2. Cable Connections (0-10 output) 3. OBE Module Installation
Wire the 15-pin “D” connector according with the wiring table using
24AWG or other suitable wire.
Using a unipolar DC power supply, set the desired operating voltage
within the range of 11.5 VDC to 30 VDC.
1 ............ Green ................. Setpoint #1 Output (Hi)
2 ............ Blue .................... Setpoint #2 Output (Lo)
3 ............ Red..................... Power Input (9-36 VDC)
4 ............ Black .................. Power Ground
5 ............ White.................. Analog Out (1000 Torr)
6 ............ White/Black ........ Analog Ground
7 ............ Red/White .......... Analog Out (1000 mTorr)
8 ............ Black/White ........ Digital Ground
9 ............ Orange ............... Remote Zero
10 .......... Green/White ....... NC
11 .......... Blue/White.......... NC
12 .......... Red/Black........... NC
13 .......... Orange/Black...... NC
14 .......... Green/Black ....... Rx In (Internal use)
15 .......... Blue/Black .......... Tx Out (Internal use)
Pin No. Color
(suggested) Description
Connect the OBE module to the HPM-2002 sensor tube.
Note that the connector is keyed.
A finger tight connection is all that is required for adequate operation.