Fungilab VISCOLED ADVANCE Series User manual

VISCOLEAD SERIES
ADVANCE

ADVANCE Manual
2/69
VISCOLEAD SERIES
ADVANCE
Rotational Viscometer
Software Version: 5.2
Manual Version:1.3
Instruction Manual
FUNGILAB S.A.
C/ Constitució, 64
Pol. Ind. Les Grases
08980 Sant Feliu de Llobregat
Barcelona, SPAIN
Phone: +34 93 685 35 00
Fax: + 34 93 685 37 50
Email: sales@fungilab.com
www.fungilab.com
FUNGILAB INC
89 K Cabot Ct. Hauppauge
11788 NY (US)
Phone: .+1 631-750-6361
Fax: +1 631-750-6362
Email: sales@fungilab.us
www.fungilab.com

ADVANCE Manual
3/69
INDEX
INDEX.................................................................................................................................................................. 3
1.
Introduction...................................................................................................................................................5
2.
Safety Instructions....................................................................................................................................... 5
3.
Safety Symbols and Precautions..............................................................................................................6
4.
Utilities........................................................................................................................................................... 6
5.
Specifications............................................................................................................................................... 6
6.
Conditions for use........................................................................................................................................ 7
7.
Maintenance................................................................................................................................................. 7
8.
Equipment presentation.............................................................................................................................. 8
9.
Equipment Description.............................................................................................................................. 10
9.1
Equipment set-up.................................................................................................................................... 12
9.2
The keyboard and screen ...................................................................................................................... 13
10.
Menu system............................................................................................................................................ 14
10.1
Start-up................................................................................................................................................... 14
10.2
Autotest .................................................................................................................................................. 14
10.3
Main Menu ............................................................................................................................................. 16
10.4
Instrument Setup menu........................................................................................................................ 16
10.4.1
Language............................................................................................................................................ 17
10.4.2
Units..................................................................................................................................................... 18
10.4.3
Calibration........................................................................................................................................... 18
10.4.3.1
Reset................................................................................................................................................ 19
10.4.3.2
Viscosity calibration........................................................................................................................ 20
10.4.3.3
Temperature calibration................................................................................................................. 22
10.4.4
Time Settings...................................................................................................................................... 23
10.5
Measurement Configuration................................................................................................................ 24
10.5.1
Measurement Screen........................................................................................................................ 25
10.6
Test Profiles........................................................................................................................................... 27
10.6.1
Edit Test Profile.................................................................................................................................. 27
10.6.1.1
Viscometer programming.............................................................................................................. 28
10.6.1.2
Output options................................................................................................................................. 29
10.6.1.3
Measuring configuration options.................................................................................................. 29
10.6.2 Select Profile ...................................................................................................................................... 29
10.8
Options ................................................................................................................................................... 32
10.8.1
Output.................................................................................................................................................. 32
10.8.2
Information.......................................................................................................................................... 33
10.8.3
Comunication ..................................................................................................................................... 33
11.
Important Rheological Information........................................................................................................ 35
12.
Accessories.............................................................................................................................................. 40
12.1.
Low viscosity adapters (LCP and LCP/B) ........................................................................................ 40
12.1.1
Mounting ............................................................................................................................................. 41
12.1.2
Dismounting and cleaning................................................................................................................ 42
12.1.3
Technical specification for LCP accessories ................................................................................. 42
12. 2. Small sample adapters APM and APM/B........................................................................................ 43
12. 2. 1 Assembly .......................................................................................................................................... 44
12. 2. 2 Dismounting and cleaning.............................................................................................................. 45
12.
2. 3 Technical specifications of APM and APM/B .............................................................................. 45
12.3 HELDAL UNIT –Helicoidal Movement Unit...................................................................................... 47
12.3.1 Heldal unit mounting.......................................................................................................................... 48
12.4. Thermosphere...................................................................................................................................... 50
12.4.1 Connecting Thermosphere to viscometer...................................................................................... 50
13.
Model/Spindle correspondence tables................................................................................................. 51
14.
Model/spindle/oil calibration tables....................................................................................................... 53
15.
ADVANCE L standard spindle selection table.................................................................................... 54
16.
ADVANCE L Special aerial spindle selection table............................................................................ 55
17.
ADVANCE L Special spindle selection table....................................................................................... 56

