Fungilab EVO Series User manual

EVO SERIES
Rotational Viscometer
Instruction Manual

EVO EXPERT Manual 2/105
EVO SERIES
Rotational Viscometer
Software Version: 1.0
Manual Version 1.0
(Updated February 2014)
Instruction Manual
FUNGILAB S.A.
C/ Constitució, 64
Pol. Ind. Les Grases
08980 Sant Feliu de Llobregat
Barcelona, Spain
Phone: +34 93 685 35 00
Fax: +34 93 685 37 50
www.fungilab.com

EVO EXPERT Manual 3/105
0. Table of Contents
0. Table of Contents.................................................................................................................. 3
1. Introduction......................................................................................................................... 5
2. Safety Instructions ............................................................................................................... 5
3. Symbols used in this manual ................................................................................................. 5
4. Conditions for use ................................................................................................................ 6
5. Maintenance ......................................................................................................................... 6
6. Equipment presentation......................................................................................................... 7
7. Equipment Description........................................................................................................... 8
7.1 Equipment set-up ................................................................................................................ 9
7.2 The keyboard and screen................................................................................................... 10
8. Menu system ...................................................................................................................... 13
8.1 The Main Menu................................................................................................................. 13
8.2 Instrument Setup menu ..................................................................................................... 14
8.2.1 Language (language change submenu)............................................................................ 15
8.2.2 Units. (Unit change submenu) ......................................................................................... 15
8.2.3 Density. (Default density change submenu)...................................................................... 16
8.2.4 Calibration (Calibration submenu) .................................................................................... 17
8.2.4.1 Reset .......................................................................................................................... 18
8.2.4.2 Viscosity Calibration ..................................................................................................... 19
8.2.4.3 Temperature calibration ............................................................................................... 26
8.2.5 Time Settings ................................................................................................................. 28
8.3 Measurement Configuration ............................................................................................... 29
8.3.1 Measurement Screen ...................................................................................................... 31
8.4 Test Profile ....................................................................................................................... 32
8.4.1 Writing Tests Profile (Edit Profile) .................................................................................... 32
8.4.1.1 Viscometer programming ............................................................................................. 33
8.4.1.1.1 TTT and TTS............................................................................................................. 33
8.4.1.1.2 Storage .................................................................................................................... 35
8.4.1.2 Measurement Configurations ........................................................................................ 36
8.4.2 Select Profiles................................................................................................................. 36
8.5 Programming .................................................................................................................... 38
8.5.1 TTT (Time to Torque) and TTS (Time to Stop).................................................................. 39
8.5.2 Speed settings................................................................................................................ 40
8.5.3 Multistep........................................................................................................................ 43
8.5.4 Ramp............................................................................................................................. 47
8.6 Options............................................................................................................................. 49
8.6.1 Storage.......................................................................................................................... 49
8.6.2 Communications ............................................................................................................. 50
8.6.3 Information.................................................................................................................... 54
8.6.4 Graphic Mode ................................................................................................................. 55
9. Important rheological information ........................................................................................ 57
10. Accessories ....................................................................................................................... 62
10.1. Low viscosity adapters (LCP and LCP/B)........................................................................... 62
10.1.1 Mounting...................................................................................................................... 63
10.1.2 Dismounting and cleaning ............................................................................................. 64
10.1.3 Technical specification for LCP accessories ................................................................... 64
10. 2. Small sample adapters APM and APM/B......................................................................... 65
10. 2. 2 Dismounting and cleaning ........................................................................................... 66
10. 2. 3 Technical specifications of APM and APM/B................................................................... 67
10.3 HELDAL UNIT –Helicoidal Movement Unit......................................................................... 68
10. 3. 1 Heldal unit Mounting................................................................................................... 69
11. Model/Spindle correspondence tables ................................................................................. 71
12. Model/spindle/oil calibration tables ..................................................................................... 73
Table 9. EVO L standard spindles selection ............................................................................... 74
Table 10. EVO L special spindle selection.................................................................................. 75
Table 11. LCP Adaptor for EVO L.............................................................................................. 76
Table 12. EVO R standard spindle selection .............................................................................. 77

