FMS EMGZ309.EIP Series User manual

Operating Manual
EMGZ309._.EIP
Tension Measuring Amplifier with EtherNet Interface
Version 2.11 07/2012 ff
Firmware Version: V2.11 (from S/N 1125523)
Diese Bedienungsanleitung ist auch in Deutsch erhältlich.
Bitte kontaktieren Sie Ihre nächstgelegene FMS Vertretung.
This operation manual is also available in German.
Please contact your local FMS representative.
© by FMS Force Measuring Systems AG, CH-8154 Oberglatt –All rights reserved.

Operating Manual EMGZ309._.EIP
2
Table of contents
1Safety Instructions..................................................................................4
1.1 Description conditions 4
1.2 List of safety instructions 4
2Product Description ................................................................................5
2.1 Product Range 5
2.2 Block Diagram 5
2.3 System Description EMGZ309 5
2.4 System Description EMGZ309._.EIP 6
3Quick Installation Guide .........................................................................7
3.1 Preparations for Set-up 7
3.2 Installation Procedure 7
3.3 Installation and wiring 7
3.4 Mounting the Force Sensors 8
3.5 Wiring the Amplifier 8
4Configuring the Amplifier......................................................................10
4.1 Power up the EMGZ 309 10
4.2 Operating Panel 10
4.3 Change the Measuring Units 11
4.4 Change the Device Mode 11
4.5 Offset Compensation 11
4.6 Offset Compensation over the Operating Panel 12
4.7 Calibration 12
4.8 Calibration over the Operating Panel 13
4.9 Analytic Calibration Method 14
5Network Connection over EtherNet/IP ................................................16
5.1 EtherNet/IP Specification: 16
5.2 Amplifier Functionality over EtherNet/IP 16
5.3 Integration of EMGZ309._.EIP in an EtherNet Network 16
5.4 Configuration of the Amplifier over EtherNet/IP 17
5.5 Calibration over EtherNet/IP 17
6Integration in an Allen-Bradley PLC .....................................................18
6.1 Hardware Environment 18
6.2 IP Configuration 18
6.3 Integration in a Project 18
6.4 Access to the Tension Values 21
7Operation...............................................................................................22
7.1 Operating the EMGZ309 over the Operating Panel 22
7.2 Display Value Selection State 23
7.3 Displaying of Overload, Over and Underflow 23

Operating Manual EMGZ309._.EIP
3
8Parameterization over the Operating Panel ........................................24
8.1 Reset to Default Parameter Set 26
8.2 Parameter List 27
8.3 Description of Parameters 28
9Parameter Setting via a PC ..................................................................33
9.1 Parameterization in an Ethernet Network via Web Browser 34
9.2 Offsetkompensation über einen Webbrowser 38
9.3 Calibration via the Web Browser 38
9.4 Parameterization via a PC (peer-to-peer connection) 40
10 Dimensions ...........................................................................................44
11 Technical Specification ........................................................................47

Operating Manual EMGZ309._.EIP
4
1Safety Instructions
1.1 Description conditions
a) High danger of health injury or loss of life
Danger
This symbol refers to high risk for persons to get health injury or loss of life. It has
to be followed strictly.
b) Risk of damage of machines
Caution
This symbol refers to informations, that, if ignored, could cause heavy mecanical
damage. This warning has to be followed absolutely.
c) Note for proper function
Note
This symbol refers to an important information about proper use. If not followed,
malfunction can be the result.
1.2 List of safety instructions
Proper function of the Tension Measuring Amplifier is only guaranteed with the
recommended application of the components. In case of other arrangement,
heavy malfunction can be the result. Therefore, the installation instructions on
the following pages must be followed strictly.
Local installation regulations are to preserve safety of electric equipment. They
are not taken into consideration by this operating manual. However, they have to
be followed strictly.
Bad earth connection may cause electric shock to persons, malfunction of the
total system or damage of the electronic unit! It is vital to ensure that proper
earth connection is done.
Improper handling of the electronic boards may cause damage to the fragile
equipment! Don’t use rough tools such as screwdrivers or pliers! Operators
handling the electronic boards must wear a well earthed bracelet in order to
discharge static electricity.