ADVANCE Manual
4/69
18.
ADVANCE L LCP Adaptor table ........................................................................................................... 57
19.
ADVANCE L special Vane spindle selection table............................................................................. 58
20.
ADVANCE L special Heldal spindle selection table........................................................................... 59
21.
ADVANCE R standard spindle selection table ................................................................................... 60
22.
ADVANCE R Special spindle selection table...................................................................................... 61
23.
ADVANCE R LCP Adaptor table........................................................................................................... 62
24.
ADVANCE R special Vane spindle selection table............................................................................ 63
25.
ADVANCE R special Heldal spindle selection table .......................................................................... 64
26.
ADVANCE H standard spindle selection table ................................................................................... 65
27.
ADVANCE H Special spindle selection table...................................................................................... 66
28.
ADVANCE H special Vane spindle selection table............................................................................ 67
29.
ADVANCE H special Heldal spindle selection table .......................................................................... 68
WARRANTY CERTIFICATE ....................................................................................................................... 69

ADVANCE Manual
5/69
1. Introduction
Thank you for acquiring the ADVANCE rotational viscometer model from Fungilab.
The ADVANCE is a rotational viscometer, based on the torque measurement of a rotating spindle in the
sample at a specified velocity. Three different models (type L, R and H), as well as various accessories, allow
it to cover a wide range of viscosity measurement.
2. Safety Instructions
It is not the purpose of this manual to outline all of the safety instructions recommended for the
use of the rotational viscometer, its accessories and samples. It is the user’s responsibility to
establish health and safety practices and to determine the application’s limits before use.
Fungilab guarantees the satisfactory operation of the viscometers and its accessories if there has
not been any unauthorized adjustment to the mechanical pieces, the electronic components and
the software.
The operator should follow all the instructions and warnings of this manual to ensure the safe and
proper operation of the equipment.
Do not use the equipment for any other purpose than those described in this manual.
Do not use any accessory that is not supplied or approved by Fungilab.
Do not use the viscometer or its accessories if there is any suspicion of malfunction. Do not use the
equipment in situations or conditions that can cause personal injuries or material damage.
The rotational viscometer is not a flameproof or an intrinsically safe (ATEX) instrument. Therefore, it should
not be used in areas where there is an explosion risk.
Before using the viscometer, read and observe carefully the following precautions:
Not following these instructions may cause serious harm or personal injuries.
To avoid an electric shock:
The socket by which the viscometer will be connected should have a ground. Verify that the voltage
and the frequency match with the specifications of the power supply. Before switching on the
equipment, let it sit for some time so that it acclimates to the surrounding temperature in order to
avoid a short-circuit caused by condensation. The fluctuations of the power source should not
surpass ±10 % of the nominal voltage.

ADVANCE Manual
6/69
3. Safety Symbols and Precautions
Safety Symbols: The following symbols are used in this instruction manual:
This symbol warns of an operational, practical, or similar procedure that, if it is not
carried out properly, may damage the equipment
This symbol indicates hazardous voltages may be present
This symbol indicates that additional information has to be considered
Precautions
If this instrument is used in a not specified by the manufacturer manner, the protection
provided by the instrument may be impaired.
This instrument is not intended for use in a potentially hazardous environment.
In case of emergency, switch off the instrument and, after that, disconnect the electrical
cord from the wall outlet.
The user should ensure that the substances placed under test do not release
poisonous, toxic or flammable gases at the temperatures which they are subjected to
during testing.
4. Utilities
Input Voltage: Universal Power Supply (100-240V)
Input Frequency: 50 / 60 Hz
Power Consumption: 15 VA
Main supply voltage fluctuations cannot exceed ±10% of the nominal supply voltage
5. Specifications
Speeds: 0.3 –100 RPM
Viscosity Accuracy: ±1.0 % of full scale range
The use of accessory items will have an effect on the measurement
accuracy
Viscosity Repeatability: ±0.2 % of full scale range
Temperature Accuracy: ±0.1 °C | -40 °C to 300 °C (-40 °F to 572 °F)
Operating Environment: +5 °C to 40 °C temperature range (41 °F to 104 °F)
Non condensing humidity

