invicon TwinLux Dual User manual

Instructions for use
Informations concernant l‘utilisation
Foglio illustrativo
Gebrauchsinformation

2
Manufacturer:
Invicon chemical solutions GmbH
Schweizer Straße 96
A 6830 Rankweil
Austria
Tel.: +43 (5522) 45301 – 0
Fax: +43 (5522) 45301 – 10
www.invicon.at
User manual created:
04/2017
English
Contents
1 Safety
1.1 General instructions
1.2 Labels in the user manual
1.3 Safety information
1.4 Intended use
1.5 Standards and classifications
1.6 Labels on the device
2 Device description
3 Start-up operation
3.1 Checking the charge level
3.2 Charging the battery
4 Operation
4.1 Applying light
4.2 Increasing the exposure time during
exposure
5 Maintenance and cleaning
6 Troubleshooting
3
3
3
4
5
5
5
6
9
9
11
11
12
13
14
14
7 Technical data
7.1 Light sources
7.2 Battery
7.3 Power supply
7.4 Weight
7.5 Environment / Storage conditions
7.6 Risk Group
8 Scope of delivery and spare parts
9 Guarantee
9.1 Guarantee conditions
9.2 Guarantee exclusion
9.3 Guarantee claim procedure
EC Declaration of Conformity
10 Guarantee certificate
information
15
15
15
15
15
15
15
16
16
16
16
17
18
19

3
1| Safety
1.1 General information
Please read the user manual for your device carefully before using
it for the first time. It contains important information on safety and
using your device and its components.
1.2 Labels in the user manual
Special warning information can be found in the user manual before
dangerous tasks. This information is broken down into the following
danger classes:
Warning!
Dangers which could lead to serious injury or
significant property damage.
Caution!
Dangers which could lead to injury or property
damage.
Note!
Contents of the note.
In addition to these warnings, there are also general notes
containing useful information.
Danger!
Dangers which could lead to serious injury
or death.

4
1.3 Safety information
Danger!
Light from the LED light source
Risk of explosion in areas with highly flammable air conditions.
-> Do not operate the device in areas where there is a risk
of explosion.
Danger!
Electromagnetic disturbances
May induce malfunctions in pacemakers or other implanted
electronic devices.
-> Do not use the device in the vicinity of a pacemaker or
any other implanted electronic device.
Warning!
Light from the LED light source
–Do not look directly into the LED light source.
–Only use the device when the protective shield is attached
and functioning correctly.
–Always ensure the protective shield is attached correctly.
–Never point the beam of the LED light source directly at eyes.
–Always aim the beam directly at the polymerisation surface.
Warning!
Defective device
Risk of injury due to parts of the device not functioning
correctly.
Check the following points before each use:
–The device and its components are in good working order.
–The protective shield is attached correctly.
–The ventilation slots are not obstructed.
Caution!
Exposure of the device to water
Damage to the device due to water exposure.
–Do not hold the device under running water or immerse it
in water.
–Keep the device out of the reach of children.
–Only use genuine accessories supplied by the manufacturer.
–In the event of malfunction, request assistance from an
authorised customer support service or from Invicon chemical
solutions immediately.

5
1.4 Intended use 1.6 Labels on the device
1.5 Standards and classifications
The device is intended to polymerise light-curing materials of
HyCeram®Bond II or Bond THX NF, LuxArt®and Luxon® design
materials.
–The device may only be used by trained specialists.
–Always use light protection (the protective shield) when operating
the device.
The following device information can be found on the device itself:
–Model
–Serial number
–Supply voltage
–Power consumption
Please always provide this information if you contact an authorised
customer support service or Invicon chemical solutions directly.
This data allows us to ensure that problems are solved quickly and
efficiently.
EC standard: EN62471
IEC classification Risk group 2
CIE classification Risk group 2
> We recommend wearing protective goggles.
(Optional available)

