InnoLux Yushan Series User manual

1
© INNOLUX Corporaon │ X-Ray Products
Yushan Series
X-Ray Flat Panel Detector
Operation Manual
INNOLUX Corporation

© INNOLUX Corporaon │ X
Radiologist or any other practitioners licensed
adequate knowledge of the operation prior using the
Flat Panel Detector.
This Operation
Manual should be studied and understood before proceeding
to op
erate the equipment on patients; it describes deta
the INNOLUX Yushan X
when operating it.
After reading this manual, store it nearby the
Panel Detector
so that you can see it whenever necessary
"Caution: Fede
ral Law restricts this device to sale by or
radiologist or any other practitioners licensed by the law of the state in which
that person practices to use or order the use of the device.
_____________________________________________________________
Caution
This Operation
Manual contains confidential and proprietary information of the
Manufacturer.
Before using the device, please check local regulations. If any local legislation
is
violated, use cannot be authorized.
If additional training or material for training is needed, please directly
our official dealer or INNOLUX Representative.
_____________________________________________________________
Trademarks
Company names
described in this manual are the trademarks or registered
trademarks of INNOLUX Corporation or their respective holders.
Windows is the registered trademark of US Microsoft Corporation in the U.S.A.
and other countries.
Copyright © 2013
INNOLUX Corporatio
Document Number: YS
-
Version: A1
Issued date: 2018/09/1
8
© INNOLUX Corporaon │ X
-Ray Products
Radiologist or any other practitioners licensed
should ensure that they have
adequate knowledge of the operation prior using the
INNOLUX
Y
Manual should be studied and understood before proceeding
erate the equipment on patients; it describes deta
ils
on how to operate
the INNOLUX Yushan X
-Ray Flat Panel Detector
and cautions to be observed
After reading this manual, store it nearby the
INNOLUX Yushan X
so that you can see it whenever necessary
.
ral Law restricts this device to sale by or
on
the order of a
radiologist or any other practitioners licensed by the law of the state in which
that person practices to use or order the use of the device.
"
_____________________________________________________________
Manual contains confidential and proprietary information of the
Before using the device, please check local regulations. If any local legislation
violated, use cannot be authorized.
If additional training or material for training is needed, please directly
our official dealer or INNOLUX Representative.
_____________________________________________________________
described in this manual are the trademarks or registered
trademarks of INNOLUX Corporation or their respective holders.
Windows is the registered trademark of US Microsoft Corporation in the U.S.A.
INNOLUX Corporatio
n. All rights reserved.
-
OM-01
8
2
should ensure that they have
Y
ushan X-Ray
Manual should be studied and understood before proceeding
on how to operate
and cautions to be observed
INNOLUX Yushan X
-Ray Flat
the order of a
radiologist or any other practitioners licensed by the law of the state in which
_____________________________________________________________
Manual contains confidential and proprietary information of the
Before using the device, please check local regulations. If any local legislation
If additional training or material for training is needed, please directly
contact
_____________________________________________________________
described in this manual are the trademarks or registered
trademarks of INNOLUX Corporation or their respective holders.
Windows is the registered trademark of US Microsoft Corporation in the U.S.A.

3
© INNOLUX Corporaon │ X-Ray Products
Table of Content
Chapter 1 Introduction ............................................................................................. 6
1.1 Product Introduction ............................................................................ 7
1.2 Product Series ....................................................................................... 7
1.3 Product Intended Use .......................................................................... 7
1.4 Product Significant Performance Characteristics ....................... 7
1.5 Product Intended Patient Population ..............................................
1.6 Product Intended Part of the body or type of tissue applied to
or interacted with ..............................................................................................
1.7 Product Intended user profile ............................................................ 9
1.8 Product Intended conditions of use ................................................ 9
1.9 Product Features ................................................................................. 10
1.10 Clinical Guide ....................................................................................... 10
Chapter 2 Safety and Regulatory ........................................................................ 11
2.1 Compliance Standards and Classification ........................................ 11
2.1.1 Compliance Standards ................................................................ 11
2.1.2 Classification.................................................................................. 12
2.2 Contraindications ..................................................................................... 12
2.3 Power Supply Hazard .............................................................................. 12
2.4 Electric Shock Hazard ............................................................................. 12
2.5 Abnormalities Hazard .............................................................................. 14
2.6 Installation Precautions .......................................................................... 14
2.7 Connection Instructions ......................................................................... 15
2.8 External Network Connection ............................................................... 15
2.9 System Isolation Instructions ............................................................... 16
2.10 Software Precautions ............................................................................ 17
2.11 Disinfection Instructions ...................................................................... 17
2.12 Charging the Battery Pack ................................................................... 1
2.13 Battery Pack Instructions .................................................................... 19
2.14 Other Precautions .................................................................................. 22
2.15 Disposal of Waste .................................................................................. 24
2.16 Product Labels ........................................................................................ 24
2.16.1 Identification Label ..................................................................... 25
2.16.2 Battery Polarity Label ................................................................ 25
2.16.3 Battery Label ................................................................................ 26
2.16.4 Packing Carton Label ................................................................ 26
2.16.5 Adapter Specification Label ..................................................... 26