ADVANCE Manual
7/69
Noise emitted: Set less than 70 dB (A)
Vibrations emitted: Less than 2.5 m/s²
Certifications:
Conforms to CE Standards:
EN 61010-1:2010 Safety requirements for electrical equipment, for measurement, control
and laboratory use.
EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use
Directive RoHS: 2011/65/UE + 2014/IUE a 2014/6/UE +2014/8/UE a 2014/16/UE
Notice to customers:
The product is made up of various components and various materials that
must be recycled or, failing that, deposited in the corresponding debris
removal sites when the product's life has been completed or when
otherwise it is necessary to dispose of it. To do this, the end user who
acquires the product must know the current regulations of each
municipality and / or locality based on the waste electrical and electronic
equipment. The user who acquires this product must be aware of and
responsible for the potential effects of the components on the environment
and human health as a result of the presence of hazardous substances.
Never place the product in a conventional container of citizen scope if a
previous dismantling and knowledge of the components that incorporates.
If you do not know the procedure to follow, consult with the city council for
more information.
6. Conditions for use
-
Indoor use
-
Maximum altitude 2000 m.
-
Surrounding temperature range: from +5 to 40ºC.
-
The equipment temperature must be kept above the dew point so moisture doesn't condense on
or in it. The power source fluctuations should not surpass 10% of the nominal voltage
-
Installation category II
-
Pollution level II
7. Maintenance
Always clean all of the parts after each use! Clean and dry the spindles and the spindle guard well.
Make sure that there is not any sample remaining, especially in the delicate zones such as the
spindle connector.
Use detergents or solvents to clean the spindles and the protector:
-
For cleaning food samples, use lukewarm water and if necessary, use soft household
detergents
-
Other solvents that generally provide good results are acetone, gasoline, or any solvent
with a high percentage of alcohol
-
For the use of any other solvent, make sure that it does not corrode the spindles or the
protector. The spindles are made in AISI 316.
Warning: Handle the volatile and flammable solvents with proper cautions. It is the
user’s responsibility to establish safety conditions at work.
Check regularly the spindle’s thread and the viscometer shaft.
During the viscometer’s lifespan, regular maintenance is important. FUNGILAB recommends annual
check-ups by the technical service of your local distributor.

ADVANCE Manual
8/69
The viscometer is powered by a MEAN WELL GS25A12-P6J power supply. Do not open, expose,
modify or touch the internal circuitry of the power supply.
8. Equipment presentation
-
Once the equipment package is received, check and confirm the delivery
note. If some discrepancy or problem is found, notify immediately the
supplier.
-
Verify that the viscometer model corresponds to the one that was
ordered.
-
Read carefully the instruction manual.
-
The manufacturer is not responsible for any damages that may result from
modifications or lack of maintenance of any of the machine’s mechanisms
(directive 89/655/CEE).
Fungilab recommends using the carry-case provided with the equipment for
making any kind of delivery. Please, keep the carry-case in a safe location. In case
of transporting the equipment or during long storage periods, always place each
part of the equipment as shown in figure 1. It can be seen that the position of each
piece inside the equipment’s carry-case is presented. In case of an incorrect
packaging, the pieces of equipment can suffer some damage. This damage will not
be covered by Fungilab’s guarantee.
Parts included within the equipment standard delivery:
-
Viscometer head with serial number
-
Foot or base, 3 height adjustable knobs for the base
-
Nut
-
Fastening rod
-
Standard spindles
-
Spindle guard
-
Spindle support
-
Calibration Certificate
-
Conformity declaration
-
Oil certificate document copy
-
USB-Memory containing the User Manual (PDF file). The USB-Memory might contain also the
company catalogues.
-
Power cable
-
PT100 Temperature probe
-
Clip for holding the PT100 probe
-
MEAN WELL GST25A12-P1J power supply
Do not open the power supply due to electrical shock risk. There are not
serviceable parts inside. In case of suspecting of a power supply malfunction,
please contact FUNGILAB for assistance.
It is mandatory to leave enough free space around the equipment ON/OFF switch,
needs to be reachable at any time, especially in case of an emergency or
malfunction.
It is very important to treat the silkscreen printed logos carefully when cleaning
the equipment. Please use a soft cloth, with isopropyl alcohol (70%).