6
2| Device description
The TwinLux Dual is a mains-independent, portable device with a
built-in battery. It is used to cure the following light-curing materials:
–HyCeram®Bond II or Bond THX NF materials
–LuxArt®design composite
–Luxon®design acrylate
More detailed descriptions of HyCeram®, LuxArt® and Luxon®can
be found in the corresponding instructions for use.
Two high-efficiency Royal Blue LED diodes with a wavelength range
of 425 - 470 nm (peak 450 nm) are used as light sources.
One UV LED diode with a wavelength range of 350 - 375 nm
(peak 365 nm) is also used.
The total radiant power is 1.5 W.
This means that the entire LED light spectrum emitted by TwinLux
Dual is used to full effect in the activation of the HyCeram®Bond II or
Bond THX NF, LuxArt®or Luxon®initiators.This produces excellent
polymerisation results with low energy consumption and minimal
heat loss.
The LED light sources can be switched on and off using the multi-
function button (4).
Three different preset exposure times can be selected, depending on
how the button is pressed.
The following exposure times can be chosen:
–30 seconds
–60 seconds
–Continuous exposure
As soon as the function button is pressed and the light is active,
the exposure indicator will light up or start to blink.
The exposure times can be activated as follows:
Exposure time
30 seconds
60 seconds
Continuous exposure
Multifunction button
action (7)
Short tap
Short press
Long press
Exposure time
indicator (11)
Fast blinking
Slow blinking
Steady light

7
The following illustration always refers to the TwinLux Dual device.
Mains connection (3)
The mains connection supplies power to the TwinLux Dual and charges
the battery.
Overheat protection (4)
The TwinLux Dual is designed for intensive, continuous use. The device
features overheat protection with ventilation slots (4) in order to prevent
overheating.
LED light source (5)
Light from the LED light source.
Risk of eye injury from direct exposure to light.
–Do not look directly into the LED light source.
–Only use the device when the protective shield is attached and functioning
correctly.
–Always ensure the protective shield is attached correctly.
–Never point the beam of the LED light source directly at eyes.
–Always aim the beam directly at the polymerisation surface.
1Power supply unit plug
2Power connector on device
3Mains connection
4Overheat protection with ventilation slots
5LED light source
6Stand
7Function button on/off
8Protective shield
TwinLux Dual with accessories:
3
2
1
7
8
5
6
4
Warning!
Note!
Illustration in the user manual

8
The LED light source (5) contains two Royal Blue LEDs and one UV LED
behind a protective Plexiglas pane.
Royal Blue LED:Wavelength range of 425 - 470 nm (peak 450 nm).
UV LED:Wavelength range of 350 - 375 nm (peak 365 nm).
Stand (6)
The stand (6) is used to secure the unit.
Function button (7)
The multifunction button (7) is used to switch the LED light source (5) on
and off. As soon as the function button is pressed and the light is active,
the exposure time indicator (11) will light up or start to blink.
Protective shield (8)
Attaching the protective shield prevents uncontrolled light emission
from the LED light source (5).
LED indicators
Power supply indicator (9)
The red LED (9) on the back of the device shows the power
supply status:
9 Power supply indicator
10 Charge level indicator
11 Exposure time indicator
9
10
11
TwinLux Dual
No power supply
Battery is charging
Power supply indicator
LED off
Steady light

9
Charge level indicator (10)
The red LED (10) on the top of the device indicates the battery charge
level of the light unit (see also 3.2).
Accurate charge level indication
The charge level will be indicated more accurately if you carry out 2 - 3
exposures beforehand.
Exposure time indicator (11)
The green LED (11) on the top of the device indicates the selected
exposure time. The device will automatically switch off after the se-
lected exposure time is up (unless continuous exposure is selected).
Note!
Exposure time
30 seconds
60 seconds
Continuous exposure
Multifunction button
action (7)
Short tap
Short press
Long press
Exposure time
indicator (11)
Fast blinking
Slow blinking
Steady light
3| Start-up operation
> Place the device on a horizontal surface.
> Connect the power supply unit plug (1) to the power connector.
3.1 Checking the charge level
> Check the battery charge level.

