WRIST BLOOD PRESSURE
MONITOR MANUAL
Email: info@greatergoods.com
Phone: (866) 991-8494
VIDEO TUTORIAL:
greatergoods.com/0605
Model 0605

1
TABLE OF CONTENTS
MONITOR COMPONENTS / INSTALLING BATTERIES p. 3-6
SETTING TIME, DATE, AND MEASUREMENT UNIT p. 7-8
TAKING MEASUREMENTS p. 9-10
MEMORY / MANAGING RECORDS p. 11-12
TIPS / MAINTENANCE p. 13-14
ABOUT BLOOD PRESSURE p. 15-16
IRREGULAR HEARTBEAT DETECTOR / BATTERY CAUTIONS p. 17-18
TROUBLESHOOTING p. 19-20
DEVICE SPECIFICATIONS p. 21-22
COMPLIED STANDARDS LIST, FCC STATEMENT, EMC GUIDANCE p. 23-26
SAFETY INFORMATION / INDICATIONS FOR USE p. 27-32
WARRANTY p. 33

2
OUR PROMISE
We’re committed to creating 5-star products.
If we haven’t delivered on our promise, please contact us.
For the best possible experience with your product,
please visit greatergoods.com/0605.
WARNING: No modication of this equipment is allowed.

3
MONITOR COMPONENTS
LCD DISPLAY SYSTOLIC
DIASTOLICGRADE
TIME
ARROW / MEMORY
BUTTON
SET / POWER
BUTTON
PULSE RATE
NOTE: While the monitor only has two buttons, each button has two functions.
Throughout the manual, you will see the buttons referred to by the function
being described in that step (e.g. Press and hold the “POWER” button to turn
on, then press the “SET”button to select your user and begin measuring).

4
Systolic blood pressure High pressure result
Diastolic blood pressure Low pressure result
kPa (kiloPascal) Measurement unit of blood pressure
mmHg
(millimeters of mercury) Measurement unit of blood pressure
Pulse per minute Beats per minute (BPM)
Irregular Heartbeat
Detection Irregular heartbeat
Blood Pressure Grade Displays grade of blood pressure
Low batteries Batteries are low and need to be replaced
Current time Month / Day / Year – Hour / Minute
User ID Start measurement and save the results
for USER 1 or USER 2
Shaking alert Shaking will result in inaccurate measurements

5
MONITOR COMPONENTS (cont.)
BATTERY
COMPARTMENT
CUFF
(Type BF applied part)
Package Contents:
1. Wrist Blood
Pressure Monitor
2. 2×AAA batteries
3. User manual

6
-
+
+
-
INSTALLING AND REPLACING THE BATTERIES
• Slide o the battery cover.
• Install the batteries by matching the
correct polarity, as shown here. Always use
the correct battery type (2 x AAA batteries).
• Replace the cover.
Replace the batteries whenever the
following happen:
appears.•
The display dims.•
The display does not light up.
•

7
SETTING TIME, DATE, AND MEASUREMENT UNIT
It is important to set the clock before using your blood pressure monitor,
so that a time stamp can be assigned to each record that is stored in the memory.
(year: 2014 - 2054 , time format: 24 H /12 H)
1. When the monitor is o, press and hold
the “POWER” button for about 3 seconds
to enter into setup mode. The blinking
numeral represents YEAR.
2. Press the “ARROW” button to change
the YEAR. Each press will increase
the numeral by 1 in a cycling manner.
Then, press the “SET” button to conrm
your selection and it will turn to the
next step.

8
3. Use the “ARROW” and “SET”buttons like you did in the previous step to
conrm MONTH, DAY, TIME FORMAT [24 hours or 12 hours], HOURS,
MINUTES, and the MEASUREMENT UNIT.
IMPORTANT!
Most doctors in the United States use mmHg (millimeters of mercury)
as the measurement unit.
4. After conrming the measurement unit, the LCD will display all the
settings you have done once again and then turn o.

9
FASTEN THE CUFF
1. Remove all accessories (watch, bracelet, etc.) from your wrist. If your
physician has diagnosed you with poor circulation in one wrist,
use the other one.
2. Roll or push up your sleeve to expose the skin.
3. Apply the cu to your wrist with your palm facing up.
4. Position the edge of the cu about 1.5 cm away from your palm.
5. Fasten the wrist cu around your wrist, leaving
no extra room between the cu and your skin.
If the cu is too loose, the measurement will not
be accurate. Sit upright comfortably and relax as
much as possible with your legs uncrossed.
Make sure your arm is supported and the wrist
monitor is level with the right atrium of the heart.