EVO EXPERT Manual 4/105
Table 13. EVO R Special spindle selection ................................................................................. 78
Table 14. LCP Adaptor for EVO R ............................................................................................. 79
Table 15. EVO H Standard spindle selection .............................................................................. 80
Table 16. EVO H special spindle selection ................................................................................. 81
Table 17. HELDAL special spindle selection for EVO L ................................................................ 82
Table 18. HELDAL special spindle selection for EVO R................................................................ 83
Table 19. HELDAL special spindle selection for EVO H ............................................................... 84
Table 20. VANE special spindle selection for EVO L.................................................................... 85
Table 21. VANE special spindle selection for EVO R ................................................................... 86
Table 22. VANE special spindle selection for EVO H ................................................................... 87
Appendix A: Wireless Datalogger configuration......................................................................... 88
WARRANTY CERTIFICATE ......................................................................................................105

EVO EXPERT Manual 5/105
1. Introduction
Thank you for acquiring the Evo Expert rotational viscometer model from Fungilab.
The Evo is a rotational viscometer, based on the measurement of the torque of a rotating spindle in
a sample at a specified velocity. Three different models, as well as various accessories, allow it to
cover a wide range of viscosity measurement.
2. Safety Instructions
It is not the purpose of this manual to outline all of the safety instructions recommended
for the use of the rotational viscometer, its accessories and samples. It is the responsibility
of the user to establish health and safety practices and to determine the application’s limits
before use.
Fungilab guarantees the satisfactory operation of the viscometers and its accessories only if
there have not been any unauthorized adjustments to the mechanical pieces, the electronic
components and the software.
The operator should follow all of the warnings and instructions of this manual to ensure the
safe and proper operation of the equipment.
Do not use the equipment for any other purpose that is not described in this manual.
Do not use any accessory that is not supplied and approved by Fungilab.
Do not use the viscometer or its accessories if there is any suspicion of malfunction. Do not
use the equipment in situations or conditions that can provoke personal injuries or material
damage.
The rotational viscometer is not an inflammable, non-hazardous instrument and therefore should
not be used in areas where there is an explosion risk.
Before using the viscometer, carefully read and observe the following
precautions: those who do not follow them may cause serious harm or
personal injuries.
To avoid an electric shock:
Do not use the rotational viscometer without a solid connection to the ground.
3. Symbols used in this manual
The following symbols are used in this instruction manual:
This symbol warns us of an operational, practical, or similar
procedure that, if it is not carried out properly, may damage
the equipment.

EVO EXPERT Manual 6/105
This arrow indicates additional information that should be
used by the user.
This symbol warns us of an operational, practical, or similar
procedure that, if it is not carried out correctly, may
irreparably damage the equipment. Do not proceed further
unless the indicated conditions are fulfilled and have been
perfectly understood.
4. Conditions for use
-Indoor use
-Maximum altitude 2000 m.
-Surrounding temperature range: from +5 to 40ºC.
-80% maximum relative humidity for up to 31ºC and going as low as 50% of relative
humidity for up to 40ºC.
-The power source fluctuations should not surpass 10% of the nominal voltage
-Installation category II
-Pollution level II
5. Maintenance
Always clean all of the parts after each use! Clean and dry the spindles and the spindle
guard well. Make sure that there is not any sample remaining left, especially in the delicate
zones like the spindle connector.
Use detergents or solvents to clean the spindles and the protector:
-For cleaning food samples, use lukewarm water and if necessary, use soft
household detergents
-Other solvents that generally give good results are acetone, gasoline, or any
solvent with a high percentage of alcohol
-For the use of any other solvent, make sure that it does not corrode the spindles or
the protector. The spindles are made in AISI 316.
.
Warning: Handle the volatile and inflammable solvents with proper cautions.
It is the user’s responsibility to establish safety conditions at work.
Regularly check the spindle’s thread and the viscometer shaft.
During the viscometer’s lifespan, the equipment might require certain check-ups to perform
as expected. For this service, please contact the local distributor.
Regular maintenance is important. As the manufacturer, we advise annual check-
ups by the technical service of your local distributor.
The viscometer is powered by the MEAN WELL GS25A12-P6J power supply. Do not
open, expose, modify or touch internal circuitry of the power supply.

