4SDL710-en-GB_V1.0 12/16
SetupMode
Basicsettingsataglance
Toviewthecurrentconfigurationofthemeterwithregardtotime,date,anddatalogging
samplingratepresstheSETbuttonmomentarily.Themeterwillnowdisplaytheconfigurationin
quicksuccession.Iftheinformationismissedonthefirsttry,simplypresstheSETbuttonagain
untilalloftheinformationisnoted.
AccessingtheSetupmode
1. PressandholdtheSETbuttonforatleast1.5secondstoaccesstheSetupmenu.
2. PresstheSETbuttonmomentarilytostepthroughtheavailableparameters.Theparameter
typeisshownonthebottomoftheLCDandthecurrentselectionforthattypeisshown
aboveit.
3. Whenaparameterisdisplayedthatistobechanged,usethearrowkeystochangethe
setting.PresstheENTERbuttontoconfirmachange.
4. PressandholdtheSETbuttonforatleast1.5secondstoexittheSetupmode.Notethatthe
meterautomaticallyswitchesoutoftheSetupmodeifnokeyispressedwithin7seconds.
5.
TheavailableSetupparametersarelistedbelow.Additionaldetailedinformationis
providedbelowthislist:
dAtE
Settheclock(Year/Month/Date;Hours/Minutes/Seconds)
SP‐t
Setthedataloggersamplingrate(Hours/Minutes/Seconds)
PoFF
Automaticpower‐offmanagement(Enableordisabletheauto‐poweroff
function)
bEEP
SetthebeepersoundON/OFF
dEC
Setthenumericalformat;USA(decimal:20.00)orEuropean(comma:20,00)
SdF
FormattheSDmemorycard
SettingtheClockTime
1. AccessthedAtEparameter.
2. Usethearrowkeystochangeavalue
3. UsetheENTERbuttontostepthroughtheselections
4. PressandholdtheSETbuttonforatleast1.5secondstoexittothenormaloperationmode
(orsimplywait7secondsforthemetertoautomaticallyswitchtothenormaloperating
mode).
5. Theclockwillkeepaccuratetimeevenwhenthemeterisswitchedoff.However,ifthe
batteryexpirestheclockwillhavetoberesetafterfreshbatteriesareinstalled.
SettingtheDataloggerSamplingTime(Rate)
1. AccesstheSP‐tparameter.
2. Thesamplingratecanbesetfrom‘0’seconds(manuallogmode)upto8hours,59minutes,
and59seconds.
(Datalosemayoccurwhenloggingata1secondrate)
3. UsetheENTERbuttontomovethroughtheHours,Minutes,andSecondsdigitgroupsand
usethearrowkeystochangethedigitvalues.
4.
PresstheENTERbuttontoconfirmtheentry.
5. PressandholdtheSETbuttonforatleast1.5secondstoexittothenormaloperationmode
(orsimplywait7secondsforthemetertoautomaticallyswitchtothenormaloperating
mode).