10
START THE MEASUREMENT
1. When the blood pressure monitor is o, press the “POWER” button to turn
it on. The LCD will display the user ID rst. Press the “ARROW” button to change
the user ID between User 1 and User 2. Press the “SET” button to conrm,
and it will nish the whole measurement for the selected user.
2. Press the “POWER” button to power o, otherwise the device will turn o
within 1 minute.
LCD display Adjust to zero Inating &
measuring
Display and save
the result.

11
RECALLING RECORDS
1. When the monitor is o, press the “MEMORY” button to show the user
selection screen. Use the “ARROW” button to switch between users. Use the
“SET” button to select the user you want to view. This will display the average
of the latest three measurements.
2. Press “ARROW” again. It will display the latest measurement of result, date,
and time. Press the “ARROW” button again, and it will display the next record,
and so on. During the process of displaying the results, if there is no
operation, the blood pressure monitor will turn o after one minute.
Or you can press the “POWER” button to turn it o.
NOTE: The record number, date, and time will display alternately.
NOTE: For a meaningful comparison, try to measure under similar
conditions. For example, take daily measurements at approximately
the same time, on the same wrist, or as directed by a physician.

12
DELETE THE RECORDS
You can delete all results of the selected user by following the steps below:
1. When the monitor is o, press the “MEMORY” button to show the user
selection screen. Use the “ARROW” button to switch between users.
Use the “SET” button to select the user you want to view.
2. Press and hold the “SET” button for 3 seconds, the display will show “dEL”
3. Press “ARROW” to toggle between YES and NO. Press “SET” to conrm the
selection. If YES is selected, the unit will delete the records and turn o.
4. If there is no record,
this display will show:

13
MAINTENANCE
• Store monitor in a dry place.
• Avoid contact with water; clean it with a dry cloth if needed.
• Do not attempt to clean the cu with water.
• Never submerge monitor or cu in water.
TIPS FOR MEASUREMENT
It can cause inaccuracy if the measurement is taken under
the following circumstances:
• Measuring within 1 hour of eating or drinking.
• Measuring after tea, coee, or smoking.
• Measuring within 20 minutes after taking a bath.
• Measuring when talking or moving your ngers.
• Measuring in a very cold environment.
• Measuring when needing to use the restroom.

14
STANDARD CLASSIFICATION

15
ABOUT BLOOD PRESSURE
Q: What is the dierence between systolic pressure and diastolic pressure?
A: When ventricles contract and pump blood out of the heart, the blood
pressure reaches its maximum value in the cycle, which is called systolic
pressure. When the ventricles relax, the blood pressure reaches its
minimum value in the cycle, which is called diastolic pressure.
Q: Does it matter which wrist I use
when measuring?
A: It is acceptable to use either wrist
when measuring. However, to obtain
the most consistent readings possible,
it is suggested that you use the same
wrist each time you measure your
blood pressure.

16
ABOUT BLOOD PRESSURE (cont.)
Q: Why does my blood pressure uctuate throughout the day?
A: Individual blood pressure varies throughout the day. It is also aected
by the way you fasten your cu and your measurement position. Please
make sure you measure your blood pressure under the same conditions
each time you take a measurement. Variations in pressure can also
occur due to certain medicines being taken. Waiting at least 3 minutes
between measurements is recommended.
Q: Why is the blood pressure I get from the hospital dierent from home?
A: Your blood pressure can vary throughout the day due to many factors such
as weather, your emotions, exercise, etc.

17
IRREGULAR HEARTBEAT DETECTOR
This Blood Pressure Monitor is equipped with an intelligent function of
Irregular Heartbeat Detector (IHB). During each measurement, this equipment
records the heartbeat intervals and works out the standard deviation. If the
calculated value is larger than or equal to 15, this equipment will light up the
IHB symbol on the screen when displaying the results.
NOTE: The appearance of the IHB icon indicates that a pulse irregularity
consistent with an irregular heartbeat was detected during measurement.
Usually this is NOT a cause for concern. However, if the symbol appears often,
we recommend you seek medical advice. Please note that the device does not
replace a cardiac examination, but serves to detect pulse irregularities at an
early stage.

18
BATTERY CAUTIONS
Remove batteries if the device is not likely to be used for some time.
The old batteries are harmful to the environment; do not dispose with other
daily trash. Remove the old batteries from the device and follow your local
recycling guidelines. Do not dispose of batteries in re. Batteries may
explode or leak.

19
TROUBLESHOOTING
Display will not
light up.
Batteries are exhausted
Batteries are inserted
incorrectly
Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Display is dim or
icon is showing
Batteries are low
E1 shows on screen The ination is too low Refasten the cu and then
measure again
E2 shows on screen The cu is not secure Refasten the cu and then
measure again
Refasten the cu and then
measure again
Can’t inate the air
during the process of
measurement
E3 shows on screen
E4 shows on screen The cu is too tight Refasten the cu and then
measure again
Table of contents