Operating Manual EMGZ309._.EIP
5
2Product Description
2.1 Product Range
The tension measuring amplifier series EMGZ309 is available in two versions:
EMGZ309._ with standard analogue outputs
EMGZ309._.EIP with analogue outputs and integrated EtherNet/IP interface
Both versions are available in three different housing / mounting options:
EMGZ309.R rail mount housing
EMGZ309.S panel mount housing
EMGZ309.W wall mount housing
For mechanical dimensions refer to charter 10.
2.2 Block Diagram
Fig. 1: Block diagram EMGZ309 Tension Measuring Amplifier E309011e
2.3 System Description EMGZ309
The digital tension measuring amplifiers EMGZ309 are suitable for tension
measurement with all FMS force sensors. The amplifier can handle 1 or 2 force
sensors. Three different analogue output modes can be selected (0…20mA,
4…20mA or 0…10 VDC). As a plug & play unit the setup through the operating panel
is straightforward and quickly accomplished. Alternatively, more advanced users
can access all the internal parameters through a laptop or local PC via a web
browser. That allows a fine tuning of the amplifier even to the most demanding
system requirements.

Operating Manual EMGZ309._.EIP
6
Fig. 2: System Configuration EMGZ309 Amplifier using the analogue output E309019e
2.4 System Description EMGZ309._.EIP
The amplifier EMGZ321._.EIP offers in addition to his standard features the
possibility to be used as an EtherNet/IP device in a corresponding network.
EtherNet/IP bases on the Ethernet TCP/IP and the UDP technology as well as on the
CIP (Common Industrial Protocol). The amplifier can be embedded in a standard
Ethernet infrastructure. Its fast data rate with cycle times down to 2ms makes the
device ideal for applications in automated assembly and production lines. It allows
the measurement of web tension and its transmission over Ethernet to a controller
in real time. Configuration of the amplifier takes place either over the EtherNet/IP, a
web browser or over the operating panel.
Fig. 3: System Configuration EMGZ309.EIP integrated in EtherNet network E309020e

Operating Manual EMGZ309._.EIP
7
3Quick Installation Guide
In a Plug & Play configuration the set-up of the EMGZ309 and force sensor is limited to only
the offset compensation and the calibration.
3.1 Preparations for Set-up
1. Read the Operation Manual of your force measuring sensors
2. Check your requirements such as:
–Display unit of nominal force
–additional output configuration 0…10V or 0/4… 20mA on display
3. Draw the wiring diagram for your configuration (ref. to 3.5 “Wiring the Amplifier“)
4. If required, determine special parameters (ref. to 8.2 “Parameter List”)
3.2 Installation Procedure
1. Mount your force maesuring sensors to the machine frame
2. Wire the Force Measuring Sensor(s) (ref. to Fig. 4)
3. Connect the amplifier to the power supply. The power supply voltage must be in
the range 18 to 36V DC.
4. Perform offset compensation and calibration (ref. to chapter 4.5 to 4.9)
5. If required, do additional settings (ref. to 8“ Parameterization over Operating
Panel”)
6. Integration of amplifier in an Ethernet network and PLC (see chapters 5.3 and 6)
3.3 Installation and Wiring
Caution
In order to improve the natural convection and keep the temperature of the
amplifiers that are mounted in a cabinet, to a minimum, a distance of at least
15mm (0.6”) between the devices must be kept.
Caution
Proper function of the Tension Measuring Amplifier is only guaranteed with the
recommended application of the components. In case of other arrangement,
heavy malfunction can be the result. Therefore, the installation instructions on
the following pages must be followed strictly.
Caution
Local installation regulations are to preserve safety of electric equipment. They
are not taken into consideration by this operating manual. However, they have to
be followed strictly.

Operating Manual EMGZ309._.EIP
8
3.4 Mounting the Force Sensors
Mounting of the force sensors is done according to the FMS Installation manual
which is delivered together with the force sensors.
3.5 Wiring the Amplifier
One or two force sensors can be connected to the measuring amplifier. When using
two force sensors, the sensors are internally connected in parallel. The output
signal of the measuring amplifier will then correspond to the average value of the
two sensors. The connection between force sensors and measuring amplifier has to
be done using a 2x2x0.25mm2[AWG 23] shielded twisted-pair cable
Fig. 4: Wiring Diagram EMGZ309 E309010e

Operating Manual EMGZ309._.EIP
9
Caution
Bad earth ground connection may cause electric shock to persons, malfunction of
the total system or damage of the electronic unit! It is vital to ensure that proper
earth connection is done. The cable must be installed separate from power lines
Note
The shield should be connected only to the electronic unit. On the force sensor
side the shield should stay open.

