Aqua TROLL 500/600 pH/ORP Sensor Overview
The In-Situ pH sensor measures pH levels in natural water, surface water,
groundwater, produced water and aquaculture applications.
Getting Started (3 steps)
Replace the reference lling solution.
1
Before calibration and deployment, replace the lling solution according to the instructions
below. Repeat lling procedure between deployments or when measurements begin to drift.
Instruction Sheet
Remove sensor from
sonde and unscrew
reference junction.
Turn the cap 90° more
(one quarter of a turn) to
secure.
Lightly shake the bottle
of reference lling
solution to mix.
Overll slightly. Reinstall
reference junction cap
and tighten until it
touches sensor body.
Don’t pour solution
down the drain. Pour it
onto a paper towel and
discard.
Insert the ll tube into
the bottom of reservoir.
Squeeze a steady stream
of solution into the
reservoir while slowly
pulling out the tube.
You must have a Temperature or Conductivity/Temperature sensor installed to use the
pH/ORP sensor.