Icon Process Controls ProCon C300 Series User manual

C300 Series Controller
Conductivity |TDS |Resistivity |Salinity
Operating Manual

C300 Series Operating Manual
2◀◀
|
Contents
Preface .........................................................................................................................................................................................................................................3
Features...........................................................................................................................................................................................................................................3
Technical Specications ...................................................................................................................................................................................................... 4
Typical Application.................................................................................................................................................................................................................. 4
Instrument Connection........................................................................................................................................................................................................5
Dimension......................................................................................................................................................................................................................................5
Instrument Installation .........................................................................................................................................................................................................6
Keypad Descriptions...............................................................................................................................................................................................................7
Display Descriptions ...............................................................................................................................................................................................................8
Menu Structure ..........................................................................................................................................................................................................................9
Measurement Settings (Quick Start).........................................................................................................................................................................12
Calibration Settings (Quick Start) ................................................................................................................................................................................ 13
Alarm Settings (Quick Start)............................................................................................................................................................................................14
Current Settings (Quick Start)........................................................................................................................................................................................ 15
Communication Settings (Quick Start) ...................................................................................................................................................................16
System Settings (Quick Start).........................................................................................................................................................................................17
Maintenance Settings (Quick Start)...........................................................................................................................................................................18
Measurement Setup.............................................................................................................................................................................................................19
pH Buffer Reading vs Temperature...........................................................................................................................................................................20
Calibration Setup....................................................................................................................................................................................................................20
Alarm Setup...............................................................................................................................................................................................................................24
Current Setup ...........................................................................................................................................................................................................................24
System Setup............................................................................................................................................................................................................................26
Maintenance..............................................................................................................................................................................................................................26
Troubleshooting......................................................................................................................................................................................................................27
Warranty.......................................................................................................................................................................................................................................27

C300 Series Operating Manual
3◀◀
|
Preface
Please read this manual carefully before use.
When receiving the instrument, please open the package carefully, check whether the instrument and accessories are damaged by trans-
portation and whether the accessories are complete. If any abnormalities are found, please contact our after-sales service department or
regional customer service center, and keep the package for return processing.
This instrument is an analytical measurement and control instrument with highly precision.Only skilled,trained or authorized person should
carry out installation, setup and operation of the instrument.Ensure that the power cable is physically separated from the power supply
when connection or repair. Once the safety problem occurs, make sure that the power to the instrument is off and disconnected.
For example,
1. Apparent damage to the analyzer
2. The analyzer does not work properly or provides specied measurements
3. The analyzer has been stored for a long time in an environment where the temperature exceeds 70°C
The analyzer must be installed by licensed professionals in accordance with relevant local specications, and instructions are included in
the operation manual. Comply with the technical specications and input requirements of the controller.
Features
The instrument is equipped with different types of pH or ORP sensors. Widely used in power plants, petrochemical industry, metallurgical
electronics, mining, paper industry, biological fermentation engineering, medicine, food and beverage, environmental protection water
treatment, aquaculture, modern agricultural planting and other industries. The pH (acidity and alkalinity) value, ORP (redox potential) value
and temperature value of water solution were continuously monitored and controlled.
Large LCD display
Intelligent menu operation
Multiple automatic calibration
Manual + automatic temperature compensation
Two (2) Relay Control Switches
4-20mA & RS485 Multiple output modes
Multi Parameter Display Simultaneously shows – pH, Temp, mV, etc.
Password protection

C300 Series Operating Manual
4◀◀
|
Technical Specifications
Measuring Range pH: -2 – 16pH | ORP: -2000 – +2000mV
Unit pH | mV
Resolution pH: 0.001pH | ORP: 1mV
Basic Error pH: 0.001pH | ORP: 1mV
Temperature 14 – 302°F | -10 – 150°C (Sensor Selection Dependant)
Temperature Resolution 0.18°F | 0.1°C
Temperature Accuracy ±0.54°F | ±0.3°C
Temp. Compensation 32 – 302°F | -10 – 150°C
Temp. Compensation Manual or Automatic
Stability pH: 0.01pH/24h | ORP: 1mV/24h
Output Two 4–20mA |20–4mA |0–20mA
Communication RS485 MODBUS RTU
Relay Control Contact 5A 240VAC | 5A – 28VDC
Power Supply 85 – 265VAC | 9 – 36VDC | Power Consumption 3W
Working Temperature 14 – 140°F | -10 – 60°C
Relative Humidity ≤ 90%
Waterproof Rating IP65
Weight 0.5kg
Dimensions 98 × 98 × 130mm
Installation Opening Size 93 X 93mm
Installation methods Panel |Pipe |Wall
Typical Application

