRECAUTIONS
The gas detector is personal safety device. You are responsible for
further actions in case of warning signal.
Avoid mechanical shock of the gas detector.
Avoid mechanical influence on sensors. Do not press on sensors or pierce them
by any objects. Avoid influence of pressurized airflow on the gas detector sensor
during cleaning process of the housing.
Do not operate the gas detector with broken components, housing or other
defects.
Do not replace the device components without special authorisation.
Do not charge the gas detector, replace sensors and open the device in explosion
hazard areas.
Violation of operation rules specified by the manufacturer could lead to decrease
of protection level provided by the housing and to degradation of explosion proof
used in this equipment.
Calibration shall be carried out only in safe place in the absence of dangerous
gases.

1 Safety precautions
6
To prevent battery damage due to deep discharge during long
intervals between gas detector operations or in case of long storage
the battery shall be charged at least once per 6 months.
Calibration shall be carried out according to the schedule depending on influence
of toxic and contaminating substances on the sensor. It is recommended to perform
calibration at least once per 6 months.
PG ERIS-414 shall be fully charged before first operation.
For safety reasons the gas detector PG ERIS-414 shall be operated and
maintained only by the qualified personnel. Carefully study the operation manual
before operation and maintenance of the device.
Use BUMP TEST to check capability of sensors to detect gases.

2 Application and functions
7
2APPLICATION AND FUNCTIONS
2.1 PG ERIS-414 application
The portable gas detector PG ERIS-414 is designed for measurement of
concentration of toxic flammable, hydrocarbon gases and oxygen in air of working
area, industrial premises and open spaces of production facilities.
Operating principle of the gas detector is based on physical-chemical methods of
analysis:
- electrochemical - measurement of volume concentration of toxic gases and
oxygen content;
- optical or thermocatalytic - measurement of flammable and hydrocarbon gases
content.
The gas detector simultaneously makes analysis from one to four components of
the gas mixture, perform continuous monitoring and indication of the measured
values and gas detector status on OLED display.
2.2 Gas detector functions:
digital indication of all detected gases content in real time on OLED display of
the gas detector;

2 Application and functions
8
visual, audible and vibration alarm if detected gas content reaches limits of
alarm response ALARM 1 and ALARM 2 of each measured gas;
digital indication of the set alarm limits under the selected measuring channel;
calculation of STEL values (maximum permissible concentration)* and TWA
(average changeable value of maximum permissible concentration)*;
*optionally;
event archive record and storage in non-volatile memory;
PC data exchange via USB interface;
current time indication;
indication of device operation faults / errors;
Energy safety mode - switching over to power standby mode during normal
operation. Functions of the gas detector available during device setting-up by
software installed on PC are specified in clause 17.

3 Scope of supply
9
3SCOPE OF SUPPLY
3.1 Scope of supply of the gas detector shall correspond to information
indicated in the table
Designation
Quantity
Notes
Portable gas detector PG ERIS-414
1
-
Certificate
1
-
Verification procedure MP 83-221-2016
1
For lot
Software
1
Available on the
website
www.eriskip.com
Calibration cap
1
-
Dock station
1
-
USB cable
1
-
Power adapter (Charger)
1
-
Operation manual
1
-
Sensor plug*
0-3
*Upon request
Conformity certificate TR TS
1
For lot
Conformity certificate PB
1
*Upon request
Packing
1

4 Storage
10
4STORAGE
4.1 Original package
The gas detector with scope of supply and operational documentation is
delivered to the customer originally packed in carton box. Packing method,
preparation for packaging, transport package and materials used during packaging,
positioning order correspond to the manufacturer drawing.
4.2 Storage
Storage of the gas detector corresponds to conditions of the second group
according to GOST 15150-69:
- air temperature –from minus 30 up to plus 50 °С;
- relative air humidity (not more than) - 95 % at 25 °С.
The gas detectors shall be placed on racks during storage. Storage environment
of the gas detector shall not contain harmful impurities causing corrosion.
During long intervals between gas detector operations or in case of
long storage the battery shall be charged at least once per 6 months.
Refer to clause 14 of the present manual for rules of battery charging
procedure.

5 First operation (switching on) of the gas detector
11
5FIRST OPERATION (SWITCHING ON)OF THE GAS DETECTOR
Follow the below steps before first operation of the gas detector:
Check scope of supply.
Carefully study the operation manual and certificate of the gas detector.
Perform visual check of the gas detector:
- compare the serial number indicated on the rear part of the device housing and
in accompanying documents;
- check housing integrity;
- check availability of all fasteners and quality of their joints.
Switch on the device and wait till the end of warming up, refer to clause 9.1.
In case the gas detector was stored at negative temperature, it should
be kept switched off at temperature from 0 to plus 30°С within 2-4
hours.
If battery of the device discharged after long storage period it shall be
fully charged, refer to clause 14.
If sensor zero self-calibration function is turned on, perform the
procedure in definitely clean atmosphere.
Check operability. In this regard use BUMP TEST as per clause 13.
Then start operation of the device.

