Innovive Innorack IVC Mouse User manual

RACK SETUP
Innorack®IVC Mouse
Quick Start Guide
Supply blower(s) Exhaust blower(s)
Controller mount
with screws
Wall mount
bracket(s)
Blower port plugs
Controller with cable
Power supply with cables Power cable
RACK COMPONENTS
Single-sided rack only
Single-sided rack (1-sm, 1-lrg)
Double-sided rack (3-sm, 2-lrg)
Before proceeding please check the package to see if
the items below have been included in the shipment. F3
F1
F4
F5
F6
F7
F2
Controller Port
Rack to Blower Cable Port
Power Supply
Por t
Single-sided rack (1 power cable attached)
Double-sided rack (2 power cables attached)
1. INSTALLING THE POWER SUPPLY
Verify the power switch is set to the off
position ( O ).
Double-sided racks mount the power supply
on “Supply A” or “Exhaust B”. Single-sided
racks mount the power supply on the Supply
Blower end.
Secure power supply box to the top of
the rack by placing the L-shaped bosses
extending from the power supply box into the
mating feature of the plastic blower supports
(see F1).
2. MOUNTING THE CONTROLLER
NOTE - Mounting instructions are for
single-sided racks. Optional for double-
sided racks.
Attach controller mount bracket to the top
of blower support section using the screws
provided and a #2 Phillips Head screwdriver
(see F2).
Mount controller to the Controller Bracket
(see F3).
3. INSTALLING THE BLOWERS
Match the label on the front of the blower with
the corresponding label on the side of the rack.
Tilt blower back and slide the two grooves in
the back of the blower onto the two hooks at
the back of the blower support (see F4).
Check to make sure blower feet sit properly in
feet cavities. Pull the front of the blower down
until both latches on the blower engage with
the blower support. Press down on the front of
the blower to ensure that it is locked in place
Connect the Rack to Blower Cable Port
(see F5).
4. CONNECTING THE CONTROLLER
CABLE AND POWER SUPPLY CABLES
Attach the controller cable at the Controller
Port and then connect the power supply
cables at Power Supply Port as shown in F5 &
F6. Single-sided systems will only attach one
power supply cable into the Power Supply Port.
Attach the provided power cable to the power
supply box (see F7) and then plug the power
cord into a wall outlet or a power strip
(100-240 VAC).
Finally, ensure all the blowers and the
controller are plugged into the rack prior to
turning the system on and insert the Blower
Port Plugs into any unused ports.
5. TURNING THE SYSTEM ON
Press the switch on the power box located
on the top of the rack to the on position ( | ).
*Remove the wash caps before continuing with the setup process.

1 2
3 4
5 6
DIGITAL CONTROLLER
BLOWER CALIBRATION
Tutorials • www.innovive.com/resources/video-training.html
F 10
F9
F8
Network Info
Sound Volume
Screen Brightness
Blower Calibration
Blower Calibration
This will calibrate airflow
sensors on all blowers. Press
'Start' to begin
Current values:
Updated values:
Calibrate blower airflow
Up
Done
Down
Start
6/6
Network Info
Sound Volume
Screen Brightness
Blower Calibration
Blower Calibration
Ensure rack is disconnected
from HVAC. Press Calibrate
to begin
Current values: 350 351 352 353
Updated values: • • • •
Calibrate blower airflow
Up
Done
Down
Calibrate
6/6
Network Info
Sound Volume
Screen Brightness
Blower Calibration
Blower Calibration
Please confirm rack is
disconnected from HVAC
Current values: 350 351 352 353
Updated values: • • • •
Calibrate blower airflow
Up
Done
Down
Confirm
6/6
F 12
F 11
Network Info
Sound Volume
Screen Brightness
Blower Calibration
Blower Calibration
Calibrating side A...
Supply
Current values: 350 351 352 353
Updated values: • • • •
Calibrate blower airflow
Up
Done
Down
6/6
Network Info
Sound Volume
Screen Brightness
Blower Calibration
Blower Calibration
Calibration complete,
restoring setpoints.
Success!
Current values: 350 351 352 353
Updated values: 380 358 361 387
Calibrate blower airflow
Up
Done
Down
6/6
BEFORE OPERATING THE INNORACK, LEAVE THE RACK
RUNNING IN ITS OPERATING ENVIRONMENT FOR 24 HOURS.
ALL RACKS MUST BE CALIBRATED PRIOR TO OPERATING.
DO NOT PRESS ANY BUTTONS WHILE BLOWER
IS CALIBRATING.
DO NOT DISCONNECT BLOWERS WHILE RACK
IS POWERED ON!
IT IS RECOMMENDED TO REPEAT THE BLOWER
CALIBRATION PROCEDURE EVERY 6 MONTHS.
STEP 1. GETTING TO THE BLOWER CALIBRATION SCREEN
• Press and hold bottom buttons 5 and 6 on the digital controller until
the menu screen appears.
• Scroll UP (button 1) or DOWN (button 5) until you reach the Blower
Calibration screen
(see F8).
• Press START to begin
(see F8)
.
The controller will confirm the number of rack sides and current
blower calibration values
(see F9)
.
STEP 2. DISCONNECT HVAC BEFORE CALIBRATING
• Make sure the rack is disconnected from the HVAC system.
• Press CALIBRATE to begin
(see F9)
.
• Press CONFIRM to ensure the rack is disconnected from all HVAC
systems
(see F10).
STEP 3. CALIBRATING THE BLOWER
After confirming the rack is disconnected from the HVAC, calibration of
each blower will begin
(see F11)
.
During the calibration process, the following will take place:
1. Airflow set points will be saved.
2. Blower fans will be turned off, and the controller will wait 30 seconds for
the fans to stop spinning.
3. Each blower will be calibrated in sequence.
As each blower is calibrated, the updated calibration values will be
displayed at the bottom right of the calibration screen (
see F12).
STEP 4. BLOWER CALIBRATION IS COMPLETE
SUCCESS: Upon successful recalibration of all blowers, airflow set points
will be restored. Blower calibration screen will remain displayed. To exit
blower calibration screen, press DONE
(see F12)
.
PROBLEMS: If any problems are encountered during the blower
calibration process, airflow set points will be restored. Problem message
will appear on the screen. Contact Innovive support for assistance.
If failure occurs, restart the blower calibration process, press RESTART.
To cancel the blower calibration process, press DONE.
Table of contents
Popular Rack & Stand manuals by other brands

Kioti
Kioti CK20 Assembly manual

Rackmount
Rackmount 5URAIL-2900-EARS installation instructions

Smart Inventions
Smart Inventions FSSBID 100 Installation and integrator's guide

Honeywell Home
Honeywell Home ADEMCO 5828DM Installation and setup guide

OTC
OTC 1750 Assembly & operating instructions

Denon
Denon FS500DSD Assembly guide