INTRODUCTION TO THE LDC-3916370 SERIES CONTROLLER MODULE CHPT 1
PAGE 4
LDC-3916370 SERIES CONTROLLER MODULE SPECIFICATIONS
Laser Current Source Specifications
MODEL NUMBER: 3916372 500mA/9W 3916374 1A/9W 3916376 1.5A/9W
LASER CURRENT DRIVE OUTPUT1
Output Current Range: 0 to 500 mA 0 to 1000 mA 0 to 1500 mA
Set-Point Resolution: 10 µA20 µA40 µA
Set-Point Accuracy: ±0.1% of full scale ±0.1% of full scale ±0.1% of full scale
Compliance Voltage: 6 V (adjustable voltage limit) 6 V (adjustable voltage limit) 4.75 V (adjustable voltage limit)
Temperature Coefficient: ≤50ppm/°C≤50ppm/°C≤50ppm/°C
Short-Term Stability (1 hr.):2≤20 ppm ≤20 ppm ≤20 ppm
Long-Term Stability (24 hr.):3≤50 ppm ≤50 ppm ≤50 ppm
Noise and Ripple:4
High Bandwidth Mode: < 10µA rms < 10µA rms < 12µA rms
Low Bandwidth Mode: < 5.0µA rms < 5.0µA rms < 8.0µA rms
Transients:
Operational:5< 3 mA < 3 mA < 3 mA
1kV EFT / Surge:6< 4 mA / < 8 mA < 5 mA / < 10 mA < 5 mA / < 10 mA
LASER DRIVE LIMIT SETTINGS
Current Limit Range: 0 to 500 mA 0 to 1000 mA 0 to 1500 mA
Current Limit Resolution 0.2 mA 0.4 mA 0.6 mA
Current Limit Accuracy: ±0.7 mA ±1.4 mA ±4.5 mA
Voltage Limit Range: 0 to 7.5 V 0 to 7.5 V 0 to 7.5 V
Voltage Limit Resolution: 0.1 V 0.1 V 0.1 V
Voltage Limit Accuracy: ±0.2 V ±0.2 V ±0.2 V
PHOTODIODE FEEDBACK
Type: Differential 10ΩInput,
Selectable Zero or 5V Reverse Bias Differential 10ΩInput,
Selectable Zero or 5V Reverse Bias Differential 10ΩInput,
Selectable Zero or 5V Reverse Bias
PD Current Range: 0 to 5,000µA0 to 5,000µA0 to 5,000µA
Stability7:±0.01% ±0 .01% ±0 .01%
Accuracy, setpoint (% of FS): ±0.1% ±0.1% ±0.1%
EXTERNAL ANALOG MODULATION
Input :80 to 10V, 50Ω0 to 10V, 50Ω0 to 7.5V, 50Ω
Transfer Function: 50 mA / V 100 mA / V 200 mA / V
High Bandwidth Mode, Small Signal Bandwidth:9DC to 1.2 MHz DC to 1.0 MHz DC to 0.9 MHz
High Bandwidth Mode, Large Signal Bandwidth:10 DC to 1.0 MHz DC to 1.0 MHz DC to 0.9 MHz
Low Bandwidth Mode: DC to 30 kHz DC to 30 kHz DC to 30 kHz
LASER CURRENT MEASUREMENT
(DISPLAY)
Output Current Range: 0 to 500.00 mA 0 to 1000.0 mA 0 to 1500.0 mA
Output Current Resolution: 0.01 mA 0.01 mA 0.01 mA
Output Current Accuracy (@25°C): ± 0.05% of full scale ± 0.05 % of full scale ± 0.07 % of full scale
Photodiode Current Range: 0 to 5,000µA0 to 5,000µA0 to 5,000µA
PD Current Resolution: 0.1 µA0.1 µA0.1 µA
PD Current Accuracy (@25°C): ±2µA±2µA±2µA
PD Responsivity Range:11 0.00 to 1000.00µA/ mW 0.00 to 1000.00µA/ mW 0.00 to 1000.00µA/ mW
PD Responsivity Resolution: 0.01 µA/ mW 0.01 µA/ mW 0.01 µA/ mW
Optical Power Range: 0.00 to 500.00 mW 0.00 to 500.00 mW 0.00 to 500.00 mW
Optical Power Resolution: 100 µW100 µW100 µW
Forward Voltage Range: 0.000 to 7.5 V 0.000 to 7.5 V 0.000 to 7.5 V
Forward Voltage Resolution: 10 mV (1 mV through GPIB) 10 mV (1 mV through GPIB) 10 mV (1 mV through GPIB)
Forward Voltage Accuracy:12 ±7 mV (±2 mV through GPIB) ±7 mV (±2 mV through GPIB) ±7 mV (±2 mV through GPIB)
CURRENT SOURCE NOTES:
1. All values relate to a one-hour warm-up period.
2. Over any 1-hour period, half-scale output.
3. Over any 24-hour period, half-scale output.
4. Measured optically, evaluating noise intensity of a laser diode into a photodetector with 150 kHz bandwidth.
5. Maximum output current transient resulting from normal operational situations (e.g., power on-off, current on-off), as well as accidental situations (e.g., power
line plug removal).
6. Maximum output current transient resulting from a 1000V power-line transient spike.
7. Maximum monitor photodiode current drift over any 30 minute period. Assumes zero drift inresponsivity of photodiode.
8. Modulation input is 50Ωterminated inside the mainframe.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com