F&P Airvo 3 User manual

www.fphcare.com
USER MANUAL

SECTION
English A
Español B
繁體中文 C
Français Canadien D

Intellectual property information
Fisher & Paykel Healthcare products:
F&P, Airvo, AirSpiral, Optiflow, WigglewiNG and Wigglepads are trademarks of
Fisher & Paykel Healthcare Limited.
For patent information refer to: www.fphcare.com/ip
For more information, please contact your local Fisher & Paykel Healthcare representative.
Compatible third-party products:
Nonin:
Nonin™, Xpod®, PureLight®, PureSAT®, FlexiWraps®, Flexi-Form®are trademarks of Nonin Medical Inc.
For patent information refer to: www.nonin.com
Using any sensors other than Nonin-branded PureLight® sensors with the Nonin Xpod USB connector
may result in inaccurate performance (of the Airvo™ 3 and/or Nonin products) and will void the Nonin
product warranty.
English

Airvo 3 - USER MANUAL | A - 1
English
Before you start
• This user manual is for instructions on using the Airvo 3.
• This user manual is intended for healthcare professionals. While the information provided is believed to be accurate, it is not
asubstitute for exercising professional judgement.
• Read this user manual, including all warnings, before using the Airvo3.
• Before the Airvo 3 is used for the first time, it must be set up according to the instructions in the Airvo3 Technical Manual.
• Some accessories may not be available in certain countries. Please contact your local Fisher & Paykel Healthcare representative for
more information.
• If any device or accessory label is damaged or unreadable, contact your Fisher & Paykel Healthcare representative for a replacement.
Additional resources
• If using the Disinfection Kit to reprocess the Airvo 3, refer to the Disinfection Kit Manual provided with the Disinfection Kit
(900PT600).
• Refer to user instructions supplied with individual accessories for correct use and additional safety information.
• Refer to the Airvo3 Technical Manual for initial setup, maintenance, servicing and additional troubleshooting instructions.
• Visit the Airvo 3 website at: www.fphcare.com/airvo3 to download user instructions including this usermanual.
• For assistance from your Fisher & Paykel Healthcare representative contact us at: www.fphcare.com/contact-us.
Conventions used in this manual
Warning
A warning alerts the user to a potential hazard with use or misuse of the device which, if not avoided, could result in death or
seriousinjury.
Caution
A caution alerts the user to a potential hazard with use or misuse of the device which, if not avoided, could result in minor or
moderateinjury.
Note
A note emphasizes information important for using the Airvo 3 correctly.

A - 2 | Airvo 3 - USER MANUAL
English
Contents
Before you start A - 1
1. Introduction A - 4
1.1 Intended use/indications for use................................................................................A - 4
1.2 Contraindications..............................................................................................A - 4
1.3 Side-eects...................................................................................................A - 4
2. Safety information A - 4
2.1 General.......................................................................................................A - 4
2.2 Supplementary oxygen ........................................................................................A - 6
2.3 Pulse oximetry ................................................................................................A - 6
3. Overview A - 7
3.1 Identifying system components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 7
3.2 Identifying device components .................................................................................A - 8
3.3 Navigating the user interface ...................................................................................A - 9
4. Preparing the Airvo3 A - 11
4.1 Equipment required ...........................................................................................A - 11
4.2 Airvo 3 setup .................................................................................................A - 13
4.3 Supplementary oxygen ........................................................................................A - 15
5. Using the Airvo 3 A - 16
5.1 Getting started................................................................................................A - 16
5.2 Optiflow high flow therapy settings .............................................................................A - 18
5.3 Starting Optiflow high flow therapy .............................................................................A - 19
5.4 During therapy ...............................................................................................A - 22
5.5 Mobility and battery operation.................................................................................A - 23
5.6 Stopping therapy.............................................................................................A - 24
6. Monitoring data A - 25
6.1 Patient data..................................................................................................A - 26
6.2 Live value graphs.............................................................................................A - 26
6.3 Long term graphs ............................................................................................A - 26