C300 Series Operating Manual
5◀◀
|
Instrument Connection
N/C NO 1
N/C COM 1
Temp NC 1
Temp NO 2
Input COM 2
Ref NC 2
N/C pH/ORP
N/C
N/C Pow DC+
(9–36V)
Cur 1+ Pow DC-
Com - Pow PE
Cur 2+
RS-485 B
RS-485 A
Pow ACL
(85–265V)
Pow AC N
Pow PE
Terminal Function Terminal Function
N/C No Connection NO 1 High set relay working position
N/C No Connection COM 1 Alarm relay common
Temp Temperature Sensor NC 1 High set relay resting position
Temp Temperature Sensor NO 2 Low set relay working position
Input Measurement COM 2 Alarm relay common
Ref Reference NC 2 Low set relay resting position
N/C No Connection
N/C No Connection Pow DC + (9 – 36V) DC power + (9 – 36V)
N/C No Connection Pow DC - DC power -
Cur 1 + Current output 1+ Pow PE Earth wire
Com - Current output common
Cur 2 + Current output 2+ Pow ACL (85 – 265V) Live wire
RS485 B Communication Terminal Pow AC N Neutral wire
RS485 A Communication Terminal Pow PE Earth wire
Dimension

C300 Series Operating Manual
7◀◀
|
Keypad Descriptions
Keypad Operation Tips:
Short Press: press the button and release immediately. if not specied below,
means short press.
Long Press: press the button for 3 seconds and then release it.
Press and hold: hold the button for a while and press the button until the right data is displayed then release the button.

C300 Series Operating Manual
9◀◀
|
Menu Structure
Press [MENU] key to enter programming mode:
If the monitor prompts you to enter the calibration security password, press the or key to set the calibration security password, and
then press the [ENT] key to conrm the calibration security password. No initial password here, please enter directly by pressing the
[ENT] key.
Item Display Selection Sub Menu
Measurement
Electrode pH
ORP
Temperature
Type
NTC2.252 kΩ
NTC10 kΩ
Pt100
Pt1000 (Default)
Offset 0.00 (Default)
Input
Automatic (Default)
Manual
N/A
Unit °C (Default)
°F
Calibration
S-S-C
(Standard solution
calibration pH)
pH
pH 2 points 4/6/9
pH 3 points 4/6/9
pH 2 points 4/7/10
pH 3 points 4/7/10
S-S-C query
(Standard solution
calibration query pH)
Offset 1
Slope 1
Offset 2
Slope 2
On-site calibration Offset Adjustment 1
On-site query Linear
Offset
S-S-C
(Standard solution
calibration ORP)
ORP S-P (single point)
S-S-C
(Standard solution
calibration ORP)
ORP slope
On-site calibration
On-site query Offset
Linear

C300 Series Operating Manual
10◀◀
|
Item Display Selection Sub Menu
Alarm
Relay 1 (pH)
Channel pH
Alarm Mode Low Alarm
High Alarm
Limit
Hysteresis 0.00 (Default)
Relay 2 (pH)
Channel pH
Alarm Mode Low Alarm
High Alarm
Limit
Hysteresis 0.00 (Default)
Relay 1 (ORP)
Channel ORP
Alarm Mode Low Alarm
High Alarm
Limit
Hysteresis 100 (Default)
Relay 2 (ORP)
Channel ORP
Alarm Mode Low Alarm
High Alarm
Limit
Hysteresis 100 (Default)
Current setting
Current 1
Channel Major measured value (Default)
Temperature
Output Type 4-20mA |20-4mA |0-20mA
Upper Limit 14.00 (Default for pH)
2000 (Default for ORP)
Lower Limit 0.00 (Default for pH)
-2000 (default for ORP)
Current 2
Channel Major measured value (Default)
Temperature
Output Type 4-20mA |20-4mA |0-20mA
Upper Limit 100 (Default)
Lower Limit 0.0 (Default)
Communication
Baud Rate
4800 Bps
9600 Bps (Default)
19200 Bps
Parity bit
No Parity (Default)
Even Parity
Odd Parity
Stop bit 1 bit (Default)
2 bits
Address 1 (default) 1 – 247

