Box Contents
1. Documentation
2. Aqua TROLL 500 sonde
3. Maintenance kit
4. Instrument tools
5. Sensor
6. pH/ORP or ISE sensor maintenance supplies
7. RDO cap
Getting Started
Installing the RDO Cap and pH/ORP Sensor...
1
Remove the restrictor
from the instrument.
Use included Allen
wrench to loosen set
screw on RDO sensor.
Place Allen wrench in the
small hole at the bottom
of the sensor.
Lever the Allen wrench
towards the top of the
sonde to slide the sensor
out.
Pull the plastic cap o of
the RDO sensor.
Install the RDO cap on
the sensor.
Insert RDO sensor in
instrument sensor port.
Sensor tongue must
slide into blue motor
interlock groove.
1
2
5
4
3 6 7
Interlock
grooves
Your Aqua TROLL 500 ships with all sensors installed except for pH. If your instrument does not
include a pH/ORP sensor, all four sensors will be installed by the factory.
Aqua TROLL 500 Quickstart Guide
Information subject to change without notice. In-Situ, In-Situ logo, Baro Merge, BaroTROLL, HERMIT, HydroVu™, iSitu, Pocket-Situ, RDO, RuggedCable, RuggedReader, SmarTROLL™, TROLL,
VuSitu, and Win-Situ are trademarks or registered trademarks of In-Situ Inc.©2016. All rights reserved. This product may be covered by patents identied at www.in-situ.com/patents
Connecting the Rugged Cable...
2
Remove protective caps
from instrument and
cable.
Flat edge inside cable
end must align with
at edge on instrument
connector.
Hold textured section of
sleeve in one hand and
instrument in other.
Push and twist until you
hear a click.
Ensure O-ring on
instrument connector is
clean.
Apply vacuum grease to
the O-ring.
Connecting the communication device...
3
If desiccant is present,
remove it from cable.
Align TROLL Com
connector with cable
end. Push and twist until
you hear a click.
Remove desiccant pack
from sensor port.
Insert pH/ORP sensor in
empty sensor port.
Tighten screw at
base of sensor with
Allen wrench. Do not
overtighten.
Replace restrictor
with vent holes at
base of instrument for
calibration.
After calibration, ip the
restrictor 180 degrees for
deployment.
Apply a pea-sized drop
of lubricant to sensor
O-rings.
Remove caps from pH/
ORP sensor.
O-ring
2019-09-10 | 0085822
Connecting to the software...
Getting more information...
4
5
Get complete instructions about calibration, logging and working with data in the product manual.
Download it from www.in-situ.com.
VuSitu’s on-screen instructions will guide you through instrument calibration, live readings and
data sharing.
Tap the serial number of
your Wireless TROLL Com.
VuSitu will display the
Connected Instrument
screen when pairing is
complete.
c d
Launch VuSitu and tap
Connect.
The app will locate and
display nearby In-Situ
devices.
iOS Android
An iOS device will
automatically connect
to the closest In-Situ
instrument.
To connect to another
instrument, press
Disconnect and then
Choose or Add Device.
VuSitu will display a list of
available connections.
aa bb
You must have the VuSitu mobile app to use the RDO Blue with a mobile device. Download
VuSitu from the Google Play Store or the Apple App Store.

Other manuals for Aqua TROLL 500

6

Other In-situ Test Equipment manuals

In-situ TROLL RDO 9500 Manual

In-situ

In-situ TROLL RDO 9500 Manual

In-situ SMARTROLL User manual

In-situ

In-situ SMARTROLL User manual

In-situ Aqua TROLL 500 User manual

In-situ

In-situ Aqua TROLL 500 User manual

Popular Test Equipment manuals by other brands

WIKA CPH6600 operating instructions

WIKA

WIKA CPH6600 operating instructions

Presco AG PD-CLSB-CD00 instruction manual

Presco AG

Presco AG PD-CLSB-CD00 instruction manual

Tietzsch MultiSafe DSP TT1 User instructions

Tietzsch

Tietzsch MultiSafe DSP TT1 User instructions

Sigfox SNT1.1-1 user guide

Sigfox

Sigfox SNT1.1-1 user guide

Seismic Source DAQlink III user manual

Seismic Source

Seismic Source DAQlink III user manual

EXFO EXFO MaxTester DSL user guide

EXFO

EXFO EXFO MaxTester DSL user guide

Ametek Teseq CDN 3153-S63 user manual

Ametek

Ametek Teseq CDN 3153-S63 user manual

Tektronix 7934 Operator's manual

Tektronix

Tektronix 7934 Operator's manual

Fluke 709 Quick reference guide

Fluke

Fluke 709 Quick reference guide

Atlanta BYK Cupping Tester manual

Atlanta

Atlanta BYK Cupping Tester manual

Kett HB-300 operating manual

Kett

Kett HB-300 operating manual

REED R8802 instruction manual

REED

REED R8802 instruction manual

Martel LC-110 Reference manual

Martel

Martel LC-110 Reference manual

Kingfisher KI 2400 Series Training manual

Kingfisher

Kingfisher KI 2400 Series Training manual

Hanna Instruments HI 98121 user manual

Hanna Instruments

Hanna Instruments HI 98121 user manual

Perfomance Tool W2990 owner's manual

Perfomance Tool

Perfomance Tool W2990 owner's manual

NEC TTS 26B instruction manual

NEC

NEC TTS 26B instruction manual

Hioki IR4051 instruction manual

Hioki

Hioki IR4051 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.