Operation
Installation 2
2-1
Section 2
Operation
2.1 Installation
Manual is written in English. Contact factory for
help with the translation of this manual.
a. Disconnect the converter used with the Primalert or Primalarm; it is no longer needed. However,
keep it in the event that the Primalert or Primalarm are ever used without the PrimaPak.
b. Select the most favorable location for the PrimaPak, bearing in mind that the interconnecting cable
supplied is 36” long. Mount the PrimaPak, using screws.
c. Connect the interconnecting cable to the “Primalert Power Output” port on the PrimaPak and to the
power input of the Primalert or Primalarm.
d. Connect the converter to the “PB Battery Charger Input” port of the PrimaPak.
CAUTION
N
TE
The converter supplied with the PrimaPak should
NEVER be connected directly to a Primalarm or
Primalert.
e. Plug the converter into the power line.
2.2 Operation
The PrimaPak contains a gel battery pack that provides a 12 VDC output for 1200 mAh with an end-of-life
of 8.2 volts. When used in the standby mode, a minimum charging current is provided, ensuring full
capacity upon demand. The system’s design ensures a full charge in approximately 4 hours after a
complete discharge. A small Bicolor LED is located on the Pb charger. The red color indicates fast
charge operation. The Bicolor LED shall change to green to indicate a trickle charge mode.
2.3 Testing
Periodic testing should be performed by disconnecting the plug at the “Pb battery charger input” jack
while placing the Primalert in an alarm condition, using a test source. If the buzzer and the alarm lights
operate, the PrimaPak is capable of providing standby battery power. This test indicates the system is in
working condition but does not test for the life and capacity of the internal battery pack.