Gossen MetraWatt SECULIFE SP BASE User manual

Operating Instructions
SECULIFE SP BASE
Digital Pulse Oximetry Simulator
3
3-447-093-03
1/12.20

SECULIFE SP BASE Gossen Metrawatt GmbH
Contents
1 Safety Precautions ....................................................................1
2 Applications...............................................................................3
2.1 Intended Use / Use for Intended Purpose................................................... 3
2.2 Use for Other than Intended Purpose ......................................................... 3
2.3 Liability and Guarantee............................................................................... 3
2.4 Opening the Instrument / Repairs............................................................... 3
3 Documentation ..........................................................................4
3.1 Identifiers ................................................................................................... 4
4 First Steps..................................................................................5
5 The Instrument ..........................................................................6
5.1 Scope of Delivery ....................................................................................... 6
5.2 Instrument Overview .................................................................................. 7
5.3 Technical Data ........................................................................................... 8
5.4 Applied Standards ..................................................................................... 9
5.5 Combined Products ................................................................................... 9
6 Menu and Functions ...............................................................10
6.1 User Interface........................................................................................... 10
6.2 Manual Mode ........................................................................................... 12
6.3 Preset Mode ............................................................................................ 13
6.4 Setup Menu ............................................................................................. 14
7 Installation ...............................................................................16
7.1 Unpacking the Instrument ........................................................................ 16
7.2 Power Supply........................................................................................... 16
8 Operation .................................................................................17
8.1 Switching on/off ....................................................................................... 17
8.2 Connecting the Instrument to a DUT ........................................................ 17
8.3 Adjusting Simulation Environment ............................................................ 19
8.4 Changing Setup Configuration ................................................................. 20
8.5 Displaying Instrument Information............................................................. 22
8.6 Updating Firmware................................................................................... 22
9 Maintenance ............................................................................23

Gossen Metrawatt GmbH SECULIFE SP BASE
9.1 Cleaning................................................................................................... 23
9.2 Maintenance ............................................................................................ 23
9.3 Troubleshooting ....................................................................................... 23
9.4 Repair ...................................................................................................... 24
10Returns and Environmentally Sound Disposal.......................25
11Service and Support................................................................26
11.1Product Support ...................................................................................... 26
11.2Repair and Replacement Parts Service
Calibration Center * and Rental Instrument Service .................................. 26
12CE Declaration.........................................................................28

1
SECULIFE SP BASE Gossen Metrawatt GmbH
1 Safety Precautions
Observe this documentation, in particular all included safety information, in order to
protect yourself and others from injury, and to prevent damage to the instrument.
General
• Carefully and completely read and adhere to these operating instructions.
The document can be found at http://www.gossenmetrawatt.com. Retain this
document for future reference.
• Use only the specified accessories (included in delivery or listed as optional) with
the instrument.
• Carefully and completely read and adhere to the product documentation of the
optional accessories. Retain these documents for future reference.
• Observe and comply with all safety regulations which are applicable for your
work environment.
Handling
• The instrument may only be used as long as it’s in good working order.
• Inspect the instrument before use. Pay particular attention to damage, broken
insulation or kinked cables.
• Damaged components must be replaced immediately.
• Accessories and cables may only be used as long as they’re fully intact.
• Inspect all cables and accessories before use. Pay particular attention to dam-
age, broken insulation or kinked cables.
• If the instrument or an accessory doesn’t function flawlessly, remove it from
operation and secure it against inadvertent use.
• If the instrument or accessory is damaged during use, e.g. through falling,
remove it from operation and secure it against inadvertent use.
Operating Conditions
• Do not use the instrument after long periods of storage under unfavorable con-
ditions (e.g. humidity, dust or extreme temperature).
• Do not use the instrument after extraordinary stressing due to transport.
• The instrument must not be exposed to direct sunlight.
• Only use the instrument and the accessories in compliance with the specified
technical data and under the specified conditions (ambient conditions, IP pro-
tection class, measuring category, etc.).
• Do not use the instrument in potentially explosive atmospheres.
• The instrument and the included accessories may only be used for the tests/
measurements described in the instrument‘s documentation.

2
Gossen Metrawatt GmbH SECULIFE SP BASE
Electricity
• Do not use the instrument if the battery compartment cover has been removed.
Otherwise you risk touching dangerous voltages.
• Inspect the batteries before use. Pay particular attention to leaking and damage.