Airvo 3 - USER MANUAL | A - 3
English
7. Troubleshooting A - 27
7.1 Alarms ......................................................................................................A - 27
7.2 Alarm priority ................................................................................................A - 27
7.3 Auditory information signals ..................................................................................A - 27
7.4 Viewing alarm details .........................................................................................A - 28
7.5 Checking the alarm system....................................................................................A - 28
7.6 Airvo 3 alarms ...............................................................................................A - 28
8. Reprocessing A - 33
8.1 Airvo 3 device exterior reprocessing............................................................................A - 33
8.2 Outlet elbow reprocessing ....................................................................................A - 34
8.3 Schedule for changing accessories .............................................................................A - 36
8.4 Replacing the air filter ........................................................................................A - 36
8.5 Servicing ....................................................................................................A - 36
9. Pulse oximetry A - 37
9.1 Setup for pulse oximetry ......................................................................................A - 37
9.2 During therapy ..............................................................................................A - 37
9.3 Description of measurements .................................................................................A - 39
9.4 Description of settings and alarms .............................................................................A - 40
9.5 Alarm and measurement settings ..............................................................................A - 42
9.6 Troubleshooting..............................................................................................A - 43
Specifications A - 44
Glossary A - 48
Appendix 1. Patient consumables A - 49
Appendix 2. Parts and accessories A - 50
Appendix 3. Pulse oximetry accessories A - 51
Appendix 4. Humidification behavior during battery operation A - 52

A - 4 | Airvo 3 - USER MANUAL
English
1. Introduction
The Airvo 3 is designed to deliver Optiflow™ high flow therapy to spontaneously breathing patients.
A blower inside the Airvo 3 entrains flows of room air of 2–70 L/min, which may be blended with oxygen from high-pressure sources
(such as wall supplies or bottles) or low-pressure sources (such as flowmeters). The air-oxygen mixture is then warmed and humidified in
the water chamber, before being transported through the heated breathing tube to a nasal, tracheostomy or mask patient interface.
The Airvo 3 is powered by wall power supply, with internal battery backup to provide continuity of therapy during intra-hospital transport.
1.1 Intended use/indications for use
The Airvo 3 is for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified
respiratory gases. This includes patients who have had upper airways bypassed. The flow may be from 2–70 L/min depending on the
patient interface. The Airvo 3 is for patients in hospitals and sub-acute facilities.
The Airvo 3 can deliver these high flow gases through nasal cannula to augment the breathing of spontaneously breathing neonate, infant,
child, adolescent and adult patients suering from respiratory distress and/or hypoxemia in the hospital setting. The Airvo 3 is not
intended to provide total ventilatory requirements of the patient and is not for use during field transport.
1.2 Contraindications
Contraindications are therapy-specific. Refer to instructions of patient interfaces and/or tube and chamber kits for therapy-specific
contraindications.
1.3 Side-eects
Side-eects are therapy-specific. Refer to instructions of patient interfaces and/or tube and chamber kits for therapy-specific side-eects.
2. Safety information
The Airvo 3 and accessories are to be operated by, or under the supervision of, qualified personnel only. Read this manual and the
instructions for use supplied with all accessories (particularly all warnings, cautions and notes) before using the device.
2.1 General
Warnings
• The Airvo 3 is not intended for life support. Do not use Airvo 3 on patients who cannot tolerate a brief interruption of therapy.
• Appropriate patient monitoring is required for all patients using the Airvo 3.
• Delivery of respiratory gases may generate positive airway pressure. This must be considered where positive airway pressure could
have adverse eects on a patient. To avoid serious injury, appropriately monitor the patient for risk factors of airway and lung pressure
injury.
• Anybody connecting patient consumables, accessories or spare parts to the Airvo 3 is accountable for the compatibility of the device
and those patient consumables, accessories and/or spare parts.
• Do not use any patient consumables, accessories or spare parts that are not listed in this user manual, or the Airvo3 Technical Manual.
Incompatible consumables, parts or accessories could aect the quality of therapy, injure the patient, decrease electromagnetic
immunity or increase electromagnetic emissions.
• Use only patient interfaces, heated breathing tubes, water chambers and filters specified in this manual to prevent disconnection during
use, especially when moving the Airvo 3.
• Do not use antistatic or electrically conductive hoses or tubing with the Airvo 3.
• Do not connect the Airvo 3 to the battery of a battery-powered wheelchair, which may compromise device performance and therapy
delivered.
• Carefully route accessories, cords and cables, including the breathing tube, to reduce the possibility of patient entanglement or
strangulation.
• Visually inspect the Airvo 3 and accessories before use and replace if damaged or suspected to be damaged. Using a damaged device
or accessories may impair performance and/or compromise safety.
• Make sure the auditory alarm signal is audible to the operator who will respond to alarms by following the instructions in section 7.5 to
test the alarm before starting therapy.
• Do not use an Airvo 3 on more than one patient at any one time.
• Do not use accessories beyond the maximum period of use specified in this manual. Exceeding the maximum use period can result in
serious injury, including infection.