10
Note!
Validity of time indications
The time indications are only accurate at room temperature (around 18 – 25° C) and after around five minutes of continuous use.
Device status Battery status Battery capacity
Charge level indicator (10) blinks 1 x Full Around 30 minutes of use remaining
Charge level indicator (10) blinks 2 - 3 x 2/3 full Around 20 minutes of use remaining
Charge level indicator (10) blinks 4 - 5 x 1/3 full Around 15 minutes of use remaining
Charge level indicator (10) blinks
continuously Battery nearly
empty
Around 10 minutes of use remaining
The following exposure cycles are still possible:
>10 - 20 exposure cycles of 30 seconds each
>5 - 10 exposure cycles of 60 seconds each
Light does not activate when button is
pressed Battery drained Connect the power supply unit and charge the battery for 7 to 12 hours
Device turns off automatically Battery defective Have the device examined by an authorised specialist or directly by Invicon
chemical solutions.

11
Note!
Note!
3.2 Charging the battery
Increased battery lifespan
Charging the battery fully and then discharging it fully will increase the
lifespan of the battery.
Battery charge level on delivery
The battery is not fully charged on delivery.
–Fully charge the battery for 12 - 14 hours before first use.
–Always fully charge the battery for 7 - 12 hours after the first discharge.
4| Operation
Warning!
Warning!
Light from the LED light source
Risk of eye injury from direct exposure to light.
–Do not look directly into the LED light source.
–Only use the device when the protective shield is attached and
functioning correctly.
–Always ensure the protective shield is attached correctly.
–Never point the beam of the LED light source directly at eyes.
–Always aim the beam directly at the polymerisation surface.
Defective device
Risk of injury due to parts of the device not functioning correctly.
Check the following points before each use:
–The device and its components are in good working order.
–The protective shield is attached correctly.
–The ventilation slots are not obstructed.

12
4.1 Applying light
7 7
8
8
5
5LED light source
7Function button
8Protective shield
TwinLux dual
Attach the protective shield (8) over the LED light source (5).
Press the function button (7).
Exposure
Exposure time
30 seconds
60 seconds
Continuous exposure
Multifunction button
action (7)
Short tap
Short press
Long press
Exposure time
indicator (11)
Fast blinking
Slow blinking
Steady light
Once light exposure has started, briefly press the function button (7)
again to stop the exposure.

13
4.2 Increasing the exposure time during exposure
It is also possible to increase the exposure time by one level during
exposure:
–To increase the exposure time to 60 seconds, when an exposure time
of 30 seconds has been chosen, hold down the function button (7) for
two seconds.
–To increase the exposure time to continuous exposure, when an
exposure time of 60 seconds has been chosen, hold down the function
button (7) for longer than two seconds.
Exposure time
120 seconds
120 seconds
120 seconds
Material
HyCeram®Bond II or Bond THX NF
LuxArt® Design
Luxon® Design
Distance
2 cm
2 cm
2 cm
Please observe the following exposure times and distances
Note!
Ensure that all material surfaces are exposed to light carefully.
(Exposure Time, distance).

14
5| Maintenance and cleaning
If the surface of the device or the LED light source (5) gets dirty:
> Switch off device
> Disconnect the power supply unit plug (1)
6| Troubleshooting
Cleaning the surface of the device
Damage to the device due to water exposure.
–Do not hold the device under running water or immerse it in water.
–Do not let liquids enter the ventilation slots (4).
–Leave the device to dry for at least five minutes after cleaning.
–Clean the surface with a damp cloth (alcohol-based cleaner).
If the ventilation slots (4) are very dirty:
–Switch off the device.
–Disconnect the power supply unit plug (1).
–Clean ventilation slots (4) with compressed air.
Caution!
Problem Potential
cause Solution
Device turns off automatically Battery
Have the device
examined by an
authorised specialist
or directly by Invicon
chemical solutions.
Device turns off automatically Malfunction
Have the device
examined by an
authorised specialist
or directly by Invicon
chemical solutions.