EVO EXPERT Manual 7/105
6. Equipment presentation
-Once the equipment package is received, verify and confirm the
delivery note. If some discrepancy or problem is found, immediately
notify the supplier.
-Check that the viscometer model corresponds to the one that was
ordered.
-Carefully read the instruction manual.
-All modifications, eliminations, or lack of maintenance of any of the
machine’s mechanisms, defy directive 89/655/CEE and the
manufacturer is not responsible for any damages that may result.
In the attached photograph (Figure 1) the position of each piece inside the
equipment’s carry-case is presented. Please, keep the carry-case in a safe
location. In case of requiring transport the equipment or for long storage
periods, always use the carry-case by placing each part as shown in the
picture. In the case of incorrect packing, the pieces of equipment can suffer
some damage, this damage will not be covered by Fungilab’s guarantee.
Fungilab recommends using the carry-case provided with the equipment for
making any kind of delivery.
Parts included with the equipment for standard delivery:
-Viscometer head, with serial number label and a plastic or metal cap protecting the
spindle connector
-Foot or base, 3 height adjustable knobs for the base
-Nut
-Fastening rod
-Standard spindles
-Spindle guard
-Spindle support
-USB Cable
-Carrying case
-Calibration Certificate
-Temperature probe and clip
-Power cable
-MEAN WELL GS25A12-P6J power supply
-USB-Memory containing the User Manual (PDF file) the
WiFi-config
application and the
Fungilab Data Boss program (optional)
Standard spindles
Model L: L1, L2, L3, L4
Models R and H: R2, R3, R4, R5, R6, R7

EVO EXPERT Manual 8/105
Fig 1. Viscometer in its carry-case
7. Equipment Description
Fig. 2 Frontal view of the equipment
1. Nut
2. Temperature probe
3. Spindle
4. Base (viscometer stand)
5. Screen
6. Keyboard
7. Fastening rod
8. Spindle guard
9. Levelling knob
1
7
8
6
5
3
4
9
2

EVO EXPERT Manual 9/105
Fig. 3 Back view of the equipment
1. Power switch
2. Power cable slot
3. Warning Label
4. Serial number label
5. USB Connector
6. USB Temperature probe
connector
7. Level
Fig. 4. Equipment identification label
Description of the equipment identification label:
1. Viscometer model
2. Viscometer code
3. Electronic equipment (specifies throw in trash)
4. Serial number of the equipment
5. Voltage and power of the equipment
7.1 Equipment set-up
Remove all of the parts from the carry-case. Note the figure below (fig 5).
Correctly place the three height adjustable knobs (B) on the Y-shaped base (A).
1
1
2
3
4
5
2
3
4
5
6
7

EVO EXPERT Manual 10/105
Mount the fastening rod (C) with the holding screw (D) at the base (A).
Attach the nut (F) to the fastening rod. The viscometer should be connected to the nut (F) by
means of its rod (E).
Note:
The following process should be done carefully in order to not harm to the shaft of the
viscometer. Immediately remove the shaft’s plastic protector before beginning to use
the viscometer.
Insert the horizontal rod of the viscometer (E) into the nut (F).
Fig. 5 Set-up for the viscometer base
The viscometer should be placed on a stable surface free of vibrations (i.e. caused by other
machines or equipment). Do not put the viscometer in direct contact with sunlight or in the
middle of any air flow (the temperature of the sample can be easily influenced by the
surrounding conditions). The viscometer is designed for indoor use.
Use the height adjustment knobs until the height of the viscometer (located in rod E) is
correctly adjusted.
Plug the power cable into the connector located on the back of the equipment (Fig. 3
position 5) and plug it into the power source.
WARNING:
The socket by which the viscometer will be connected should have a ground. Always use
a power cable with a ground connection! Verify that the voltage and the frequency
coincide with the specifications for the viscometer (look at the identification label Fig. 4,
for more information). Before turning on the machine, let it sit for some time so that it
acclimates to the surrounding temperature in order to avoid a short-circuit caused by
condensation. The fluctuations of the power source should not surpass 10% of the
nominal voltage.
7.2 The keyboard and screen
Before starting up the machine, it is recommended to become familiar with the viscometer controls
seen in the previous section. The instrument has a 12 key keyboard (number 2 Fig. 2) and a colour
TFT screen (number 1 Fig. 2) on the frontal to allow the user to interact with the viscometer. The
E
F
C
D
B
A

