Hygiena SystemSURE Plus v2 User manual

I
This Manual Copyright © 2012 by Hygiena LLC
All rights reserved. No part of this manual may be reproduced,
stored in a retrieval system, or transmitted, by any means or in any
form, without the prior permission of Hygiena LLC.
The information contained in this manual is subject to change
without notice. Hygiena LLC assumes no responsibility for any errors
that may appear in this document.
This manual will describe the specifications and general operations
of the SystemSURE Plus v2 and EnSURE v2 luminometers.
For older luminometers refer to the operators manual that shipped
with the luminometer for specific operations.
Document: SystemSURE Operators Manual V3.0. 2013
Copyrights and Trademarks
Worldwide Patents Pending.
Hygiena, EnSURE, SystemSURE Plus, SystemSURE II,
Ultrasnap, SureTrend and dataSURE II are trademarks of
Hygiena LLC.
Operating Precautions and Limitations of Use
The following symbol is used in this manual:
Description: CAUTION / WARNING
IMPORTANT
Hygiena’s products are designed and constructed to be safe
and without risk to health when properly used in accordance
with the supplied documentation and when the operating
precautions outlined in this manual are fully observed.
It is essential that the user of this manual is aware of the
potential hazards associated with the unit and its accessories
All operators should be familiar with the safety precautions
and warnings given in this manual before attempting to
operate the unit.
If the unit is used in a manner which is not specified by the
manufacturer, the protection provided by the equipment may
be impaired.
Please read this manual in its entirety and the “Operating
Precautions and Limitations of Use” at the end of the manual
before using the Luminometer.

II
Table of Contents
1. INTRODUCTION _____________________________________________________ 1
1.1 WHAT IS ATP? _____________________________________________________ 2
1.2 HOW DOES THE LUMINOMETER WORK?____________________________________ 2
1.3 WHAT IS AN RLU? __________________________________________________ 2
1.4 TEST DEVICES AND ACCESSORIES _________________________________________ 3
2. LUMINOMETER DESCRIPTION _________________________________________ 4
2.1 KEYPAD SYMBOLS ___________________________________________________ 4
2.2 INSTALLING BATTERIES ________________________________________________ 5
2.3 TURNING THE LUMINOMETER ON AND OFF __________________________________ 6
2.4 DISPLAY LAYOUT AND ICONS ____________________________________________ 7
2.5 LOW BATTERY INDICATOR______________________________________________ 7
2.6 POWER SAVING STANDBY MODE _________________________________________ 8
2.7 INTERNAL CALIBRATION _______________________________________________ 8
2.8 LOW MEMORY WARNING _____________________________________________ 8
3. SET-UP MENU OPTIONS ______________________________________________ 9
3.1 USER ID_________________________________________________________ 10
3.2 SETTING UP PROGRAM TEST POINTS AND PASS/FAIL LIMITS _____________________ 10
3.3 TEST PLANS ______________________________________________________ 11
3.4 ESTABLISHING ATP PASS/FAIL LIMITS ____________________________________ 12
3.5 VIEWING STORED TEST RESULTS ________________________________________ 12
3.6 VIEWING STATISTICAL DATA AND TEST FAILURES _____________________________ 12
3.7 ERASING THE TEST RESULTS MEMORY ____________________________________ 13
3.8 SETTING THE CLOCK TIME AND DATE _____________________________________ 13
3.9 ADJUSTING THE LCD CONTRAST ________________________________________ 14
3.10 QUICK BOOT ____________________________________________________ 14
3.11 STANDBY MODE __________________________________________________ 14
3.12 BOLD TEXT ______________________________________________________ 15
3.13 SAVE TESTS _____________________________________________________ 15
4. SAMPLE MEASUREMENTS AND TEST RESULTS ___________________________ 17
4.1 TAKING A SAMPLE MEASUREMENT_______________________________________ 18
4.2 PERFORMING A RETEST ______________________________________________ 19
5. INSTALLING SURETREND DATA ANALYSIS SOFTWARE _____________________ 19
5.1 CONDITIONS OF USE ________________________________________________ 19
5.2 HARDWARE AND SOFTWARE REQUIREMENTS________________________________ 20
5.3 STARTING THE INSTALL PROCESS ________________________________________ 20