Airvo 3 - USER MANUAL | A - 5
English
• Do not expose the Airvo 3 battery to water, fire or excessive heat. Do not crush, disassemble orpuncture the battery, or short-circuit the
connector terminals.
• In the event of a battery leaking, do not allow the liquid to come in contact with the skin or eyes. If contact has been made, wash the
aected area with copious amounts of water and seek medical advice.
• Seek medical advice immediately if a cell or a battery has been swallowed.
• Changes or modifications not expressly approved by Fisher & Paykel Healthcare voids the user’s authority to operate the device.
• Do not use any solutions, suspensions, emulsions, anesthetic or respirable gases that are not identified in these user instructions.
Theymay not be compatible with the patient consumables, device or accessories.
• Use only genuine F&P replacement battery modules to prevent damage to the Airvo 3, excessive temperatures, fire or explosion.
Operating environment
• Do not use the Airvo 3 above the altitude range listed in the specifications section of the manual.
• Do not use the Airvo 3 when outside the operating conditions listed in the specifications section. Therapy may be compromised outside
this range.
• Do not use the Airvo 3 in a magnetic resonance imaging (MRI) environment.
• Do not use the Airvo 3 with, or in the presence of, a flammable anesthetic mixture with air or oxygen.
• Do not use the Airvo 3, or accessories, during defibrillation.
• Do not use the Airvo 3, or accessories, near any ignition source, including electrosurgery, electrocautery, or laser surgery instruments.
Exposure to oxygen increases the risk of fire that may result in patient injury.
• Do not use the Airvo 3 in a hyperbaric chamber.
• Avoid using the Airvo 3, or accessories, adjacent to, or stacked with, other equipment, which could result in improper operation. If such
use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
• The Airvo 3 is not designed for use in the home.
Caution
• The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in
a residential environment (for which CISPR 11 class B is normally required) this equipment might not oer adequate protection to
radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the
equipment.
To avoid burns
• Do not touch the hot surface of the heater-plate or chamber base.
• Never operate the Airvo 3 if:
– the heated breathing tube has been damaged in any way including holes, tears or kinks,
– it is not working properly, or
– water has entered the device.
• Do not restrict ventilation around the Airvo 3, which may cause it to overheat.
• Do not block the flow of air through the Airvo 3 or breathing tube.
To avoid electric shock
• Do not store or use the Airvo 3 where it can fall, or be pulled, into water. Disconnect the power cord and stop using the Airvo 3 if water
has entered the case.
• Never operate the Airvo 3 if it has, or is suspected of having:
– been dropped or damaged,
– a damaged power cord or plug, or
– been dropped into water.
• See the Airvo3 Technical Manual for instructions to replace a damaged power cord.
• Do not attempt to adjust, repair, open, disassemble or modify the Airvo 3 except as described in this user manual or the Airvo3
Technical Manual. Return the Airvo 3 to your Fisher & Paykel Healthcare representative for servicing, if necessary.
• Do not touch the patient at the same time as any conductive parts of the device, such as USB ports.
Notes
• If a serious incident has occurred while using this device please inform your local Fisher & Paykel Healthcare representative and
Competent Authority in your country.

