Golden engineering XR200 User manual

JUNE 2017

1
2
2
DUTY CYCLE 2
STORAGE 2
3
HIGH VOLTAGE PULSER/TUBEHEAD 3
BASE 3
BATTERY PACK 3
BATTERY CHARGER 3
CONTROL MODULE 4
CABLE CONNECTOR DIAGRAM 5
6
BLOCK DIAGRAM 6
7
OPERATING PRECAUTIONS 8
EXCLUSION ZONE 8
PULSE SETTINGS 8
9
11
DOSE MEASUREMENT 11
TUBE REPLACEMENT 11
11
12
REMOVING BOARDS 12
BOARD INSTALLATION 12
HEAD REPLACEMENT 12
BATTERY DISPOSAL 13
13
13
14
PHYSICAL DIMENSIONS 14
X-RAY OUTPUT 14
ELECTRICAL & THERMAL CHARACTERISTICS 14
14

2
The XR200 is a small, lightweight x-ray generator that operates on its own removable battery pack. The XR200 is
a pulsed X-ray device that produces X-ray pulses of very short duration (50 nanoseconds). It produces a low dose
rate comparable to a 0.5 ma constant potential machine. The energy produced by the XR200 is up to 150KVP,
which makes it possible to radiograph up to one (1/2) inch (1.27 cm) of steel.
XR200 standard accessories are two keys, two battery packs, and one battery charger. Remote cable, carrying
case, are also common accessories.
The XR200 is a pulsed X-ray generator that emits hazardous ionizing radiation when pulsing. The XR200 should
only be by who are properly trained to safely operate the generator. The XR200
must be with proper authorities prior to use and should be used to intentionally expose humans.
Develop and closely follow a safe operating system for using the XR200. The safe operating system must ensure
that no one is exposed to radiation above the permissible limits which are 2 mR (0.02 mSv) per hour for a member
of the public. The safe operating system must ensure the XR200 is used within federal and state guidelines.
All operators and users of the XR200 X-ray machine must wear a personal radiation monitoring device, such as
a TLD (thermoluminescent dosimeter), lm badge, and/or a pocket dosimeter
(note: an electronic dosimeter will not detect the XR200 radiation
pulses).
Due to the short pulse width of the XR200, survey meters of the Geiger-Mueller and scintillator type do not
accurately detect the radiation emitted from the X-ray source.
Survey meters should be of the ionization type and should be used in the integration modeSurvey meters must not
be used in the rate mode because the XR200 does not produce constant radiation. The XR200 produces very high
rates of radiation for very short periods of time resulting in either unrealistically high readings or no readings for
a survey meter in rate mode.
Upside down storage or use may cause
premature failure including no output dose.
The XR200 is a light duty machine that is not made to pulse continuously. The maximum
duty cycle for the XR200 is 200 pulses every four minutes (3000 pulses per hour). Exceeding the duty cycle will
shorten the life of the tube and head.

3
Figure 1: XR200 X-ray Unit
The main body of the XR200 is the tube head
which contains the tube cavity, cold cathode type X-ray tube, spark gap, high voltage
capacitor, and transformer. The standard collimator located on the front of the head
limits the X-ray beam to 40 degrees. Special order collimators up to 85 degrees are
available.
. The base of the XR200 contains an identication label and a ¼-20 brass
insert compatible with standard camera tripods. The base also accommodates a
quick release external tripod mount. The identication label located on the bottom
of the generator lists the manufacturer’s name and address, model number, serial
number, weight, volt, amp, and production date.
. The standard battery pack
is a DeWalt® 20V 2 amp hour Li Ion battery
(DCB203).
The standard battery
charger is the DeWalt® DCB107 110V charger
or DCB105 220V charger. (Note DeWalt part
numbers may change). Battery charge time is
less than one hour. See battery charger manual
for additional instruction and warnings.
The XR200 has a 21 mm picatinny rail located on each side of
the housing.
Figure 2: Base

4
Illuminates when battery voltage is applied to control module.
Blinks after time delay button or remote cable button is pressed to warn that the
XR200 is going to pulse. The light stays on continuously while the XR200 is pulsing. This is a failsafe warning
light. If the light does not work the X-ray unit will not pulse. See settings menu for fail safe override in emergency
situations.
80 digit LCD. Home displays number of pulses selected and delay time before
the rst pulse. Home will also indicate if multiple pulse trains have been entered. Arrows allow operator to scroll
through various menu options.
Used to maneuver through the menu.
Left and Right arrow buttons pressed simultaneously Initiates the delay mode.
Stops the unit before it begins pulsing or stops the unit in the middle of a pulse
train. Also used as the enter button to select desired option.
Figure 3: Control Module

