Operating instructions
Pressure transducer | DS01+

FSM AG | Erich-Rieder-Straße 2 | D-79199 Kirchzarten | messt[email protected] | www.fsm.ag 27.04.2021
2 | 6
Table of Content
1 Introduction ................................................................................................................................................................................... 3
1.1 Notes on operating instructions................................................................................................................................................3
1.2 Explanation of icons and symbols.............................................................................................................................................3
1.3 Intended use and foreseeable misuse......................................................................................................................................3
1.4 Functions.........................................................................................................................................................................................3
2 Notes for your safety.................................................................................................................................................................... 3
3 Installation and start-up .............................................................................................................................................................. 4
3.1 Scope of delivery...........................................................................................................................................................................4
3.2 Assembly.........................................................................................................................................................................................4
3.3 Start-up ...........................................................................................................................................................................................4
3.3.1 Connection ........................................................................................................................................................................4
3.3.2 Setting the measuring range ........................................................................................................................................4
3.3.3 Setting the output signal ...............................................................................................................................................5
3.3.4 Setting the time constant ..............................................................................................................................................5
4 Maintenance and service............................................................................................................................................................. 6
4.1 Zero point correction ...................................................................................................................................................................6
4.2 Cleaning...........................................................................................................................................................................................6
4.3 Repair ..............................................................................................................................................................................................6
5 Accessories and spare parts....................................................................................................................................................... 6
6 Technical data ................................................................................................................................................................................ 6
7 Disposal ................................................................................................................................................................................... 6

FSM AG | Erich-Rieder-Straße 2 | D-79199 Kirchzarten | messt[email protected] | www.fsm.ag 27.04.2021
3 | 6
1 Introduction
1.1 Notes on operating instructions
These operating instructions are intended to ensure optimum installation, commissioning, operation and maintenance of the
device and must be read prior to performing any of these actions. Keep this documentation handy and accessible to all users
so that it can be refered to whenever necessary. Pass this documentation on to subsequent users of the product. All necessary
settings are described in this manual.
If any difculties should occur during commissioning or operation, please do not carry out any unauthorized manipulations.
Doing so could void your warranty.
Instead, please contact us immediately:
FSM AG
Erich-Rieder-Strasse 2
D-79199 Kirchzarten
+49 7661 9855 0
1.2 Explanation of icons and symbols
Danger
Indicates a hazard that could lead to personal injury.
Notice
Indicates important information which, if not observed, may result in damage to property, for example.
1.3 Intended use and foreseeable misuse
The DS01+ differential pressure transmitter is used to detect low pressures and convert them into a pressure-proportional
measuring signal. The device may only be used in the specied measuring range. Only differential pressures of non-aggressive
gases may be measured with the DS01+.
Notice
Do not use the device in explosive atmospheres or for measurement of aggressive gases.
No liability is accepted for damages resulting from non-intended use. If so, any warranty claims shall be rendered
void. Unauthorized modications, additions or alterations to the instrument are prohibited. Modications to the
device are prohibited.
1.4 Functions
Depending on the selected measuring range, the DS01+ pressure transmitter can be used to record and display positive,
negative and differential pressures.
2 Notes for your safety
Danger
In order not to endanger any person involved in the installation and commissioning of the device, such work
should only be carried out by qualied and trained personnel.
Persons using the device must also:
>be conscious of the dangers present when working next to live parts.
>take measures to protect themselves against direct contact with live parts.
>have read and understood the operating instructions.

FSM AG | Erich-Rieder-Straße 2 | D-79199 Kirchzarten | messt[email protected] | www.fsm.ag 27.04.2021
4 | 6
3 Installation and start-up
3.1 Scope of delivery
The scope of delivery includes:
>1 x DS01+ pressure transducer
>1 x quick reference guide
>optional: 1 x calibration certicate
3.2 Assembly
The DS01+ pressure transmitter can be attached to the wall via the side lugs using dowels and screws.
Notice
>Mount the device on a vertical, smooth surface
>Do not mount the device in the immediate vicinity of sources of interference or heat
3.3 Start-up
3.3.1 Connection
Connect the supply voltage as well as the output wiring according to the connection diagram. For this purpose run the cables
through the cable glands attached to the housing and connect them to the screw terminals provided.
In order to guarantee the specied accuracy, the sensor should only be fully put into operation after a run-in time of one
hour. After the run-in time has elapsed, a zero point correction should be carried out to adjust the pressure transmitter to the
ambient pressure (see 4.1 Zero point correction).
3.3.2 Setting the measuring range
+ -
Clamps
0V
V+
I
U
0V
DIP-Switches
S1
S2
S3
S4
S5
S6
Button S
On
Off