Operating Manual EMGZ309._.EIP
10
4Configuring the Amplifier
The amplifier EMGZ309._.EIP can be configured over the operation panel (chapter
8), over a PC (web interface, chapter 9) or over EtherNet/IP. The basic configuration
is performed over the operation panel.
4.1 Power up the EMGZ 309._.EIP
1. Connect the first force sensor (see Fig. 4 Wiring Diagram EMGZ309)
2. Check whether applying a force in measuring direction (in the direction of the red
point) on the first sensor results in a positive output signal. If not, exchange the
two signal wires of the corresponding force sensor (terminals 12/13).
3. If applicable, connect the second force sensor.
4. Check whether applying a force in measuring direction on the second sensor
results in a positive output signal. If not, exchange the two signal wires of the
corresponding force sensor in the terminal block (terminals 22/23)
4.2 Operating Panel
Fig. 5: Operating Panel EMGZ309.W.EIP E309007e
The keys on the operating panel have the following functions:
Key
Key Function
Change the selections or increase / decrease the values
Enter
>0< + PARA
Enter Offset Compensation Procedure. Press the two keys at
the same time for longer then 3 sec.
Cal + PARA
Enter Calibration Procedure. Press the two keys at the same
time for longer then 3 sec.
PARA
Enter the Parameter Setting Mode. Press the key at the same
time for longer then 3 sec.

Operating Manual EMGZ309._.EIP
11
4.3 Change the Measuring Units
The default tension measuring units is N (Newton). The user can, however, select
another unit before the calibration process. To select the units of measure, follow
the steps below:
1. Hold the PARA key until the display changes to [Dis. Filt]
2. Scroll the menu by pressing the key until the display shows [Unit]
3. Press the and the word [Unit] will start flashing indicating that the unit of
measure can now be changed.
4. With every press of the keys the display will change from N, kN, then
to lb, g, kg.
5. Stop at the desired unit and press the again to confirm the selection.
6. Press the PARA key for longer then 3 sec. to save the setting and return to the
Operating State. Or if no action is taken after 30 seconds, the value will be
stored in the system automatically.
4.4 Change the Device Mode
The EMGZ309 amplifiers have two Device Modes:
●Display of Force Gauge: Tension data are displayed in a force unit
●Display of Voltage Gauge: Tension data are displayed in Volt (V)
1. Go to Parameter Selection Mode (press PARA key longer than 3 sec).
2. Go to Parameter [DevicMod]
3. With the key select desired Device Mode [ForcGaug] or [VoltGaug] and
confirm selection with the key .
Note
Each change of the Device Mode requires a re-calibration of the system (see
chapter 4.8).
4.5 Offset Compensation
With the Offset Compensation you can compensate the roller weight. It is always
performed before the calibration. The Force Measuring Sensor should not be loaded
while the Offset Compensation is being done. There are two possible ways
performing the offset compensation.
A) Offset compensation over the operating panel (see chapter 4.5)
B) Offset compensation via EtherNet or a Web Browser.
For this method proceed to chapter 9.1 (Fig. 23).

Operating Manual EMGZ309._.EIP
12
4.6 Offset Compensation over the Operating Panel
1. Make sure the measuring roller is unloaded.
2. Press the two keys >0< and PARA at the same time for longer then 3 seconds.
3. The display will show [Offset Complete] when the process is finished (see Fig. 6
State diagram offset compensation).
Fig. 6: State diagram offset compensation EMGZ309014e
4.7 Calibration
By calibrating (setting the Gain), amplifier and
load cells are adjust so that the display gives
the actual tension value. There are two
methods of calibrating the system. The first
method uses a defined weight (see Fig. 7). This
method is the preferred one since it delivers
the most accurate results. The second method
is based on a calculation method in
conjunction with the FMS Calculator. This
Calculator can be down loaded from the FMS
web page. Please consult the factory for this
option.
Note
If the Device Mode is set to [VoltGaug], the force corresponding to the calibration
weight must be input in Volt (V) in the following calibration procedure. The
parameters System Force [SysForce] and Force at maximum output [F@OutMax]
won’t be requested (see also chapter 9.3, section B).
Fig. 7: Calibrating the
measuring amplifierC431011e

