T20.0.0X.6C-07 page 4 of 4
Configuration of the device
In the configuration following parameters can be changed:
- Display unit of temperature measuring
- Scaling of the humidity output and of the temperature output
- Adjusting of humidity and temperature display (offset and scale correction)
By means of the scaling the analogue signal outputs can be adopted to Your needs.
The adjusting by means of offset and scale is intended to be used to compensate
errors of the measurings. It is recommended to keep the scale correction deactivated
(“oFF”). The display value is given by following formula:
Display = measured value - offset
With a scale correction (just for calibration laboratories, etc) the formula changes:
Display = (measured value - offset) * ( 1 + scale adjustment/100)
To configure the parameters proceed like follows:
1. Press the key 1 (SET) for more than 4 sec’s until 'unit' with temperature arrow
appears in the display.
I.) 'unit' with Temp-arrow: Temperature unit
Enter the desired Temperature unit. All referring settings and displays are done in this unit.
2. Choose the desired value by pressing 2 (down) or 3 (up) key. Choice between °C and °F (ex works: °C)
3. Enter by key 1 (SET), select next parameter by pressing key 1 again: 'Out.0' without temperature arrow appears.
II.) 'Out.0': Display at zero output of humidity measuring (output scaling)
Enter the humidity value at which the output should have 4mA (respectively 0V).
4. Choose the desired value by pressing 2 (down) or 3 (up) key. Max. input range: 0.0...100.0 % (ex works: 0.0%)
5. Enter by pressing key 1 (SET), select next parameter: 'Out.1' without temperature arrow appears in the display.
III.) 'Out.1': Display at maximum output for humidity measuring (output scaling)
Enter the humidity value at which the output should have 20mA (respectively 10V).
6. Choose the desired value by pressing 2 (down) or 3 (up) key. Max. input range: 0.0...100.0 % (ex works: 100.0%)
7. Enter by pressing key 1 (SET) , select next parameter: 'Out.0' with temperature arrow appears in the display.
IV.) 'Out.0' with Temp-arrow: Display at zero output of temperature measuring (output scaling)
Enter the temperature value at which the output should have 4mA (respectively 0V).
8. Choose the desired value by pressing 2 (down) or 3 (up) key. (ex works: 0.0°C / 32.0°F)
9. Enter by pressing key 1 (SET), select next parameter: 'Out.1' with temperature arrow appears in the display.
V.) 'Out.1' with Temp-arrow: Display at maximum output for temperature measuring (output scaling)
Enter the temperature value at which the output should have 20mA (respectively 10V).
10.Choose the desired value by pressing 2 (down) or 3 (up) key. (ex works: 100.0°C / 212.0°F)
11.Enter by pressing key 1 (SET), select next parameter: 'OFFS' without temperature arrow appears in the display.
VI.) 'OFFS': Offset of humidity measuring (correction of measuring deviations):
The offset of the measuring will be shifted by this value, the input is in %RH. Calculation: see above.
12.Choose the desired value by pressing 2 (down) or 3 (up) key.
Max. input range: -5.0...5.0 %RH or 'oFF': offset is deactivated (=0.0%, ex works)
13.Enter by pressing key 1 (SET), select next parameter: 'SCAL' without temperature arrow appears in the display.
VII.) 'SCAL': Scale of humidity measuring (correction of measuring deviations):
The scale of the measuring is changed by this value. Calculation: see above.
14.Choose the desired value by pressing 2 (down) or 3 (up) key.
Max. input range: -5.00...5.00 or 'oFF': scale is deactivated (=0.00, ex works)
15.Enter by pressing key 1 (SET), select next parameter: 'OFFS' with temperature arrow appears in the display.
VIII.) 'OFFS' with Temp-arrow: Offset of temperature measuring (correction of measuring deviations):
The offset of the measuring will be shifted by this value, the input is in °C. Calculation: see above.
16.Choose the desired value by pressing 2 (down) or 3 (up) key.
Max. input range: -5.0...5.0 °C / -9.0...9.0°F or 'oFF': offset is deactivated (=0.0, ex works)
17.Enter by pressing key 1 (SET), select next parameter: 'SCAL' with temperature arrow appears in the display.
IX.) 'SCAL' with Temp-arrow: Scale of temperature measuring (correction of measuring deviations):
The scale of the measuring is changed by this value. Calculation: see above.
18.Choose the desired value by pressing 2 (down) or 3 (up) key.
Max. input range: -2.00...2.00 or 'oFF': scale is deactivated (=0.00, ex works)
19.Enter by pressing key 1 (SET). After pressing key 1 again, the instrument will restart (segment test).
If the jumper is removed from the
shown contacts, the configuration is
inaccessible, values are protected.
Never connect other contacts!