G3610/G3611 Gas Sampling Boards for N2Generators
Version 1.0 –August 2019 9
3.3 Safety Aspects
Hot sensor
The sensor is hot and can cause severe burning of personnel if not handled with
care.
Analyzer
Before working with the analyzer, please read the oxygen analyzer manual in its
entirety.
Installation and operation
It should be pointed out that installation and operation of the sampling board for
nitrogen generators and associated equipment must be carried out by skilled,
trained, and certified personnel, and that Green Instruments A/S does not take any
responsibility of the operation of the system and associated equipment whatsoev-
er.
The successful and safe operation of this equipment is dependent upon proper
handling, installation, operation, and maintenance.
Recycling
Do not dispose any part of the sampling board with regular refuse. Disposal
should be in accordance with the requirements of the current statutory regulations.
3.4 Sampling Board
The sampling boards are designed for mounting with the G36pOxygen Analyzer
panel mounted version or the G36aOxygen Analyzer.
3.4.1 Mounting Panel
The mounting panel has two angle iron mounting brackets. The brackets are made
of ordinary mild steel and can be welded or bolted directly to a chosen structure as
required. The two mounting brackets are mounted horizontal and parallel with a
distance of 550 mm. See Figure 3-3 for the dimensions and layout of the narrow
board and Figure 3-4 for the wide sampling board.
The sampling board needs to be installed as close as possible to the nitrogen gas
supply line in order to reduce dead time of test gas flowing through the piping
(see section 3.2). When the sampling board is installed in an appropriate location,
the sample gas, calibration gas, and power are connected as described below.