FSM AG | Erich-Rieder-Straße 2 | D-79199 Kirchzarten | messt[email protected] | www.fsm.ag 27.04.2021
5 | 6
You can adapt the respective desired measuring range to your application via the DIP switches S1, S2 and S3. The respective
setting options can be found in the following table:
Basic measuring range Measuring range Position S1 Position S2 Position S3
10 hPa
0… 10 hPa Off Off Off
0… 7,5 hPa On Off Off
0… 5,0 hPa Off On Off
0… 2,5 hPa On On Off
-10… 10 hPa Off Off On
-7,5… 7,5 hPa On Off On
-5,0… 5,0 hPa Off On On
-2,5… 2,5 hPa On On On
100 hPa
0… 100 hPa Off Off Off
0… 75 hPa On Off Off
0… 50 hPa Off On Off
0… 25 hPa On On Off
-100… 100 hPa Off Off On
-75… 75 hPa On Off On
-50… 50 hPa Off On On
-25… 25 hPa On On On
3.3.3 Setting the output signal
You can adapt the respective desired analog output to your application via the DIP switch S4. The respective setting options
can be found in the following table:
Output signal Position S4
0… 10 V / 4… 20 mA Off
2… 10 V / 0… 20 mA On
3.3.4 Setting the time constant
You can adapt the respective desired time constant to your application via the DIP switches S5 and S6. The respective setting
options can be found in the following table:
Time constant Position S5 Position S6
50 ms Off Off
500 ms On Off
2000 ms / 2 s Off On
4000 ms / 4 s On On

FSM AG | Erich-Rieder-Straße 2 | D-79199 Kirchzarten | messt[email protected] | www.fsm.ag 27.04.2021
6 | 6
4 Maintenance and service
4.1 Zero point correction
The zero position of a pressure sensor shifts over time due to external inuences and the condition of the piezoresistive
sensor element. It is therefore recommended to carry out a zero point correction before commissioning and at annual inter-
vals thereafter. To do this, connect the supply voltage to terminals 1 and 2 and then press the „S“ key. The current measured
value is now zeroed.
Danger
When working on the open device, always ensure that either the supply voltage is interrupted or suitable
protective measures have been taken against contact with live parts.
4.2 Cleaning
Cleaning of the equipment can be done with a damp cloth. Remove external soiling from the housing, the connections and
the display.
Notice
Do not use sprays, solvents, alcohol-based cleaners or abrasive cleaners to dampen the cloth.
Remove soiling of the housing, the connectors and the display.
4.3 Repair
Notice
Repairs may only be carried out by the manufacturer or appropriately authorized and qualied personnel.
If changes are made to the device independently, the warranty will expire.
5 Accessories and spare parts
The scope of accessories may vary depending on the order variation. All possible accessories are listed in Chapter 3.1 Scope
of delivery. Spare parts are to be obtained exclusively from the manufacturer. The replacement of the spare parts may only be
done by the manufacturer or an authorized technician.
6 Technical data
Technical data can be found in the corresponding data sheet, which is available at www.fsm.ag, or can be downloaded
using the QR code below.
7 Disposal
Danger
Incorrect disposal can result in hazards for the environment. Device components and packaging materials must
be disposed of in an environmentally friendly manner in accordance with the country-specic waste treatment and
disposal regulations.
Table of contents
Popular Transducer manuals by other brands

Siedle
Siedle novotechnik T Series manual

Measurement Computing
Measurement Computing KPSI 351 Series user manual

Electro-Voice
Electro-Voice DH7-16 Technical specifications

Balluff
Balluff BTL7-S5 B-M Series user guide

Fairchild
Fairchild T6000 instructions

SIAP+MICROS
SIAP+MICROS t001 TTEP User manual and maintenance