15
7| Technical data
LEDs 2 x LED Royal Blue 425 - 470 nm (peak 450 nm)
1 x LED UV 350 - 375 nm (peak 365 nm)
Total radiant
power 1,5 W Temperature -10° C - +70° C
Rel. humidity 10 % - 90 %
Air pressure 500 hPa - 1060 hPa
Initial charge time 12 - 14 hours
Recharge time 7 - 12 hours
Number of cells 4
Type NiMH, HHR, 210AAB2B
Voltage 4.8 V
Power supply unit DC 12V/1A
Power supply unit 90 - 260 VAC / 50 - 60 Hz
7.1 Light sources
7.5 Environment/Storage conditions
7.4 Weight: 394 g (not including power supply unit)
7.6 Risk group: 2 (Classification in accordance with DIN EN 62471
300 nm - 1400 nm)
7.2 Battery
7.3 Power supply

16
8| Scope of delivery and spare parts
1x TwinLux Dual light unit
1x power supply unit
1x protective shield
1x user manual
1x guarantee certificate
TwinLux Dual Power supply unit
Protective shield
9| Guarantee
9.1 Guarantee conditions
Invicon chemical solutions guarantees all TwinLux Dual light units
sold (including by an Invicon reseller or importer) against material or
production faults, effective for two years from the date of purchase.
During the guarantee period, Invicon chemical solutions agrees to
repair parts it considers defective free of charge or to replace them at
its own discretion. In general, the relevant valid version of the General
Terms and Conditions of Business and Delivery shall apply.
Replacement of the entire TwinLux Dual light unit is not possible.
9.2 Guarantee exclusion
In the following cases, guarantee claims shall be void:
•The device or the associated components were not used in
accordance with the safety rules and instructions described in these
instructions for use.
•The device or the associated components were not used in
accordance with the intended use.
•The device or components were structurally modified without
authorisation.

17
•Assembly, settings, modifications, extensions or repairs were not
carried out by Invicon specialist personnel or authorised specialist
companies.
•The electrical system in the rooms in which the device is operated
does not correspond to the valid laws or rules.
•The ambient conditions in which the device is stored or used do not
correspond to the conditions listed in the “Technical data” section.
•Disaster situations as a result of foul play, accident, vandalism or
force majeure.
The following components are excluded from the guarantee:
•LED light sources
•LED indicators
•Switches
•Accessories
The following damage is excluded from the guarantee:
•Transport damage
•Damage resulting from improper use or neglect
•Damage resulting from connection to a voltage that the device is not
designed for
9.3 Guarantee claim procedure
The guarantee is only valid if the guarantee certificate provided with
the unit is completed in full, including proof of purchase with date of
purchase, and delivered to our company headquarters or an autho-
rised Invicon chemical solutions sales partner within 15 days of the
date of purchase.
In the event of a guarantee claim, the customer must provide the
device to Invicon chemical solutions or the Invicon sales partner, and
must supply the following information:
•Name, address, phone number and e-mail address of the buyer
•Name and contact details of the Invicon distributor
•Serial number of the device
•Proof of purchase for the device (delivery note)
•Detailed description of the malfunction
The information contained in this description is non-binding and may
be changed at any time without prior notice.