4
© INNOLUX Corporaon │ X-Ray Products
2.16.6 Safety Symbols ........................................................................... 27
3.1 How to Connect FPD ............................................................................... 2
3.2 Unit Names and the Functions ............................................................. 29
3.2.1 Flat Panel Detector ....................................................................... 29
3.2.2 Accessories .................................................................................... 30
3.3 LED indicators ........................................................................................... 33
Chapter 4 Basic Operation ................................................................................... 34
4.1 Preparing the Flat Panel Detector ........................................................ 34
4.1.1 Type of Flat Panel Detector ........................................................ 34
4.1.2 Number of the Connectable Flat Panel Detector .................. 34
4.1.3 Connecting/Disconnecting the Flat Panel Detector ............ 34
4.1.4 Charging the Battery Pack for the Flat Panel Detector ....... 35
4.1.5 Installing/Removing the Battery Pack for the Flat Panel
Detector ..................................................................................................... 35
4.1.6 LED Indicator to show battery capacity .................................. 36
4.2 Detector Connection................................................................................ 37
4.2.1 Wire connection ............................................................................ 37
4.2.2 Wireless connection ..................................................................... 3
4.2.3 AP mode connection .................................................................... 39
4.3 Starting Up and Shutting Down the Yushan FPD ............................ 41
4.3.1 Starting Up the Yushan ............................................................... 41
4.3.2 Overview of User Interface ......................................................... 41
4.3.3 Search and connect FPD ............................................................ 42
4.3.4 Enter the operation and select the mode ............................... 44
4.3.5 Performing an Exposure and Image Acquistion .................. 45
4.3.6 Download image from FPD ......................................................... 4
4.3.7 Shutting Down the Yushan ......................................................... 49
4.4 Correction procedure .............................................................................. 50
4.4.1 Collect images ............................................................................... 50
4.4.2 Correction process ....................................................................... 51
4.4.3 Storage of calibration data ......................................................... 53
5.1 Daily User Inspection and Maintenance ............................................. 54
5.1.1 Periodical Inspection ................................................................... 54
Chapter 6 Troubleshooting .................................................................................. 55
Appendix A Specifications of FPD ..................................................................... 56
A.1 Specifications ........................................................................................... 56
A.1.1 Reduced Equivalent ..................................................................... 56
A.1.2 Power Supply Conditions when Using Back-up Cable ...... 56

5
© INNOLUX Corporaon │ X-Ray Products
A.1.3 Power Supply to Battery Charger ............................................ 57
A.1.4 Power Consumption and Battery Specification ................... 57
A.1.5 Environmental Conditions ......................................................... 57
A.1.6 Image Performance ...................................................................... 59
A.1.7 Radio Waves .................................................................................. 60
A.1.8 Reliability and Lifetime ............................................................... 60
Appendix B Electromagnetic Compatibility (EMC) ........................................ 62
B.1 Yushan FPD ............................................................................................... 62
B.2 Further Information for IEC 60601-1-2 (EN 60601-1-2) ................... 63
Appendix C Radio Frequency (RF) Compliance Information ...................... 64

© INNOLUX Corporaon │ X
Chapter 1
This manual is intended to provide the operator with an overview of the
operation and safety requirements for the
Detector. This manual is not intended to provide instructions on actual
treatment procedures and it is expected that users should have official
radiologist or any other practitioners
Manufacturer and Di
stribution organization assume no liability thro
of the device.
All care has been taken in the preparation and checking of this manual
however there is no guarantee provided that all information is correct. The
information provided in this manua
Only CE or FDA approved or authorized
device
. The Manufacturer and Distribution organization shall not be held liable
or responsible for damages or injury caused as a result of
accessories.
Installation may only be conducted by authorized service
personal.
Some maintenance and service work
authorized service agents and only those procedures outlined in the operator
and service manual are all
person will void all warranties. No circuit diagrams or component part lists are
to
be provided for this device
provided in this manual then please conta
distributor in writing with your reasons for wanting them and then a copy of the
service manual may be provided. Before using the
Flat Panel Detector,
the operator should read this manual carefully and
particular attention to the sections of Safety, Operation and Maintenance.
________________________________________
Caution
INNOLUX Corporation shall not be liable for malfunctions and damages
resulting from use under
conditions for this product such as power supply, installation environment, etc.
contained in this manual.
INNOLUX Corporation shall not be liable for malfunctions and damages
resulting from natural disaste
etc.
_____________________________________________________________
© INNOLUX Corporaon │ X
-Ray Products
Chapter 1
Introduction
This manual is intended to provide the operator with an overview of the
operation and safety requirements for the
INNOLUX Yushan X-
Ray Flat Panel
Detector. This manual is not intended to provide instructions on actual
treatment procedures and it is expected that users should have official
radiologist or any other practitioners
license prior to using the
stribution organization assume no liability thro
All care has been taken in the preparation and checking of this manual
however there is no guarantee provided that all information is correct. The
information provided in this manua
l is subject to change without notice.
Only CE or FDA approved or authorized
accessories may be used in this
. The Manufacturer and Distribution organization shall not be held liable
or responsible for damages or injury caused as a result of
using no
Installation may only be conducted by authorized service
Some maintenance and service work
also
must be carried out by
authorized service agents and only those procedures outlined in the operator
and service manual are all
owed. Any service work carried out by unauthorized
person will void all warranties. No circuit diagrams or component part lists are
be provided for this device
. If you require technical documentation that is not
provided in this manual then please conta
ct the manufacturer or your local
distributor in writing with your reasons for wanting them and then a copy of the
service manual may be provided. Before using the
INNOLUX
Yushan X
the operator should read this manual carefully and
particular attention to the sections of Safety, Operation and Maintenance.
________________________________________
____________________
INNOLUX Corporation shall not be liable for malfunctions and damages
resulting from use under
environment conditions outside the range of using
conditions for this product such as power supply, installation environment, etc.
contained in this manual.
INNOLUX Corporation shall not be liable for malfunctions and damages
resulting from natural disaste
rs such as fires, earthquakes, floods, lightning,
_____________________________________________________________
6
This manual is intended to provide the operator with an overview of the
Ray Flat Panel
Detector. This manual is not intended to provide instructions on actual
treatment procedures and it is expected that users should have official
license prior to using the
device. The
stribution organization assume no liability thro
ugh the use
All care has been taken in the preparation and checking of this manual
however there is no guarantee provided that all information is correct. The
l is subject to change without notice.
accessories may be used in this
. The Manufacturer and Distribution organization shall not be held liable
using no
n-approved
Installation may only be conducted by authorized service
must be carried out by
authorized service agents and only those procedures outlined in the operator
owed. Any service work carried out by unauthorized
person will void all warranties. No circuit diagrams or component part lists are
. If you require technical documentation that is not
ct the manufacturer or your local
distributor in writing with your reasons for wanting them and then a copy of the
Yushan X
-Ray
the operator should read this manual carefully and
pay
particular attention to the sections of Safety, Operation and Maintenance.
____________________
_
INNOLUX Corporation shall not be liable for malfunctions and damages
environment conditions outside the range of using
conditions for this product such as power supply, installation environment, etc.
INNOLUX Corporation shall not be liable for malfunctions and damages
rs such as fires, earthquakes, floods, lightning,
_____________________________________________________________