III
5.4 USING THE SURETREND SOFTWARE ______________________________________ 20
5.5 INSTALLING THE USB SMART CABLE DRIVER ________________________________ 21
5.6 CONNECTING THE LUMINOMETER TO THE PC________________________________ 22
6. OPERATOR MAINTENANCE___________________________________________ 22
6.1 CLEANING THE LUMINOMETER CASE______________________________________ 22
6.2 REPLACING THE BATTERIES ____________________________________________ 22
6.3 CLEANING AND REPLACING THE PROTECTIVE POCKET __________________________ 23
7. TROUBLESHOOTING ________________________________________________ 24
7.1 LUMINOMETER BEEPS _______________________________________________ 24
7.2 TROUBLESHOOTING TIPS _____________________________________________ 24
7.3 LUMINOMETER ERROR CODES __________________________________________ 26
8. UNIT WARRANTY AND RETURNS ______________________________________ 27
8.1 WARRANTY DURATION_______________________________________________ 27
8.2 PARTICULAR EXCLUSION ______________________________________________ 27
9. GLOSSARY OF TERMS AND ABBREVIATIONS _____________________________ 27
10. TECHNICAL SPECIFICATION __________________________________________ 28
11. CAUTIONS AND WARNINGS _________________________________________ 29

1
1. Introduction
The luminometer is part of an ATP Hygiene/Sanitation Monitoring
System intended to detect ATP found in organic matter and
microorganisms. The Hygiena luminometer, in conjunction with
Hygiena’s ATP and Micro-snap test devices, measures levels of
contamination on surfaces, water and product samples.
1. Remove swab from test tube and collect sample. Refer
to instructions included with testing device.
2. Place swab back in test tube.
3. Break Snap Valve at top of swab by bending bulb.
Squeeze bulb twice, pushing liquid reagent down swab
shaft.
4. Shake test for 5 seconds.
5. Place test in the luminometer and close lid.
6. Hold the luminometer upright and press “OK”. Reading
will appear in 15 seconds.

2
1.1 What is ATP?
Adenosine Triphosphate (ATP) is an
organic molecule that is used by living
cells as the main source of energy.
Animal, plant, bacterial, yeast, and mold
cells produce and break down ATP in
order to drive a number of biological
processes. It is the main energy-transfer
molecule in the cell. The presence of ATP on a surface indicates that
the surface has not been adequately cleaned and has the potential
to harbor and support bacterial growth. The residue may also
contain hazardous material including potential allergens. Therefore,
ATP is an ideal indicator of surface cleanliness.
1.2 How Does the Luminometer Work?
The luminometer detects ATP using a chemical reaction.
Luciferin/Luciferase + ATP = Light
A sample is collected using a Hygiena ATP device (see section 1.4).
Surface tests come with a buffer moistened swab bud that aids in
the removal of any biological material (ATP) on either a wet or dry
surface. Once the sample is collected and exposed to the chemistry
(buffer) in the Hygiena device, the ATP contained inside the
collected cells will be released. ATP released from inside these cells,
along with any free ATP picked up from the surface by the swab, is
now available to react with the chemistry in the bulb of the device.
When the Luciferin/Luciferase enzymes are exposed to ATP, light is
produced. The light is detected and measured by the luminometer.
The amount of light detected is directly proportional to the quantity
of ATP collected and accordingly is proportional to the amount of
residue in the tested sample. The higher the reading, the higher the
amount of contamination in the sample.
1.3 What is an RLU?
The luminometer displays results in Relative Light Unit (RLU) values.
The light produced from the Luciferin/Luciferase and ATP reaction in
the swab is emitted in the form of photons. A photon is an
elementary particle and the basic unit of light. The luminometer
detects these photons and displays them directly as RLU values. The
more light detected by the luminometer, the greater the RLU value.
The quantitative RLU reading is then compared against user
programmable thresholds to provide an overall qualitative pass ,
caution or fail result.
The luminometer detects total ATP, not just ATP from bacteria,
yeast, and mold but also the ATP from anything organic in the
sample. Therefore, an RLU value is not the same as a microbial