C300 Series Operating Manual
11◀◀
|
Item Display Selection Sub Menu
System
Language English
Display
Speed
Low
Standard
Medium
High
Backlight
30 sec
1 min
5 min (Default)
15 min
30 min
Always on
Password No password (Default)
Series No XXXXXXXXXX
Version VER6.1.00
Restore Calibration
Parameter
Maintenance
Current2 4mA
Current
Current1 4mA
Current2 20mA
Current2 4mA
Current2 20mA
Relay Test Relay 1: off
Relay 1: off

C300 Series Operating Manual
19◀◀
|
Measurement Setup
Electrode Setup
When the pH electrode connected to this instrument, you need to select the electrode type as “PH” in this item. It’s the same when the
electrode becomes OPR.
Temperature Setup
There are four kinds of temperature types: NTC2.252K , NTC10K , PT100 and PT1000, user needs to choose the right temp comp for the
electrode to display the right temperature. “Err” means no temp comp.
The user can compare the temperature measured by other standard instruments with the temperature measured by the D300. If there is
an error, you can use the “temperature offset” to modify it, assuming that the temperature measured by D300 is 25°C, and the temperature
measured by other standard instruments is 24.8°C. Then enter “-0.2°C” here, then press “Esc”, and then select “Save”.
You can choose Auto, Manual and None in `Temperature input`. Choosing `Auto` the temperature is from electrode. Choosing “Manual”,
user is able to input the temperature. Choosing “None” there will be no display of temperature, the default temperature of the system is
25°C as comp.
User can choose (°C) and (°F)
Temperature Offset Setting
Temperature Input Setting
Temperature Units Setting

C300 Series Operating Manual
20◀◀
|
pH Buffer Reading vs Temperature
Temp(°C) pH 4.01 pH 6.86 pH 9.18 pH 4.00 pH 7.00 pH 10.01
04.01 6.98 9.47 4.01 7.12 10.32
54.01 6.95 9.38 4.00 7.09 10.25
10 4.00 6.92 9.32 4.00 7.06 10.18
15 4.00 6.90 9.27 4.00 7.04 10.12
20 4.00 6.88 9.22 4.00 7.02 10.06
25 4.01 6.86 9.18 4.00 7.00 10.01
30 4.01 6.85 9.14 4.01 6.99 9.97
35 4.02 6.84 9.10 4.02 6.98 9.93
40 4.03 6.84 9.07 4.03 6.97 9.89
45 4.04 6.83 9.04 4.04 6.97 9.86
50 4.06 6.83 9.01 4.06 6.97 9.83
55 4.08 6.83 8.99 4.07 6.97 9.91
60 4.10 6.84 8.96 4.09 6.98 9.97
70 4.12 6.85 8.92 4.12 6.99 9.76
80 4.16 6.86 8.89 4.16 7.00 9.74
90 4.20 6.88 8.85 4.20 7.02 9.73
Calibration Setup
Press [ESC] and [ENT] “Calibration Setup” If you are prompted to enter the password, you need to enter the password set in advance.
If the password is not set, you can directly select “OK” and press the [ENT] key to enter:
Standard
liquid calibration
pH 2 points calibration NIST: 4.00, 6.86, 9.18 Calibrate middle point rst
USA: 4.01, 7.01, 10.01 Calibrate middle point rst
pH 3 points calibration NIST: 4.00, 6.86, 9.18 Calibrate in order
USA: 4.01, 7.01, 10.01 Calibrate in order
ORP 1 Point Input ORP standard liquid value
Onsite calibration Input the onsite value
Calibration with Standard Liquid (pH, 2 points)
There are two groups of 2-point standard liquid USA: 4.01, 7.01, 10.01 and NIST:
4.00, 6.86, 9.18, The following example uses the NIST standard solution set:
Choose pH standard liquid, press [ENT] to enter the mode
Table of contents
Other Icon Process Controls Controllers manuals
Popular Controllers manuals by other brands

Johnson Controls
Johnson Controls CVM03050 installation guide

hydrawise
hydrawise HWC-012-US user guide

ESYLUX
ESYLUX SmartDriver Series operating instructions

SMC Networks
SMC Networks LECP7-XB54 Series Operation manual

NANOTEC
NANOTEC SMCP33 Technical manual

Hitachi
Hitachi PSC-A32MN Installation and operation manual