7
© INNOLUX Corporaon │ X-Ray Products
1.1 Product Introduction
INNOLUX Yushan X-Ray Flat Panel Detector is an X-ray image acquisition
device that is based on flat-panel. This device should be integrated with an
operating PC and an X-ray generator. It can do to utilize as digitalizing X-ray
images and transfer for radiography diagnostic. Each X ray photon will be
converted to electronic signal by scintillator and the sensor array. Electronic
signal in the pixel will be readout by driver IC and then become an image
before send to PC.
1.2 Product Series
Model Description
Yushan V14C Glass version 14”x 17” FPD with CsI
Yushan V14G Glass version 14”x 17” FPD with GOS
1.3 Product Intended Use
The Wireless/Wired INNOLUX Yushan X-Ray Flat Panel Detector is intended
to capture for display radiographic images of human anatomy. It is intended for
use in general projection radiographic applications including pediatric and
neonatal exams wherever conventional film/screen or CR systems may be
used. The INNOLUX Yushan X-Ray Flat Panel Detector is not intended for
mammography, fluoroscopy, tomography, and angiography applications.
INNOLUX Yushan series provide either raw X ray image or corrected image
for system integrator to do further image process.
1.4 Product Significant Performance
Characteristics
Items Characteristics
Image resolution 2500x3052
Data 16 bit data
MTF (Modulation transfer
function) (Typical)
CsI 0.69 @ 1 lp/mm
0.39 @ 2 lp/mm
0.22 @ 3 lp/mm
GOS 0.52 @ 1 lp/mm
0.23 @ 2 lp/mm

© INNOLUX Corporaon │ X-Ray Products
0.11 @ 3 lp/mm
DQE (Detector quantum
efficiency) (Typical)
CsI
0.51 @ 0 lp/mm
0.48 @ 1 lp/mm
0.39 @ 2 lp/mm
0.29 @ 3 lp/mm
GOS
0.31 @ 0 lp/mm
0.27 @ 1 lp/mm
0.16 @ 2 lp/mm
0.09 @ 3 lp/mm
Maximum linear dose (Typical) 88 uGy
Sensitivity (Typical) CsI 574 lsb/uGy
GOS 592 lsb/uGy
Dark noise (Typical) CsI 4.5 LSB
GOS 7.9 LSB
Lag <0.5% after 1 minute
1.5 Product Intended Patient Population
Considerations Requirement Description
Age -Not restricted
Height -Not restricted
Weight
-The uniform load of the device is 300kg
-The local load on 4 cm diameter of the
device is 120kg.
*Note: Please evaluate the patient’s
condition before using.
Health -Not restricted
Nationality -Multiple
Sex -Not restricted
1.6 Product Intended Part of the body or type of
tissue applied to or interacted with
-Measurement site: body
-Condition: Intact or wound skin

9
© INNOLUX Corporaon │ X-Ray Products
1.7 Product Intended user profile
Considerations Requirement Description
Education Minimum -Graduate of radiology college
Maximum -No maximum
Knowledge Minimum
-Read and understand 'westernized
Arabic' numerals when written in Arial
font
-Can distinguish of human body
-Understands hygiene
Maximum -No maximum
Language
Understanding
Minimum -Local language
Maximum -Understand the operation manual that
is writing in English
Experience Minimum -Physician or legally certified operator
Maximum -No maximum
_____________________________________________________________
Note
Patient should not operate the device by its own, in case result in malfunction
of the equipment.
_____________________________________________________________
1.8 Product Intended conditions of use
Considerations
Condition
Environment including
hygienic requirements
-Non-sterile
-Multiple patient use
-Less than ten minute contact
-Indoor use only
-Ambient luminance range: 100 to 1500 lx
-Viewing angle: normal to the display ±20°
Frequency of use -Reusable
-1day: 200 shot
Location -In hospital environment
Mobility -Portable ME equipment to be used on a
patient