18
EC Declaration of Conformity
pursuant to EU Low Voltage Directive 2014/35/EU, dated 26/2/2014
Manufacturer: oehri electronic AG | Industriering 35 | 9491 Ruggell, Liechtenstein
Description and identification of interchangeable equipment:
Product: TwinLux-Dual
Task: NA16-002406 and following
Function: Blue/UV portable exposure unit
It is expressly declared that the interchangeable equipment complies with all relevant conditions of the following EU directives:
2014/35/EU: (Low Voltage Directive) Directive of the European Parliament and of the Council of 26 February 2014 on the harmonisation of
the laws of the Member States relating to the making available on the market of electrical equipment designed for use within
certain voltage limits
2011/65/EC RoHS: Restriction of the use of certain hazardous substances
Reference of applied harmonised standards:
EN61000-6-1:2007 Electromagnetic compatibility (EMC) - Part 6.1 Generic standards
EN61000-6-3:2006-3 Emission standard and immunity for residential, commercial and industrial environments
EN 62471 300 to 1400 nm, risk group 2 (moderate risk):
Labelling is required.
EC Declaration of Conformity
pursuant to EU Low Voltage Directive 2014/35/EU, dated 26/2/2014
oehri electronic AG
Ruggell, 22 November 2016
EG-Konformitätserklärung
im Sinne der EU-Niederspannungsrichtlinie 2014/35/EU, vom 26.2.2014
Hersteller:
oehri electronic AG
Industriering 35
9491 Ruggell
Liechtenstein
Beschreibung und Identifizierung der auswechselbaren Ausrüstung:
Produkt: TwinLux-Dual
Auftrag: NA16-002406 und folgende
Funktion: Blau / UV Handbelichtungsgerät
Es wird ausdrücklich erklärt, dass die auswechselbare Ausrüstung allen einschlägigen
Bestimmungen der folgenden EG-Richtlinien entspricht:
2014/35/EU: (Niederspannungsrichtlinie) Richtlinie des Europäischen Parlaments und des Rates
vom 26.Februar 2014 zur Angleichung der Rechtsvorschriften der Mitgliedsstaaten
betreffend elektrische Betriebsmittel zur Verwendung innerhalb bestimmter
Spannungsgrenzen
2011/65/EG RoHS : Beschränkung der Verwendung bestimmter gefährlicher Stoffe
Fundstelle der angewandten harmonisierten Normen:
EN61000-6-1:2007 Elektromagnetische Verträglichkeit (EMV) - Teil 7-01 Fachgrundnormen;
EN61000-6-3:2006-3 Störaussendung und Störfestgkeit für Wohnbereich, Geschäfts- und Gewerbebereiche
sowie Kleinbetriebe
EN 62471 300 bis 1400 nm, Risikogruppe 2 (mittleres Risiko):
Eine Kennzeichnung ist erforderlich.
EG-Konformitätserklärung
im Sinne der EU-Niederspannungsrichtlinie 2014/35/EU, vom 26.2.2014
oehri electronic AG
Ruggell
, 22. November 2016
Manfred Laim
Geschäftsführer
Manfred Laim
Managing Director

19
Guarantee certificate information
Date of purchase
Serial number
Company
Contact person
Address
Postcode
Location
Street and no.
Country
Telephone
E-Mail
Invicon sales partner
Manufacturer
Invicon chemical solutions GmbH
Schweizer Straße 96
A 6830 Rankweil
Austria
Tel.: +43 (5522) 45301 – 0
Fax.: +43 (5522) 45301 – 10
www.invicon.at

20
Fabricant:
Invicon chemical solutions GmbH
Schweizer Straße 96
A 6830 Rankweil
Autriche
Tel.: +43 (5522) 45301 – 0
Fax: +43 (5522) 45301 – 10
www.invicon.at
Création du guide d’utilisation:
04/2017
Table des matières
1 Sécurité
1.1 Instructions générales
1.2 Terminologie du guide d’utilisation
1.3 Consignes de sécurité
1.4 Utilisation conforme à l’usage prévu
1.5 Normes et classifications
1.6 Caractéristiques de l’appareil
2 Description de l’appareil
3 Mise en service
3.1 TwinLux Dual
3.2 Chargement de la batterie
4 Utilisation
4.1 Mise en œuvre d’une exposition lumineuse
4.2 Augmentation de la durée d’exposition au
cours d’une exposition
5 Maintenance et nettoyage
6 Assistance en cas de problème
21
21
21
22
23
23
23
24
27
27
29
29
30
31
32
32
7 Données techniques
7.1 Sources lumineuses
7.2 Batterie
7.3 Alimentation
7.4 Poids
7.5 Conditions ambiantes/ de stockage
7.6 Groupe de risque
8 Contenu et pièces de rechange
9 Garantie
9.1 Conditions de garantie
9.2 Exclusion de garantie
9.3 Procédure en cas de recours à la
garantie
Déclaration de conformité CE
Informations sur le bon de garantie
33
33
33
33
33
33
33
34
34
34
34
35
36
37
Français
Other manuals for TwinLux Dual
1
Table of contents
Languages:
Popular Power Tools manuals by other brands

STEINEL
STEINEL HG 2320 E Translation of the original operating instructions

Buehler
Buehler SimpliMet 3000 manual

Draper
Draper PT142K instruction manual

Far Tools
Far Tools P 13F Original manual translation

Milwaukee
Milwaukee M18 BCT Operator's manual

Fast Mover
Fast Mover FMT7560 Product instruction manual