3
colony forming unit (CFU). Since the luminometer is detecting total
ATP, it is unknown whether the RLU result displayed by the
luminometer is due to the detection of microbial ATP, residual ATP,
or a combination of both. Therefore, a comparison cannot be drawn
between RLU values from ATP and standard plate counts (SPC); or
rather RLU does not equal CFU. True microbial contamination can
only be determined by a standard plate count method that can take
up to 48 hours for results. The luminometer is capable of providing
results in 15 seconds offering a rapid, real-time approach to
hygiene/sanitation monitoring. The luminometer is not meant to
replace microbial testing. It can be an effective tool when
incorporated into a Hazard Analysis Critical Control Point (HACCP),
Good Manufacturing Practices (GMP), or sanitation plan.
1.4 Test Devices and Accessories
Device
Cat No.
Description
UltraSnap
US2020
Surface total ATP test device
SuperSnap
SUS3000
High sensitivity total ATP test suited
for allergen prevention applications
AquaSnap Total
AQ100X
Water total ATP test device
AquaSnap Free
AQ100FX
Water Free ATP test device
Contamination
Control Kit
CCK-4
Cuvette based water testing kit
MicroSnap
Enterobacteriaceae
MS-EB
Enterobacteriaceae test
MicroSnap
Coliform Test
MS-COLIFORM
Coliform test
MicroSnap E. coli
Test
MS-ECOLI
E. coli test
MicroSnap Total
MS-TOTAL
Total Viable Count test
ZymoSnap ALP
ZS
Alkaline Phosphatase test
CrossCheck
CX3000
Acid Phosphatase test
Calibration Kit
PCD4000
Positive and negative rod for
validating luminometer calibration
ATP Positive
Control Kit
CK25
25 vials of freeze-dried ATP to verify
efficacy of ATP test devices
See the Hygiena website for a complete list of test devices,
accessories, and other products. www.hygiena.com

4
2. Luminometer Description
Luminometer components:
2.1 Keypad Symbols
The keypad is arranged with the following buttons:
Power on/off button
Menu select and Exit button
Up/increase button
Start/accept button
Down/decrease button
Program select button
Statistics shortcut button
The functions of the keypad are explained in detail in the following
sections of this manual.
TIP: Pressing and holding down the or button will make it
auto-repeat. The longer the buttons are held down, the faster it
goes.

5
Opening the lid reveals the following internal features:
RS232 connector cover
Protective pocket and
sample insertion port
For details on the RS232/USB connector and protective pocket,
refer to sections 5.6 and 6.1 respectively.
2.2 Installing Batteries
The unit is designed to operate from both non-rechargeable alkaline
batteries and rechargeable Nickel Metal Hydride (NiMH) or Nickel
Cadmium (NiCD) batteries:
Battery
Technology
Nominal
Cell Voltage
Relative
Capacity
Alkaline
1.5V
1.0
NiMH
1.2V
0.6
NiCD
1.2V
0.5
The unit requires two batteries of size AA, LR6 or E91.
TIP: For best results, always use a quality brand of Alkaline battery
and replace as soon as the batteries become low (see section 2.5).
WARNING: Do not mix batteries of different technologies, or
use recharged alkaline batteries as these are prone to overcharging
and leaking, which may cause permanent unit damage and
malfunction.
The batteries are fitted by unclipping the battery cover on the back
of the unit, and inserting both batteries with their positive terminals
pointing towards the top of the unit.

6
CAUTION: Take care not to insert the batteries in the wrong
direction, as this may cause permanent damage to the unit’s
internal electronics.
When the batteries are inserted correctly, the unit will automatically
turn on and enter the clock set-up mode. (see section 3.8).
2.3 Turning the Luminometer On and Off
To turn the unit on, press the button. The unit
will beep once and display the power-up self-
check screen:
NOTE: If the batteries are too low, the unit may
turn on and back off again automatically, or may
not turn on at all. If this happens, replace the
batteries immediately.
The unit will then perform an internal 15 second
calibration (see section 2.7).
NOTE: If the clock is not set, the unit will automatically enter the
time and date set-up mode (see section 3.8) before performing its
calibration.
TIP: It is recommended you synchronize the unit after changing the
batteries before use. This will allow SureTrend software to set the
date, time, and format to the computer’s date, time, and format.
TIP: To activate the backlight features press and hold down the
button. The backlight uses more power and will drain the battery
faster.
To turn the luminometer off, press the button. The luminometer
will beep once and the display will go blank.
NOTE: To avoid accidental turn off, the button is disabled when
the luminometer is performing a sample measurement.