5
REMOTE CONNECTOR: LEMO EEG.0K.305.CLN
MATING CABLE PLUG: LEMO FGG.305.CYCC50Z
PIN 5
PIN 2
Remote switch inputs are activated when grounded.
3 4
5
2
1
Lemo “K” series ve pin connector located on the back of the control module beneath the
battery receives the remote cable or imaging system cable.
PIN # DESCRIPTION
1 +5 VOLTS 100 ma MAXIMUM
2 REMOTE SWITCH
3 REMOTE SWITCH - NO DELAY
4 X-RAY ON SIGNAL
5 COMMON 0 VOLTS
Remote cable connector

6
The block diagram below illustrates how the XR200 functions. The following sequence of events takes place
each time the XR200 is red:
1. User initiates operation of the machine.
2. The control section sends a signal to the converter section to begin oscillating.
3. Once oscillating, the converter section changes the 20 volts DC to 22Khz AC.
4. The transformer charges the High Voltage Capacitor to about 8000 volts.
5. The spark gap arcs after the High Voltage Capacitor reaches proper voltage.
6. The pulse detector signals the control block that the unit has pulsed.
7. As the High Voltage Switch is closed, a high voltage transient of approximately 150,000 volts and
50 nanoseconds in duration is applied across the x-ray tube generating x-rays.
The closing of the High Voltage Switch produces an audible pulsing sound.
This unit generates X-rays through high voltage bombardment of a tungsten target. The XR200 does not contain
radioactive materialsAll the high voltage is contained within the aluminum canister and as long as the canister
is not punctured the operator is not exposed to dangerous voltages.
PULSE DETECTOR
USER INTERFACE
X-RAY TUBE
HIGH VOLTAGE
CAPACITOR
CONVERTER
BATTERY
SPARK GAP
CONTROLLER

7
The following are basic operation instructions to take an X-ray image using the XR200. Certain applications may
require modications to these basic procedures.
DETERMINE PULSE INITIATION METHOD:
DR, REMOTE CABLE, TIME DELAY
DR REMOTE CABLE TIME DELAY
Attach Battery Attach Battery Attach Battery
Connect System Attach Remote Cable Position XR200*
Position XR200* Position XR200*
Turn on XR200 using key and
turning clockwise 1/4 turn
Turn on XR200 using key and
turning clockwise 1/4 turn
Turn on XR200 using key and turning
clockwise 1/4 turn
Set Pulse Count
Set pulse count to 100 Set Pulse Count
Set Delay Time
Retreat to imaging system
controller (computer)**
Retreat length of cable behind the
XR200**
Press left and right buttons
simultaneously to initiate time
delay
Follow imaging system
instructions to set pulses and
initiate pulse train
Press and hold cable button down to
initiate and complete pulse train
Retreat safe distance behind
XR200**
View Image Retreive image plate or film cassette
Retreive image plate or film
cassette
Turn off XR200 View Image View Image
Turn off XR200 Turn off XR200

8
The chart below lists pulses necessary to penetrate various materials. Settings vary greatly
depending on imaging system used. Refer to imaging system instructions for more information.
MATERIAL PULSE SETTING (24” BETWEEN X-RAY & IMAGER)
CARDBOARD / LIGHT WOOD / PLASTIC 2-5
LIGHT METAL 10
STEEL 1/4” 25
STEEL ½” 50
06.1
06.1
Figure 4: Exclusion Zone
The operator should always stand at least 10 feet (3m) behind the X-ray unit and clear
all personnel at least 10 feet (3m) behind the unit or at least 100 feet (30m) from the front of the unit before pulsing.
The exclusion zone (below) should be a controlled area free of all personnel while X-ray pulses.
*XR200should be positioned directly in front of the object to be X-rayed and the imager placed directly behind
the object to be X-rayed. Imager should be placed as close to the object as possible. Distance between XR200
and imager is usually 24 to 72 inches (30 to 180 cm). During operation XR200 should be stabilized on a at
surface, a tripod, or a custom xture suitable for holding the 11 pound (5 Kg) XR200.
24 to 72 inches (30 to 180 cm)

9
The following is true when using lm or digital systems that generate a positive image. If the radiograph is too
dark, the lm is underexposed. If the radiograph is too light the lm is overexposed. can be
corrected by increasing the number of pulses and/or decreasing the distance between the imaging medium and
the XR200. can be corrected by reducing the number of pulses and/or increasing the distance
between the imaging medium and XR200.
Underexposed
Overexposed Correct exposures (pulse setting)
Correct exposures (pulse setting)
Low Battery Battery voltage is too low to pulse the XR200. Charge the battery.
Slow Pulsing
More than one second between pulses in a pulse train. Either low
battery, problem with transformer, or electronics. Charge battery or
replace oscillator board.
No Feedback
Current not detected. No current ow going to the head. Occurs at
the beginning of pulse train. Check 20 amp fuse on the oscillator
board. If fuse is ok problem could be in head or electronics. Replace
board or send back for repair.
Cycle Reached Displayed if duty cycle is reached (200 pulses in 4 minutes).