ADVANCE Manual
9/69
Standard spindles
Model L: L1, L2, L3, L4
Models R and H: R2, R3, R4, R5, R6, R7
Fig. 1.a First level of the viscometer in its carry-case
Fig. 1.b Second level of the viscometer in its carry-case

ADVANCE Manual
10/69
9. Equipment Description
Fig. 2 Frontal view of the equipment
1.
Screen
2.
Capacitive keyboard
3.
Nut
4.
Spindle guard
5.
Fastening rod
6.
Temperature probe
7.
Spindle
8.
Sample container (not included)
9.
Base (viscometer support)
10.
Height adjustable knob

ADVANCE Manual
11/69
Fig.3 Back view of the equipment
1.
Serial number label
2.
Warning label
3.
Level
4.
Power switch
5.
Power cable slot
6.
USB Temperature probe connector
7.
USB connector
8.
Thermosphere connection
Fig.4. Equipment identification label
Description of the equipment identification label:
1.
Viscometer model
2.
Viscometer code
3.
Serial number of the equipment
4.
Voltage and power of the equipment
5.
Electronic equipment (specifies throw in trash)
5
1
3
2
4
2
8
4
7
1
6
5
3

ADVANCE Manual
12/69
9.1 Equipment set-up
Remove all of the parts from the carry-case or the standard package. Note the Figure below (fig 5).
Place correctly the three height adjustable knobs (B) on the Y-shaped base(A).
Mount the fastening rod (C) with the holding screw (D) at the base (A).
Attach the nut (F) to the fastening rod. The viscometer should be connected to the nut (F) by means of
its rod (E).
Note:
The following process should be done carefully in order to not harm to the shaft of the
viscometer. Immediately remove the shaft’s plastic protector before beginning to use the
viscometer.
Insert the horizontal rod of the viscometer (E) into the nut (F).
Fig. 5 Set-up for the viscometer base
The viscometer should be placed on a stable laboratory table or on a stable surface free of
vibrations (i.e. caused by other machines or equipment). Do not put the viscometer in direct
contact with sunlight or in the middle of any air flow (the temperature of the sample can be easily
influenced by the surrounding conditions). The viscometer has been designed to work indoor!
Turn the height adjustment knobs until the height of the viscometer (located in rod E) is correctly
adjusted.
Plug the power cable into its correct slot located on the back of the equipment (Fig. 3 position 4)
and plug it into the power source.
WARNING:
Verify that the voltage and the frequency coincide with the specifications of the power supply
(look at the identification). Before turning on the machine, let it sit for some time so that it
acclimates to the surrounding temperature in order to avoid a short-circuit caused by
condensation. The fluctuations of the power source should not surpass
10% of the nominal
voltage.
F
E
C
A
D
B