A - 6 | Airvo 3 - USER MANUAL
English
2.2 Supplementary oxygen
Warnings
• You must take special care when using supplementary oxygen to reduce the risk of fire. Keep all sources of ignition away from the
Airvo3 and, preferably, not in the same room as the Airvo 3 during use.
• Do not use supplementary oxygen while smoking, near sparks or open flames.
• A spontaneous and violent ignition may occur if oil, grease or greasy substances contact oxygen under pressure. Keep these substances
away from all oxygen equipment.
• The Airvo 3 is a high flow device. Ensure the oxygen supply is designed to provide enough oxygen flow for all connected equipment,
particularly when the supply is shared by multiple devices.
• Only connect pure oxygen to the oxygen inlet ports on the Airvo 3. The oxygen concentration displayed will be wrong if any other gas,
or mixtures of gases, is connected.
• Only use lotions and/or salves that are labeled as oxygen-compatible to avoid the risk of fire and burns.
2.3 Pulse oximetry
Warnings
• Do not adjust, repair, open, disassemble, or modify the pulse oximetry sensor, cable or adapter (pulse oximetry accessories). Injury to
personnel or equipment damage could occur. Return the device for servicing if necessary.
• In line with the indications for use of the Airvo 3, the monitoring functionality of the Airvo 3 is intended for use on spontaneously
breathing patients and not intended for patients requiring life support. It is the responsibility of the clinician to select the appropriate
level of monitoring for their patient and to be prepared to deal with alarms and equipment malfunction. Additional, independent
monitoring equipment may be necessary.
• Explosive hazard: Do not use this device in an explosive atmosphere or in the presence of flammable anesthetics or gases.
Nonin:
–Operation of the Nonin Xpod USB connector below the minimum amplitude of 0.3% modulation may cause inaccurate results.
Cautions
• Before cleaning the pulse oximetry accessories, disconnect the device from the Airvo 3 to avoid electrical shock and flammability
hazards.
• Do not place the pulse oximetry accessories on electrical equipment that may aect the device, preventing it from working properly.
• To minimize radio interference, other electrical equipment that emits radio frequency transmissions should not be used near the pulse
oximetry accessories.
• A functional tester cannot be used to assess the accuracy of a pulse oximeter or sensor.
Nonin:
–The Nonin Xpod USB connector has motion tolerant software that minimizes the likelihood of motion artefact being misinterpreted
as good pulse quality. In some circumstances, however, this device may still interpret motion as good pulse quality. This covers all
available outputs (i.e. SpO₂, PR, PLETH, PPG).
Note
• For more information about required safety and regulatory requirements for pulse oximeters, refer to ISO 80601-2-61, and IEC 60601-1.
Additional safety information can be found in the labelling provided with each Nonin sensor.

Airvo 3 - USER MANUAL | A - 7
English
3. Overview
This section shows the Airvo 3 system and compatible accessories.
3.1 Identifying system components
3.1 Identifying system components
Auto-fill water
chamber
Dual oxygen
input manifold
Water bag
Wall oxygen
supply
Mobile pole
stand handle
Storage basket
Airvo 3
AirSpiral™
breathing tube
E.g Optiflow+ nasal interface
Oxygen bottle
Mobile pole stand
Oxygen-bottle
holder
Pulse oximetry
sensor
Patient interface
The Airvo 3 System

A - 8 | Airvo 3 - USER MANUAL
English
3.2 Identifying device components
Touch screen
Filter cover
Air-filter
release
Finger guard
Integrated mounting
point for mobile pole
stand
High-pressure
oxygen (HPO)
port*
Low-pressure
oxygen (LPO) port Heated
breathing tube
port
Outlet
elbow
Chamber ports
Heater
plate
USB port 1
Power cord
Battery
USB port 2
Cable tidy
*HPO connection may vary depending on regional selection of connector type (DISS, NIST or SIS)

Airvo 3 - USER MANUAL | A - 9
English
Signal light
Touch screen
Data and Graphs
Power on/o
System menu
3.3 Navigating the user interface
The Airvo 3 touch screen provides access to therapy and device status, settings and alarms. You interact with the user interface by:
• touching elements on the screen to open setting screens, make selections and change values, and
• swiping up/down to scroll through menus that are only partly displayed.
3.3.1 Home screen
System
menu
Power
on/o
Settings tiles
Displays therapy settings and
measurements. Tap tiles to
change settings or view more
information
Start therapy
Tap to start
delivering therapy
Messages and alarms
Displays therapy
information and alarms
Status indicators
Data and Graphs
information
Target
value
Therapy on
Stabilizing to target value
Standby