3
SECULIFE SP BASE Gossen Metrawatt GmbH
2 Applications
Please read this important information!
2.1 Intended Use / Use for Intended Purpose
SECULIFE SP BASE is a digital pulse oximetry simulator. It is intended as a testing
instrument for patient monitors. It produces pulses of infrared and red light to simu-
late various combinations of oxygen saturations, heart rates, and perfusion indices.
At the same time it reads the signals from a connected patient monitor.
SECULIFE SP BASE is also designed to function as an accessory for the patient
simulator SECULIFE PS300. SECULIFE SP BASE adds the simulation of SpO2 to
the patient simulator‘s range of parameters.
This instrument must not be used as a medical device. That means, the instrument
is intended for testing only and must never be used in diagnostics, treatment or any
other capacity where it would come in contact with a patient.
Safety of the operator, as well as that of the instrument, is only assured when it’s
used for its intended purpose.
2.2 Use for Other than Intended Purpose
Using the instrument for any purposes other than those described in the con-
densed operating instructions or these instrument operating instructions is contrary
to use for intended purpose.
2.3 Liability and Guarantee
Gossen Metrawatt GmbH assumes no liability for property damage, personal injury
or consequential damage resulting from improper or incorrect use of the product, in
particular due to failure to observe the product documentation. Furthermore, all
guarantee claims are rendered null and void in such cases.
2.4 Opening the Instrument / Repairs
The instrument may only be opened by authorized, trained personnel in order to
ensure flawless, safe operation and to assure that the guarantee isn’t rendered null
and void. Even original replacement parts may only be installed by authorized,
trained personnel.
Unauthorized modifications to the instrument are prohibited.
If it can be ascertained that the instrument has been opened by unauthorized per-
sonnel, no guarantee claims can be honored by the manufacturer with regard to
personal safety, measuring accuracy, compliance with applicable safety measures
or any consequential damages.
Nor does Gossen Metrawatt GmbH assume any liability for data loss.

4
Gossen Metrawatt GmbH SECULIFE SP BASE
3Documentation
3.1 Identifiers
The following identifiers are used in this documentation:
Identifier Meaning
Attention!
!
Warning
Safety information that must be complied with.
Note!
Important
Important information which must be taken into consider-
ation and complied with.
Prerequisite A condition etc. which must be fulfilled before a given ac-
tion can be taken.
1. Procedural step Steps of a procedure which must be completed in the
specified order.
Result Result of a procedural step.
•Enumeration
–Enumeration
Bullet lists
Figure 1: Caption Description of the content of a figure.
Tabelle 1: Description of the content of a table.
Footnote Comment

5
SECULIFE SP BASE Gossen Metrawatt GmbH
4FirstSteps
Read and adhere to the product documentation. In particular observe all safety
information in the documentation, on the product and on the packaging.
"Safety Precautions" 1
"Applications" 3
"Documentation" 4
Familiarize yourself with the product.
"The Instrument" 6
"Menu and Functions" 10
"Installation" 16
"Operation" 17

6
Gossen Metrawatt GmbH SECULIFE SP BASE
5 The Instrument
5.1 Scope of Delivery
Please check for completeness.
1 SECULIFE SP BASE, product no. M696B
1 USB plug-in power supply unit incl. international adapters US/Japan, Euro,
GB, China/AUS/NZL
1 Cable micro-USB to USB-A
1 Cable micro-USB to PS2
2 LR6 AA batteries
1 Carrying case
1 Information CD
1 Factory calibration certificate

7
SECULIFE SP BASE Gossen Metrawatt GmbH
5.2 Instrument Overview
Figure 2: Front Panel
5.2.1 Symbols on the Instrument and the included Accessories:
Warning concerning a point of danger
(attention, observe documentation!)
European conformity marking
The device may not be disposed of with household trash "Returns
and Environmentally Sound Disposal" 25.
Light transmitter
Finger
Color touchscreen
Key on/off
Bottom:
Micro USB slot
Back:
Battery
compartment
Back:
Receiver
!