Operating Manual EMGZ309._.EIP
13
4.8 Calibration over the Operating Panel
1. Load a rope with a defined weight corresponding to your calibration force on the
roller. The roller configuration must correspond to the real configuration in the
machine (wrap angle, distance of the rollers etc.).
2. The calibration mode can be entered by pressing the two keys Cal and PARA at
the same time for longer than 3 seconds. Display changes to
The System Force [SysForce] determines the measuring
capability of your measuring roller. E.g. if two 500 N sensors are
installed in the measuring roller, enter 1000N. If only one 500N sensor is used,
enter 500N (see 8.3 “Descriptions of Parameter”, [SysForce]).
3. The parameter [SysForce] flashes indicating that the force can now be changed.
Use keys to enter the system force.
4. Press the key to confirm your input. Display will change to
5. Enter the force corresponding to your calibration weight with the
keys.
6. Press the key to confirm your input. Display will change to
7. With the parameter [F@OutMax] the force corresponding to the maximum
amplifier output (10V or 20mA) can be assigned (see 6.3 “Descriptions of
Parameter”, [F@MaxOut]). Enter the value with the keys.
8. Press the key to confirm your input. Display will change to
The calibration procedure has been completed successfully.
Fig. 8: State diagram calibration EMGZ309015e
1) The display shows the unit of measure that was previously selected.
3) Permanent key pressing expedites the changing speed.
4) If the Device Mode [VoltGauge] is set, these parameters are deactivated
SysForce
1000.0 N
Calibra.
1000.0 N
F@OutMax
1000.0 N
Calibra.
Complete

Operating Manual EMGZ309._.EIP
14
4.9 Analytic Calibration Method
Determining the gain by means of the FMS-calculator:
This method is used when the traditional calibration method with a weight is
difficult or impossible at all due to machine accessibility, weight adequateness etc.
This calculation method requires the FMS-calculator V2.1 or higher.
If necessary download and install FMS-calculator from the FMS homepage at:
http://www.fms-technology.com/en/technical-information.php
Gain-calculation-tool
The gain calculator is an optional feature of the FMS-calculator. It delivers the gain
factor for easy weightless calibration of FMS amplifiers.
1. First, the application related parameters must be entered in the FMS-calculator
e.g. min./max. force values, wrap angle and other sensor specifications. Follow
the instructions of the FMS-calculator manual.
2. Choose the appropriate sensor type.
3. If the sensor evaluation is correct, the gain calculation tool can be started with a
click on the corresponding symbol. The date will be taken over from the previous
input (field (1) Fig. 9). The gain calculation tool will only start-up, if all data have
been entered and the appropriate sensor defined.
4. Choose the amplifier type (field (2) Fig. 9).
5. The calculated gain factor is available in field (3) (see Fig. 9)

Operating Manual EMGZ309._.EIP
15
Fig. 9: Gain calculation tool E309_gain_e.jpg
Calibration Procedure:
1. Calculate the gain factor with the FMS-calculator as described.
2. Press the PARA key for longer then 3 seconds to entered the parametrisation
mode.
3. Scroll up or down the parameter menu with the keys until the display
show [Gain].
4. Confirm the parameter with the key
5. The display will start flashing indicating that the gain factor can now be entered.
6. Use the keys to enter the calculated value.
7. Press the key to confirm the input. Or if no action is taken after 30 seconds,
the value will be stored in the system automatically.
1
2
3

Operating Manual EMGZ309._.EIP
16
5Network Connection over EtherNet/IP
The amplifier EMGZ309._.EIP is able to communicate with other devices over an
Ethernet network. CIP (Common Industrial Protocol) is used as the protocol for the
data link. The amplifier acts as a so called “Producer”and the machine control (or
PLC) as the “Consumer”.
5.1 EtherNet/IP Specification:
Physical layer: Ethernet IEEE 802.3.
Transport protocol: TCP/IP und UDP
Standard: CIP IEC 61158
Conformity: CPF 2 (Communication Profile Family 2) IEC 61784-1
5.2 Amplifier Functionality over EtherNet/IP
Basically all functions for configuration and calibration can be performed over
EtherNet/IP. Furthermore, the measured tension values are cyclically propagated
over the network and made available to the Consumer (PLC).
5.3 Integration of EMGZ309._.EIP in an Ethernet Network
In order to integrate the amplifier in an Ethernet network and enable data transfer
over it following conditions must be fulfilled:
✔An EMGZ309 amplifier with the option EtherNet/IP (EIP) must be used.
✔The IP-Address must be assigned in the fix address block and must be
unambiguously. It may only exist once in the network.
✔The network mask must correspond with the used device configuration.
If these prerequisites are given, the controller will recognise the amplifier as an
EtherNet/IP device in the network and be able to communicate with it. The interface
information for EMGZ309._.EIP is defined in an EDS-file.
EDS-files for EMGZ309._.EIP can be downloaded from the FMS webpage.
Link: http://www.fms-technology.com/en/eds-files.php