A - 10 | Airvo 3 - USER MANUAL
English
3.3.2 Message bar
The Message bar shows the current state of therapy delivery, confirms settings changes and displays alarms.
Example messages are shown in the table below.
Message bar Description
Breathing gases are not being delivered to the patient.
Tap the Start button to begin therapy.
Breathing gases are being delivered.
Tap the Stop button, then confirm action to return to standby mode.
Active alarms are displayed on top of other messages.
Tap the alarm for details or press to temporarily pause the alarm audio.
See section 7 for troubleshooting alarms.
3.3.3 Status indicators
The following icons may be displayed in the Message bar.
Icon Description
Audio pause
The Airvo 3 is being powered from the wall power supply
Status of the internal battery:
50% of the battery charge is remaining
Battery is charging and 50% of the charge is remaining
Battery is not charging properly*
Battery is missing or faulty*
Battery is due for replacement*
Touch display is locked to prevent accidental changes
An Airvo 3 USB storage device is connected to one of the USB ports
*Check the battery is properly installed. Replace the battery if the problem persists.
3.3.4 Signal light Signal light
The signal light flashes when any alarm is active. Its color indicates the highest priority
alarm that is active. See section 7 for troubleshooting alarms.

Airvo 3 - USER MANUAL | A - 11
English
3.3.5 System menu
The system menu provides access to additional settings and information. Tap to open
the system menu when the Home screen is displayed.
Menu item Description
Pulse Oximeter Alarms
and Settings Configure the pulse oximetry settings including SpO2alarms.
Lock Screen The lock screen can prevent accidental settings changes.
Device Info Displays the version, disinfection, and battery information.
System Settings Change advanced Airvo 3 settings, limits and behaviors. Refer to the Airvo 3 Technical Manual for
more information.
3.3.6 Data and Graphs screen
The Data and Graphs screen displays current and previous measurements and settings for the current patient.
Tap to open the Data and Graphs screen when the Home screen is displayed. The values available depends on the active therapy
mode.
4. Preparing the Airvo3
Review the safety information in section 2 before proceeding. Refer to appendices 1–3 for a list of consumables and accessories that have
been validated for use with the Airvo3.
4.1 Equipment required
You will need:
• Airvo3 attached to a mobile pole stand,
• clean and disinfected outlet elbow,
• bag of USP sterile/distilled water for inhalation (or equivalent).
If supplementary oxygen is prescribed for your patient, you will need either:
• high-pressure oxygen hoses to connect the Airvo3 to the wall oxygen supply or an oxygen-bottle regulator, or
• low-pressure oxygen tubing to connect the Airvo3 to a flowmeter.
Outlet elbows can be processed in two dierent ways:
Disinfection kit (900PT600)
For hospitals using the disinfection kit for reprocessing: A clean and disinfected outlet elbow will already be installed in the Airvo 3.
Remove the clean storage cover and/or the red disinfection tube before use.
Washer-disinfector
For hospitals using a washer-disinfector for reprocessing: obtain a clean and disinfected outlet elbow, e.g. from your Central Sterile
Services Department (CSSD) system.

A - 12 | Airvo 3 - USER MANUAL
English
4.1.1 Optiflow high flow therapy
To provide Optiflow high flow therapy, you will need a:
1. Breathing tube and chamber kit.
2. Optiflow patient interface.
Refer to Appendix 1 for a list of compatible consumables.
Nasal interface
Optiflow+
Optiflow 3S
Optiflow+ Duet
Optiflow Junior 2
Optiflow Junior 2+
Tracheostomy interface Mask interface adapter
Optiflow+ tracheostomy interface Optiflow+ mask interface adapter
Warnings
Only use patient consumables and accessories that are compatible with the Airvo 3 (see Appendix 1-3).
Do not modify patient consumables or accessories in any way.