8
Gossen Metrawatt GmbH SECULIFE SP BASE
5.3 Technical Data
Power
Supply
Battery LR6 AA alkaline battery
Voltage: 1.5 V
Capacity: 2900 mAh
Battery life: 10 h (full backlight)
USB plug-in power
supply unit
Input:
100 ... 240 V AC, 50/60 Hz, 0.18 A
Output:
5VDC,1A
Ambient
Conditions
Operating temperature: +15 … +40 °C
Storage temperature: −20 … +60 °C
Relative humidity: max. 80 %, no condensation allowed
Elevation: Not specified
Electro-
magnetic
Compatibil-
ity (EMC)
Interference emission: EN 61325-1:2013 class A
Interference immunity: EN 61325-1:2013
Mechanical
Design
Protection: Not specified (IPX0)
Housing (W × H × D): Approx. 6.35 x 9.53 x 2.54 cm
(without finger)
Weight: Approx. 170 g (with batteries)
Display: Color touchscreen, 6 cm
Interfaces Micro USB
Simulation
Parameters
SpO280, 85, 90, 95, 98, 99 %
SpO2Accuracy 2 %
Heart Rate 30, 40, 60, 80, 100, 120, 140, 150,
245 BPM
Rate Accuracy 1 %
Perfusion index 0.2 %, 2.0 %, 10 %
Simulation
Modes
Manual mode
5 factory presets
Supported
manufac-
turer Type
Nellcor, Masimo, Nonin, Nihon Kohden, Mindray, GE-Ohmeda,
Philips/HP, BCI

9
SECULIFE SP BASE Gossen Metrawatt GmbH
5.4 Applied Standards
The instrument has been manufactured and tested in accordance with the follow-
ing safety regulations:
5.5 Combined Products
2014/30/EU Electromagnetic compatibility (EMC)
2014/35/EU Low voltage (LVD)
IEC 61 010-1
EN 61 010-1
VDE 0411-1
Safety requirements for electrical equipment for measure-
ment, control and laboratory use –
General requirements
DIN EN 61 326-1
VDE 0843-20-1
Electrical equipment for measurement, control and laboratory
use – EMC requirements –
Part 1: General requirements
EN 60 529
VDE 0470 Teil 1
Test instruments and test procedures
Degrees of protection provided by enclosures (IP code)
EN 61000-3-2 Electromagnetic compatibility (EMC). Part 3-2: Limits. Limits
for harmonic current emissions (equipment input current =16
A per phase)
EN 61000-3-3 Electromagnetic compatibility (EMC) - Part 3-3: Limits - Lim-
itation of voltage changes, voltage fluctuations and flicker in
public low-voltage supply systems, for equipment with rated
current = 16 A per phase and not subject to conditional con-
nection
Product Item no.
SECULIFE PS300 M695N

10
Gossen Metrawatt GmbH SECULIFE SP BASE
6 Menu and Functions
The instrument is operated via touchscreen.
• Touch a button to open a menu.
• Touch a button repeatedly to advance through a list of options.
• Slide your finger up or down to scroll a scrollable list.
6.1 User Interface
6.1.1 On-screen Indicators
6.1.2 On-screen Icons and Buttons
Indicator Description
Battery life indicator
– Green: Battery level ok
– Red: Battery level below 20 %
Input light intensity meters
Icon/Button Description
Mode selection
Selection area
Selection field %SpO2
Selection field BPM
Selection field %Perfusion
Selection field Monitor

11
SECULIFE SP BASE Gossen Metrawatt GmbH
Enter setup menu
Scroll through a list of options
Back to previous menu level
Increase a value
Decrease a value
Exit without saving
Save new settings
Control bar
Function on/off
Icon/Button Description

12
Gossen Metrawatt GmbH SECULIFE SP BASE
6.2 Manual Mode
In manual mode, the user can select individual combinations of SpO2%, heart rate,
and perfusion index.
Figure 1: Manual mode
Selectable parameters in manual mode
Parameters Options
SpO2
80 %
85 %
90 %
95 %
98 %
99 %
Monitors
Nellcor
Nonin
Nihon Kohden
Masimo
Mindray
GE-Ohmeda
Philips/HP
BCI