EVO EXPERT Manual 11/105
keyboard gives the user the mobility throughout all of the menus and the selection of different
options and configurations. The screen presents informative menus in which the user operates.
These menus are detailed later in this manual. The measurements collected by the instrument will
also be explained later on.
The keyboard has the following configuration:
Fig. 6 The keyboard for the Evo Expert viscometer
The twelve keys available have many assigned functions depending on the operations that need to
be carried out. Some of these functions or operations can be carried out from any screen.
The different numbered keys will always allow you to type in the proper numerical value (if a
modifiable field has been selected).
Key
Function
‘’
Go to the previous option; increase a value when a field has
been selected.
‘’
Go to the next option; decrease a value when a field has been
selected.
‘►’
Change the selected field on some menus.
‘◄’
Return to the previous screen.
‘ENTER’
Accept an option or value in a field. It also allows editing to
fields that can be modified. Access to special functions.
‘MEM/CLEAR’
Stop the motor during measurements and returns to the main
menu screen. Erase the information present in a field when it
is highlighted. Shortcut to a test profile from the main menu
screen.
‘0’/ON
Start the motor and pause it during measurements. It also
allows running the measurement from its configuration
screen.
Keys 1M1 to 9M9 are used for recordings and their functions are detailed in section 8.4 of this
manual.
In the following sections, the function of each key in the corresponding menus will be explained in
full detail, including the exceptions to the general operation.

EVO EXPERT Manual 12/105
7.3 Start-up
Turn on the switch on the back of the machine (number 4, Fig. 3). If after doing this, the machine
does not turn on:
Verify that the power cable is connected to both the power and the Power Supply and that
the Power Supply is also connected to the equipment (back part, number 5, Fig. 3).
The machine will beep, indicating that it has started and it will show the AUTOTEST screen:
The equipment initially comes configured with:
-English
-Temperature units in Celsius (ºC)
-Viscosity units in centipoises (cP).
If these are not the desired basic configurations, the equipment can be reconfigured in the
‘INSTRUMENT SETUP’ (section 8.2). Any changes made to the machine configuration in the
previous menu will remain even after restarting the equipment.
7.4 Autotest
The AUTOTEST process allows you to verify the proper operation of the viscometer, in a way that
allows detection of motor malfunctions in a simple and practical way.
VERY IMPORTANT: The AUTOTEST should be carried out without a spindle.
Once the AUTOTEST screen is shown on the screen, we should confirm that the spindle is not
connected. Afterwards, press ‘ENTER’ and the auto-check process will begin. While this test is
running, the screen will show this message:
The progress bar that appears below the word ¨AUTOTESTING¨ displays the status of this process
accompanied by a textual representation of the progress in a percent format.

EVO EXPERT Manual 13/105
Once the AUTOTEST process finishes, two possible messages will appear, depending on the result
of the diagnostic.
If the viscometer detects an anomaly, it will show the following message on the screen while it
emits an acoustic warning:
If you press ‘ENTER’, the viscometer will present the contact information of the technical service.
The format of the menu should appear similar to the one in the following picture:
If there is a system error, the equipment will stay blocked, meaning the motor is not working
properly. If the machine is turned off and restarted, the same screen will reappear.
In the case of a successful check, the main menu will be displayed.
8. Menu system
8.1 The Main Menu
Fungilab viscometers work with a system of menus that allow the user to go through the instrument
in a quick and simple way. The basic actions in the menus are: moving through the options (‘’ and
‘’ keys), selecting an option (‘ENTER’ key) or returning to the previous menu (‘MEM/CLEAR’ key).

EVO EXPERT Manual 14/105
The main menu is the one that appears after the AUTOTEST screen. It is accessed by turning on
the machine normally and after a satisfactory result from the test run.
The main menu screen will show:
The menu can be navigated with the ‘’ and ‘’ keys. The current selection will be highlighted and
by pressing ‘ENTER’ you will access to the selected submenu (for more information about each
function in particular see the corresponding sections).
The first time the machine is used, it is advisable to access the ‘INSTRUMENT SETUP’ option as the
first step in order to establish the values for certain parameters of the viscometer such as language
and measurement units.
In the following sections, each of the 5 submenus of the main menu can be seen beginning with the
configuration submenu.
8.2 Instrument Setup menu
The configuration menu contains those functions that are not standardized and that modify the
state and/or operations of the instrument. Once the ‘INSTRUMENT SETUP’ option is selected by
pressing the ‘ENTER’ key, the following screen will appear:
Move through the options using the ‘’ and ‘’ keys and select a submenu with the ‘ENTER’ key. By
pressing the ‘MEM/CLEAR’ key, the user can return to the main menu and by pressing the ‘◄’ key,
the user can return the previous screen.
The main menu provides the possibility of:
-Changing the working language
-Selecting the measurement units (viscosity and temperature)
-Changing the value of the sample density (by default 1 g/cm3)
-Carrying out calibrations (the machine comes calibrated from factory, therefore it is not
necessary to do any calibrations when the machine is received)
-Adjusting the date and time.
The language, time and units should be selected by the user before beginning to work with the
equipment so that it functions properly.