7
2.4 Display Layout and Icons
The liquid crystal display (LCD) has the following
layout. The upper half of the display contains the
status icons and the large RLU result digits.
The following is a list of icons and their definitions.
Low battery warning icon
Lid icon - flashes when the lid needs to be closed, or the
RS232 connector cover needs to be properly tucked under
the lid
Insert icon –arrow flashes when the testing device should
be inserted
Remove icon –arrow flashes when a testing device needs
to be removed
Pass result icon
Caution result icon
Fail result icon
Sample measurement reading in Relative Light Units (RLU)
Retested icon –identifies the failed test result as having
been retested (see section 4.2)
Retest result icon –identifies the Program location or test
result as a retest result
2.5 Low Battery Indicator
The low battery icon indicates the state of the batteries:
Icon
Battery State
Not visible
Good
Visible
Low –replace batteries soon
Flashing
Flat –replace batteries now!
When the batteries are completely low, the luminometer will flash
the icon, beep three times, and automatically turn off.
CAUTION: Never leave low batteries in the luminometer, as
they are prone to leaking/corrosion, which will damage the
luminometer.
TIP: Always store the luminometer in a cool dry place when not in
use because elevated temperatures will shorten the battery life.

8
2.6 Power Saving Standby Mode
If the luminometer is on but has not been used for 19 minutes, it
will beep every 10 seconds for 1 minute, and then turn off. If any
key is pressed during that time the luminometer will reset the
Standby Mode countdown. The Standby Mode settings can be
changed (see section 3.11).
To turn the luminometer back on, simply press the button.
2.7 Internal Calibration
When the luminometer is turned on (see section
2.3), it performs an internal calibration, as the
display counts down from 15 to 0 seconds.
NOTE: During the calibration sequence, there must
not be a testing device in the luminometer and the
lid must remain closed. If the icon is shown with the arrow
flashing, open the lid and remove the testing device from the
luminometer. If the icon is flashing, close the lid. If it is still
flashing, make sure the rubber flap covering the RS232 port is
tucked under the lip and is flat.
TIP: The button can be pressed at any time to access the set-up
menu options (see section 3).
When the calibration is complete, the luminometer is ready to
perform a sample measurement.
For details of how to perform a sample
measurement refer to section 4.
NOTE: When the luminometer is ready to perform a
measurement, with the lid closed and no testing
device present, the luminometer will automatically
perform a recalibration cycle under any of the following
circumstances:
The user presses and holds down the button for 1 second.
The luminometer is used in an environment where the
temperature changes significantly (typically more than +/-5oC).
2.8 Low Memory Warning
The luminometer can store up to 2000 test results
in its internal memory. When the memory is more
than 95% full (i.e. less than 100 tests remaining)
the following warning symbol is displayed.
When the memory becomes completely full, no
additional tests can be performed until the memory is either erased
or uploaded to the PC (see sections 3.6 and 5respectively). Test
can be performed if Save Test is turned off (see section 3.13)

9
3. Set-up Menu Options
With the luminometer turned on, the Set-up Menu can be accessed
by pressing the button.
Use the and buttons to scroll through the menu list, and then
press the button to select the required menu option, or the
button the exit the menu.
The following 11 menu options are available:
User ID
Test Plan
Program
Statistics
Memory
Clock
LCD Contrast
Quick Boot
Standby Mode
Bold Text
Save Tests

10
NOTE: SureTrend software is required for sections 3.1 to 3.3. For
further details, refer to the separate SureTrend User's Manual
located on the SureTrend software CD that is packaged with the
luminometer.
3.1 User ID
The luminometer can store a list of 201 User IDs (User: 0 to 200).
Each test result will be tagged with the currently selected User ID.
Select the User ID menu option, then use the and buttons to
scroll through the list of User IDs, then press the button to select
a User ID, or the button to cancel the selection.
TIP: User:0 is the default selection if no User ID list has been
defined.
3.2 Setting up Program Test Points and Pass/Fail Limits
The luminometer has the ability to store a list of 5001 unique
Program test points (Prog# 0 to 5000), each with its own upper ()
and lower () pass/fail result threshold values.
Program test point location names can only be programmed into the
luminometer through the SureTrend software. Program pass/fail
thresholds can either be set-up on the luminometer, or set using the
SureTrend software. If a location name is entered and threshold set
in SureTrend, and then downloaded to the luminometer, the
threshold values can’t be changed via the luminometer. This is a
security feature and prevents inadvertent changes to the thresholds
after being set in SureTrend.
TIP: Prog:0 has a special function, whereby the threshold values
defined for Prog:0 are used as the initial values when setting-up
the other Prog thresholds. Setting Prog:0 with your typical
threshold values will save time when setting-up the other Programs.
The luminometer comes with pass/fail thresholds of 10 RLU for pass
and 30 RLU for fail. These pass/fail thresholds were determined
based on a comprehensive study of ATP, food residue, and
microorganism samples. To receive the report or more information
on setting up pass/fail thresholds contact Hygiena or visit our
Question and Answer section at www.hygiena.com.
To set the Program result thresholds on the luminometer, first select
the PROGRAM menu option by pressing . Scroll to “PROGRAM”
and press . Then use the and buttons to scroll through the
list of Programs, followed by the button to select the Program to
be changed:

11
Prog:
Next use the and buttons to first change the value of the
upper threshold (), followed by the button; to select the lower
threshold use (). Use the and buttons to change the lower
value, followed by the button to store the new values:
upper threshold ()
lower threshold ()
NOTE: If the selected Program location does not have any
thresholds defined, it must be set-up using the PROGRAM menu
option before the Program can be used.
3.3 Test Plans
The luminometer can also store up to 100 individual Test Plans.
Each Test Plan can contain a list of up to 251 test locations,
specified in the order in which they would normally be tested.
Having selected the TEST PLAN menu option, use the and
buttons to scroll through the list of Test Plans, and then press the
button to select the new Plan, or the button to cancel the
selection.
TIP: Plan:0 allows any of the 5001 test locations to be tested in
any order, and is the default selection if no Test Plans have been
defined or selected.
An example set of Test Plans for a typical week might be:
Plan: 1
Monday
Plan: 2
Tuesday
Plan: 3
Wednesday
Plan: 4
Thursday
Plan: 5
Friday
Prog:1
Prog:1
Prog:2
Prog:3
Prog:4
Prog:2
Prog:7
Prog:8
Prog:9
Prog:5
Prog:3
Prog:6
Prog:5
Prog:4
Prog:6
Prog:7
Prog:10
Prog:9
Prog:8
Prog:8
Prog:9
Prog:7
Prog:6
Prog:3
Prog:5
Prog:2
Prog:4
Prog:1
Prog:10

12
Test Plans are defined by using the SureTrend Windows software.
Please refer to the SureTrend User’s Manual located on the
SureTrend software CD for full details (see section 5).
To select a Test Plan, use the TEST PLAN menu
option.
NOTE: If Test Plan Plan:0 is selected, or no Test
Plans have been defined, any of the 5001 Program
locations can be selected.
To select a Program location press the button,
then use the and buttons to change the Program Prog: to the
desired Program location, followed by the button to accept the
new location, or the button to cancel the selection.
TIP: When selecting a Prog:, the display also shows the time and
date that the Program location was last tested.
3.4 Establishing ATP Pass/Fail Limits
For instructions on establishing Pass/Fail limits, refer to the
implementation guide included with your system, or visit the
resources available on www.hygiena.com.
3.5 Viewing Stored Test Results
The luminometer can store a maximum of 2000
test results in its internal memory, even when the
batteries are low or removed.
To view these stored test results, select the
MEMORY menu option by pressing the button.
Scroll to “MEMORY” and press . Then use the and buttons to
scroll through the test results, pressing the button to exit:
TIP: After reviewing the stored test results, a failed Program
location can be selected for retesting by pressing the button (see
section 4.2).
3.6 Viewing Statistical Data and Test Failures
The test results database can be analyzed by
selecting the STATISTICS menu option by pressing
the button. Scroll to “STATISTICS” and press .
This will display a simple comparison of all the
stored results. The button also provides a direct
shortcut to this function.
The and buttons can then be used to scroll through all of the
failed test results. Press the button to exit.

13
3.7 Erasing the Test Results Memory
WARNING: Once the test results have been erased
from memory they are permanently deleted and
can no longer be viewed or uploaded to SureTrend.
NOTE: Once started, the erase function cannot be
stopped, and will take about 30 seconds to erase a
full 2000 results.
The entire test results memory can be cleared using the MEMORY
ERASE function, by first selecting the , then using the up or down
arrows to select the MEMORY menu option, then press to select,
and then pressing and holding down the button for 2 seconds.
The display will show the total number of stored results to be
erased. To accept and start the erase function, press and hold down
the button for 1 second, or press any other button to exit the
option.
3.8 Setting the Clock Time and Date
To set or change the clock time, date, and format, select the
CLOCK menu option by pressing the button. Scroll to “CLOCK”
and press . Then use the and buttons to change the time and
date value, followed by the button to accept each new value.
First the time is set (style hours minutes), which can be
configured as either a 12-hour clock (select AM or PM) or a 24-hour
clock (select 24H):
AM, PM or 24H
hours
minutes
Next set the date format, which can be configured for either
European format, select D/M/Y for day, month, year, or American
format, select M/D/Y for month, day, year):
D/M/Y or M/D/Y
day (D/M/Y) or month (M/D/Y)
month (D/M/Y) or day (M/D/Y)
year
TIP: Pressing the button at any point will exit the clock set-up
mode, leaving the time and date unchanged.
NOTE: The clock does not have automatic daylight saving
adjustment. If this is required, the time must be manually changed
when necessary.