10
© INNOLUX Corporaon │ X-Ray Products
1.9 Product Features
-140 micron pixel pitch
-Wide image
-16-bit dynamic range output
-Exposure times up to 3.0 seconds
-Wireless communication mode or wired communication mode is available.
When used in wireless communication mode, an access point and battery
pack (optional) are required.
1.10 Clinical Guide
Users may require to attend the traning courses held by INNOLUX before
using. A complete training should consist of the following elements:
-All safety precautions for practitioners and patients when using the device.
-A review of published clinical literature should be conducted to understand
the indications and effects of the X-Ray Flat Panel Detector.
_____________________________________________________________
Note
It is strongly recommended for INNOLUX Yushan practitioners to attend
organized meetings, workshops, seminars, and conferences on use of X-Ray
Flat Panel Detector.
_____________________________________________________________

11
© INNOLUX Corporaon │ X-Ray Products
Chapter 2 Safety and Regulatory
This INNOLUX Yushan X-Ray Flat Panel Detector has been designed and
tested to function in a safe and correct when used as indicated in this manual.
Do not use this device before reading and completely understand this
Operation Manual. Always observe precautions for safety and only operates
the device in a qualified room that provides protection.
2.1 Compliance Standards and
Classification
2.1.1 Compliance Standards
-FDA Standards 21 CFR 1020.31 for ionizing radiation emitting products
-FDA Standards 21 CFR 892.1680 for stationary x-ray system
-European Medical Devices Directive (93/42/EEC)
-EN ISO 13485:2016
-EN ISO 14971:2012
-EN 60601-1: 2006+A1: 2013+A12: 2014
-ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012
and A2:2010/(R)2012
-CAN/CSA C22.2 No. 60601-1:14
-EN 60601-1-2:2015
-IEC 62304:2006/Amd 1:2015
-EN 60601-1-6: 2010/A1:2015
-EN 62366-1: 2015
-EN ISO 10993-1:2009/AC:2010
-EN ISO 10993-5:2009
-ISO 10993-10:2010
-EN 1041:2008
-EN ISO 15223-1:2016
-RED Directive (2014/53/EU)
-RoHS Directive (2011/65/EU)
-REACH Directive (EC/1907/2006)
-
WEEE Directive (2012/19/EU)

© INNOLUX Corporaon │ X
2.1.2
Classification
-
Type of protection against electrical shock: Class I Equipment
-
Degree of protection a
-
Degree of protection against harmful ingress of water: IPX6
rating does not indicate
tested for compliance with, any other IPX_ ratings.
-
Degree of safety of application in the presence of a flammable anesthetics
mixture with air or with oxygen or nitrous oxide: Equipment not suitable for
us
e in the presence of a flammable anesthetics mixture with air or with
oxygen or nitrous oxide.
2.2 Contraindications
No contraindications present.
2.3 Power Supply Hazard
_____________________________________________________________
Warning
The power
supply to the Yushan is AC 100 to 240V.
To avoid electric shocks, users should always take the following precautions:
-
Do not open any covers when it is not necessary.
-
Install the equipment in a location where it will not be exposed to water.
-Check that the
equipment is securely earthed.
-
Check that all of the cables are completely and securely connected.
-
Keep the control cabinet out of reach of patients.
_____________________________________________________________
2.4
Electric S
_________________
____________________________________________
Warning
© INNOLUX Corporaon │ X
-Ray Products
Classification
Type of protection against electrical shock: Class I Equipment
Degree of protection a
gainst electrical shock: Type B
Equipment
Degree of protection against harmful ingress of water: IPX6
rating does not indicate
compliance with, and the detectors have not been
tested for compliance with, any other IPX_ ratings.
)
Degree of safety of application in the presence of a flammable anesthetics
mixture with air or with oxygen or nitrous oxide: Equipment not suitable for
e in the presence of a flammable anesthetics mixture with air or with
oxygen or nitrous oxide.
2.2 Contraindications
No contraindications present.
2.3 Power Supply Hazard
_____________________________________________________________
supply to the Yushan is AC 100 to 240V.
To avoid electric shocks, users should always take the following precautions:
Do not open any covers when it is not necessary.
Install the equipment in a location where it will not be exposed to water.
equipment is securely earthed.
Check that all of the cables are completely and securely connected.
Keep the control cabinet out of reach of patients.
_____________________________________________________________
Electric S
hock Hazard
____________________________________________
12
Type of protection against electrical shock: Class I Equipment
Equipment
Degree of protection against harmful ingress of water: IPX6
(An IPX6
compliance with, and the detectors have not been
Degree of safety of application in the presence of a flammable anesthetics
mixture with air or with oxygen or nitrous oxide: Equipment not suitable for
e in the presence of a flammable anesthetics mixture with air or with
_____________________________________________________________
To avoid electric shocks, users should always take the following precautions:
Install the equipment in a location where it will not be exposed to water.
Check that all of the cables are completely and securely connected.
_____________________________________________________________
____________________________________________

13
© INNOLUX Corporaon │ X-Ray Products
Observe the following precautions when using the cables:
-Do not touch the plug and connector with wet hands. Otherwise, electric
shock may result causing death or severe injury.
-Hold the plug or connector when removing the cable.
-Pulling the cable or carrying by holding it may damage the cable, causing
fire or electric shock.
-Do not damage or remodel the cable.
-Do not place a heavy object on the cable or lay it under the flat panel
detector. Do not step on, pull, forcibly bend, or bundle the cable.
Otherwise, fire or electric shock may result.
-Do not use the flat panel detector for the
radiographic examination stand if its cable
becomes overloaded. Otherwise, the cable may
be damaged, causing fire or electric shock.
-Do not touch the back-up cable connector of
Yushan, otherwise it may result in electric shock
or malfunction of the equipment.
Do not use the equipment in a location where metal particles could come into
the equipment. This may cause an electric shock.
Do not disassemble or remodel the equipment. Otherwise, fire or electric
shock may result. Keep away from the parts inside the product, which may
cause electric shock. If you touch them accidentally, death or severe injury
may result.
Do not hit or drop the equipment or subject it to severe shock. Otherwise, the
equipment may be damaged. If the damaged equipment is used, fire or electric
shock may result.
Do not use the flat panel detector to lift up patient.
In addition, do not apply strong pressure onto the
flat panel detector. If applied, the flat panel detector
deforms and the waterproof function may be
invalid.
Have the patient take a fixed posture and do not let the patient touch parts
unnecessarily. If the patient touches connectors or switches, it may result in
electric shock or malfunction of the equipment.
Do not use the flat panel detector without the battery packs. If the battery