10
>HOME MENU PRESETS
PULSES: _ _ _ _
DELAY: _ _ _ _
HOME >MENU PRESETS
PULSES LIFE PC
DELAY RESET PC
TRAINS SETTINGS
HOME - MENU PRESETS SET PULSE COUNT
>PULSES LIFE PC _ _ _ PULSES
DELAY RESET PC
TRAINS SETTINGS BACK SAVE
HOME - MENU PRESETS SET DELAY TIME
PULSES LIFE PC _ _ _ SECONDS
>DELAY RESET PC
TRAINS SETTINGS BACK SAVE
HOME - MENU PRESETS SET PULSE TRAINS
PULSES LIFE PC TRAINS: _ _ _ _
DELAY RESET PC TIME BETWEEN: _ _ _
>TRAINS SETTINGS BACK SAVE
HOME - MENU PRESETS LIFETIME PULSES
PULSES >Life PC _ _ _
DELAY Reset PC
HOME - MENU PRESETS RESETTABLE COUNTERS
PULSES LIFE PC #1: _ _ _ _ RESET
DELAY >RESET PC #2: _ _ _ _ RESET
TRAINS SETTINGS BACK
HOME - MENU PRESETS EXTRA SETTINGS BACKLIGHT
PULSES LIFE PC >BACKLIGHT FEEDBACK ON OFF
DELAY RESET PC INFO FAILSAFE
TRAINS >SETTINGS BACK
EXTRA SETTINGS FEEDBACK SIGNAL
BACKLIGHT >FEEDBACK OUT OF REMOTE CABLE
FAILSAFE
BACK >PER PULSE ON/OFF
EXTRA SETTINGS X-RAY INFORMATION
BACKLIGHT FEEDBACK PROGRAM: _ _ _
>INFO FAILSAFE HEAD # _ _ _ _ _ _ _
BACK S/N: _ _ _ _ _
EXTRA SETTINGS WARNING LED
BACKLIGHT FEEDBACK FAILSAFE CIRCUIT
X-RAY INFO >FAILSAFE
BACK >ENABLE DISABLE
HOME MENU - PRESETS PRESET #1
> PRESET #1 PULSES: _ _ _ _ DELAY: _ _ _ _
PRESET #2 TRAINS: _ _ _ _ BTWN: _ _ _ _
PRESET #3 BACK LOAD SAVE
HOME MENU - PRESETS PRESET #2
PRESET #1 PULSES: _ _ _ _ TRAINS: _ _ _ _
>PRESET #2 TRAINS: _ _ _ _ TIME BETWEEN: _ _ _ _
PRESET #3 BACK LOAD SAVE
HOME MENU - PRESETS PRESET #3
PRESET #1 PULSES: _ _ _ _ TRAINS: _ _ _ _
PRESET #2 DELAY: _ _ _ _ TIME BETWEEN: _ _ _ _
> PRESET #3 BACK LOAD SAVE

11
Using a dosimeter, the average X-ray dose for new tube can be established.
With the dosimeter located 1 foot from the front of the case and in line with the center of the beam angle
label, the reading for 10 pulses should be 26 mR to 36 mR.
The leakage sheet illustrates the X-ray dose and maximum allowable radiation leakage levels for each
X-ray unit. A completed copy of this form accompanies each X-ray.
The XR200 tube should last at least 100,000 pulses. Under normal conditions the tube’s
output will decrease slowly with use. If the tube is broken or glass cracks the tube output will cease immediately.
The following are tube replacement instructions.
1. Remove the battery before unscrewing the collimator.
2. To replace the tube unscrew the collimator.
3. Using needle nose pliers or your ngers grab the front of the tube and pull straight out.
No “power on” light -Check battery voltage
-Check battery connection
- Replace or charge battery
- Make sure battery is securely
attached and battery clips are not
bent or broken.
Power on lights, X-ray pulsing light
does not illuminate, X-ray does not
pulse
- Go to settings menu
failsafe
disable
- To x light replace processor board
X-ray pulses, but no image or black
image. -Test X-ray output. -Replace the tube.
Unit stops pulsing in the middle of a
pulse train and LCD displays 00.
-Check the battery voltage.
-Check 15 amp fuse.
-Check feedback line connection.
- Charge battery if necessary.
- Replace the fuse if blown.
- Make sure the screw holding
wire to the oscillator board.
Unit makes loud popping noise while
pulsing. -
.
Oil leaking from unit. Return for repair.
Unscrew the collimator Grab the front of the tube and pull straight out