ADVANCE Manual
13/69
9.2 The keyboard and screen
Before starting up the machine, it is recommended to become familiar with the viscometer controls seen in
the previous section. The instrument has a 6 capacitive keyboard (Fig. 6) and a 4-lined Alphanumeric Display
screen (Fig. 2, number 1) on the frontal part ready to use and they allow the user to interact with the
machinery. The screen always shows the operations that the user is carrying out by showing menus that will
be explained later on. The measurements collected by the instrument will also be explained in this manual.
The keyboard gives the user the mobility throughout all of the menus, the selection of different options and
the creation and/or modification of viscosity measurement configurations to suit the user’s needs.
The keyboard has the following configuration:
Fig. 6 The keyboard for the ONE viscometer
The different numbered keys will always allow you to type in the proper numerical value (if a modifiable
field has been selected).
Key
Function
‘’
Go to the previous option; increase a value when a field has been
selected.
‘’
Go to the next option; decrease a value when a field has been
selected.
‘TAB’
The field selected changes in some menus.
‘QUIT’
Return to previous screen and stop the motor during
measurements.
‘ENTER’
Accept an option or value in a field. It also allows editing to fields
that can be modified.
‘ON’
Stop/Start the motor during measurements.
In the following sections, the function of each key in the corresponding menus will be explained in full detail,
including the exceptions to the general operation.

ADVANCE Manual
14/69
10. Menu system
Fungilab viscometers work with a system of menus that allow the user to go through the instrument in a
quick and simple way. The basic actions in the menus are: moving through the options (‘’ and ‘’ keys),
selecting an option (‘ENTER’ key) or returning to the previous menu (‘QUIT” key).
10.1 Start-up
Turn on the switch placed at the back of the machine (number 3, Fig. 3). If after doing this, the machine
does not turn on:
Verify that the power cable is connected to both the power and the Power Supply and that the
Power Supply is also connected to the equipment (back part, number 4, Fig. 3).
The machine will beep, indicating that it has started and it will show the following screen:
FUNGILAB S.A.
V.5.2
ADVANCE SERIES
English
The screen informs the user of the version and the instrument model in addition to the selected language.
After a few seconds, the Start-up screen will disappear and the Autotest screen for the viscometer is shown
(section 10.2 of this manual).
The equipment initially comes configured with:
-
English
-
Temperature units in Celsius (ºC)
-
Viscosity units in centipoises (cP).
If these are not the desired basic configurations, the equipment can be configured and changed to meet the
user’s needs. The method of configuring the apparatus by varying these and other parameters is explained
in detail in a later section of this manual called ‘Instrument setup menu’ (section 10.4). Any changes made to
the machine will stay configured to the last modification made at the configuration menu and will not return
to the factory settings after a restart.
Once the configuration information is given will submit the system to a Autotest.
10.2 Autotest
The Autotest menu allows the user to verify the operation of the viscometer in a way that allows detection
of motor malfunctions in a simple and practical way.
The following message will appear on the screen:
AUTOTEST
Remove the
Spindle and
Press <ENTER>

ADVANCE Manual
15/69
VERY IMPORTANT: The Autotest should be carried out without a spindle.
Once this message is shown on the screen, we should confirm that the spindle is not connected.
Afterwards, press ‘ENTER’ and the auto-check process will begin. While this test is running, the screen will
show this message:
Autotesting
...
The dots that appear below the Word ¨Autotesting¨ will appear continuously every half second.
Once the Autotest process finishes, two possible messages will appear, depending on the result of the
diagnostic.
If the viscometer detects an anomaly, it will show the following message on the screen while it emits an
acoustic warning:
AUTOTEST ERROR
The system is not
working properly,
press <ENTER>
If this message appears, the technical service from the supplier or manufacturer should be contacted. To
get the manufacturer’s contact information, press the <ENTER> key and the following format will appear.
TECHNICAL SERVICE
FUNGILAB, S.A.
+34 93 685 35 00
www.fungilab.com
If there is a system error, the equipment will stay blocked, meaning the motor is not working properly. If
the machine is turned off and restarted, the same screen will reappear.
In the case of a successful check, the main menu will be displayed.
Instrument Setup
Measurement
Test profiles
↓ Programming

