8
• Read the operating instructions in full
before starting up and follow the safety
instructions.
• Keep the operating instructions in a place whe-
re they can be accessed by everyone.
• Ensure that only trained staff work with the ap-
pliance.
• Follow the safety instructions, guidelines, occu-
pational health and safety and accident preven-
tion regulations.
• The stand pillar must be inserted into the base-
plate so that the guide groove is at the back.
Other alignments are not permissible, as other-
wise the stand can topple over when a unit is
mounted on it.
•
Set up the stand in a spacious area on an even,
stable, clean, non-slip, dry and fireproof surface
.
• Any small unevennesses that cause the stand to
rock must be compensated for with the height-
adjustable feet.
• The T 65 D ULTRA-TURRAX®dispersion unit
must be lifted on to the guide head of the te-
lescopic stand with the aid of an assistant, and
secured with the two screws and toothed lock
washers (supplied), in accordance with the T 65
D ULTRA-TURRAX®operating instructions. The
unit must be mounted in front, vertically abo-
ve the baseplate; a different alignment is not
permissible, as otherwise the assembly could
topple over.
• The overhead stirrer IKA®RW 47 D must be
lifted on to the guide head of the telescopic
stand with the aid of an assistant, and secured
with the four screws (supplied), in accordance
with the IKA®RW 47 D operating instructions.
The unit must be mounted in front, vertically
above the baseplate; a different alignment is
not permissible, as otherwise the assembly
could topple over.
Safety instructions
•
This telescopic stand must be fitted with a sa-
fety disconnection device, which stops the unit
if the stand is moved upwards when the unit
is operating.
• Switch the unit off before moving the telesco-
pic stand upwards.
• When lowering the telescopic stand, take care
that no parts of the body get between the sa-
fety switches (danger of crushing).
• Only switch the unit on when the tool (agitator
or dispersion unit) is in the vessel.
• The handwheel bolt on the stand can come
loose due to vibration, which would lead to
the unit moving down. So, check the tightness
of the handwheel bolt regularly when using
the unit, and retighten it if necessary.
• Raise or lower the stand (after slackening the
handwheel bolt) by applying the manual force
as close as possible to the stand pillar. The
further the force is applied from the pillar, the
more force you need to apply, plus there is an
increased risk of the equipment toppling over.
• Secure the vessel against spinning round, by
using a clamp, for example.
• Beachten Sie die Betriebsanleitung des Disper-
giergerätes T 65 D ULTRA-TURRAX®oder des
Rührgerätes IKA®RW 47 D!
• Check the appliance and accessories before-
hand for damage each time you use them. Do
not use damaged components.
• Protect the appliance and accessories from
bumps and impacts.
• Unpacking
- Please unpack the device carefully
- In the case of any damage a detailed report
must be set immediately (post, rail or forwar-
der)
Unpacking
• Delivery scope
Telescopic stand
- Telescopic stand corresponding to
the type ordered T 653 oder R 474
- Operating instructions
For technical reasons associated with shipping,
the telescopic stand is supplied stripped down,
as pre-assembled modules.