© INNOLUX Corporaon │ X
packs are not attached, an electric shock may result.
Make sure to use
the optional parts and accessories
Failure to use the optional parts and accessories recommended by us may
result in damage to the equipment and/or electric shock and injury.
Keep the equipment away from patient’s body fluids, chemical
Otherwise, it may be damaged, causing fire or electric shock. If necessary,
protect the flat panel detector by covering it with a disposable bag.
_____________________________________________________________
2.5 Abnormalities Hazard
_____________________________________________________________
Warning
If any of the following occurs, immediately turn off the power of each unit,
unplug the power cable from the outlet, and then contact our official dealer or
INNOLUX Representative.
●W
hen smoke, strange odor, or abnormal sound is present.
●When a foreign object (such as a metal object) or liquid enters the product.
●When the equipment is dropped or hit and is damaged.
_____________________________________________________________
2.6
Installation Precautions
_____________________________________________________________
Cautions
Do not install the equipment in a locatio
-
Where the temperature changes sharply.
-
Close to heat sources such as a heater.
-Where
the equipment may be exposed to water due to water leakage or
ingress.
-
Where corrosive gas may be generated.
-
Where there is excessive dust.
-
Where the equipment is subject to frequent or excessive vibration/shock.
-
Where the equipment is exposed to direct
© INNOLUX Corporaon │ X
-Ray Products
packs are not attached, an electric shock may result.
the optional parts and accessories
provided by innolux.
Failure to use the optional parts and accessories recommended by us may
result in damage to the equipment and/or electric shock and injury.
Keep the equipment away from patient’s body fluids, chemical
Otherwise, it may be damaged, causing fire or electric shock. If necessary,
protect the flat panel detector by covering it with a disposable bag.
_____________________________________________________________
2.5 Abnormalities Hazard
_____________________________________________________________
If any of the following occurs, immediately turn off the power of each unit,
unplug the power cable from the outlet, and then contact our official dealer or
INNOLUX Representative.
hen smoke, strange odor, or abnormal sound is present.
When a foreign object (such as a metal object) or liquid enters the product.
When the equipment is dropped or hit and is damaged.
_____________________________________________________________
Installation Precautions
_____________________________________________________________
Do not install the equipment in a locatio
n with the following conditions:
Where the temperature changes sharply.
Close to heat sources such as a heater.
the equipment may be exposed to water due to water leakage or
Where corrosive gas may be generated.
Where there is excessive dust.
Where the equipment is subject to frequent or excessive vibration/shock.
Where the equipment is exposed to direct
sunlight.
14
provided by innolux.
Failure to use the optional parts and accessories recommended by us may
result in damage to the equipment and/or electric shock and injury.
Keep the equipment away from patient’s body fluids, chemical
s, water, etc.
Otherwise, it may be damaged, causing fire or electric shock. If necessary,
protect the flat panel detector by covering it with a disposable bag.
_____________________________________________________________
_____________________________________________________________
If any of the following occurs, immediately turn off the power of each unit,
unplug the power cable from the outlet, and then contact our official dealer or
When a foreign object (such as a metal object) or liquid enters the product.
_____________________________________________________________
_____________________________________________________________
n with the following conditions:
the equipment may be exposed to water due to water leakage or
Where the equipment is subject to frequent or excessive vibration/shock.

