
1.0. Installation, Operation, and Maintenance Warnings
This section contains warnings regarding the installation, operation, and maintenance of the VHM Arm.
This section must be read in its entirety before installing, operating, and maintaining the VHM Arm. Failure
to follow these warnings may result in damage to equipment or injury to personnel.
Warnings
• Do not position VHM Arm or mounted instrument above a patient. Note that the VHM ARM has a
wide range of motion both up/down and side to side. Please consider carefully the instrument
being mounted and the proximity of the mounting assembly to other equipment, hospital
personnel, and the patient. GCX recommends that the hospital’s risk management personnel
verify that the application is appropriate prior to installation and use of the VHM Arm.
• If VHM Arm is wall mounted, verify that the channel has been installed and approved in
accordance with channel installation guide.
• Ensure that the weight of the instrument being mounted does not exceed the load rating of the
VHM Arm. Check the bottom surface of the Arm for "Maximum Load" label (see 2.1 for label
location). Mounted instrument must be within "Optimal Weight Range" for best performance.
• The Height Locking Lever or Knob must be engaged when removing a mounted instrument from
the VHM Arm.
• Stand to the side of the VHM Arm and mounted instrument and use
caution when disengaging the Height Locking Lever. The total
mounted load may have changed due to the addition or removal of
equipment (other than primary instrument) while the Height Locking
Lever was engaged. A change in the total load can cause a sudden
downward or upward movement of the Arm when the Height Locking
Lever is disengaged.
• The mounted device may move suddenly due to normal wear or
improper adjustment of the tilt and swivel functions (see Sections 4.0,
5.2, 5.5, and 5.6) or ultimately, gas spring end of life. The VHM Arm
must be inspected and adjusted at least two times per year. This
inspection must include the following steps:
1) With the device mounted, release the lock and move the arm through its entire vertical range of
motion. The load should maintain its position at every point in the travel of arm. If necessary, the
counterbalance mechanism may be adjusted in accordance with section 4.0.
2) Grasp the mounted device and swivel it from side to side. The device should swivel with some
tension or resistance, not loosely. If necessary, the swivel tension may be adjusted in accordance
with section 5.5.
3) Grasp the Arm and pivot it from side to side at the channel attachment. The arm should pivot with
some tension or resistance, not loosely. If necessary, the pivot tension may be adjusted in
accordance with section 5.4.
4) Grasp the mounted device and tilt it forward and back, through its entire range of motion. There
should be enough tension or resistance in the tilt mechanism to prevent the device from tilting
forward unexpectedly when in use. If necessary, the tilt tension may be adjusted in accordance
with section 5.6.
5) Inspect all remaining fasteners for looseness. Tighten as required for optimal operation and
safety.
6) The VHM Arm housing must never be disassembled by non-GCX personnel.
Failure to periodically inspect and adjust the Arm as instructed may result in damage to equipment or
injury. If properly maintained, the VHM Arm should function properly for 5 to 10 years. If you are unable to
adjust the arm or require service, contact GCX technical support at 800-228-2555.
Install your VHM Arm in the sequence presented in this manual (Sections 2.0, 3.0, and 4.0).
DU-WS-0001 Rev G 2/24/10 GCX Corp. Page 2 of 11