Airvo 3 - USER MANUAL | A - 13
English
4.2 Airvo 3 setup
Standard aseptic techniques should be followed to minimize contamination when handling the Airvo3 and accessories.
1. Check Airvo 3 height
Check that the Airvo 3 is attached securely to the mobile pole stand and is
below the patient’s head height.
Position the Airvo 3 so that the power cord connection to the wall power supply
is easily accessible and can be disconnected if necessary.
Caution
Do not place the Airvo 3 where controls can be changed by the patient.
2. Connect the outlet elbow (if applicable)
This step applies if your hospital uses a washer-disinfector to clean and disinfect
the outlet elbow. This step does not apply if your hospital uses the disinfection
kit (900PT600).
Insert the clean, disinfected outlet elbow into the slot on the top of the Airvo 3.
3. Assemble the water chamber
Open the tube and chamber kit and remove the MR290
auto-fill water chamber and chamber adapter.
Remove the blue port caps from the chamber by pulling the
tear tab upwards then remove the bracket holding the water
supply tube.
Fit the supplied adapter over the two vertical ports on the
chamber and push on fully then clip the water supply tube into
position.
4. Insert the water chamber
Fit the water chamber to the Airvo 3, sliding the chamber past the finger guard
onto the heater-plate. Take care to align the port adapter with the blue ports on
the Airvo 3.
Ensure the water chamber is fully inserted by pushing firmly on the front of the
chamber until it slides past the finger guard.
To remove the water chamber, grip the port adapter and pull the chamber away
from the Airvo 3.
Warnings
To avoid burns:
Do not start therapy without the water chamber in place.
Do not touch the heater-plate, water chamber or chamber base during
use.
Exercise caution when removing and emptying the chamber.
The water in the chamber is hot during use.
To avoid electrical shock:
When handling the Airvo 3 with the water chamber in place, avoid
tilting the device to prevent any chance of water entering the unit
enclosure
Do not use the MR290 auto-fill water chamber if it has been dropped,
allowed to run dry or damaged in any way. This could lead to the
chamber overfilling.
Warning
Make sure the Airvo 3 is o when connecting the outlet elbow.

A - 14 | Airvo 3 - USER MANUAL
English
H2O
>50 cm
5. Connect the water bag
Attach the sterile water bag to the hanging bracket 50 cm
above the Airvo 3. Remove the spike from the chamber bracket and push the
bag spike into the fitting at the bottom of the bag.
Open the vent cap on the side of the bag spike.
Caution
Only use USP sterile/distilled water, suitable for inhalation, to fill the water
chamber. Adding other substances can adversely aect the humidifier and
therapy delivered.
6. Check the water level
Check that water flows into the chamber and remains below the maximum
water-level line.
The chamber will automatically maintain the correct water level until the
water bag is empty.
Caution
Do not use the MR290 auto-fill chamber if the water level rises above the
maximum water-level line. This may lead to water entering the patient’s
airway.
7. Install the breathing tube
Connect the breathing tube by lining up the pins on top of the Airvo3,
pushing down until you hear a click and the tube locks into place.
To remove the breathing tube, squeeze the sides of the connector and pull
up.
Warnings
To avoid burns:
Do not use an insulating sleeve or any similar accessories which are not
recommended by Fisher & Paykel Healthcare.
Note
Make sure the outlet elbow is installed in the Airvo 3 before attaching the
heated breathing tube.
See step 2 “Connect the outlet elbow (if applicable)” above.

Airvo 3 - USER MANUAL | A - 15
English
4.3 Supplementary oxygen
Low-pressure
oxygen (LPO) port
High-pressure
oxygen (HPO) port
The Airvo 3 provides two options for connecting
supplementary oxygen:
1. A high-pressure oxygen (HPO) inlet port, and
2. A low-pressure oxygen (LPO) inlet port.
The high-pressure oxygen inlet port is connected to the wall oxygen supply or
to the pressure regulator on an oxygen bottle. The ability of the Airvo 3 to
provide the target FiO2is limited by the line pressure of the high-pressure
inlet port (HPO). If the Airvo 3 is unable to maintain the target FiO2, the device
will generate a “FiO2Below Target” alarm.
The low-pressure oxygen inlet port is connected to an external flowmeter,
typically a rotameter.
Warnings
You must take special care when using supplementary oxygen to reduce the risk of fire. Keep all sources of ignition away from the
Airvo 3 and, preferably, not in the same room as the Airvo 3 during use.
Do not use supplementary oxygen while smoking, near sparks or open flames.
When using bottled oxygen, ensure the volume remaining in the bottle is sucient for the therapy planned.
Connect only pure oxygen gas to the oxygen inlet ports on the Airvo 3. The oxygen concentration displayed will be wrong if any
other gas, or mixtures of gases, is connected.
The oxygen concentration delivered to the patient can be aected by changes to the oxygen setting, patient interface or
obstructions in the air path.
Only use lotions and/or salves that are labeled as oxygen-compatible to avoid the risk of fire and burns.
Appropriate patient monitoring must be used at all times.
Make sure that all oxygen connectors are tightened suciently to prevent leaks.
As the low-pressure oxygen (LPO) inlet port uses an alternative small-bore connector design dierent from those specified in the
ISO 80369 series, there is a possibility that a misconnection can occur with a medical device using a dierent alternative
small-bore connector, which can result in a hazardous situation causing harm to the patient. Special measures need be taken by
the user to mitigate these reasonably foreseeable risks.
During Optiflow high flow therapy, the fraction of oxygen inspired by the patient will be lower than the value displayed on the FiO2
tile if the patient’s peak inspiratory demand exceeds the flow delivered.
Caution
Do not connect an oxygen supply to both the high-pressure oxygen inlet port and the low-pressure oxygen inlet port at the same time.
Using the low-pressure inlet at the same time as the high-pressure inlet may cause incorrect oxygen delivery and a FiO2Above Target
alarm.
Note
The built-in oxygen analyzer uses ultrasonic measurement technology. It does not require in-field calibration.
4.3.1 High-pressure oxygen (HPO) source
When oxygen is connected to the HPO port, the Airvo 3 directly controls the oxygen input to meet the target FiO2setting.
Wall oxygen
supply