© INNOLUX Corporaon │ X
Use the equipment on a flat place. If the equipment falls, it may cause damage
to the equipment or personal injury.
When you move the equipment, place it in the cassette storage box of a mobile
X-
ray unit or hold it by hand to prevent it from fal
the equipment, place it horizontally.
For veterinary or mobile applications, contact our official dealer or INNOLUX
Representative.
When the devices are used outdoors in wireless communication mode, contact
our official de
aler or INNOLUX Representative.
Do not place any object in a place where removal of the power cable is
prevented.
To ensure optimal image quality, it is recommended that you do not use the flat
panel detector
near devices (motor, transformer, switching supply, etc.) that
generate electromagnetic noise.
To ensure optimal image quality, it is recommended that you do not place the
cables (power cable,
back
(motor, tra
nsformer, switching supply, etc.) that generate electromagnetic
noise and their cables.
_____________________________________________________________
2.7 Connection Instructions
_____________________________________________________________
Warning
Make
sure that the devices to be connected to the equipment are authorized
for connection.
Connect the panel unit
Yushan
_____________________________________________________________
2.8
External Network Connection
_____________________________________________________________
© INNOLUX Corporaon │ X
-Ray Products
Use the equipment on a flat place. If the equipment falls, it may cause damage
to the equipment or personal injury.
When you move the equipment, place it in the cassette storage box of a mobile
ray unit or hold it by hand to prevent it from fal
ling. If the cart is used to move
the equipment, place it horizontally.
For veterinary or mobile applications, contact our official dealer or INNOLUX
When the devices are used outdoors in wireless communication mode, contact
aler or INNOLUX Representative.
Do not place any object in a place where removal of the power cable is
To ensure optimal image quality, it is recommended that you do not use the flat
near devices (motor, transformer, switching supply, etc.) that
generate electromagnetic noise.
To ensure optimal image quality, it is recommended that you do not place the
back
-up
cable, etc.) of the equipment near devices
nsformer, switching supply, etc.) that generate electromagnetic
_____________________________________________________________
2.7 Connection Instructions
_____________________________________________________________
sure that the devices to be connected to the equipment are authorized
Yushan
only to the
access point or DR system.
_____________________________________________________________
External Network Connection
_____________________________________________________________
15
Use the equipment on a flat place. If the equipment falls, it may cause damage
When you move the equipment, place it in the cassette storage box of a mobile
ling. If the cart is used to move
For veterinary or mobile applications, contact our official dealer or INNOLUX
When the devices are used outdoors in wireless communication mode, contact
Do not place any object in a place where removal of the power cable is
To ensure optimal image quality, it is recommended that you do not use the flat
near devices (motor, transformer, switching supply, etc.) that
To ensure optimal image quality, it is recommended that you do not place the
cable, etc.) of the equipment near devices
nsformer, switching supply, etc.) that generate electromagnetic
_____________________________________________________________
_____________________________________________________________
sure that the devices to be connected to the equipment are authorized
access point or DR system.
_____________________________________________________________
External Network Connection
_____________________________________________________________

© INNOLUX Corporaon │ X
Caution
When a setting of the network to which the equipment is connected has been
changed,
check that the change does not affect the system operation and take
measures if necessary.
The
setting change may include the following:
-
Change of connection destination
-Addition of devices
-
Removal of devices
-Update of devices
-
Upgrade of devices
After connecting this system to the network with other systems, confirm that
the other systems are not af
such as network separation.
Warning
Make sure to use the optional parts, accessories and networks recommended
by us. Failure to use the optional parts, accessories and networks
recommended by us may resu
shock and injury.
Connect to the Ethernet Network of 100BASE
the IEEE standard 802.3. Do not connect telephone lines to LAN connector.
Only UTP-
type straight LAN cables of 4
higher are appropriate for connection to this connector.
_____________________________________________________________
2.9
System Isolation Instructions
_____________________________________________________________
Warning
To
ensure complete system isolation, never install any unauthorized
accessories or other such items.
When it is necessary to install authorized accessories or optional items,
© INNOLUX Corporaon │ X
-Ray Products
When a setting of the network to which the equipment is connected has been
check that the change does not affect the system operation and take
setting change may include the following:
Change of connection destination
Removal of devices
Upgrade of devices
After connecting this system to the network with other systems, confirm that
the other systems are not af
fected. If they are affected, take countermeasures
such as network separation.
Make sure to use the optional parts, accessories and networks recommended
by us. Failure to use the optional parts, accessories and networks
recommended by us may resu
lt in damage to the equipment and/or electric
Connect to the Ethernet Network of 100BASE
-TX or 10BASE-
T prescribed in
the IEEE standard 802.3. Do not connect telephone lines to LAN connector.
type straight LAN cables of 4
-pair Ca
tegory 5 cable (CAT 5E) or
higher are appropriate for connection to this connector.
_____________________________________________________________
System Isolation Instructions
_____________________________________________________________
ensure complete system isolation, never install any unauthorized
accessories or other such items.
When it is necessary to install authorized accessories or optional items,
16
When a setting of the network to which the equipment is connected has been
check that the change does not affect the system operation and take
After connecting this system to the network with other systems, confirm that
fected. If they are affected, take countermeasures
Make sure to use the optional parts, accessories and networks recommended
by us. Failure to use the optional parts, accessories and networks
lt in damage to the equipment and/or electric
T prescribed in
the IEEE standard 802.3. Do not connect telephone lines to LAN connector.
tegory 5 cable (CAT 5E) or
_____________________________________________________________
System Isolation Instructions
_____________________________________________________________
ensure complete system isolation, never install any unauthorized
When it is necessary to install authorized accessories or optional items,