12
1. Use T10 Torx driver to remove 6 screws holing control module in place.
2. Remove the control module.
3. Remove cable connecting the boards.
4. Control Module Removed
1 2 3 4
Remove keyswitch connector Remove 3 screws holding processor board to top Remove processor board.
Disconnect ribbon cable. Remove the 8 screws that hold the housing together Remove half the housing.
Remove the head. Loosen the four screws that hold the feedback wires in place.

13
s for disposal of lithium-io
the end user. . To make a claim under this limited
warranty, customer must ship the entire unit (or the component believed to be defective) to Golden Engineering,
post-paid. Golden Engineering assumes no liability for units or components shipped until they are actually in
the custody of Golden Engineering, Inc. Provided Golden Engineering, in its sole discretion, is satised that the
failure is not the result of excessive use, abuse, misuse, accident, modication or improper disassembly or repair,
Golden Engineering will provide parts and labor required for the repair. Golden Engineering reserves the right
to use reconditioned and remanufactured components that meet original specications. The unit or component
will be shipped to customer at customer’s expense. THIS EXPRESS LIMITED WARRANTY IS IN LIEU OF ALL OTHER
WARRANTIES AND GUARANTEES, EITHER EXPRESS OR IMPLIED OR CREATED BY OPERATION OF LAW.
at and include a copy of the
printed form with the repair. If you do not have internet access prior to sending repair then include a letter
containing a brief description of the problem, contact name, phone number, and return address.
Remove battery before shipping the unit.
Accessories are not required
Be sure the unit is securely packaged for shipment and wrap in plastic bag if there is an oil leak.
Ship to address:
Remove feedback wires.
Golden Engineering, Inc. warrants XR200 X-ray unit made and sold by it or its authorized representatives to be
free of for a period of twelve (12) months from the date of shipment to
Remove the screws that connect the
oscillator board to the head.
Remove the oscillator board.
The XR200 X-Ray Source is
manufactured by:
GOLDEN ENGINEERING, INC.
PO BOX 185
CENTERVILLE, IN 47330 USA
Phone: 1-765/855-3493
Fax: 1-765/855-3492
Web: www.goldenengineering.com
Email: [email protected]
European Representative:
Certification Experts Europe
Nieuwstad 100
1381 CE Weesp,
The Netherlands
Follow all federal, state, and local laws for disposal of Li-Ion batteries.

14
LENGTH 11.95 inches (30.35 cm) with battery
WIDTH 4.26 inches (10.82 cm)
HEIGHT 5.83 inches (14.81 cm)
WEIGHT 11 pounds (5 Kg) with battery
X-ray dose per pulse 2.6 mR to 4.0 mR (12 inches in front of unit)
Pulses per battery charge 6000
Pulses per second 10 (Nominal)
Expected tube life (glass tube) 100,000 pulses
X-ray source size 1/8 in. (3mm)
Maximum Photon Energy 150 KVP
X-ray pulse width 50 nanoseconds
Battery voltage 18-20 volts
Battery type Li Ion
Battery recharge time One Hour
Current draw 9 amps @ 20 volts
Average X-ray Tube Current 0.5 mA
Temperature range 0 to 120 degrees F (-18 to 50 degrees C)
Maximum duty cycle 200 pulses every 4 minutes (3000 pulses per hour)
Warm-up None required
Thumbwheel Key 2002000
Flat key 5951040
Tube 2200020
DeWalt® Battery 20V (2 Amp Hour) DCB203 1800106
DeWalt® Battery Charger (110V) DCB107 1800152
DeWalt® Battery Charger (220V) DCB105 or DCB115 1800164
Remote Cable 1809013
Tripod Mount 4000352
Handle 4000153
Carrying case (holds X-ray, 2 batteries, charger, cable) 1701641
11.0 SPECIFICATIONS
Airborne Noise Emissions 80 dB at 10 cm

Table of contents
Other Golden engineering Scanner manuals