ADVANCE Manual
16/69
10.3 Main Menu
The main menu is the one that appears after the Autotest screen. It is accessed by turning on the machine
normally and after a satisfactory result from the test run. The main menu screen will show:
Instrument Setup
Measurement
Test profiles
↓ Programming
By default, the cursor ‘’ is placed on the ‘Instrument Setup’ option.
The menu can be navigated with the ‘’ and ‘’ keys. The current selection will be highlighted and by
pressing ‘ENTER’ you will access to the selected submenu (for more information about each function in
particular see the corresponding sections).
The first time the machine is used, it is advisable to access the ‘Instrument Setup’ option as the first step in
order to establish the values for certain parameters of the viscometer such as language and measurement
units.
The ‘↓’ symbol indicates that the menu is more extensive than shown. When you are in the first option in
menu and press ‘’ or in the last option of the same menu and press ‘’ the screen will change and show:
The ‘↑’ symbol indicates that using the ‘’ and ‘’ keys you can return to the first screen of the main menu.
In the following sections, each of the 5 submenus of the main menu can be seen beginning with the
configuration submenu.
10.4 Instrument Setup menu
The configuration menu contains those functions that are not standardized and that modify the state and/or
operations of the instrument. Once the ‘Instrument Setup’ option is selected by pressing the ‘ENTER’ key,
the following screen will appear:
---Instrument Setup---
Language
Units
↓Calibration
Move through the options using the ‘’ and ‘’ keys and select a submenu with the ‘ENTER’ key.
Once again, the ‘↓’ symbol indicates that the menu is larger than the part shown. Using the ‘’ or ‘’ keys
you can change to the following screen:
↑ ---Instrument Setup---
Time settings
↑ Options

ADVANCE Manual
17/69
Using the ‘’ and ‘’ keys you can return to the original submenu.
The main submenu provides the possibility of:
-
Changing the working language
-
Selecting the measurement units (viscosity and temperature)
-
Carrying out calibrations (the machine comes calibrated from factory, therefore it is not necessary
to do any calibrations when the machine is received)
-
Adjusting the date and time.
The language, time and units should be selected by the user before beginning to work with the equipment
so that it functions properly.
10.4.1 Language
Once the configuration menu has been accessed, the first option that the cursor ‘’ points to is ‘Language’.
To change the language, this option must be selected by hitting the ‘ENTER’ key.
When we enter in this submenu, the viscometer will show a screen like the next one:
---Select language--
English
By using ‘’ and ‘’ the different working languages for this equipment can be seen, which are:
English
French
German
Italian
Japanese
Portuguese
Spanish
Dutch
Polish
Catalan
Once the language has been selected, press ‘ENTER’ and it will automatically change the language of the
menus and return to the configuration main menu screen.

ADVANCE Manual
18/69
10.4.2 Units.
The ADVANCE-type viscometer allows the user to select the units that are used for measuring viscosity and
temperature.
The possible choices for dynamic viscosity are:
-
International system of units (Pa·s or mPa·s)
-
Centimetre-gram-second system of units (Poise or centipoises)
And those of temperature units are:
-
Celsius (ºC)
-
Fahrenheit (ºF)
When the cursor key, ‘’, points to the units submenu, it can be accessed by pressing the ‘ENTER’ key and
the viscometer will show the following screen:
-Select the units-
Temperature
ºC
→
By default, this submenu screen for ‘Units’ comes configured with the temperature unit’s field selected.
Press the ‘TAB’ key to change to the viscosity units. Press ‘TAB’ again to return to the temperature units.
-Select the units-
Viscosity
cP/P (CGS)
→
Once the desired field has been selected, the units to be used with the viscometer can be varied by using
the ‘’ and ‘’ keys to switch the options.
After the desired units have been selected, press the ‘ENTER’ key to save the changes and return to the
configuration main menu screen.
10.4.3 Calibration
This submenu contains the viscosity calibration options that the user can exploit to recalibrate the viscometer.
IMPORTANT:
The viscometer contains a default calibration element, which is installed during the
manufacturing process. It is for this reason that it is unnecessary to calibrate the
equipment when using it for the first time. Nevertheless, certain norms of quality
recommend that the equipment be recalibrated once a year, which is why we offer the
user the possibility of realizing this calibration without needing to send the viscometer
back to the usual provider, or to FUNGILAB.
FUNGILAB cannot be held responsible for the measurements taken by an
independently recalibrated viscometer and it is essential to follow the instructions
given by Fungilab carefully when recalibrating.