EVO EXPERT Manual 15/105
8.2.1 Language (language change submenu)
Once the configuration menu has been accessed, the first option that the cursor ‘>’ points to is
‘LANGUAGE’. To change the language, this option must be selected by pressing the ‘ENTER’ key.
When we enter in this submenu, the viscometer will show a screen like the next one:
By using ‘’ and ‘’ the different working languages for this equipment can be seen, which are:
English
French
German
Italian
Spanish
Catalan
Once the language has been selected, press ‘ENTER’ and it will automatically change the language
of the menus and return to the configuration main menu screen.
If you want to leave without changing the language, the ‘MEM/CLEAR’ keys will take him to the
main menu or the ‘◄’ key will take you to the configuration menu.
8.2.2 Units. (Unit change submenu)
The Evo viscometer allows the user to select the units that are used for measuring viscosity and
temperature.
The possible choices for temperature units are:
-Celsius (ºC)
-Fahrenheit (ºF)
And those of dynamic viscosity are:
-International system of units (Pa·s or mPa·s)
-CGS (Poise or centipoises)
When the ‘UNITS’ submenu is highlighted, it can be accessed by pressing the ‘ENTER’ key and the
viscometer will show the following screen:

EVO EXPERT Manual 16/105
By default, the unit for the viscosity is cP and the unit for the temperature is ºC. Moreover, the
‘VISCOSITY’ field appears with light blue background, which means that its value can be changed
by using the ‘’ and ‘’ keys. Press ‘ENTER’ to save the selected viscosity unit and the field
‘TEMPERATURE’ will appear highlighted with light blue background. The light blue background
indicates that the value of the field can be modified by using the ‘’ and ‘’ keys. Press ‘ENTER’ to
save the selected temperature units.
After the desired units have been selected, hit the ‘ENTER’ key with the ‘SAVE’ option highlighted in
light blue background. The viscometer will save the selected units and it will return to the
‘Instrument setup’ menu.
If the ‘MEM/CLEAR’ key is pressed, it will cancel the new selections made for viscosity and
temperature, returning back to the previously used settings.
8.2.3 Density. (Default density change submenu)
The value assigned to the density of the fluid being measured can be changed by means of this
submenu. By default we consider the density of water as a reference point, but you can select any
other value. The default units will be g/cm3 of the Centimetre-gram-second system of units (CGS).
The field of the whole numbers appears highlighted in light blue background, which means that it
can be edited. Use the numerical keyboard to introduce the value desired for the density whole
numbers.
Once the digits of the whole numbers are introduced, press ‘ENTER’ to skip to the next field. Then,
the field of the decimal numbers will appear highlighted indicating that this field can be modified.
Use the numerical keyboard to introduce the decimal numbers of the density and press ‘ENTER’ to
save these numbers. In order to save the value of the density press ‘ENTER’ with the ‘SAVE’ option
highlighted in light blue background. The viscometer will return to the ‘Instrument setup’ menu.
NOTE:
If you modify the density, the viscometer will give its measurements in
cSt (centiStokes), whereas if you conserve the initial density (considered the
density by default), the measurements will be in cP (centipoises), P (Poise) or
mPa·s, Pa·s.

