• The patient can measure, transmit data, and change batteries under 
normal circumstances and maintain the device and its accessories 
according to the accompanying documents.
• To avoid measurement errors, please avoid strong electromagnetic 
fields, radiated interference signals, or electrical fast transient/burst 
signals.
• The blood pressure monitor and the cuff are suitable for use within 
the patient environment. If you are allergic to polyester, nylon, or 
plastic, please don't use this device.
• During use, the patient will be in contact with the cuff. The materials 
of the cuff have been tested and found to comply with requirements 
of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any 
potential sensation or irritation reaction.
• Adapter is specified as a part of ME EQUIPMENT.
• If you experience discomfort during a measurement, such as pain in 
the arm or other complaints, press the START/STOP button to release 
the air immediately from the cuff. Loosen the cuff and remove it from 
your arm.
• If the cuff pressure reaches 40 kPa (300 mm Hg), the unit will 
automatically deflate. Should the cuff not deflate when pressure 
reaches 40 kPa (300 mm Hg), detach the cuff from the arm and press 
the START/STOP button to stop inflation.
• Before use, make sure that the device functions safely and is in proper 
working condition. Do not use the device if it is damaged in any way. 
The continuous use of a damaged unit may cause injury, improper 
results, or serious danger.
• Do not wash the cuff in a washing machine or dishwasher.
• The service life of the cuff may vary by the frequency of washing, skin 
condition, and storage state. The typical service life is 10,000 times.
• It is recommended that the performance should be checked every 
two years and after maintenance and repair, by retesting at least the 
requirements in limits of the error of the cuff pressure indication and 
air leakage (testing at least at 50 mm Hg and 200 mm Hg).
• Please dispose of ACCESSORIES, detachable parts, and the ME 
EQUIPMENT according to the local guidelines.
• Manufacturer will make available on request circuit diagrams, 
component part lists, descriptions, calibration instructions, etc., to 
assist to service personnel in parts repair.