ADVANCE Manual
19/69
Calibration Norms:
To execute a viscosity calibration, it is necessary to have on hand at least a
little standard calibration oil and a thermo-statization system to maintain the
sample at a constant temperature. If you do not possess this equipment, then
you will not be able to guarantee good post-calibration measurements.
FUNGILAB provides upon request the standard oils necessary for the
calibration, as well as the accessories need to thermo-statize theoils.
The calibration of any spindle will only modify the values of that individual
spindle. The rest of the equipment’s spindles will not be affected by this
calibration. If you want to calibrate more than one spindle it will have to be
calibrated one by one. The oils used for each spindle will also be different, so
for calibration you should have standard silicon oil for each spindle you’re
calibrating.
Tables 6, 7 and 8 (page 54) specify the standard oils necessary for each
spindle.
This submenu is accessed through the main configuration menu, by choosing the Calibration menu and
pressing ‘ENTER’. Once at the submenu, the following screen will appear:
----Calibration----
Reset
↓Viscosity
Using the ‘’ and ‘’ keys, you can select the different options of this submenu, placing the ‘’ cursor over
each option and pressing ‘ENTER’ to choose it.
Using the same ‘’ and ‘’ keys you can also switch between the original calibration menu screen and the
next one:
↑----Calibration----
Temperature
10.4.3.1 Reset
This submenu contains the equipment’s RESET option.
After resetting, the equipment will restore the original viscosity calibration.
Upon entering this submenu, the following screen will appear:
WARNING:
RESET THE EQUIPMENT
<ENTER> <QUIT>
Press ‘ENTER’ to continue with this procedure.
Once the ‘ENTER’ key is pressed, a second confirmation will be asked as a security measure. The following
screen will appear:

ADVANCE Manual
20/69
Are you sure?
<ENTER> <QUIT>
If you press ‘ENTER’ here, the factory-stage calibration will be restored (calibration, language), the memory
will be erased as well as the programming and you will return to the main configuration screen.
10.4.3.2 Viscosity calibration
When selecting the viscosity option (moving through the menu with the ‘’ and ‘’ key) and pressing
‘ENTER’ the user will access to the following screens, depending on the model of the viscometer:
Model L
Spindle L1
v 100.0 cP
Models R and H
Spindle R1
v 100.0 cP
The possible spindles’ list to use depends on the model of your viscometer (L, R or H). Thus, in tables 1 to 5
(page 50 and 51) you can see the different available spindles for each model.
Once this field is selected and situated in the list of corresponding spindles, you can select the spindle that
you wish to calibrate using the ‘’ and ‘’ keys.
Once the spindle is selected, go to the “Viscosity” field using the ‘TAB’ key. Pressing ‘ENTER’, accept the field
and introduce the value of the standard oil corresponding to the viscosity calibration. To introduce the data,
use the ‘’ and ‘’ keys to increase or decrease the value of each digit. Then, press ‘TAB’ again to go from
one-digit place to another.
Once the value of the oil is determined, press ‘ENTER’ to continue with the calibration process.
Afterwards, press the ‘ON’ key and the following screen will be shown:
Attach the
spindle and
press <ENTER>
This manual suits for next models
4
Table of contents
Other Fungilab Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Fleetminder
Fleetminder MDVR8AHD-4G Installation and user guide

IFM
IFM SD25 Series operating instructions

Martini Instruments
Martini Instruments MI 413 user manual

KBR
KBR multimess F144-0-LED-EP 4 Series User Manual Technical Parameters

YOKOGAWA
YOKOGAWA FLXA402T user manual

BD
BD FACSCelesta Flow Cytometer user guide