1. Installation
1.1 Feature
1.Water cooling for ballasts installed inside the cover hat of chamber.
2.Water right rate: IP56
3.Outdoor and indoor installation available
4.Start up and off according the signal from flow switch.
5. Easily installation and simple maintenance.
1.2 Installation of UV sterilizer
The UV sterilizer should be placed Vertically. Filtration should be applied before the UV Sterilizer
system to provide high quality pretreated water for UV disinfection. Procedure for UV Sterilizer
installation is as follow:
(1) Unloose and take off the nut(s) on the end(s) of chamber.
(2) Insert the quartz sleeve into the chamber carefully, then cover the O-ring on the end of the quartz
sleeve.
(3) Screw the nut on the end of chamber and tighten it. Note: Do not use force implement to avoid
breaking up the quartz sleeve.
(4) Check all the junctions and open the tap to see if leaks happen.
(5) Please install water flow controller before UV sterilizer if the water flow is too fast.
(6) Once it is determined that there are no leaks, connect the lamp with ballast , then insert into quartz
sleeve. Screw the waterproof connector into the nut, and tighten it.
(7) Open the water valve first, then power on the unit..
(8) Power on and to check whether the lamp works well or not. Warning: Do never look at the UV
Lamps with naked eyes when the UV lamp is power on.
1.3 Operation
1.Open the Inlet and Outlet valve.
2.Plug the power, and the unit will be powered on automatically.
When the water level can not reach the set level, the alarm will be triggered.
3.When the ballast and lamp fail, the alarm will be triggered.
4.Power off the unit, closed the Inlet and Outlet valve, then unplug the power.