© INNOLUX Corporaon │ X
contact our official dealer or INNOLUX Representative.
Keep equipment other than tho
appropriate system isolation.
In normal use, have a patient take a proper positioning for exposure. The
operator should operate the system in a place where safety from radiation is
ensured. The operator sho
the patient is in the exposure area and the operating area of the system
_____________________________________________________________
2.10
Software Precautions
_____________________________________________________________
Caution
Do not install additional software to the system. Do not uninstall any of the
software
preinstalled in the system. The system is preinstalled with the
appropriate software. If other so
uninstalled, various operational errors may result.
Open-
Source Software Contained in This Product
-
This product contains third party’s software that is made available as open
source software or free softwa
-
This software is provided “as is” with no warranty of any kind as to its
merchantability or fitness for any particular purpose.
-
If you would like to receive such source codes, please contact INNOLUX
dealer or the service representatives at the agency
purchased this product. (Please be noted that any inquiries concerning
the contents of source codes should be directed to original licensors of
open source software.)
_____________________________________________________________
2.11 Disinfe
ction Instructions
_____________________________________________________________
Warning
Do not use the following disinfectants or sterilize
© INNOLUX Corporaon │ X
-Ray Products
contact our official dealer or INNOLUX Representative.
Keep equipment other than tho
se used for patients out of their reach to ensure
appropriate system isolation.
In normal use, have a patient take a proper positioning for exposure. The
operator should operate the system in a place where safety from radiation is
ensured. The operator sho
uld also make sure before exposure that no one but
the patient is in the exposure area and the operating area of the system
_____________________________________________________________
Software Precautions
_____________________________________________________________
Do not install additional software to the system. Do not uninstall any of the
preinstalled in the system. The system is preinstalled with the
appropriate software. If other so
ftware is installed or if the existing software is
uninstalled, various operational errors may result.
Source Software Contained in This Product
This product contains third party’s software that is made available as open
source software or free softwa
re.
This software is provided “as is” with no warranty of any kind as to its
merchantability or fitness for any particular purpose.
If you would like to receive such source codes, please contact INNOLUX
dealer or the service representatives at the agency
from which you
purchased this product. (Please be noted that any inquiries concerning
the contents of source codes should be directed to original licensors of
open source software.)
_____________________________________________________________
ction Instructions
_____________________________________________________________
Do not use the following disinfectants or sterilize
rs at the time of disinfection,
17
se used for patients out of their reach to ensure
In normal use, have a patient take a proper positioning for exposure. The
operator should operate the system in a place where safety from radiation is
uld also make sure before exposure that no one but
the patient is in the exposure area and the operating area of the system
.
_____________________________________________________________
_____________________________________________________________
Do not install additional software to the system. Do not uninstall any of the
preinstalled in the system. The system is preinstalled with the
ftware is installed or if the existing software is
This product contains third party’s software that is made available as open
This software is provided “as is” with no warranty of any kind as to its
If you would like to receive such source codes, please contact INNOLUX
from which you
purchased this product. (Please be noted that any inquiries concerning
the contents of source codes should be directed to original licensors of
_____________________________________________________________
_____________________________________________________________
rs at the time of disinfection,

© INNOLUX Corporaon │ X
which the quality,
performance and safety of the equipment cannot be
assured.
-
Chloric disinfectant which is strongly corrosive to metals and rubber parts.
-
Disinfectant whose uses on metals, plastics, and coating are forbidden
according to the
instructions supplied with the disinfectant.
-
Formalin gas and disinfectant sprays that may
-
Ultraviolet sterilizers
Disinfectant ethanol is recommended for disinfection. Carefully read the
instructions and cautions supplied with the disinfectant before use.
For details on the disinfectant, contact a INNOLUX dealer or th
representatives at the agency from which you purchased the disinfectant.
Caution
If flat panel detector
is not disinfected, it may lead secondary infection.
Be sure to disinfect with ethanol after use.
Clean the flat panel
detector
to prevent infection.
_____________________________________________________________
2.12
Charging the Battery Pack
_____________________________________________________________
Caution
Charging
battery only by the back
battery is installed on Yushan flat panel detector or u
provided
by INNOLUX Corporation.
instruction manual for the battery charger.
Do
not charge the battery pack near fire or under strong sunsh
protection
mechanisms are activated by a high temperature, the batter
cannot be charged. Also,
the b
attery pack may be
abnormal chemical
reactions may occur inside the
overheat, emit smoke, explode or ignite.
© INNOLUX Corporaon │ X
-Ray Products
performance and safety of the equipment cannot be
Chloric disinfectant which is strongly corrosive to metals and rubber parts.
Disinfectant whose uses on metals, plastics, and coating are forbidden
instructions supplied with the disinfectant.
Formalin gas and disinfectant sprays that may
get inside the equipment.
Ultraviolet sterilizers
.
Disinfectant ethanol is recommended for disinfection. Carefully read the
instructions and cautions supplied with the disinfectant before use.
For details on the disinfectant, contact a INNOLUX dealer or th
representatives at the agency from which you purchased the disinfectant.
is not disinfected, it may lead secondary infection.
Be sure to disinfect with ethanol after use.
detector
with ethanol for disinfection, etc. for each patient
_____________________________________________________________
Charging the Battery Pack
_____________________________________________________________
battery only by the back
-
up cable provided by INNOLUX when
battery is installed on Yushan flat panel detector or u
se the battery charger
by INNOLUX Corporation.
For details on operations, refer to the
instruction manual for the battery charger.
not charge the battery pack near fire or under strong sunsh
ine. If the built
mechanisms are activated by a high temperature, the batter
cannot be charged. Also,
if the built-
in protection mechanisms are damaged,
attery pack may be
charged with
extremely high current and voltage, and
reactions may occur inside the
battery pack, causing it to
overheat, emit smoke, explode or ignite.
1
performance and safety of the equipment cannot be
Chloric disinfectant which is strongly corrosive to metals and rubber parts.
Disinfectant whose uses on metals, plastics, and coating are forbidden
get inside the equipment.
Disinfectant ethanol is recommended for disinfection. Carefully read the
instructions and cautions supplied with the disinfectant before use.
For details on the disinfectant, contact a INNOLUX dealer or th
e service
representatives at the agency from which you purchased the disinfectant.
is not disinfected, it may lead secondary infection.
with ethanol for disinfection, etc. for each patient
_____________________________________________________________
_____________________________________________________________
up cable provided by INNOLUX when
se the battery charger
For details on operations, refer to the
ine. If the built
-in
mechanisms are activated by a high temperature, the batter
y pack
in protection mechanisms are damaged,
extremely high current and voltage, and
battery pack, causing it to