EVO EXPERT Manual 17/105
8.2.4 Calibration (Calibration submenu)
This submenu contains the viscosity and temperature calibration options that the user can exploit to
recalibrate the viscometer. Moreover, it also contains the ‘RESET’ option to restore the factory-stage
calibration and erase the memory and the programming.
IMPORTANT:
The viscometer contains a default calibration element, the Factory Calibration,
which is installed during the manufacturing process. It is for this reason that it
is unnecessary to calibrate the equipment when using it for the first time.
Nevertheless, certain norms of quality recommend that the equipment be
recalibrated once a year. This is why it is offered the user the possibility of
realizing this calibration, the User Calibration, without needing to send the
viscometer back to the usual provider, or to Fungilab.
Fungilab cannot be held responsible for the measurements taken by an
independently recalibrated viscometer and it is essential to follow the
instructions given by Fungilab carefully when recalibrating.
Calibration Norms:
To execute a viscosity calibration it is necessary to have on hand at
least a little standard calibration oil and a thermo-statization system to
maintain the sample at a constant temperature. If you do not have this
equipment then you will not be able to guarantee good post-calibration
measurements. Fungilab provides upon request the standard oils
necessary for the calibration, as well as the accessories need to
thermostatize the oils.
There are two types of calibration:
oCalibration of reference spindle: These spindles are coaxial
spindles, with which the accessories APM or APM/B must be
used. By calibrating these spindles, you’re changing the
calibration of all of the viscometer’s spindles. Reference
spindles:
Model L TL5
Model R TR8
Model H TR8
oCalibration of the rest of the spindles: The calibration of any
spindle, which is different from the reference spindle, will only
modify the values of that individual spindle. The rest of the
equipment’s spindles will not be affected by this calibration. If
you want to calibrate more than one spindle and you don’t do it
with the reference spindle, the spindles will have to be
calibrated one by one. The oils used for each spindle will also
be different, so for calibration you should have standard silicon
oil for each spindle you’re calibrating.
Tables 5, 6 and 7 (page 67 and 68) specify the standard oils necessary
for each spindle.
This submenu is accessed through the main configuration menu, by choosing the Calibration menu
and pressing ‘ENTER’. Once at the submenu, the following screen will appear:

EVO EXPERT Manual 18/105
Using the ‘’ and ‘’ keys, you can select the different options of this submenu, highlighting each
option and pressing ‘ENTER’ for choosing it. Using the ‘◄’ key, you can return to the previous
screen and with the ‘MEM/CLEAR’ key you will return to the main menu. If you hit ‘ENTER’, you will
select the option indicated by the cursor.
8.2.4.1 Reset
This submenu contains the equipment’s RESET option.
After resetting, the equipment will recuperate the original viscosity calibration.
Upon entering this submenu, the following screen will appear:
If you want to continue with this process, hit ‘ENTER’ and you will be brought to the following
screen. Otherwise, hit the ‘MEM/CLEAR’ key, which will bring you back to the main menu. In this
submenu, the keys ‘’ and ‘’ have no function.
If you press ‘ENTER’ here, the factory-stage calibration will be restored (calibration, language), the
memory will be erased as well as the programming and you will return to the main configuration
screen. If you hit ‘MEM/Clear’, you will return to the main menu and by hitting ‘◄’, no configuration
will be restored and you will also return to the main configuration screen. If you press ‘ENTER’ with
the field ‘QUIT’ highlighted the system will return to the ‘CALIBRATION’ menu.

EVO EXPERT Manual 19/105
8.2.4.2 Viscosity Calibration
It will be described first the procedure to perform the Factory Calibration. Once the ‘FACTORY
CALIBRATION’ option is chosen you will be prompted for a password, as it is shown in the following
screen:
Once introduced the correct password the following screen will appear:
If you select the viscosity option (moving through the menu with the the ‘’ and ‘’ keys) and you
press ‘ENTER’ you will access to the following screens, depending on the model of your viscometer:
Model L
Models R and H

EVO EXPERT Manual 20/105
Upon entering this screen, the spindle field is highlighted in light blue background. Using the ‘’ and
‘’ keys you can change the Spindle.
The list of possible spindles to use depends on the model of your viscometer (L, R or H). Thus, in
tables 8 through 22 (page 65 and on) you can see the different spindles available for each model.
Note: It is recommended to use the following combinations of Spindle and standard
oil depending on the viscometer model:
Model L: TL5 Spindle (LCP adapter) and RT-50 (50 cP) standard
Model R: TR8 Spindle (LCP adapter) and RT-500 (500 cP) standard
Model H: TR8 Spindle (LCP adapter) and RT-500 (500 cP) standard
Once you’ve selected your spindle press ‘ENTER’ and the ‘VISCOSITY’ field will be highlighted. Press
‘ENTER’ again and the following screen will appear:
Use the numerical keyboard to introduce the value of the viscosity of the standard oil used for
calibration (the standard oils provided by Fungilab provide viscosity tables according to different
working temperatures). There is a field for entire numbers and other one for the decimal figures.
Once introduced the density value, press ‘ENTER’ with the ‘SAVE’ option highlighted to confirm the
modification. Next, the following screen will appear:
Remove the Spindle if it is connected to the viscometer and press ‘ENTER’. Then, the viscometer
will perform the Offset calibration, showing the following screen:
The screen shows the progression of this step of the calibration with a status bar. Once the Offset
calibration is completed, the following screen will be on:
This manual suits for next models
3
Table of contents
Other Fungilab Measuring Instrument manuals