14
3.9 Adjusting the LCD Contrast
The contrast of the LCD screen is factory set at its optimum level for
normal operating conditions. However, in extremes of temperature
the display may appear too dark or light.
This setting can be manually adjusted by selecting the LCD
CONTRAST menu option by pressing the button. Scroll to
“CONTRAST” and press . Then use the or button to increase
or decrease the contrast level, followed by the button to store the
new setting:
darker (+1 to +5)
Normal (0)
lighter (-1 to -5)
3.10 Quick Boot
The luminometer will turn on when you press the button. The
luminometer will perform a standard calibration in 15 seconds and
will then be ready to perform readings. Quick Boot can be turned off
and the luminometer will perform a more in-depth system check the
next time it is turned on. The more in-depth system check takes 60
seconds.
NOTE: The more in-depth system check is typically used for
diagnostic purposes and not necessary for normal use.
To change the Quick Boot setting, select the QUICK BOOT menu
option by pressing the button. Scroll to “QUICK BOOT” and press
button to switch between ON and OFF:
OFF
ON
Exit
3.11 Standby Mode
The Standby Mode is used to control when the luminometer should
automatically turn off when not in use. This is designed to conserve
battery power.
The default is 20 minutes. If no buttons are pressed within the time
period, the luminometer will turn off. If any button is pressed, or a
reading is performed, the Standby Mode countdown timer is reset to
zero.

15
If the bell icon is enabled the luminometer will begin beeping
every 10 seconds for 1 minute as a warning before turning off. For
example, if the Standby Mode time is set to 20, then if the
luminometer is not used for 19 minutes it will start to beep every 10
seconds for 1 minute. If no action is taken the luminometer will turn
off. If any button except the power button is pressed the countdown
will restart.
To change the Standby Mode setting, select the STANDBY MODE
menu option by pressing the button. Scroll to “STANDBY MODE”
and press button. Then use the or button to increase or
decrease the Standby Mode duration, followed by the button to
store the new setting. Standby Mode duration can be from 5 to 60
minutes in increments of 5 minutes. Then use the or button to
enable or disable the warning beeps. This will toggle the bell icon
from enabled to disabled . Press the button to accept the
settings:
increase or decrease duration (5to 60)
set duration
enable or disable warning beeps
set warning state
3.12 Bold Text
The luminometer displays key information on the LCD in bold to
improve legibility. In some cases the number and size of the text
will exceed the allotted space on the LCD. If bold is set to OFF all
text on the luminometer will be displayed normally. This will allow
all text to be displayed even if the text is maximum and all wide
characters.
To change the Bold Text setting, select the BOLD TEXT menu
option by pressing the button. Scroll to “BOLD TEXT” and press
button to switch between ON and OFF:
OFF
ON
Exit
3.13 Save Tests
By default any test performed on the luminometer is recorded to
the luminometer’s non-volatile memory. Even if you take the
batteries out the results of the test are retained in the
luminometer’s memory.

16
In some cases, such as training, calibration checks, etc. users may
want to discard test results. If Save Test is set to OFF, all tests
taken in this mode will not be recorded.
NOTE: The Save Tests mode will be set back to ON
when the luminometer is turned off and back on
again. When Save Tests mode is set to OFF
“RESULTS NOT SAVED”will appear at the bottom of
the LCD
To change the Save Tests setting, select the SAVE TESTS menu
option by pressing the button. Scroll to “SAVE TESTS” and press
button to switch between ON and OFF:
OFF
ON
Exit

17
4. Sample Measurements and Test Results
With the luminometer turned on, and having completed its internal
calibration self-checks, it is ready to perform a new sample
measurement, and the display shows the following details:
Various keypad options are now available, as detailed in each of the
following sections:
Button
Action
Section
Access set-up menu options
3
Select different User ID
3.1
Select program number
3.2
Select different Test Plan
3.3
Perform calibration
2.7
Start new measurement
4.1
View previous test results
3.5
Show statistical result data
3.6
Turn luminometer off
2.3
This manual suits for next models
1
Table of contents