A - 16 | Airvo 3 - USER MANUAL
English
4.3.2 Low-pressure oxygen (LPO) inlet port
When using the LPO port, the amount of oxygen taken in by the Airvo 3 is controlled by an external
flowmeter. Connect a tube from the external flowmeter to the LPO port. Make sure that the flowmeter
is turned o whenever the Airvo 3 is not delivering therapy.
When using the low-pressure oxygen inlet port, monitor the oxygen concentration displayed on the
Home screen. The oxygen flow-regulator must be adjusted manually to maintain the prescribed
oxygen concentration when changing the respiratory gas flow rate.
Clinicians may configure a High FiO2alarm to discourage use of high FiO2values in particular clinical
environments.
The High FiO2alarm can be disabled or a threshold between 30% and 95% can be selected when the
Airvo 3 is initially set up for your environment (see Oxygen high alarm threshold, Airvo 3 Technical
Manual). The alarm threshold is displayed on the Titrate FiO2screen, if enabled. Tap the FiO2tile to
open the Titrate FiO2screen.
5. Using the Airvo 3
5.1 Getting started
Turn on the Airvo 3
Plug the Airvo 3 power cord into the wall power supply.
Lock the wheels of the mobile pole stand to prevent the Airvo 3 from
moving.
Turn on the Airvo 3 by holding down the Power on/o button for
2seconds.
Note
If the Airvo 3 has been unused and disconnected from the wall power
supply for some time, the device will not power on without being
plugged in.
Wall oxygen
supply
Flowmeter
Warning
Turn o the low-pressure oxygen source whenever the Airvo 3 is not delivering therapy, to ensure that oxygen does not build up
inside the device.
Warning
Make sure the Airvo 3 is dry before plugging the power cord into the
wall power supply to avoid a potential electric shock.
Warnings
The Airvo 3 must be cleaned and disinfected between patients. Refer to section 8 for the steps required to reprocess the Airvo 3
between patients.
Do not exceed the maximum use period for single-patient-use accessories and consumables (see section 8.3 for the schedule for
changing accessories).

Airvo 3 - USER MANUAL | A - 17
English
Review disinfection state
The Airvo 3 will ask you if it will be used on:
the same patient who last used the device (tap Same Patient)
OR
a new patient (tap New Patient).
For a new patient, check that:
1. The outlet elbow has been cleaned and disinfected.
2. A new tube and chamber have been installed.
Review disinfection state
(if the disinfection method is set to Disinfection kit only)
For a new patient, check that:
1. The outlet elbow has been cleaned and disinfected.
The Airvo 3 will indicate the outcome of the last disinfection cycle:
The number of successful disinfection cycles completed by the Airvo 3 is
displayed in the lower left hand corner under 'Disinfection count'.
2. A new tube and chamber have been installed.
Green: The previous disinfection cycle was completed successfully.
Orange: A successful disinfection cycle has not been performed.
Please run a successful disinfection cycle before use on a new
patient.
Red: The previous disinfection cycle failed to complete. Please run
a successful disinfection cycle before use on a patient.
Table of contents
Languages:
Other F&P Medical Equipment manuals