Operating Manual EMGZ309._.EIP
17
5.4 Configuration of the Amplifier over EtherNet/IP
EMGZ309._.EIP may, but need not be configured via EtherNet/IP. The Amplifier can
be configured by means of a machine control unit, provided this unit is integrated in
the same network. The communication between machine control and amplifier is
carried out with dedicated software (e.g. RSLogix5000 from Rockwell). How exactly
this is done, it can be found in the operation manual of the respective controller.
As mentioned earlier there are three possibilities to configure the EMGZ309._.EIP. If
the method over EtherNet/IP is used, the other two options (operation panel or web
interface) are still possible but not meaningful. The EtherNet/IP method has then
always priority and overwrites any other configurations set by the other two
methods. If the configuration over EtherNet/IP is preferred, there are two different
ways of doing it:
1. Initialisation of all parameters at connection establishment with controller.
2. Configuration of individual parameters as needed
The Bit cfgApplyConfAsmData determines which of these two methods will be used
(see EDS-file). If the Bit is set, the first method will be applied and the controller will
initialise the amplifier at each connection. If the Bit is not set, all parameters remain
unchanged at a connection. They can then be individually programmed via
EtherNet/IP, the operation panel or the web interface.
5.5 Calibration over EtherNet/IP
Usually the calibration procedure for amplifier and load cells is carried out at the
commissioning of the machine. The two commands Offset and Calibration are
available for the calibration of the amplifier over EtherNet/IP. From case to case it is
to consider whether a calibration via web interface or operation panel is more
favourable. With difficult accessibility of the amplifier it might make sense to do the
calibration over EtherNet/IP.

Operating Manual EMGZ309._.EIP
18
6Integration in an Allen-Bradley PLC
Following chapters deal with the integration and operation of the EMGZ309._.EIP
amplifier with an Allen-Bradley PLC. The project software RSLogix5000 from Allen-
Bradley is required to program the PLC. Furthermore, it is assumed that a
development system with PC and corresponding development software RSLogix
from Rockwell Automation is available and the programmer is familiar with these
tools..
6.1 Hardware Environment
The EMGZ309._.EIP must be connected to an Ethernet switch which in turn is in the
same network as the PLC. Only in this way a communication between amplifier and
PLC can take place. Several EMGZ309._.EIP amplifiers or other ErtherNet/IP
devices can be connected to the same network.
.
6.2 IP Configuration
In an industrial environment the IP addresses are normally assigned fix to the EIP
devices. This holds also for the EMGZ309._.EIP amplifier that needs a fix IP address
in the network. The IP address configuration must take place before the amplifier is
integrated in a network. This is done over the operation panel or over the web
interface.
The procedures are described in the state diagrams chapter 8 (Fig. 18; 19) and
chapter 9 (Fig.31)
6.3 Integration in a Project
Before starting the development system RSLogix5000 the PLC and amplifier need
to be connected to the network and powered-up. Now a new project or an existing
project can be opened in RSLogix5000. Search in the tree structure Controller
Organizer for the listing Ethernet and make a right mouse click (see Fig. 10).

Operating Manual EMGZ309._.EIP
19
In this example CompactLogixL23E-QB1B is used. If a different hardware is used,
one may find the listing Ethernet in another place.However, the principal approach
is always the same. After a right mouse click on Ethernet a dialog window will open
up. Left mouse click on New Module...and a new window will open up (see Fig.11).
Fig. 10: Controller Organisation E309021

Operating Manual EMGZ309._.EIP
20
Open plus-sign [+] in
front of Communications
to see all the other
listings.
Select ETHERNET-
MODULE Generic
Ethernet Module and
confirm with OK.
Following window opens
up in which the Name
EMGZ309_91, the
Comm Format DINT and
the IP-Address (eg.
192.168.0.91) need to
be entered. On the
Connection Parameters
side the values
Input=100 with Size 5,
Output= 101 and
Configuration= 102 with
Size 48 must be entered.
Confirm with OK
Fig. 11: Module Selection E309022
Fig. 12: : Module Selection open E309023
Fig. 13: New Module E309024
This manual suits for next models
4
Table of contents
Other FMS Measuring Instrument manuals