6 Operation and maintenance
12
6OPERATION AND MAINTENANCE
6.1 While operating the gas detector PG ERIS-414 shall be maintained.
6.2 Recommended maintenance types and schedule:
visual inspection - before every use,
surface cleaning - once per month,
operability check (BUMP TEST), refer to clause 13,
gas detector inspection, refer to clause 16.
6.3 During visual inspection check the following:
–integrity of the housing;
- availability of all fasteners and their components, quality of fasteners;
6.4 While operating the gas detector shall be cleaned in case of contamination.
The gas detector shall be cleaned by wet cotton cloth or paper tissue continuously
changing cloth / tissue contacting the surface to avoid scratches. If required water or
compressed air with pressure up to 0.15 MPa could be used with the following
wiping by cloth / tissue.
Do not direct air / water flow to the device sensor.

7 Gas detector description
13
7GAS DETECTOR DESCRIPTION
7.1 The external view of the gas detector and its characteristic features:

7 Gas detector description
14
7.2 Gas detector components and external view
Position
Description
1
Visual alarm indicators (light conductors)
2
Flammable gas sensors * (CT/IR)
3
Oxygen sensor (O2)
4
Toxic gases sensor* (EC1)
5
Toxic gases sensor* (EC2)
6
Audible alarm
7
Control button
8
OLED-display
9
Gas detector charger connector
10
Clip for clamping to clothing
11
Sticker with gas detector marking
12
Housing mounting screws
Notes:
1) *the full list of possible gases is specified in Annex А.
2) Quantity of sensors is specified by the order.
3) CT-thermocatalytic sensor, IR-optical sensor, EC-electrochemical sensor.

7 Gas detector description
15
7.3 Display
Gas detector is equipped with high-contrast OLED-display with wide view
angle. The following information is displayed in real time mode:
1 - exceeding of alarm threshold,
2 - molecular formula of measured gases,
3 - symbol of device operation error,
(refer to DEVICE STATUS),
4 - PC data exchange symbol,
5 - current time,
6 - symbol of external power supply
connection,
7 - battery charge level,
8 - measuring units,
9 - gas concentration.

8 Marking
16
8MARKING
Marking of the gas detector is applied to the rare part of the housing and includes
the following information:
manufacturer name and trade mark;
gas detector model;
gas detector serial number;
gas detector bar code;
explosion proof marking;
IP marking according to GOST14254-96;
operation temperature;
number of conformity certificate TR TS;
production year;
approval mark of measuring instrument type
according to PR 50.2.009-94;
unified mark of product circulation on the market of
the Member States of the Customs Union;
specific explosion safety mark in accordance with
TR TS 012/2011;
manufacturer details.

9 Menu
17
9MENU
9.1 Button control
To switch on the gas detector press and hold the button
within 4 seconds. Testing and warming up is performed
automatically within 60 seconds.
To switch off the gas detector press and hold the button
within 5 seconds till the end of countdown and display
switching off.
Note: The switching off of the device is carried out from the
main screen.
Shortly press the button to enter the menu and move
through menu bars.
Hold the button within 2 seconds to enter the submenu.
To exit the submenu select sign <<< at the end of the
list and press the button.

9 Menu
18
9.2 Main menu
Menu item
Function
BUMP ТЕSТ
Quick testing of the device, refer to clause 13.
- ZERO CAL.
Zero calibration of the sensor, refer to clause 15.
- SPAN CAL.
Span calibration of the sensor, refer to clause 15.
DEVICE
STATUS
Information on device status and errors occurred during
operation. Possible errors are listed in Annex B.
SENSORS
Information on device temperature, atmospheric
pressure and etc.
DEVICE
Information about device:
Serial number, firmware version, manufacturing date,
last calibration date.
Information about sensors:
Sensor designation, serial number, molecular formula
of the measured gas, measurement range, calibration
concentration, value of ALARM1, ALARM 2.
<<<
Return to the measurement mode / previous page.

9 Menu
19
9.3 Submenu SENSORS
The device displays the following information in real
time mode:
Temperature inside the device housing TEMP 1;
Voltage reference "U REF ";
Atmospheric pressure "P ATM." in kPa and mm
Hg;
Battery voltage "BATT." or mode CHARGING.
9.4 Submenu DEVICE
The following information is indicated at first
page of submenu DEVICE:
Device name PG ERIS-414;
Device serial number "S/N";
Firmware version "F/W";
Hardware version "H/W";
Manufacturing date MFG DATE;
Last calibration date LAST CAl DATE
DISABLED;
Amount of days until next calibration

9 Menu
20
days until next cal. DISABLED
The following pages contain information on each
sensor, namely:
Sensor name SENSOR EC1;
Molecular formula of the measured gas
meas.gas;
Sensor serial number "S/N";
Indication range DISP.RANGE TO;
Span calibration value SPAN VALUE.
This value shall be used during the device
calibration;
ALARM 1;
RISE
ALARM 2;
RISE
Table of contents