13
SECULIFE SP BASE Gossen Metrawatt GmbH
6.3 Preset Mode
In preset mode, the user can advance through 5 preset combinations of SpO2%,
heart rate, and perfusion index. These combinations are factory set and not cus-
tomizable by the user.
Figure 2: Preset mode
Preset combinations
Heart rate
30 BPM
40 BPM
60 BPM
80 BPM
100 BPM
120 BPM
140 BPM
150 BPM
245 BPM
Perfusion
0.2 %
2%
10 %
SpO2Heart Rate [BPM] Perfusion Index
85 % 80 High (2 % perfusion)
95 % 40 High (2 % perfusion)
98 % 80 High (2 % perfusion)
Parameters Options
Scroll preset com-
binations
Scroll monitors

14
Gossen Metrawatt GmbH SECULIFE SP BASE
6.4 Setup Menu
The setup menu is a scrollable list of user configurable options. The setup screen
can be accessed at any time by pressing Setup.
Figure 3: Setup menu
Configuration options
98 % 140 High (2 % perfusion)
99 % 80 Low (0.2 % perfusion)
Parameter Description Range
Display Brightness Adjusts the brightness
of the screen
5 %–100 % in 5 % increments
Sounds Adjusts the volume of
all system sounds
0 %–100 % in 5 % increments
Auto Off Timer
Adjusts the timeout
that will turn the instru-
ment off after a period
of inactivity (only appli-
cable when the instru-
ment is running on bat-
tery power)
5–100 minutes in 5 minute in-
crements
Turning this setting off will dis-
able the auto off feature.
SpO2Heart Rate [BPM] Perfusion Index

15
SECULIFE SP BASE Gossen Metrawatt GmbH
Auto Dim Timer
Adjust the timeout that
will turn the display
brightness down to
15 % after a period of
inactivity
5–100 minutes in 5 minute in-
crements
Turning this setting off will dis-
able the auto dim feature.
Custom User Offset
Adjusts the derived R-
curve value for SpO2
setting to a user's pref-
erence
+/- 100 adjustment increments
Instrument
Information
Displays the current version of the software and any
other applicable information
Update Firmware Erases the firmware from the instrument in order for the
user to update to a newer version of the firmware
Parameter Description Range

16
Gossen Metrawatt GmbH SECULIFE SP BASE
7 Installation
7.1 Unpacking the Instrument
1. Carefully remove instrument and accessories from the packaging.
2. Check delivery for completeness and possible damage.
3. In case of detected damages, hidden defects and short deliveries, document
type and scope and contact the manufacturer or supplier immediately.
4. Keep packing material for further transport.
7.2 Power Supply
The instrument can be powered by battery, via AC adapter or via a connected
monitor.
7.2.1 Establishing Power Supply via AC Adapter
1. Connect the USB cable to the Micro USB plug of the instrument.
2. Connect the other side of the USB cable to the USB-A plug of the AC adapter.
3. Connect the AC adapter to the mains socket.
7.2.2 Establishing Power Supply via connected Monitor
1. Connect the sync cable to the Micro USB plug of the instrument.
2. Connect the other side of the sync cable to the auxiliary port of a compatible
monitor.
7.2.3 Establishing Power Supply by Batteries
1. Press the two arrows on the battery compartment lid on the back of the instru-
ment.
2. Slide the battery compartment lid down.
3. Insert two AA batteries into the battery tray. Ensure the correct polarity.
4. Place the battery compartment lid onto the battery tray.
5. Push the battery compartment lid gently upwards until it audibly clicks into
place.
Note!
Decrease screen brightness and audio volume to preserve battery life.

17
SECULIFE SP BASE Gossen Metrawatt GmbH
8Operation
Attention!
!
All connections to patients must be removed before connecting the DUT to
the SECULIFE SP BASE. A serious hazard may occur if the patient is con-
nected when testing with the instrument.
8.1 Switching on/off
8.1.1 Switching on the Instrument
1. Press the key on/off.
The splash screen appears for a few seconds, then the main screen is dis-
played. The instrument is ready for operation.
8.1.2 Switching off the Instrument
1. Press the key on/off.
The instrument is switched off.
8.2 Connecting the Instrument to a DUT
8.2.1 Applying Finger Cuff
1. Attach the finger cuff to the top of the instrument placing the cuff LEDs on bot-
tom and the cuff sensor on top.
The the sensor's light pulses enter the back of the finger portion of the instru-
ment.
This manual suits for next models
1