© INNOLUX Corporaon │ X
To charge the battery pack, be sure to use the designated
to observe
the charging conditions specified by INNOLUX Corporation. If
battery pack is charged in
higher th
an specified, remodeled battery
be ov
ercharged or charg
chemical reactions may occur inside the batter
emit smoke, explode or ignite.
Immediately stop charging the battery pack, if charging is not completed within
the specified
time. Otherwise, the battery pack may overheat, emit smoke,
explode or ignite.
Do not use the flat panel
Do not use a faulty or broken battery charger or AC adapter.
_____________________________________________________________
2.13
Battery Pack Instructions
_____________________________________________________________
Warning
Battery pack requires regular checkup and replacement. Battery capacity
begins to warm
after a period of time.
If this equipment is not in use for while, store it with the battery pack removed.
Not
removing the battery pack may cause malfunction.
Caution
The battery pack is
used with the flat panel detector
combinations.
Charge the battery pack only with the designated battery charger. If the battery
pack is
charged under the charging conditions (voltage, current and charging
method) different
from those specified by INNOLUX Corporation, the battery
pack may emit
smoke, ignite,
Store the battery pack in a cool and dark place. Recharge the stored battery
pack every
six months or every year. Otherwise a decrease in battery capacity
or other problems may
result.
© INNOLUX Corporaon │ X
-Ray Products
To charge the battery pack, be sure to use the designated
battery charger and
the charging conditions specified by INNOLUX Corporation. If
battery pack is charged in
other conditions (temperature or voltage/current
an specified, remodeled battery
charger, etc.), the battery pack may
ercharged or charg
ed with extremely high current,
and abnormal
chemical reactions may occur inside the batter
y pack, causing it to overheat,
emit smoke, explode or ignite.
Immediately stop charging the battery pack, if charging is not completed within
time. Otherwise, the battery pack may overheat, emit smoke,
Do not use the flat panel
detector near the power cable.
Do not use a faulty or broken battery charger or AC adapter.
_____________________________________________________________
Battery Pack Instructions
_____________________________________________________________
Battery pack requires regular checkup and replacement. Battery capacity
after a period of time.
If this equipment is not in use for while, store it with the battery pack removed.
removing the battery pack may cause malfunction.
used with the flat panel detector
. Do not use them in other
Charge the battery pack only with the designated battery charger. If the battery
charged under the charging conditions (voltage, current and charging
from those specified by INNOLUX Corporation, the battery
smoke, ignite,
explode or leak fluid.
Store the battery pack in a cool and dark place. Recharge the stored battery
six months or every year. Otherwise a decrease in battery capacity
result.
19
battery charger and
the charging conditions specified by INNOLUX Corporation. If
the
other conditions (temperature or voltage/current
charger, etc.), the battery pack may
and abnormal
y pack, causing it to overheat,
Immediately stop charging the battery pack, if charging is not completed within
time. Otherwise, the battery pack may overheat, emit smoke,
_____________________________________________________________
_____________________________________________________________
Battery pack requires regular checkup and replacement. Battery capacity
If this equipment is not in use for while, store it with the battery pack removed.
. Do not use them in other
Charge the battery pack only with the designated battery charger. If the battery
charged under the charging conditions (voltage, current and charging
from those specified by INNOLUX Corporation, the battery
Store the battery pack in a cool and dark place. Recharge the stored battery
six months or every year. Otherwise a decrease in battery capacity

20
© INNOLUX Corporaon │ X-Ray Products
Do not leave the removed battery pack in the car or other places exposed to
high temperature. If the battery pack is used or stored in a place where it is
exposed to high temperature, the battery pack may emit smoke, ignite,
explode or leak fluid.
Use or store the battery pack only in the environmental conditions specified by
INNOLUX. If the battery pack is used or stored in a place where it is exposed
to high temperature, the battery pack may emit smoke, ignite, explode or leak
fluid.
When disposing of the battery pack, consult our official dealer or INNOLUX
Representative.
Do not disassemble or remodel the battery pack. The battery pack is equipped
with built-in safety and protection mechanisms. If they are damaged, the
battery pack may overheat, emit smoke, explode or ignite.
Be careful not to drop the battery pack. The patient may be injured.
Do not touch the terminal of the battery pack directly. There is a risk of electric
shock.
Do not connect the positive (+) and negative (-) terminals with a wire or any
metal object.
Do not carry or store the battery pack together with metal objects such as
necklaces or hairpins. Otherwise, the battery pack may short-circuit and
overcurrent may flow, causing the battery pack to overheat, emit smoke,
explode or ignite. Metal objects such as necklaces or hairpins may also
become hot.
Do not throw the battery pack into fire or expose it to excessive heat.
Otherwise, its insulator may melt, its gas release vent or safety mechanisms
may be damaged, and/or its electrolyte may catch fire, causing the battery
pack to overheat, emit smoke, explode or ignite.
Do not use or leave the battery pack in a place where it is exposed to high
temperature (60C or higher), such as fire or a heater. If the resin separator is
damaged due to heat, the battery pack may short-circuit, causing it to overheat,
emit smoke, explode or ignite.
Do not immerse the battery pack in water or seawater, and do not allow it to
become wet. If the built-in protection mechanisms are damaged, the battery
This manual suits for next models
2
Table of contents
Popular Security Sensor manuals by other brands

Burg Wächter
Burg Wächter ProfiScale LOCATE S operating instructions

Steelman
Steelman ChassisEAR 06600 Operating information

Interlogix
Interlogix advisorone user manual

Apollo
Apollo XP95A Series installation guide

CENTURION SYSTEMS
CENTURION SYSTEMS i5 Pocket installation guide

Optex
Optex HX-40 installation instructions